THORLABS HVA200 High Voltage Amplifier Instruction Manual

August 2, 2024
THORLABS

THORLABS HVA200 High Voltage Amplifier Instruction Manual

Important Safety Notice

Danger High Voltage
The HVA200 can produce hazardous voltages and currents which may be harmful or even lethal. Use caution and exercise preventative safety measures to prevent contact between these high voltages and any personnel.

Warning!
The line switch and fuse must be set to the correct mains voltage. The unit is shipped ready to operate on 230 V. To operate on 115 V, the line switch and fuse must be changed. See Section 5.1: Fuse Replacement.

Do Not Open Housing!
The HVA200 has no user-serviceable parts. Service should only be performed by trained service personnel.

Product Overview

The Thorlabs HVA200 High Voltage Amplifier is designed to directly drive the Thorlabs Electro-Optic Modulators. The amplifier features: a large, ±200 V output, a continuous current output of 100 mA, a wide, 1 MHz bandwidth, and low noise. The voltage gain of – 20 boosts the input up to the high voltages needed to drive our lithium niobate broadband modulators. An adjustable bias allows for precise DC offset control.

The HVA200 uses a high voltage, wideband, high slew rate output amplifier to achieve an output range of ±200 V at a bandwidth up to 1 MHz. The input amplifier includes a summing junction which allows an adjustable DC bias to be added to the input modulation. This composite signal is then boosted by a fixed voltage gain of 20 by the output amplifier. For added safety, a front panel HV Enable button must be pressed to connect the HV output to the output BNC. The output is automatically disabled each time the HVA200 is powered on.

The DC Bias control consists of a rotary encoder which allows precise control and repeatability. The bias adjustment is typically used to shift the DC level of the output as needed by the application.

A voltage monitor output is provided to allow real-time monitoring of the high voltage output. The monitor has a scaling of 20:1 so that an output of 200 V results in a 10 V monitor voltage.

A USB port is provided to allow remote operation through a terminal interface.

Setup and Operation

Warning
Before plugging the amplifier into an AC outlet, check that the line switch voltage matches your AC outlet. The amplifier is configured from the factory to operate on 230 V by default. The fuse and line switch will need to be changed for 115 V operation. See section 5.1 on page 12 for more information.

To setup the unit, refer to Figures 1 and 2, and the legend table on the preceding page and perform the following steps:

  1. See the warning above. Attach the supplied AC power cord to the AC connector on the rear panel and plug into a suitable AC outlet.

  2. Connect an EO-modulator to the HV Output connector (6) on the front panel.
    HV Output Warning
    The HV output is capable of producing hazardous voltages and currents. The HV output will handle an accidental short circuit without damage, but the output should not be shorted continuously. The current is internally limited to 100 mA.

  3. (Optional) Connect the HV monitor (4) to an oscilloscope or volt meter. The input impedance must be at least 50Ω; however, at this value the ratio will be 40:1. The true HV Output to monitor ratio can be calculated with the following equation:

HV Output Monitor Warning
The HV Output Monitor requires a minimum load of 50 Ω. The port can handle an accidental short circuit without damage but must not be shorted for more than five seconds to prevent excessive heating of the output resistor. The output will source up to 200 mA when a short circuit is applied.

  1. If a modulating signal is to be used, connect it to the Modulation Input connector (2) on the front panel. This signal will be amplified by a fixed voltage gain of – 20.
  2. Turn the power switch (1) on. At this point the amplifier is powered up but the HV Output is disabled. The power indicator should be illuminated. If not, check the AC fuse (see page 12). Confirm that the HV Enable LED (5) is off.
  3. After confirming that all connections are correct, the amplifier output can be enabled by pressing the HV Enable button on the front panel (5).
  4. Adjust the amplitude of the modulation signal as needed.
  5. The DC level of the output can be shifted by adjusting the DC Bias Adjust control (3).

Legend Table for Facing Page

HVA200 Controls and Features

Figure 1: HVA200 Front Panel Features

Figure 2: HVA200 Rear Panel Features

Computer Controlled Operation

The HVA200 may also be controlled by a command line language through the USB port (10). This is offered to enable operation through a terminal interface or for those who may want to write their own program to control the unit. The command language is described below. Prior to running the command line interface, the included drivers should be installed, the unit should be powered, and a USB cable should be connected between the HVA200 and the host.

The terminal emulator should be set as follows:

  • Baud Rate: 115.2k b/s (bits per second)
  • Data Bits: 8
  • Parity: None
  • Stop Bits: 1
  • Flow Control: None

If the connection is correct and after pressing the Enter key, you will see the following: Command error CMD_NOT_DEFINED followed by the command prompt, “>”

The basic structure of the interface is a keyword followed by either an equals sign (=) or a question mark (?). The (=) or (?) will determine if the string is a command or a query. All strings, commands and queries, must be terminated by a carriage return (CR) or pressing the ENTER key on the computer.

The command structure is as follows:
Keyword = argument (CR)

Where “keyword” defines the function and “argument” is a numerical value followed by a carriage return (CR).

The query structure is a follows:
Keyword? (CR)

The “keyword” defines the function and the question mark (?) indicates a query. The string is terminated with a carriage return (CR).

There are a few exceptions to this which are noted below, also noted are unique shortcut keys. The following table lists the commands and queries available with this device. The prompt symbol (>) will appear on power up and after a command is accepted by the HVA200 indicating it is ready to receive another command line.

If the keyword, format, or argument are incorrect or out of range the unit will return an error string.

Notes:

  • All commands and queries are in lower case letters.
    ¹ value= adjustment range based on 16 bit; from 0 – 65535.

    • 0 = maximum value
    • 65535 = minimum value
    • To accurately set the output voltage, use the output monitor BNC.

HVA200 Specifications

¹The voltage gain is inverted to preserve the high slew rate of the output amplifier (i.e., a -1 V input results in +20 V output).
²The HV output will handle an accidental short circuit without damage, but the output should not be shorted continuously.
³The HV monitor output will handle accidental short circuits without damage, but must not be shorted for more than 5 sec to prevent excessive heating of the output resistor. The HV monitor output will source up to 200 mA when a short circuit is applied.

Figure 3: Typical Gain Bandwidth

Maintenance

The HVA200 amplifier needs very little maintenance under normal operating conditions. There are no serviceable parts in the HVA200. The enclosure may be cleaned by wiping with a soft damp cloth. If you suspect a problem with your HVA200 please call Thorlabs and technical support will be happy to assist you.

Fuse Replacement

The AC input is protected by a fuse located in a pull out compartment drawer on the rear panel AC connector. Refer to Figure 2 on page 7. If replacement is needed, disconnect the power cord from the back of the amplifier and pull the fuse compartment drawer out to expose the fuse. A small screwdriver may be used to pry the drawer open.

Replace the fuse with the correct rating. Do not use a fuse with a current rating higher than the unit is rated for. The fuse ratings are as follows:

  • 115 V: 630 mA 5×20 mm SLO-BLO
  • 230 V: 400 mA 5×20 mm SLO-BLO
Ventilation

For proper operation and protection of the output amplifier, it is important that the ventilation passages located on the sides and rear of the unit not be obstructed from free airflow.

Troubleshooting

Declaration of Conformity

Regulatory

As required by the WEEE (Waste Electrical and Electronic Equipment Directive) of the European Community and the corresponding national laws, Thorlabs offers all end users in the EC the possibility to return “end of life” units without incurring disposal charges.

  • This offer is valid for Thorlabs electrical and electronic equipment:
  • Sold after August 13, 2005
  • Marked correspondingly with the crossed out “wheelie bin” logo (see right)
  • Sold to a company or institute within the EC
  • Currently owned by a company or institute within the EC
  • Still complete, not disassembled and not contaminated

As the WEEE directive applies to self contained operational electrical and electronic products, this end of life take back service does not refer to other Thorlabs products, such as:

  • Pure OEM products, that means assemblies to be built into a unit by the user (e.g. OEM laser driver cards)
  • Components
  • Mechanics and optics
  • Left over parts of units disassembled by the user (PCB’s, housings etc.).

If you wish to return a Thorlabs unit for waste recovery, please contact Thorlabs or your nearest dealer for further information.

Waste Treatment is Your Own Responsibility

If you do not return an “end of life” unit to Thorlabs, you must hand it to a company specialized in waste recovery. Do not dispose of the unit in a litter bin or at a public waste disposal site.

Ecological Background

It is well known that WEEE pollutes the environment by releasing toxic products during decomposition. The aim of the European RoHS directive is to reduce the content of toxic substances in electronic products in the future.

The intent of the WEEE directive is to enforce the recycling of WEEE. A controlled recycling of end of live products will thereby avoid negative impacts on the environment.

Thorlabs Worldwide Contacts

For technical support or sales inquiries, please visit us at www.thorlabs.com/contact for our most up- to-date contact information.

USA, Canada, and South America
Thorlabs, Inc.
sales@thorlabs.com
techsupport@thorlabs.com

UK and Ireland
Thorlabs Ltd.
sales.uk@thorlabs.com
techsupport.uk@thorlabs.com

Europe
Thorlabs GmbH
europe@thorlabs.com

Scandinavia
Thorlabs Sweden AB
scandinavia@thorlabs.com

France
Thorlabs SAS
sales.fr@thorlabs.com

Brazil
Thorlabs Vendas de Fotônicos Ltda.
brasil@thorlabs.com

Japan
Thorlabs Japan, Inc.
sales@thorlabs.jp

China
Thorlabs China
chinasales@thorlabs.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals