healthsmart 18-210-000 Compact Infrared Ear Thermometer Instruction Manual

June 6, 2024
healthsmart

healthsmart International Logo

healthsmart 18-210-000 Compact Infrared Ear Thermometer

healthsmart 18-210-000 Compact Infrared Ear
Thermometer

Introduction

  • Remove safety tab from battery. Open battery case and gently pull out the plastic safety tab from underneath the battery. Thermometer will not activate until this has been performed.
  • For safe and proper use of this product, be sure to read and fully understand Safety Precautions contained in this instruction manual.
  • Keep this Manual with you for a convenient reference.
  • The reference body site is core and mode of operation is adjusted mode.
  • The user can measure and change the battery under normal circumstances and maintain the device and its accessories according to the user manual.

Intended for Use

The device is intended for the intermittent measurement and monitoring of human body temperature in the ear by consumers in the home. This device is intended for use on people of all ages.

Safety Precautions

Proper technique is critical to obtaining accurate temperatures. Therefore, in order to obtain an accurate measurement, please read this manual carefully before using.

  • Use of this thermometer is not intended as a substitution for consultation with your physician. Measurement results are for reference only. Contact your physician if you have or suspect any health concerns.
  • The infrared sensor must be kept clean, dry and undamaged at all times to ensure accurate measurements.
  • Earwax and other debris in the ear canal may cause an inaccurate measure-ment. Ensure the ear canal is clean when taking a measurement.
  • Do not touch or blow on the infrared sensor. A dirty sensor may cause inaccuracy.
  • Avoid measuring the ear temperature if the ear is inflamed or inflected.
  • If the temperature of the storage area differs greatly from that of the measurement area, please wait for the thermometer temperature to equalize to the room temperature for about 30 minutes before use.
  • Keep the device out of the reach of children and pets to avoid inhalation or swallowing of small parts. Do not allow children to take their own tempera-tures unattended. Children may not be able to use the device according to the instructions in this user manual. This is not a toy.
  • Do not discard the battery into a fire. The battery is likely to explode.
  • Do not store the thermometer in extreme temperatures below -4°F (-20°C) or over 131°F (55°C) or in humidity below 15% RH or over 90% RH. Failing to do so may cause inaccuracy. Do not use the thermometer under temperature extremes (below 50°F / 10°C or over 104°F / 40°C) or humidity extremes (below 15%RH or over 85%RH). If you do so, this may cause inaccurate measurements.
  • Do not use the thermometer if the main body is damaged (for example the infrared sensor is broken) The continuous use of a damaged unit may result in false readings or injury.
  • Do not take the thermometer apart, attempt a repair or change any parts on the unit at any time.
  • Do not use a mobile phone nearby when the thermometer is working.
  • If your thermometer will not be used for a long time, it is advised to remove the battery to prevent possible damage due to chemical leakage. If the battery does leak, remove them carefully. Do not allow bare skin to touch the leaking fluid.
  • This product needs special precautions regarding EMC and needs to be in-stalled and put into service according to the EMC information provided. This unit can be affected by portable and mobile RF communications equipment.
  • The materials (e.g. ABS) of expect contact with patient had passed the ISO 10993-5 and ISO 10993-10 standards test, no toxicity, allergy and irritations reaction. However, based on current science and technology, other potential allergic reactions are unknown. If you have allergic reaction to the materials, please stop using immediately and consult your physician.
  • Store the thermometer in a dry, clean place. Do not leave the thermom-eter exposed to any chemical solvent, lint, dust direct sunshine or high temperatures.
  • When using, do not touch the battery and the patient simultaneously.
  • Do not maintain or service the device while the device is in use.
  • The patient may be the proper operator of the thermometer.

CAUTION: This device should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be used.

Recommendations

  • Don’t use the product for other purposes than what it is intended to be used. Only designed for use on the human body.
  • It is not a waterproof device. Clean it with a soft, wet cloth or cotton swab.
  • Do not leave the product exposed to any chemical solvent, direct sunshine or high temperature.
  • Don’t drop or impose any vibration or impact on the product.
  • Don’t use a mobile phone nearby when the thermometer is working.
  • Please act according to the native law in handling and disposing of the batteries.
  • Take out the batteries if you are not going to use the unit for a long period of time.

Body Temperature Chart

A person’s normal temperature will vary and trends to decrease with age. The following table shows normal temperature ranges by age:

Normal temperature ranges by age:

Age

| Temperature ( °F)| Temperature ( °C)
---|---|---
0 – 2 years| 97.5 – 100.4°F|

36.4 – 38.0°C

3 – 10 years

| 97.0 – 100.0°F| 36.1 – 37.8°C
11 – 65 years| 96.6 – 99.7°F|

35.9 – 37.6°C

65 years

| 96.4 – 99.5°F|

35.8 – 37.5°C

The range of normal temperature varies from person to person and can be influenced by many factors such as time of day, measure from different sites, level of activity, medications, emotion, etc. We recommend that you practice with the thermometer when you or family members are healthy, so you are familiar with the readings when they are ill.

Product Overview

Product Overview

  1. Infrared Sensor
  2. Battery Cover
  3. Measurement Button
  4. Probe Cap
  5. LCD Display
  6. Memory Button (MEMORY)
  7. Fever Indicator
  8. Settings Button (SET)

LCD Display

LCD Display

  1. Memory mode symbol
  2. Ear temperature symbol
  3. Temperature readout
  4. Low battery symbol
  5. Sleep mode symbol
  6. Fahrenheit degree symbol
  7. Celsius degree symbol
  8. Ambient temperature symbol
  9. Date display symbol
  10. Year Display symbol
  11. Display date and time in standby mode
  12. PM display
  13. Setup Mode

This package contains: (1) Thermometer, (2) 1.5v AAA Batteries, (1) User Manual

Temperature Taking Hints

  1. A temperature measurement taken from the right ear may differ from measurement taken from the left ear. For this reason, always try to take the temperature in the same ear.

  2. External factors may influence ear temperature, particularly when an individual has:

    • been lying on one ear or the other
    • had their ears covered
    • been exposed to very hot or very cold temperatures
    • been recently swimming or bathing
      In any of these cases listed above, remove the person from the situation or uncover their head and wait 30 minutes prior to taking a temperature.
  3. For people wearing hearing aids or ear plugs, remove the device and wait 30 minutes prior to taking a temperature.

  4. Use the untreated ear if ear drops or other medications have been placed in the ear canal.

Before Use

  1. To achieve an accurate measurement, it is important to check the infrared sensor and ensure it is clean before use. To clean the sensor, gently wipe its surface with a cotton swab slightly moistened with alcohol and immediately wipe dry with a clean cotton swab. After cleaning, allow at least 20 minutes drying time before taking temperatures.
  2. Earwax in the ear canal may cause an inaccurate measurement, so please ensure the ear canal is clean. If you clean the ear canal, wait 5~10 minutes before measuring.
  3. If the temperature of the storage area differs greatly from that of the measuring area, please wait 30 minutes for the thermometer and room temperature to equalize before measuring.
  4. Open the battery cover and remove the insulating plastic from the battery compartment. The thermometer should turn on automatically.

CAUTION: In Standby Mode, the LCD display will show the environment temperature along with the current date and time.

Switching Between Fahrenheit and Celsius

This thermometer is set to measure in Fahrenheit, but can be switched to measure in Celsius.

To Switch from Fahrenheit to Celsius
  1. Press the SCAN button to turn the thermometer ON. It will now be in Standby mode measuring ambient temperature.
  2. Press and hold the M (Memory) and S (Set) buttons for 2 seconds until the screen changes from °F to °C. Once the thermometer beeps, the unit is ready for measurement.

NOTE: To switch the mode again, wait until the unit beeps and then press and hold the SET button again until the mode switches.

Stored readings in memory will automatically be converted to the new temperature scale (°F to °C) that is selected.

How To Take A Reading

The ear is an ideal measurement site because body temperature is regulated by the hypothalamus, which shares the same blood supply as the tympanic membrane or more commonly called eardrum.

How To Take A Reading 01

NOTE: Make sure the measurement sensor is clean before taking a reading. Attempting to take a measurement anywhere else on the body may result in an inaccurate reading.

  1. Remove the probe cap before using and make sure the probe is clean before using.
    How To Take A Reading 02

  2. If the thermometer is OFF, press any button to turn it ON. The screen should display the ambient temperature.
    How To Take A Reading 03

  3. Insert the probe VERY gently and slowing into the ear canal.

    • Hold the thermometer so the probe faces straight in the direction of the eardrum.
      How To Take A Reading 04

    • It is important to point the probe tip toward the eardrum and is inserted inside the ear canal until slight resistance is felt.

    • Hold the outer ear and gently pull it up and toward the back of the head to straighten the ear canal.
      How To Take A Reading 05
      How To Take A Reading 06

    • Always insert the thermometer into the same ear with the same direction and depth.

  4. Press and hold the SCAN button. The thermometer take a reading and make a ‘beep’ sound, indicating the measure-ment is complete.

  5. Remove the thermometer from the ear canal. The thermometer will display the result on the backlit LCD screen.
    NOTE: After each measurement “000” will display above the measurement. Once the “000” disappears (approx. 6 seconds), a new measurement can be taken.
    How To Take A Reading 07

  6. After the test is complete, press the M (Memory) button or wait 1 minute for the display to return to standby mode and display the ambient temperature. After 1 minute in standby mode, the thermometer will automatically shut OFF.

NOTE: The LED light between the Memory and Settings button will illuminate according to the temperature reading.

  • GREEN = 93.2°F ~ 100.3°F (34.0°C ~ 37.9°C)
  • RED = 100.4°F ~ 109.3°F (38.0°C ~ 42.9°C)
  • LO = < 89.6°F (32.0°C)
  • HI = > 109.3°F (42.9°C)

Memory Recall

There are a total of 9 memory readings.
The current measurement is always stored in the last position (number 9).

To recall measurements in memory:

  1. If the thermometer is OFF, press any button to turn it ON.

  2. Press the M (Memory) button. It will show the position of the most current measurement. Followed by the actual temperature measurement along with the time and date.
    Memory Recall 01
    NOTE: All subsequent measurements taken will fill the memory spaces until they reach the number 1 spot. Once all of the spaces are filled, the oldest measurement will automati-cally be deleted from memory.

  3. Continue to press the MEMORY button to scroll the stored measurements.

  4. Press the SET and MEMORY button at the same time to return to the Standby mode.

Setting the Date and Time

  1. Press the SET button to enter the setting mode. In the settings mode, you can set the following items: Time Format (12hr or 24hr), Time, Date and Sleep Mode.
  2. Press the MEMORY button again to scroll through the options. Press the SET button to select the desired option.
    Note that “PM” will appear on the display in 12hr mode.

NOTE: When the Sleep Mode is activated (display shows ZZ), the

thermometer will turn OFF automatically when not in use. When the sleep mode is deactivated, the thermometer will remain ON. To conserve battery life, we recommend keeping the Sleep Mode turned ON.
Setting the Date and TimeReplacing the Battery

When the low battery indicator appears on the screen, or the thermometer does not turn ON, it is time to replace the batteries.

To Replace the Batteries
  1. Gently slide the battery cover back.

  2. Carefully remove the old batteries and properly discard.

  3. Insert 2 new AAA 1.5v batteries. Ensure they are inserted correctly according to the proper polarity.
    Replacing the Battery

  4. Slide the battery cover back on until it snaps into place.

To protect the environment, please dispose of exhausted batteries according to current federal, state and local regulations. Keep batteries out of the reach of children.

FCC Statement and EMC Compatibility

This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

  • Reorient or relocate the receiving antenna.
  • Increase the separation between the equipment and receiver.
  • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
  • Consult the dealer or an experienced radio/TV technician for help.
FCC Statement and EMC Compatibility

This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Changes or modifications to the product not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.

Maintenance, Storage and Calibration

Maintenance
  1. To ensure accurate measurements, keep the infrared sensor tip clean and free of scratches. Finger prints, ear wax or dirt will affect the accuracy of the thermometer.

  2. To clean the infrared sensor gently wipe the sensor with a cotton swab slightly moistened with alcohol and immediately wipe dry with a clean cotton swab. The thermometer can be used after the alcohol dries.
    Maintenance 01

  3. Use a soft dry cloth to clean the body of the thermometer.
    Maintenance 02

  4. Never clean the thermometer with an abrasive cleanser, thinner, benzene or submerge the thermometer into water or other liquids.

Storage
  1. After each use clean the thermometer.
  2. Store in a clean dry place at room temperature. NEVER leave the thermometer exposed to direct sun, high temperatures, humidity, fire, flames, shocks or impacts.
  3. If you do not use the thermometer for a long time, take out the batteries.
Calibration
  1. The thermometer is factory-calibrated. If the thermometer is used according to the user manual, periodic calibration is not required. Do not try to modify or reassemble the thermometer.

Troubleshooting

Problem

| Check lists|

Action

---|---|---
No response/ Automatic reset when the plastic battery insu- lator tab is pulled out| Battery used up?| Change new battery
Battery in wrong polarity?| Take out battery, reinsert battery correctly
Poor battery contact
Troubleshooting 01| The measured temperatures is lower than 89.6°F (32.0°C) or higher than 109.3°F (42.9°C). Check the method of operation| Follow the instructions for use.
Troubleshooting 02| Indicates a hardware problem| Contact your distributor
Troubleshooting 03| Operating temperatures is out of range| Use the thermometer within the range of operating temperatures
The sensor temperature has not been stabilized| Wait 10 seconds and take the measurement again
The measurement is not accurate or if there is any doubt on the measured result| Please check if the infrared sensor is clean| Clean the infrared sensor with a cotton swab according to the instructions.
Please check if the measurement method is correct and in the correct mode| Ensure you have fully read the user instructions and know how to properly use the thermometer.
The measurement is not accurate or if there is any doubt on the measured result| Please check if you have let the thermometer and patient temperature get stabilized

in the room for at least 30 minutes

| Please keep the thermometer and patient in the measuring room at least 30 minutes before using
Are you using the thermometer indoor?| Please take the measurement indoor
Please check if you held the thermometer in your hand too long and affect the accuracy| Put the thermometer on the table in the room where the measurement is taking place and let it cool down first
Troubleshooting 04| Please check if there is low battery symbol on LCD| Change new battery

This infrared thermometer meets requirements established in ASTM Standard E 1965-98 for the thermometer system.
ASTM laboratory accuracy requirements for the thermometer only in the display range of 98°F to 102.2°F (37°C to 39°C) for infrared thermometers is ± 0.4°F(±0.2°C), whereas for mercury-in-glass and electronic thermometers, the requirement per ASTM standard E 667-86 and E 1112-86 is ±0.2°F (±0.1°C)

Specifications

  • Product Name: Compact Infrared Ear Thermometer
  • Model: 18-210-000 / TH1009C
  • Power supply: DC 3V (2 x AAA battery)
  • Ear Measuring Range: 89.6°F –109.3°F (32.0°C – 42.9°C)
  • Laboratory Measuring Accuracy: ± 0.4°F (or ± 0.2°C) from 95°F –107.6°F (or 35°C – 42.0°C) ± 0.5°F ( ± 0.3°C) for other temperature range
  • Ambient Temperature Range: 41.0°F –139.8°F (5.0°C – 59.9°C)
  • Ambient Temperature Accuracy: ± 2°F (± 1°C)
  • Measurement Time: 1 Second
  • Resolution: 0.1°F (0.1°F)
  • Dimension: 5.18 (L) x 1.48 (W) x 1.46 (D) in. [131.8(L) x 37.7(W) x 37.3(D) mm]
  • Weight: 53.54g (without batteries)
  • Operation Temperature: 50.0°F – 104.0°F (10°C – 40°C)
  • Operation Relative Humidity: 15% RH – 85%
  • Storage & Transportation Temperature: -4°F – 131°F (-20°C – 55°C)
  • Storage & Transportation Relative Humidity: 15% RH – 90% RH
  • Service Life: 2 years
  • Battery Life: Approximately 6 months with 5 tests per day

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

healthsmart User Manuals

Related Manuals