GOLDENSEA UV UVL72 Healthcare Facility Disinfection System User Manual

June 6, 2024
GOLDENSEA UV

GOLDENSEA - logo

GOLDENSEA UV UVL72 Healthcare Facility Disinfection
System

UVL72
USER MANUAL
KEEP THIS MANUAL FOR FUTURE NEEDS

Thank you for purchasing a GOLDEN SEA UV product. You have acquired a powerful and versatile device. We are confident that you will be satisfied with our excellent products and service. For your own safety, please read this user manual carefully before installing and operating the device.

Voltage Input: AC 100-240V~, 50/60Hz; Power: 80W; Product Size: 491x76x240mm
This UV lamp is only for commercial use in non-residential locations, such as in hotels, schools, factories, theaters, gymnasiums, restaurants, offices, etc., it shall only be used by professional people. The warning sign as below shall be shown on the way to the fixture during operation.

**IMPORTANT SAFEGUARDS

**

When using electrical equipment, basic safety precautions should always be followed including the following:

READ AND FOLLOW ALL SAFETY INSTRUCTIONS
a) This equipment is designed for use with germicidal UV radiation sources and UV radiation can pose a risk of personal injury. Overexposure can result in damage to the eyes and bare skin. To reduce the risk of overexposure, this equipment must be used by a professional person.
b) Maintenance and servicing of this UV generating equipment shall be performed by a professional person. Service personnel must wear appropriate Personal Protective Equipment (PPE) if the equipment will be in operation during the maintenance or servicing work. Contact the equipment manufacturer for PPE recommendations and guidance.
c) When replacing lamps, replace them only with the lamps for which the equipment is marked and intended.
d) The use of accessories not recommended by the manufacturer may cause an unsafe condition.
e) Do not use this equipment other than intended use.
f) People, pets, and plants should leave the areas when using the disinfection light.
g) It should be stored well to avoid accidental damage to the sterilizer and personal injury after the sterilization is finished.

Manual Operation (control panel)

  1. Close all the doors and windows.
  2. Plugin and power on press the power switch <I/O> to , the power indicator turns on green.
  3. Select time setting <30 mins>, <60 mins> or <120 mins> by the Time Setting touch key, the touch key indicator light turns on.
  4. Press the Start touch key, the audible alarm starts beeping and the red power indicator will flash regularly for 2 minutes. The UV lamp will turn on after the 2 minutes delay when disinfection time is set, the power indicator stays on red.
  5. Once the disinfection process ends, the power indicator switches to green, an audible alarm sounds. Open the doors and windows to ventilate the disinfected area.

GOLDENSEA UV UVL72 Healthcare Facility Disinfection System - control
panel

Warning:
Don’t look at the disinfection light directly. People, pets, and plants should leave the area when using the disinfection light. Ventilate for 30-40 minutes after work.

Remote Control Operation

  1.  Close all the doors and windows.
  2. Plugin and power on press the power switch <I/O> to , the power indicator turns on green.
  3. Select time setting <30’>, <60’> or <120’> by the Time Setting button on the remote control, the touch key indicator light on the device turns on.
  4. Press the Start button on the remote control, the audible alarm starts beeping and the red power indicator on the device will flash regularly for 2 minutes. The UV lamp will turn on after the 2 minutes delay when disinfection time is set, the power indicator stays on red.
  5. Once the disinfection process ends, the power indicator switches to green, an audible alarm sounds. Open the doors and windows to ventilate the disinfected area.

Note: if the remote control loses pairing with the device, re-do the pairing by pressing the Start button on the remote control and switching on the device for 5 seconds. One remote control can be matched with multiple units of the same model.

GOLDENSEA UV UVL72 Healthcare Facility Disinfection System -
Start

Motion Sensor Operation (if the lamp with Motion Sensor)

  1. Turn on the UV lamp, select the disinfection time and start, leave the room, after 2 minutes alarming (Di….Di…Di….), the UV lamps will switch on automatically, the motion sensor will be activated.
  2. The Motion sensor can ‘see’ through non-metal materials, such as wood board, plastic, fabrics, etc. But the concrete will influence the motion sensor, and the metal board even can stop the detecting ability. So, when the UV Lamps are working, it will turn off when it detects that someone or something moving around within its detecting range even outside the rooms, and it will turn on again when someone or something is left. (detecting range: 3m in an open area without any of the above influence factors)

Note: for safety reasons, the UV lamp does not work if the motion sensor is defective.

Notes

  1. Put the UV lamp at a height of 0cm-400cm to get the best disinfection effect. Please do not put the lamp higher than 400cm, as the Motion Sensor cannot detect perfectly.
  2. The surface of the lamp should be kept clean.
  3. When disinfecting the surface of an object with ultraviolet rays, the surface to be irradiated should be directly irradiated within 1 meter of ultraviolet rays, otherwise, it’s necessary to increase the irradiation time.
  4. The lamp is fragile. Please handle it with care.
  5. According to the latest guideline on the diagnosis and treatment released by the National Health Commission, the virus is sensitive to ultraviolet (UV)  light and heat, so UV radiation is a quick and effective way of disinfecting rooms against germs and viruses.

