NC150 BT
**Non-Contact Thermometer
Instruction Manual
Microlife NC 150 BT
IB NC 150 BT S-V11 1021
Revision Date: 2021-02-08**
Microlife NC 150 BT
Microlife Non Contact Thermometer NC 150 BT
1 Measuring sensor | 15 Low battery indicator |
---|---|
2 Tracking light | 16 Changing between Celsius and Fahrenheit |
3 Display | 17 Recall mode |
4 START button | 18 Recall the last 30 readings |
5 ON/OFF button | 19 Measured temperature too high |
6 Mode switch | 20 Measured temperature too low |
7 Battery compartment cover | 21 Ambient temperature too high |
8 M-button (memory) | 22 Ambient temperature too low |
9 All segments displayed | 23 Error function display |
10 Memory | 24 Blank display |
11 Ready for measuring | 25 Flat battery |
12 Measurement complete | 26 Date/Time |
13 Body mode | 27 Beeper function setting |
14 Object mode | 28 Replacing the battery |
Read the instructions carefully before using this device. | |
--- | --- |
Type BF applied part | |
Keep dry |
This Microlife thermometer is a high-quality product incorporating the latest
technology and tested in accordance with international standards. With its
unique technology, this device can provide a stable, heat-interference-free
reading with each measurement.
The device performs a self-test every time it is switched on to always
guarantee the specified accuracy of any measurement. This Microlife
thermometer is intended for the periodic measurement and monitoring of human
body temperature.
This thermometer has been clinically tested and proven to be safe and accurate
when used in accordance with the operating instruction manual.
Please read through these instructions carefully in order for you to
understand all functions and safety information.
The Advantages of this Thermometer
Measures in a matter of seconds
The innovative infrared technology allows the measurement without even
touching the object. This guarantees safe and sanitary measurements within
seconds.
Multiple uses (wide range of measurement)
This thermometer offers a wide range of measurements from 0.1 – 99.9 °C / 32.2
– 211.8 °F, meaning the unit can be used to measure body temperature or it
also has a feature allowing it to be used to measure the surface temperature
of the following examples:
- Milk surface temperature in a baby’s bottle
- Surface temperature of a baby’s bath
- Ambient temperature
Accurate and reliable
The unique probe assembly construction incorporates an advanced infrared
sensor, ensuring that each measurement is accurate and reliable.
Gentle and easy to use
- The ergonomic design enables simple and easy use of the thermometer.
- This thermometer can even be used on a sleeping child without causing any interruption.
- This thermometer is quick, therefore child-friendly.
Multiple readings recall
Users will be able to recall the last 30 readings with a record of both time
and date when entering the recall mode, enabling efficient tracking of
temperature variations.
Safe and hygienic
- No direct skin contact.
- No risk of broken glass or mercury ingestion.
- Completely safe for use on children.
Fever alarm
10 short beeps and a red LCD backlight alert the patient that he/she may have
a temperature equal to or higher than 37.5 °C.
Bluetooth® Function
This device connects to the «Microlife Connected Health+» App by using
Bluetooth® 4.0 and enables easy monitoring of temperature.
Important Safety Instructions
-
Follow instructions for use. This document provides important product operation and safety information regarding this device.
Please read this document thoroughly before using the device and keep for future reference. -
This device may only be used for the purposes described in these instructions. The manufacturer cannot be held liable for damage caused by the incorrect application.
-
Never immerse this device in water or other liquids. For cleaning please follow the instructions in the «Cleaning and Disinfecting» section.
-
Do not use this device if you think it is damaged or notice anything unusual.
-
Never open this device.
-
A basic physiological effect called vasoconstriction can occur in the early stages of fever, resulting in a cool skin effect. The recorded temperature using this thermometer can, therefore, be unusually low.
-
If the measurement result is not consistent with the patient’s finding or unusually low, repeat the measurement every 15 minutes or double-check the result by another core body temperature measurement.
-
This device comprises sensitive components and must be treated with caution. Observe the storage and operating conditions described in the «Technical Specifications» section.
-
Ensure that children do not use this device unsupervised; some parts are small enough to be swallowed.
-
Do not use this device close to strong electromagnetic fields such as mobile telephones or radio installations. Keep a minimum distance of 3.3 m from such devices when using this device.
-
Protect it from:
– extreme temperatures
– impact and dropping
– contamination and dust
– direct sunlight
– heat and cold -
If the device is not going to be used for a prolonged period the batteries should be removed.
WARNING: The measurement results given by this device are not a diagnosis. It is not replacing the need for the consultation of a physician, especially if not match the patient’s symptoms. Do not rely on the measurement result only, always consider other potentially occurring symptoms and the patient’s feedback. Calling a doctor or an ambulance is advised if needed.
How this Thermometer measures Temperature
This thermometer measures infrared energy radiated from the forehead as well as objects. This energy is collected through the lens and converted to a temperature value.
Control Displays and Symbols
- All segments displayed 9: Press the ON/OFF button 5 to turn on the unit; all segments will be shown for 1 second.
- Ready for measuring AK: When the unit is ready for measuring, the «°C» or «°F» icon will keep flashing while the mode icon (body or object) will be displayed.
- Measurement complete AL: The reading will be shown on the display 3 with the «°C» or «°F» icon and the mode icon steady. The unit is ready for the next measurement as soon as the «°C» or «°F» icon is flashing again.
- Low battery indicator AO: When the unit is turned on, the « » icon will keep flashing to remind the user to replace the batteries.
Setting Date, Time and Beeper Functions
Setting the date and time
- After the new batteries are fitted, the year number flashes in the display BP. You can set the year by pressing the START button
- To confirm and then set the month, press the M-button 8.
- Press the START button 4 to set the month. Press the M-button 8 to confirm and then set the day.
- Follow the previously mentioned instructions to set the day, 12 or 24-hour mode, hours, and minutes.
- Once you have set the minutes and pressed the M-button 8 , the date and time are set and the time is displayed.
If no button is pressed for 20 seconds, the device automatically switches to ready for measuring AK.
Cancel time setup: Press the ON/OFF button 5 during time setup. The LCD will show Date/Time icons with «–:–». After that press the ON/OFF button 5 to start the measurement. If no further action is taken within 20 seconds, the device will automatically turn off.
Change current date and time: Press and hold the M button 8 for approx. 3 seconds until the year number starts to flash BP. Now you can enter the new values as described above.
Setting the beeper
- When the device is switched off, press and hold the ON/OFF button 5 for 5 seconds to set the beeper BQ.
- Press the ON/OFF button 5 again to either turn the beeper on or off. The beeper is activated when the beeper icon BQ is not crossed-out.
If no button is pressed for 5 seconds, the device automatically switches to ready for measuring 11.
Changing between Body and Object Mode
For changing from the body to object mode, slide the mode switch 6 at the side of the thermometer downwards. For switching back to body mode, slide the switch up again.
Directions for Use
Measuring in body mode
- Press the ON/OFF button 5. Display 3 is activated to show all segments for 1 second.
- When the «°C» or «°F» icon is flashing, a beep sound is heard and the thermometer is ready for measuring AK.
- Aim the thermometer at the center of the forehead with a distance of no more than 5 cm. Please remove any hair, sweat or dirt from the forehead before measuring to improve the accuracy of the readings.
- Press the START button 4 and ensure that the activated blue tracking light is aimed at the center of the forehead. After 3 seconds a long beep will verify the completion of measurement.
- Read the recorded temperature from the LCD display.
Measuring in object mode
- Follow steps 1-2 above, then aim the thermometer at the center of the object you want to measure with a distance of no more than 5 cm. Press the START button 4. After 3 seconds a long beep will verify the completion of measurement.
- Read the recorded temperature from the LCD display.
NOTE:
-
Patients and thermometers should stay in similar room conditions for at least 30 minutes.
-
Don’t take a measurement while or immediately after nursing a baby.
-
Don’t use the thermometer in high humidity environments.
-
Patients should not drink, eat or exercise before/while taking the measurement.
-
Don’t move the measurement device from the measuring area before hearing the termination beep.
-
10 short beeps and a red LCD backlight alert the patient that he/she may have a temperature equal to or higher than 37.5 °C.
-
Always take the temperature in the same location, since temperature readings may vary according to location.
-
Doctors recommend rectal measurement for newborn infants within the first 6 months, as all other measuring methods might lead to ambiguous results. If using a noncontact thermometer on those infants, we always recommend verifying the readings with a rectal measurement.
-
In the following situations, it is recommended that three temperatures are taken with the highest one taken as the reading:
1. Children under three years of age with a compromised immune system and for whom the presence or absence of fever is critical.
2. When the user is learning how to use the thermometer for the first time until he/she has familiarized himself/herself with the device and obtains consistent readings.
3. If the measurement is surprisingly low. -
Readings from different measuring sites should not be compared as the normal body temperature varies by measuring site and time of day , being highest in the evening and lowest about one hour before waking up.
Normal body temperature ranges:
• Axillar: 34.7 – 37.3 °C / 94.5 – 99.1 °F
• Oral: 35.5 – 37.5 °C / 95.9 – 99.5 °F
• Rectal: 36.6 – 38.0 °C / 97.9 – 100.4 °F
• Microlife NC 150 BT: 35.4 – 37.4 °C / 95.7 – 99.3 °F
Changing between Celsius and Fahrenheit
This thermometer can display temperature readings in either Fahrenheit or Celsius. To switch the display between °C and °F, simply turn OFF the unit, press and hold the START button 4 for 5 seconds; after 5 seconds, the current measurement scale ( «°C» or «°F» icon) will flash on the display AP. Change the measurement scale between °C and °F by pressing the START button 4. When the measurement scale has been chosen, wait for 5 seconds and the unit will automatically enter the «ready for measuring» mode.
How to recall 30 readings in Memory Mode
This thermometer can recall the last 30 readings with a record of both time and date.
- Recall mode AQ: Press the M-button 8 to enter recall mode when the power is off. The memory icon «M» will flash.
- Reading 1 – the last reading AR: Press and release the M button 8 to recall the last reading. Number «1» and a flashing «M» are displayed.
- Reading 30 – readings in succession: Press and release the M-button 8 consecutively to recall the last 30 readings in succession.
Pressing and releasing the M-button 8 after the last 30 readings have been recalled will resume the above sequence from reading 1.
Bluetooth® Function
This device can be used in conjunction with a smartphone running the «Microlife Connected Health+» App. The measurement results will be automatically transferred via Bluetooth®
Downloading the «Microlife Connected Health+» App
Download the «Microlife Connected Health+» App for free from Google Play™
(Android) or App Store (iOS) and install it on your smartphone.
How the Bluetooth® Function works
The Bluetooth® function on your device will automatically turn on and will be
ready to connect with the «Microlife Connected Health+» App after the device
is switched on. Your device will automatically upload the data once it is
connected to the smartphone.
Bluetooth® Icon Indicator
The Bluetooth® icon indicator on your device, located in the middle-left
portion of the display, is designed to provide information about the
connection between your device and smartphone.
NOTE:
- The Bluetooth®function is enabled: The Bluetooth® icon indicator 11 appears steady on the display.
- The Bluetooth® icon indicator AK flashes when connecting with a device or uploading data to the device.
- The Bluetooth® function is automatically switched off during measurement; after measurement, the device will switch it on and upload the data to the «Microlife Connected Health+» App.
App Tutorial
To access the tutorial, select «Tutorial» in the menu of the «Microlife
Connected Health+» App (located in the upper left corner on the screen).
Error Messages
- Measured temperature too high 19: Displays « H » when the measured temperature is higher than 43 °C /4 °F in body mode or 99.9 °C / 211.8 °F in object mode.
- Measured temperature too low 20: Displays «L» when the measured temperature is lower than 34 °C /2 °F in body mode or 0.1 °C / 32.2 °F in object mode.
- Ambient temperature too high 21 : Displays «H» and when the ambient temperature is higher than 40.0 °C /0 °F.
- Ambient temperature too low g): Displays «L» and when the ambient temperature is lower than 15.0 °C /0 °F in body mode or lower than 5.0 °C / 41.0 °F in object mode.
- Error function display 0): The system has a malfunction.
- Blank display 24: Check if the batteries have been inserted Also check the polarity ( < +> and < ->) of the batteries.
- Flat battery indicator ®: If only the « » icon is shown on the display, the batteries should be replaced immediately.
Cleaning and Disinfecting
Use an alcohol swab or cotton tissue moistened with alcohol (70% Isopropyl) to clean the thermometer casing and the measuring sensor. Ensure that no liquid enters the interior of the device. Never use abrasive cleaning agents, thinners, or benzene for cleaning, and never immerse the device in water or other cleaning liquids. Take care not to scratch the surface of the sensor lens and the display.
Battery Replacement
This device is supplied with 2 new, long-life 1.5V, size AAA batteries.
Batteries need replacing when this icon « » 25 is the only symbol shown on
the display. Remove the battery cover 28 by sliding it in the direction
shown.
Replace the batteries – ensure correct polarity as shown by the symbols in the
compartment.
Batteries and electronic devices must be disposed of in accordance with the
locally applicable regulations, not with domestic waste.
Guarantee
This device is covered by a 5-year guarantee from the date of purchase. During
this guarantee period, at our discretion, Microlife will repair or replace the
defective product free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
- Transport costs and risks of transport.
- Damage caused by incorrect application or non-compliance with the instructions for use.
- Damage caused by leaking batteries.
- Damage caused by accident or misuse.
- Packaging/storage material and instructions for use.
- Regular checks and maintenance (calibration).
- Accessories and wearing parts: Batteries, Probe cover (optional).
Should guarantee service to be required, please contact the dealer from where the product was purchased or your local Microlife service. You may contact your local Microlife service through our website: www.microlife.com/support Compensation is limited to the value of the product. The guarantee will be granted if the complete product is returned with the original invoice. Repair or replacement within the guarantee does not prolong or renew the guarantee period. The legal claims and rights of consumers are not limited by this guarantee.
Technical Specifications
Type: | Non Contact Thermometer NC 150 BT |
---|---|
Measurement | Body mode: 34.0 – 43.0 °C / 93.2 – 109.4 °F |
range: | Object mode: 0.1 – 99.9 °C / 32.2 – 211.8 °F |
Resolution: | 0.1 °C / °F |
Measurement | Body mode: |
accuracy
(Laboratory):| ±0.2 °C, 35.0 ± 42.0 °C / ±0.4 °F, 95.0~107.6 °F
±0.3 °C, 34.0 ± 34.9 °C and 42.1~43.0 °C /
±0.5 °F, 93.2 ~ 94.8 °F and 107.8 ~ 109.4 °F
Object mode:
±1.0 °C, 0.1 ~ 99.9 °C / ±2 °F, 32.2 ~ 211.8 °F
Display:| Liquid Crystal Display, 4 digits plus special icons
Acoustic:| The unit is turned ON and ready for the measurement: 1 short
beep. Complete the measurement: 1 long beep (1 sec.) if the reading is less
than 37.5 °C / 99.5 °F, 10 short «beep» sounds if the reading is equal to or
greater than 37.5 °C / 99.5 °F. System error or malfunction: 3 short «bi»
sounds.
Memory:| 30 readings recall in the memory mode with a record of both
time and date.
Backlight:| The display light will be GREEN for 1 second when the unit
is turned ON. The display light will be GREEN for 5 seconds when a
measurement is completed with reading less than 37.5 °C / 99.5 °F. The display
light will be RED for 5 seconds when a measurement is completed with a reading
equal to or higher than 37.5 °C / 99.5 °F.
Operating| Body mode: 15 – 40.0 °C / 59 – 104.0 °F
conditions:| Object mode: 5 – 40.0 °C / 41 – 104.0 °F
15 – 95 % relative maximum humidity
Storage| -25 – +55 °C / -13 – +131 °F
conditions:| 15 – 95 % relative maximum humidity
Automatic
Switch-off:| Approx. 1 minute after the last measurement has been taken.
Communication:| Bluetooth® Low Energy 4.0
Compatibility:| iOS: iOS 9.0 or newer
Android: Android OS 5 or newer
Battery:| 2 x 1.5 V alkaline batteries; size AAA approx. 1500
measurements
Battery lifetime:| approx. 1500 measurements (using new batteries)
Dimensions:| 141.1 x 43.3 x 36.9 mm
Weight:| 90 g (with batteries), 67 g (w/o batteries)
IP Class:| IP21
Reference to| ASTM E1965; IEC 60601-1; IEC 60601-1-2 (EMC);
standards:
Expected| IEC 60601-1-11
service life:| 5 years or 12000 measurements
This device complies with the requirements of the Medical Device Directive
93/42/EEC.
Technical alterations reserved.
According to the Medical Product User Act, a biennial technical inspection is
recommended for professional users. Please observe the applicable disposal
regulations.
The Bluetooth® word mark and logos are registered trademarks owned by
Bluetooth SIG, Inc. and any use of such marks by Microlife Corp. is under
license. Other trademarks and trade names are those of their respective
owners.
www.microlife.com
Detailed user information about our thermometers and blood pressure monitors as well as services can be found at www.microlife.com.
Guarantee Card
Name of Purchaser……………….
Serial Number………………………
Date of Purchase…………………
Specialist Dealer………………….
Microlife AG
Espenstrasse 139
9443 Widnau / Switzerland
www.microlife.com
EC REP
Microlife UAB
P. Lukšio g. 32
08222 Vilnius / Lithuania