ADC 760 Prosphyg Aneriods Instructions
- October 30, 2023
- ADC
Table of Contents
- ADC 760 Prosphyg Aneriods
- Contraindications
- Symbol Definitions
- Operation of Pocket Aneroids
- Measurement Procedure
- Position the Stethoscope
- Measurement
- Care and Maintenance
- Cuff Cleaning and Disinfecting
- Low-Level Disinfection
- Manometer Quality Control
- Disposal
- Warranty
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
ADC 760 Prosphyg Aneriods
Thank you for choosing an ADC® ProsphygTM Aneroid sphygmomanometer.
Device Description and Intended Use
An ADC® aneroid sphygmomanometer is used by professional healthcare providers
and individuals trained in the auscultatory blood pressure technique to
determine systolic and diastolic blood pressure in humans.
Contraindications
Aneroid sphygmomanometers are contraindicated for neonate use. Do not use it with neonatal cuffs or neonate patients. Review the size chart (below) for proper limb range.
780-790 Series Size Chart
Cuff Size Limb Range Inches CM
Sm. Adult | 10SA | 6.3 to 9.8 | 16 to 25 |
---|---|---|---|
Adult | 11A | 9 to 13.3 | 23 to 34 |
Lg. Adult | 12X | 11.8 to 15.7 | 30 to 40 |
Adcuff Series Size Chart
Cuff Size Limb Range Inches CM
Infant | 7I | 3.5 to 5.5 | 9 to 14 |
---|---|---|---|
Child | 9C | 5.1 to 7.6 | 13 to 19.5 |
Sm. Adult | 10SA | 7.4 to 10.6 | 19 to 27 |
Adult | 11A | 9 to 15.7 | 23 to 40 |
Lg. Adult | 12X | 13.3 to 19.6 | 34 to 50 |
Thigh | 13T | 15.7 to 25.9 | 40 to 66 |
Symbol Definitions
The following symbols are associated with your ADC sphygmomanometer.
General Warnings
A warning statement in this manual identifies a condition or practice which,
if not corrected or discontinued immediately could lead to patient injury,
illness, or death.
WARNING: Do not allow a blood pressure cuff to remain on the patient for
more than 10 minutes when inflated above 10 mmHg. This may cause patient
distress, disturb blood circulation, and contribute to the injury of
peripheral nerves.
WARNING: If luer lock connectors are used in the construction of tubing,
there is a possibility that they might be inadvertently connected to intra-
vascular fluid systems, allowing air to be pumped into a blood vessel.
Immediately consult a physician if this occurs.
WARNING: Safety and effectiveness with neonate cuff sizes 1 through 5 is
not established.
WARNING: If this equipment is modified, appropriate inspection and
testing must be conducted to ensure its continued safe use.
WARNING: Do not apply the cuff to delicate or damaged skin. Check the
cuff site frequently for irritation.
WARNING: Only use the cuff when the range markings indicated on the cuff
show that the proper cuff size is selected, otherwise erroneous readings may
result.
WARNING: Allow space between patient and cuff. Two fingers should fit in
this space if the cuff is correctly positioned.
WARNING: Do not apply the cuff to limbs used for IV infusions.
WARNING: The patient should remain still during measurement to avoid
erroneous readings.
WARNING: When using an infant or child cuff, extra care must be taken to
prevent over-inflation. With smaller cuffs (infant or child) the cuff can
inflate to over 300mmHg with just two full compressions of the bulb. To
prevent discomfort or injury to the patient and damage to the instrument, the
bulb should only be partially squeezed, so that each “stroke” inflates the
cuff in 40mmHg to 60mmHg increments until inflated to the desired level.
CAUTION: To obtain the greatest accuracy from your blood pressure
instrument, it is recommended that the instrument be used within a temperature
range of 50°F to 104°F (10°C to 40°C), with a relative humidity range of
15%-85% (non-condensing).
CAUTION: Extreme altitudes may affect blood pressure readings. Your
device has been designed for normal environmental conditions.
Operation of Pocket Aneroids
- This booklet contains operating and maintenance information for pocket aneroid sphygmomanometers. Please read and retain.
- Your pocket aneroid sphygmomanometer consists of an aneroid manometer (gauge), a complete inflation system, (latex-free inflation bladder, squeeze bulb, and the valve), a zippered carrying case, and operating instructions.
- Most models are preassembled and ready for use. In units requiring assembly, the bulb, and valve should connect to the tube closest to the Index Line. The gauge connects to the remaining tube. Apply a small amount of alcohol or soapy water to the tubing if needed.
Measurement Procedure
Patient Position
The patient should sit or lie comfortably. The arm should be fully supported
on a flat surface at heart level. (If the arm’s position varies, or is not
level with the heart, measurement values obtained will not be consistent with
the patient’s true blood pressure.) When seated, the patient should have their
back and arm supported, and their legs should not be crossed. The patient
should relax prior to measurement comfortably for five minutes and should
refrain from talking or moving during measurement. The observer should view
the manometer in a direct line to avoid “parallax error.”
Apply the Cuff
- Nylon cuffs are specially designed to promote the precisely accurate determination of blood pressure. Index and range markings ensure the use of the correct cuff size. The artery mark indicates proper cuff positioning.
- Place the cuff over the bare upper arm with the artery mark positioned directly over the brachial artery. The bottom edge of the cuff should be positioned approximately one inch (2-3cm) above the antecubital fold. Wrap the end of the cuff not containing
the bladder around the arm snugly and smoothly and engage adhesive strips (Figure 1).
NOTE: If the unit is equipped with a calibrated nylon cuff, featuring Index and Range markings, a correct fit may be verified by checking that the Index Line falls between the two Range Lines. (Figure 2).
Inflate the Cuff
- Close the valve by turning the thumbscrew clockwise.
- Palpate the radial artery while inflating the cuff. Be sure to inflate the cuff quickly by squeezing the bulb rapidly.
- Inflate the cuff 20-30 mmHg above the point at which the radial pulse disappears.
NOTE: Cuff pressure range is 0 mmHg to 300 mmHg.
Position the Stethoscope
- Position the chest piece in the antecubital space below the cuff, distal to the brachium. Do not place the chest piece underneath the cuff, as this impedes accurate measurement. Use the bell side of a combination stethoscope for the clearest detection of the low-pitched Korotkoff (pulse) sounds.
Deflate the Cuff
- Open the valve to deflate the cuff gradually at a rate of 2-3 mmHg per second.
Measurement
- Record the onset of Korotkoff sounds as systolic pressure and the disappearance of these sounds as diastolic pressure. (Some healthcare professionals recommend recording diastolic 1 and diastolic 2. Diastolic one occurs at phase 4.)
NOTE: It is recommended that K4 be used in children aged 3 to 12, and K5
should be used for pregnant patients unless sounds are audible with the cuff
deflated, in which case K4 should be used. K5 should be used for all other
adult patients.
After the measurement is completed, open the valve fully to release any
remaining air in the cuff. Remove cuff.
Care and Maintenance
STORAGE
Pocket Gauge: After measurement, fully exhaust the cuff then wrap the cuff around the gauge and bulb and store it in a zippered carrying case.
NOTE: This product will maintain the safety and performance
characteristics specified at temperatures ranging from 50°F to 104°F (10°C to
40°C) at a relative humidity level of 15% to 85%.
This device can be safely stored at temperatures ranging from -4°F to
131°F(-20°C to 55°C) with a relative humidity of 85% or below.
Manometer: Your pocket aneroid gauge requires minimal care and
maintenance.
The manometer may be cleaned with a soft cloth but should not be dismantled
under any circumstances.
- Should the indicator needle of the manometer rest outside the oval calibration mark (Figure 3), then the manometer must be re-calibrated to within ±3 mmHg when compared to a reference device that has been certified to national or international measurement standards. A manometer whose indicator needle is resting outside of this mark is NOT acceptable for use.
- In the event that the gauge is ever in need of calibration, simply return for service. Damaged or broken parts will be replaced as needed at a minimal charge. Refer to the warranty for specific details of warranty coverage.
- The manufacturer recommends a calibration check every 2 years.
Cuff Cleaning and Disinfecting
Use one or more of the following methods and allow to air dry:
- Wipe with mild detergent and water solution (1:9 solution). Rinse.
- Wipe with Enzol per the manufacturer’s instructions. Rinse.
- Wipe with .5% bleach and water solution. Rinse.
- Wipe with 70% isopropyl alcohol.
- Launder with a mild detergent in warm water, normal wash cycle.
- Remove the bladder first. Cuff is compatible with 5 wash cycles.