According to the guideline, indoor spaces should be disinfected with UV light with an intensity of over 1.5 watts per square meter. A UV lamp can disinfect objects within one meter for at least half an hour. Longer exposure to radiation is needed when the temperature indoors is below 20°C or above 45°C and relative humidity is over 60%. A room should be ventilated after UV disinfection, and people are suggested to enter the room half an hour later.

Suggestion For Disinfection Time

Area

| Power|

Disinfection Time

---|---|---
20m2| 80W| 30min
40m2| 80W| 60min
80m2| 80W| 120min

Warnings

The UV-C wavelength is 253.7nm, the radiation intensity in 200mm distance is 2501.2uw/cm².

WARNING – RISK OF PERSONAL INJURY

THIS EQUIPMENT IS DESIGNED FOR USE WITH GERMICIDAL LAMPS OR UV SOURCES AND MUST BE INSTALLED AND USED IN COMPLIANCE WITH COMPETENT TECHNICAL DIRECTIONS TO PREVENT THE USER’S EYE AND BARE SKIN FROM EXPOSURE TO HARMFUL UV OR OPTICAL RADIATION.PEOPLE, PETS, AND PLANTS SHOULD LEAVE THE AREAS WHEN USING THE DISINFECTION LIGHT.
DO NOT USE THIS EQUIPMENT IN
A DWELLING
EYE DAMAGE MAY RESULT FROM DIRECTLY VIEWING THE LIGHT PRODUCED BY THE LAMP IN THIS EQUIPMENT. ALWAYS  TURN OFF THE LAMP BEFORE SERVICING.

CAUTION: TO REDUCE THE RISK OF FIRE USE ONLY TUV PL-L, 36W,0.44A LAMPSATTENTION – RISQUE D’INCENDIE UTILISER UNIQUEMENT LES LAMPES _TUV PL-L, 36W,0.44A

Suitable for damp locations
CONVIENT AUX EMPLACEMENTS HUMIDES

UV-C RISK GROUP 3

WARNING: UV-C emitted from this product.
Avoid eye and skin exposure to unshielded products.
Follow installation instructions and user manual.

SAVE THESE INSTRUCTIONS

Product Overview

GOLDENSEA UV UVL72 Healthcare Facility Disinfection System - Product
Overview

  1. Handle 8) Power Switch
  2. Hanging Hook| 9) Power Cable
  3. 120 mins – touch key| 10) Lamp Tube Safety Catch
  4. 60 mins – touch key| 11) Protective Grid
  5. 30 mins – touch key| 12) UV Tube
  6. Start – touch key| 13) Grid Opening Tabs
  7. Power Indicator| 14) Indicator Lights

Dimensional Drawings

GOLDENSEA UV UVL72 Healthcare Facility Disinfection System - Dimensional
Drawings

Installation Instructions

9.1 Installation of Floor Bracket

GOLDENSEA UV UVL72 Healthcare Facility Disinfection System - Floor
Bracket

Steps:

  1. Take the device and the floor brackets out of the package.
  2. Align the screw holes of the floor brackets with the screw holes at the bottom of the device. ( Note: feet must face to the floor as shown).
  3. Fix the brackets to the device with the 4 screws (M4x10mm) provided.

9.2 Wall Mounting

GOLDENSEA UV UVL72 Healthcare Facility Disinfection System - Mount 2 screws
\(M4\) on the wall with a distance

Steps:

  1. Mount 2 screws (M4) on the wall with a distance of 422mm.
  2. Mount the device onto the 2 screws using the 2 hooks of the device (as shown).
  3. Attach a safety cable (not included) between the handle of the device and a fixing point on the wall.
    (ATTENTION! This step is essential to prevent the device from falling off by accident or misuse)

9.3 Installation of Reflector

GOLDENSEA UV UVL72 Healthcare Facility Disinfection System - locking
tabs

Steps:

  1. Pull up the 2 locking tabs (as shown) on the back of the device.
  2. Install the reflector onto the back of the device. Make sure the reflective (shiny) surface of the reflector is facing the lamp tubes. The rough side of the reflector should be facing away from the lamp tubes.
  3. Push the 2 locking tabs (as shown) down to lock the reflector into position.

9.4 Installation of Rigging Bracket (optional)

GOLDENSEA UV UVL72 Healthcare Facility Disinfection System - Rigging
Bracket

Steps:

  1. Fix the rigging brackets to the screw holes of both top and bottom of the device as shown with 8 screws (M4x10mm) provided.
  2. Fix the rigging brackets to the hanging rods on the ceiling as shown. (The hanging rods are not included)
  3. Attach a safety cable (not included) between the handle of the device and a fixing point on the ceiling or wall.
    (ATTENTION! This step is essential to prevent the device from falling off by accident or misuse).

9.5 Installation of Tripod Bracket (optional)

GOLDENSEA UV UVL72 Healthcare Facility Disinfection System - Tripod
Bracket

Steps:

  1. Align the screw holes of the tripod bracket with the screw holes at the bottom of the device.
  2. Fix the tripod bracket to the device with 2 screws (M4x10mm).
  3. Install your tripod to the tripod bracket with corresponding screws.

9.6 Installation of Rotating Bracket (optional)

GOLDENSEA UV UVL72 Healthcare Facility Disinfection System - Rotating
Bracket

Steps:

  1. Align the screw holes of the rotating bracket with the screw holes of both top and bottom of the device as shown.
  2. Fix the rotating bracket to the device with 4 screws (M4x10mm).

Replacing The UV Tube

To maintain the sterilizing efficacy, the UV tube needs to be replaced after 9,000 hours of use.

GOLDENSEA UV UVL72 Healthcare Facility Disinfection System -
replaced

Steps:

  1. Wait for the UV tube to cool down (to avoid scalding) after switching off the device.

  2. Pull up the locking tabs ① on both sides of the device, and open the protective grid as shown.

  3. Remove the safety catch ② of the UV tube, move the UV tube out of its holder from left to right ③ as shown, and take it out.
    Note: please follow your local toxic and hazardous waste regulations for lamp disposal.

  4. Insert the new UV tube into the holder of the device and move the UV tube towards the left until it fits into place correctly and tightly.
    Note: direct contact with the new UV tube should be avoided as grease from your skin may reduce the efficiency of the UV tube. It is advised to wear protective gloves when handling a new UV tube.

  5. Fix the safety catch ② back to the UV tube holder.

  6. Install the protective grid back onto the device and push down the locking tabs ① on both sides of the device. The device is now ready for use.

Troubleshooting Tips

11.1 What to do if the device doesn’t work?

  1. Ensure the device is plugged into a working wall socket.
  2. Ensure the plug is wired correctly and the terminal is well-connected.
  3. If you still have issues with the device, please contact your local dealer or our customer service for assistance.

11.2 How often should the UV tube be replaced?

  1. The UV tube is the core of the product, it should be replaced after 9000 hours of use to maintain the germicidal efficacy.
  2. The UV tube has defected if it flashes, please contact your local dealer or our customer service to order a new UV tube. For replacing the UV tube, please refer to the instructions in this user manual.
  3. After use for a period of time, it is normal to find some black spots on both sides of the UV tube.

11.3 The Audio alarm signal and LED indicator signal for this product:

Status

| Audio Alarm|

Power Indicator

---|---|---
Standby| –| Green indicator on
Delayed Start| Beeping for 2 minutes regularly (frequency increase in the countdown 10 seconds)| Red indicator flashing for 2 minutes regularly (frequency increase in the countdown 10 seconds)
Lamp Strike| UV Lamp on after 15 seconds when the beeping stops| Red indicator on after beeping stops
Emitting (working)| Single beep every 10 seconds| Red indicator on
Motion Detected| Dual beeps every 10 seconds| Red & Green indicators flashing
End of Cycle| Triple beeps every 10 seconds (in 30 seconds)| Green indicator on

Features

Light Source
Lamp:
TUV PL-L 36W/4P 1CT/25 (PHILIPS)
or
HNS L 36W 2G11 GFT36DL/2G11/SE/OF PURITEC UV-C Germicidal Low Pressure
Lamp (OSRAM)
Extremely long Life: up to 9,000 H

Features

  • Delayed start (2-min delayed start)
  • Motion sensor
  • Audio alarm
  • Visual indicator lights
  • 3 disinfection time settings
  • Power switch
  • Control panel with touch keys
  • Through wall remote control

Product Type
Industrial Ultraviolet disinfection light

Weight
Net weight: 2.2 kg

Accessories List:

UV Tube| | Included
---|---|---
Remote Control (battery type: CR2025)| | Included
UV Glasses (protective  eyewear)| | Included
Reflector| GOLDENSEA UV UVL72 Healthcare Facility Disinfection System -
Reflector| Included
Floor Bracket| | Included
Warning Sign| | Included
Tripod Bracket| | Included
Safety Cable| | Optional
Rigging Bracket| | Optional
Rotating Bracket| | Optional
UV Irradiance Meter| | Optional

Cleaning and Maintenance

  1. To ensure the device remains in good condition and does not fail prematurely, regular maintenance, cleaning and checks are required.
  2. For cleaning the device, please use a lint-free cloth.
  3. To avoid damage to the device, never use detergent or solvents.
  4. Never spray insecticide towards the device, it can discolor or damage.
  5. Should you need any spare parts, please order genuine parts from your local dealer.

Service

For all service needs, please contact your local authorized dealer or GOLDEN SEA customer service. Our contact details are:
Email: support@goldenseauv.com
Tel: +86 20 3996 6388
Address: No.109 Haiyong Road, Shiji Town, Panyu Zone, Guangzhou, China. 511450

Errors and omissions for all information given in this user manual are expected.
All information is subject to change without prior notice.