ADC 4053 Adsafe CPR Pocket Resuscitator User Manual
- June 6, 2024
- ADC
Table of Contents
4053 Adsafe CPR Pocket Resuscitator
User Manual
CONTENTS:
1. Foldable cushion mask with oxygen port. | 4. Elastic mask head strap |
---|---|
2. One-way air valve | 5. Powder-free examination gloves |
3. Disposable air filter | 6. Antimicrobial hand wipes |
Intended Use: This CPR Pocket Resuscitator is designed for the mouth to mask ventilation of a non-breathing adult, child, or infant. It is also used as a barrier that will direct expired air from the patient away from the user. This device should only be used by persons who have received adequate training.
Symbol Definitions
The following symbols are associated with your pocket resuscitator.
Symbol | Definition |
---|---|
Important warning/caution | |
General warning | |
Not made with natural rubber latex | |
Phthalate free | |
Read instructions before use | |
Device part number | |
Lot number | |
Manufacturer’s information | |
Use by date | |
Single use only |
GENERAL WARNINGS:
A warning statement in this manual identifies a condition or practice
which, if not corrected or discontinued immediately, could lead to patient
injury, illness, or death.
WARNING: This product should be used only by persons who have received
adequate training and instructions from qualified medical personnel.
WARNING: In the presence of high oxygen concentrations, an explosion may
be caused by naked flames, or by oil and grease on the resuscitator.
WARNING: If oxygen is used, do not smoke or use in the presence of
sparking equipment or open flame.
WARNING: The one-way valve is a single use only and MUST be discarded
after single patient use.
CAUTION: Connect the oxygen supply line if necessary.
Filter: During disassembly and assembly, make sure mask is in its folded
position.
Disassembly: Hold mask with one hand and clamp the filter housing between
your thumb and two fingers of the other hand. Separate the two by pulling.
Discard filter after each use.
Assembly: Push and snap new filter firmly in place.
|
---|---
1. Remove mask & apply gloves from case.| 2. Unfold mask and check that
filter is firmly in place.
|
3. For adults & children: Place rim between lower lip and chin. Position
“nose” mark over nose.| 4. Seal as illustrated and blow slowly into one-way
valve until chest rises.
|
5. Remove mouth. Allow patient to exhale. Repeat as per CPR guidelines.
Note: If patient vomits, remove mask and clear patient’s airway.| 6. For
infants: Follow adult application procedure (3),
except reverse mask direction so nose area is under the chin.
|
7. Use with oxygen: Strap mask to face as indicated for proper adult or
infant use (3 or 6).| 8. Open oxygen nipple cover & connect oxygen hose to
nipple. Select oxygen flow. Ventilate if necessary.
Cleaning: Mask is reusable and easy to clean. One-way valve can be cleaned and reused ONLY for repeated mannequin training. Use provided alcohol wipe to clean mask and valve after use.
Discard filter after each use.
Wash and scrub in warm soapy water. Rinse in clean water.
Submerge for 10 minutes in a 1:64 household bleach:water solution.
Rinse again and allow to dry.
Do not pasteurize, boil, gas, or steam autoclave mask.
Note: If mask becomes sticky, clean as recommended above.
Use provided alcohol wipe to clean mask and valve after use (with included
models).
Storage: Always keep mask in storage case when not in use.
SPECIFICATIONS
Dimensions: | 5.1” x 4.2” x 1.9” (130mm x 106mm x 48mm) |
---|---|
Operation Temperature: | 0.4°F – 104°F (-18°C – 40°C) |
Storage Temperature: | 22°F – 122°F (-30°C – 50°C) |
Storage relative humidity: | <85% |
Operation Function: | Inspiration/Expiration resistance less than 4.5” |
H²O at 25 LPM
ADC
55 Commerce Drive
Hauppauge, NY 11788
tel: 631-273-9600
toll free: 1-800-232-2670
fax: 631-273-9656|
www.adctoday.com
email: info@adctoday.com
Inspected in the U.S.A.
Made in Taiwan
IB p/n 93-4053-00 rev 6 (2/3/2021)