FS FOPM-201 Fiber Tester Optical Power Meter User Guide

June 6, 2024
FS

Fiber Testers
OPTICAL POWER METER
Quick Start Guide
V1.0

Introduction

The handheld optical power meter is a fiber optic tester with compact size and friendly operation interface. It has a wide range of power measurement and high accuracy. When combine with a light source, it offers a quick and accurate testing solution. The figures below display the product images of the six optical power meters.

Accessories

FOPM-201/FOPM-202FS FOPM-201 Fiber Tester Optical Power - FOPM-
201OPM-202

FOPM-203/FOPM-204

FS FOPM-201 Fiber Tester Optical Power - FOPM

FOPM-205

FS FOPM-201 Fiber Tester Optical Power - FOPMs FOPM-10

Installing

Inserting FC CablesFS FOPM-201 Fiber Tester Optical Power -
Installing

Install FC fiber cable.

Inserting ST Cables

1. Install ST connector.

2. Install ST fiber cable.

Inserting SC Cables

1. Install SC connector.

2. Install SC fiber cable.

Inserting MTP Cables

FOPM-201/FOPM-202

Function Introductions

Button Description
Power/Backlight Button
>2s UNIT Wavelength/Unit Shift Button
>2s SET REF Setting Button

GLINT
| VFL Control Button

Operation Instructions

  1. Power On/Off and Auto-off Function
    Press button to turn on the instrument. Press it again for 2 seconds or more to turn it off.
    This power meter has a power-saving function. If 10 minutes without any operation, the instrument will automatically shut down. If you need to disable this function, only need to press the button for 2 seconds when you turn on the instrument till it displays “PERM”.

  2. Backlight Function
    When the instrument is powered on, short press button, you can control the backlight function on or off. The backlight function supports you to use the power meter at night or darker occasions.

  3. Wavelengths
    According to the project, we need to measure optical signals of different wavelengths.
    Then we need to select a corresponding wavelength to measure the optical power. If the wavelength needs to be measured does not match with the wavelength we select
    on the optical power meter, it will lead to the measuring values meaningless. Press button, the instrument will change wavelength, and display.
    FOPM-201/FOPM-202 optical power meter calibration measured wavelengths are: 850nm, 1300nm, 1310nm, 1490nm, 1550nm, 1625nm.

  4. Unit
    button can be used to change the display unit of the measurement data to meet the different requirement. When press this button for 2 seconds, the display will
    successively show the dBm value or mW/uW value. The numeric relationship between mW value and dBm value is: 10lg (mW) = (dBm).

  5. Reference
    button is used to set or check the reference value. Short press this button, the display will show “REF” and the dBm value saved as reference value. When long press for 2 seconds or more, the instrument will save the current dBm value as a new reference REF value. Meanwhile the “REF” sign will flash 3 times on the display. After that the display will show the dB value. Each wavelength can set its reference value.

  6. Visual Fault Locator Function (Optional)
    When power meter with visual fault locator, you can short press the button to control VFL status (on, glint, or off ).

FOPM-203/FOPM-204

Function Introductions

Button Description

2s PERM
| Power Button
| Wavelength Shift Button
| Unit Shift Button
2s SAVE
| Value Storage Button
2s SET
| REF Setting Button
| Backlight Control Button

Operation Instructions

  1. Power On/Off and Auto-off Function
    Press button briefly. The instrument powers on, and backlight switches on.
    Press button briefly again. The instrument powers off, and backlight switches off.

FS FOPM-201 Fiber Tester Optical Power - NOTE NOTE: Auto-off function

  1. The instrument will power off automatically if 10 minutes without any operation.
    While in the power-on state, press button for about 2 seconds to turn off the “Auto-off” function.

  2. Set Wavelength and Activate Auto-wavelength Recognition (TWIN)
    Press button repeatedly until the desired wavelength is displayed. You can select from six possible wavelengths: 850nm, 1300nm, 1310nm, 1490nm, 1550nm, 1625nm.  The instrument defaults to the wavelength which the user set in the last test. When used with the FOLS-203 and FOLS-204 optical light source, the wavelength will shift automatically according to the output wavelength of the light source. Long press button to activate the auto-wavelength recognition and the “TWIN” will show on the LCD. Short press button to close the “TWIN”.

  3. Switch Measurement Mode
    There are 3 measurement modes you can choose by pressing the button repeatedly, dB, dBm and mW.

  4. Set Reference Level
    (1) Press button to display the stored reference level for the current wavelength and a sign of “REF” will be displayed on the screen to indicate that it is a reference value. The displayed value only lasts 1 second.
    (2) Press and hold button over 2 seconds to store the presently measured value as the new reference level for the current wavelength. During the process, the “REF” sign flashes twice on the screen and buzzer sound is heard. Once the new reference level is set, the instrument switches to the dB measurement mode. The displayed value only lasts 1 second.

  5. Switch Backlight of LCD On/ Off
    Press button. Backlight switches on.
    Press button again. Backlight switches off.

  6. Frequency Detecting
    If the tested wavelength is carrying a tone of 270Hz, 1kHz, or 2kHz, the respective frequency indicates on the screen.

  7. Overflow of Measured Power Value
    If the measured power value is too high, the LCD screen will display “HI”.
    If the measured power value is too low, the LCD screen will display “LO”.

  8. Storage of Current Test Value
    Press button over 2 seconds, the “SAVE” flashes on the screen once with the sounds of the buzzer. It indicates the setting is finished. It will display the stored value and the serial number of the storage. Then, the power meter will return to the test state automatically.

  9. Check the Storage Records
    Press button, it will display the latest record.
    Press button, it will browse the forward records.
    Press button, it will browse the afterward records.
    Press and button, it will delete all the records.

FOPM-205

Button Description
Power Button
Test Mode Button
Storage Button
Menu Mode Shift Button
Unit Shift Button
REF Setting Button

Operation Instructions

  1. Power On/Off and Auto-off Function
    Press button to turn on the instrument. It will automatically go to test menu. In test menu, short press button (less than 2 seconds) to activate or deactivate the auto-off function. The auto-off function means the instrument will shut down automatically if the instrument has not been operated for a certain period of time. The time period can be set and the default is 10 minutes.

  2. Test Mode Switch
    In the test menu, press button to switch between normal test mode and P/F test mode.
    (1) Normal Test Mode
    Normal test mode means do not setup threshold value but display optical power directly. In this mode, the LED indicators will not light. The results are displayed in 2
    units: dB and dBm, using button to switch. See Fig. 1.
    Unit dBm is to display the actual power.
    Unit dB is to display a power value relative to a reference value. In this mode, reference value needs to be preset correctly.
    (2) P/F Test Mode
    P/F test mode means the measured light power comparing to a preset threshold value to determine if the measured light power meets user’s requirement or not. LCD will display the optical power and current state. The LED indicators below the LCD display will also change color to match the current state of measured light. This test mode is very useful in some special cases required. See Fig. 2.

Fig.2 P/F Test Mode Menu

3. Threshold Setup
The user can set up the value. The steps are as following: In the menu mode, select the second “Threshold Setup” enters to threshold setup menu. See Fig. 3.
In this menu, left side displays wavelength and right side displays threshold value. Top line indicates system information (date and time) and bottom line indicates the information of threshold including threshold symbol, threshold number and name. In this menu, short press Threshold button (less than 2 seconds) to switch the wavelength among 1310nm, 1490nm and 1550nm. Press ▲ ( ) button to view the previous record, ▼ ( ) button to view the next record, button to validate the current threshold value. After the above operation, all P/F mode test results will base on this threshold value. dBm/dB REF/SEL

FS FOPM-201 Fiber Tester Optical Power - NMenuFig.3 Threshold Setup Menu

FS FOPM-201 Fiber Tester Optical Power - NOTE NOTE: The threshold value can only be preset by PC software. See PC software section for detail. After finishing the setup, press button (less than 2 seconds) to exit the setup menu.

4. Time Setup
In the menu mode, select the forth “Time Setup” enters to time setup menu. See Fig. 4.

Fig.4 Time Setup Menu

In the time setup menu, the button becomes arrow button – using it to move cursor. When the cursor moves to a number, the user can use ▲ button to increase the value and ▼ button to decrease the value. When the cursor moves to “ Yes” and press button for more than 2 seconds, the “ Yes” will flash, means the instrument accepts the time change. During the time setup, the user can press button (less than 2 seconds) to exit time menu and back to the test menu, then the system time setup remains unchanged. SAVE REF/SEL MENU

5. Unit Switch
When the instrument is in normal test mode, press button can switch the unit between dBm and dB. Here, the unit dBm is the actual power and unit dB is a power value relative to reference value. Refer to the next section about “Reference Value Setup”. When the instrument is in the P/F test mode, pressing button will automatically exit P/F test mode and switch to normal test mode. Its unit will be in dB. dBm/dB dBm/dB.

6. Reference Value Setup
In the test menu, press button for more than 2 seconds, then the LCD will display “REF” in red (see Fig. 5). This means the instrument chooses the current light power as the reference value. The test results afterward are the values after comparing to this REF/SEL reference value. Now, the unit will be in dB and LED below the LCD display will not light.

Fig.5 REF Value Setup Menu

In the test menu, short press button (less than 2 seconds), then current reference value will appear on LCD display. See Fig. 6.

FS FOPM-201 Fiber Tester Optical Power - alue Menu

Fig.6 View REF Value Menu

Backlight Setup

In the menu mode, select the third “Backlit Setup” enters to backlight setup menu. See Fig. 7.

In this menu, press ▲ ( ) button to brighten the backlight and ▼ ( ) button to weaken the backlight. After completing the setup, press button (less than 2 seconds) to save the setting, exit and back to the menu mode. Press button twice to exit and back to the main mode.

8. Description of Memory Record Function
Description of Interface:
At the main interface, it shows the current recording number on the left bottom of the page.
Displaying: num: xxxx (xxxx means the current recording number, the maximum number up to 1000 pieces). As you may refer to the Fig. 8, it means there are 24 pieces of test recording inside.

FS FOPM-201 Fiber Tester Optical Power -Fig.8

Fig.8

Press button for longer than 3 seconds to save the current testing value, as you may refer to the Fig. 9, the recording number automatically adds 1 when the interface shows “SAVE”.

Fig.9

At the “record view” interface (refer to Fig. 10), the characters on top of the horizon line are”total number of recording” and “wavelength” (from left to right). The characters
under horizon line are “the number of recording” and the optical power value in accordance with the corresponding wavelengths.
In the menu mode, select the first “record view”, and press button to enter the “record view” interface, refer to Fig. 10 for this interface.

At the “record view” interface, you may see the main menu on the bottom of the page, they are (from left to right): ESC X RST
There is an underline when you selected the menu ( the “ESC” has been chosen in Fig.10), press “Threshold” shortly to move the underline from left to right , after the underline moved to your ideal menu, press shortly to execute it.

Fig.10

Description of every individual menu:
◄ : Page up to the last 10 recording
► : Page down to the next 10 recording
◄◄ : Page up to the last 100 recording
►► : Page down to the next 100 recording
ESC: Exit the “record view” interface ( same as press )
X: Select one or more than one recording to delete
RST: Reset the recording memory, this operation is not reversible and it’ll take 5 seconds to complete with power supply can not be disconnected, otherwise it may damage the
chip of tester.
At the “record view” interface, press button to move “X” upward one step, press button to move “X” downward one step. In Fig. 10, the interface explains that
dBm/dB there are totally 3 testing results inside, “low” means the saving power value is lower than the value of the threshold setting. In Fig. 11, “high” means the saving power value is higher than the value of the threshold setting. The digital means the power value of corresponding wavelength.

Fig.11

FOPM-107

Button Description
Power Button
Wavelength Shift and Up Button
Menu Mode Shift Button
Left Button
Confirmation Button
Right Button
Function Switch Shortcut Button
Unit Shift Button
REF Setting Button

Operation Instructions
1. MPO Mode
(1) Insert Loss Test
Press button to enter the main menu as shown in Fig. 1.
Press button to select mode and press to enter. Select MPO mode and press to confirm, and then select line order. After entering, select the type to be tested (such as A-Type).
Press button to select the Threshold setting. The buttons could adjust the threshold, as shown in Fig. 2.

(2) MPO Line Sequence Test
When the type of MPO jumper under test is unknown, press button to switch to the line sequence test function, and press button to test the line sequence. The obtained line sequence will be automatically modified in the MPO test. The test results are displayed in the area after Type. If communication is interrupted, it will prompt: link timeout!
FS FOPM-201 Fiber Tester Optical Power - NOTE NOTE: please confirm that works on MPO light source interface, not menu or other interface when testing line sequence.

2. OPM Mode
This mode has two main functions, multi-channel power meter (as shown in Fig. 3) and single-channel power (as shown in Fig. 4).
Multichannel optical power meter: it is equivalent to a 12-channel optical power meter, which supports the selection and release of a single channel or all channels. The wavelength of individual channels can be edited when selected, and the global operation of all channels can be carried out when all channels are selected.

Maintenance

(1) The interface is sensitive, please carefully plug in and pull out the adapter.
(2) Please cover the dust-proof cap when it is not in operation.
(3) Choose the correct fiber connector before testing.
(4) Take out the batteries when not in use.
(5) Please disconnect the AC adapter/charger once you finish using.
(6) Please keep all optical connectors and surface free from oil, dirt or other contamination to ensure proper operation.
(7) Keep regular cleanings on optical port of an optical power meter with cotton swabs.
(8) To ensure the measurement accuracy, please send the instrument to FS for calibration once a year.

Online Resources

Download: https://www.fs.com/download.html
Help Center: https://www.fs.com/service/help_center.html
Contact Us: https://www.fs.com/contact_us.html

Product Warranty

FS ensures our customers that any damage or faulty items due to our workmanship, we will offer a free return within 30 Days from the day you receive your goods.

Warranty: All Optical Power Meters enjoy 1 year limited warranty against defect in materials or workmanship. For more details about warranty, please check at https://www.fs.com/policies/warranty.html

Return: If you want to return item(s), information on how to return can be found at https://www.fs.com/policies/day_return_policy.html

Q.C. PASSED
Copyright © 2020 FS.COM All Rights Reserved.

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals