ADC 9003BAT e-sphyg 3 High-Capacity Battery Installation Guide

June 6, 2024
ADC

ADC logo m1 e-sphyg 3 logo m1

High-Capacity Battery Installation

 #9003BAT

1 Press On/Off switch at back of unit to power off the unit.

Unplug power adapter socket from the back of the unit.

ADC 9003BAT - Step 1 - 1    ADC 9003BAT - Step 1 -
2

2 Remove both screws with a Philips screw driver.

Slide out back cover.

    ADC 9003BAT - Step 2 - 3

ADC 9003BAT - Step 2 - 2      ADC 9003BAT - Step 2 -
4

3 Remove the existing battery power pack.

Please do not dispose of battery in municipal waste.

ADC 9003BAT - Step 3

4 Insert new battery power pack.
Make sure exposed battery wires are tucked underneath the battery power pack so that the battery cover can slide back on.

ADC 9003BAT - Step 4 - 1     ADC 9003BAT - Step 4 -
2

5 Slide in the battery cover back into position and replace screws.

ADC 9003BAT - Step 5 - 1    ADC 9003BAT - Step 5 -
2

6 Plug in power adapter socket at back of device and plug into wall outlet. Charge battery until recharge indicator on main unit turns green.

ADC 9003BAT - Step 6 - 1       ADC 9003BAT - Step 6 -
2


Download full manual at www.adctoday.com/esphyg3

ADC logo m1

American Diagnostic Corporation
55 Commerce Drive
Hauppauge, NY 11788 USA
1-800-ADC-2670
www.adctoday.com

IFU p/n: 93-9003BAT-00 rev 0 (09/23/2021)