Low-Level Disinfection
Prepare Enzol enzymatic detergent according to the manufacturer’s instructions. Spray the detergent solution liberally onto the cuff and use a sterile brush to agitate the detergent solution over the entire cuff surface for five minutes. Rinse continuously with distilled water for five minutes. To disinfect, first, follow the cleaning steps above, then spray the cuff with 10% bleach solution until saturated, and agitate with a sterile brush over the entire cuff surface for five minutes. Rinse continuously with distilled water for five minutes. Wipe off excess water with a sterile cloth and allow the cuff to air dry.
CAUTION: Do not iron the cuff.
CAUTION: Do not heat or steam sterilize the cuff.
Manometer Quality Control
- A serial number and lot number are automatically assigned to every aneroid during manufacturing, ensuring every item is “controlled.”
- The serial number can be located on the faceplate of each aneroid (Figure 4).
- The lot number is located on the box end label (Figure 5).
Standards
- ANSI/AAMI/ISO 81060-1:2007 • EN/ISO 81060-1:2012.
Disposal
When your sphygmomanometer or any of its parts have reached their end of life, please be sure to dispose of them in accordance with all regional and national environmental regulations. Devices that have become contaminated should be disposed of in accordance with all local ordinances and regulations.
Warranty
- ADC® warrants its products against defects in materials and workmanship under normal use and service as follows:
- Warranty service extends to the original retail purchaser only and commences with the date of delivery.
- See the end panel label for package contents and to determine warranty duration.
Warranty duration is as follows
- SeriesManometerInflation System
- 760, 768, 76920 Years3 Years
- 780, 785, 790, 770, 77510 Years1 Year
What Is Covered: Replacement of parts and labor.
What Is Not Covered: Transportation charges. Damages caused by abuse, misuse,
accident, or negligence. Incidental, special, or consequential damages. Some
states do not allow the exclusion or limitation of incidental, special, or
consequential damages, so this limitation may not apply to you.
Implied Warranty: Any implied warranty shall be limited in duration to the terms of this warranty and in no case beyond the original selling price (except where prohibited by law). This warranty gives you specific legal rights and you may have other rights which vary from state to state.
To Obtain Warranty Service: Send item(s) postage paid to: ADC, Attn: Repair Dept., 55 Commerce Dr., Hauppauge, NY 11788. Please include your name and address, phone no., proof of purchase, and a brief note explaining the problem.
For Australian Consumers
Our goods come with guarantees that cannot be excluded under the Australian
Consumer Law. You are entitled to a replacement or refund for a major failure
and compensation for any other reasonably foreseeable loss or damage. You are
also entitled to have the goods repaired or replaced if the goods fail to be
of acceptable quality and the failure does not amount to a major failure.
For our European Customers
- On request, send to ADC by e-mail (info@adctoday.com), we can send to you this manual on paper form within 7 calendar days at no additional cost to the user.
- Our website, https://www.adctoday.com, where these instructions for use are available fulfills the requirements of personal data protection, according to Directive 95/46/EC and GDPR on the protection of individuals with regard to the processing of personal data and on the free movement of such data.
- Any serious incident that has occurred in relation to this medical device should be reported to ADC and the competent authority of the Member State in which the user and/or patient is established.
To register your product, visit us at
www.adctoday.com/support/warranty-
registration.
FOR QUESTIONS, COMMENTS, OR SUGGESTIONS CALL TOLL-FREE: 1-800-ADC-2670 or visit www.adctoday.com/feedback.
ADC
- 55 Commerce Drive
- Hauppauge, NY 11788
- U.S.A.
SC Cattus SRL
- Str. Baneasa Nr. 10 C
- Târgu-Mures, Jud. Mures
- România, EU
Inspected,
- assembled and packaged in the U.S.A.
- tel: 631-273-9600
- 1-800-232-2670
- fax: 631-273-9659 www.adctoday.com.
- email: info@adctoday.com.
References
- American Diagnostic Corporation - Core Medical Device Manufacturer. Stethoscopes, Blood Pressure, Thermometry, and EENT
- Customer Feedback Form | American Diagnostic Corporation
- Warranty Registration | American Diagnostic Corporation
- American Diagnostic Corporation - Core Medical Device Manufacturer. Stethoscopes, Blood Pressure, Thermometry, and EENT
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>