importvet 24.300 Sonda Tecnoscan BL Instruction Manual
- June 6, 2024
- importvet
Table of Contents
Tecnoscan BL
Instruction manual
Scanner menu
Power button (A)
- To switch on/off the scanner press and hold until it beeps.
- Short press to freeze the image.
- If on a congested Wi-Fi channel, double press to change the channel when no viewing devices are connected.
Up button (B)
- Adjust gain.
- Save videos.
Down button (C)
- Adjust depth.
- Save images.
1.1 Scanner menu
Once the gain or depth is visible, pressing the “Power Button” will select that setting. Use the Up and Down buttons to adjust the selected settings.
1.2 Saving images and videos using the scanner buttons
Freeze the image using “Power button” Press “Up” will bring up “Save Cine”.
Freeze the image using “Power button” Press “Down” will bring up “Save Image”.
Once the desired option is visible, pressing the Power Button will select that
option. Note: The scanner will store the last 300 frames (1012.5 seconds
depending on the selected scan depth) in its internal memory. To store on the
scanner, the scanner must be Frozen. During save, particularly cine save, the
app will pause while frames are downloading and saving the data while
scanning. It is recommended that you save images and videos directly from the
app where possible. The cine loop must be downloaded by the app before
switching off, otherwise, it will be lost. The scanner can store up to 100
images in its internal memory, these images are retained after power off and
can be downloaded by the app at a later time.
LED sequence
POWER LED|
---|---
Blue flashing| Scanner turning on
Blue| Scanner on and battery between 66-100% charge
Yellow| Scanner on and battery between 33-66% charge
Yellow flashing| Scanner on and battery between 0-33% charge
WI-FI LED|
Blue flashing| Scanner ready for smart device to connect, nothing connected
Blue| Smart device connected
BOTH LEDS|
Yellow flashing| Software updating
Activation
3.1 First-time use
Download the Tecnoscan Connect from Google Play or the App Store.
You are likely to be prompted for a scanner software update on first-time use.
Please see the Scanner Update section.
To activate the scanner, ensure that the scanner is charged and that you have
a local Wi-Fi internet connection. This is only necessary the first time you
connect the scanner.
After a few seconds, the ultrasound image will appear. Touch anywhere on the
screen to bring up the live scanning menu.
Note: If the Wi-Fi LED is flashing on the scanner it is not connected to
any smart device.
3.2 Normal start up process
-
Switch on the scanner and wait until the Wi-Fi LED illuminates. Launch Tecnoscan Connect App. App requests to connect to the scanner.
-
Select your Tecnoscan BL from the Wi-Fi page. Return to the App.
-
It can take up to 10 seconds to connect and start receiving an ultrasound. You are now ready to scan.
3.3 Scanner Updates
The app automatically checks the Tecnovet license server for scanner software
updates when connected to Wi-Fi.
After connecting to a Tecnoscan BL, you may be prompted to update the scanner
software. The scanner will only update if the battery is above 50% charge.
Press OK or Cancel. Pressing OK will cause the scanner LEDs to flash yellow –
the scanner will take roughly two minutes to update. Press cancel if you wish
to update later.
Once the update is finished the scanner will switch off. The scanner software
version number can be checked in the User Settings section.
Tecnoscan Connect
4.1 Live scanning
- Change the channel in Wi-Fi settings. No indicator: clear channel on the 5GHz band. This is the recommended state. White indicator: clear channel, but is on 2.4GHz band, it is more prone to interference. Orange/red indicator: channel has other networks or scanners on it. This is likely to cause interference issues.
4.2 Main menu
To enable frame averaging, touch the icon. To change the number of frames
being averaged, touch the icon (each touch increases the number of frames up
to 4 then returns to 2). To disable frame averaging, long-press the icon.
Frame averaging combines the current frame with the previous 2, 3, or 4
frames, this produces a smoother image with less noise, but at the cost of
blurring when the probe is moving quickly.
Note: Touch in the text area to type, up to 4 lines of text may be entered and will appear as red text with may be dragged to a preferred location. This text will be present on any saved images.
Callipers
Cattle
Note: Use your finger to move the cursors.
Gestational measurements available
Crown-Rump length (distance from the top of the skull to the end of the body)|
Trunk diameter (width of the ribcage at the widest part)| Eye diameter
(diameter of the eye socket)| Head diameter, (width of the skull at the widest
point)| Head length (distance from the top of the skull to the end of the
nose)
---|---|---|---|---
14mm – 180mm,
33 – 76 days| 14mm – 123mm,
53 – 133 days| 3mm – 28mm,
62 – 241 days| 14mm – 86mm,
55 – 135 days| 21mm – 146mm,
56 – 134 days
Sheep
Head diameter (width of the skull at the widest point)
19mm – 66mm
48 – 121 days
| Trunk diameter (width of the ribcage at the widest part)
14mm – 101mm
49 – 126 days
---|---
Horse
Eye diameter (diameter of the eye socket)
9mm – 35mm
99 – 344 days
| Equine Conceptus diameter (diameter of the embryo)
14mm – 75mm
14 – 45 days
---|---
Freedraw
Touch and drag to draw on screen can be used to make simple annotations or to highlight areas of interest. Any measurements placed on-screen remain visible.
To enable meat quality, touch the icon, to disable, touch the icon again.
When in meat quality mode the app will automatically find the fat/muscle
boundary and the reflection of the rib and draw horizontal lines at both
interfaces.
The gain control becomes a sensitivity control, with the sensitivity at
minimum, the app looks for the strongest reflection, with the sensitivity at
maximum, it uses a mixture of the average value and the maximum, which can
compensate for a poor or noisy image.
The fat thickness is the distance from the probe face to the fat line and the
meat thickness is the distance between the fat line and the rib line.
If the app considers the measurements to be good, then the fat line will be
green and the rib line purple, if not, then the lines will be red.
4.2.2 Meat Quality
4.3 Settings
Usage files
The app will generate usage files. Events logged:
- Images saved – time and date, scanner serial number and location (if permitted)
- Connecting to a scanner
- Freeze and unfreeze
- Entering and exiting standby mode
Factory reset
A scanner can be reset to factory firmware settings in the unlikely event the
scanner fails. Hold down Power and Down until both LEDs glow yellow and
release.
This will take a few minutes and the scanner will then switch off – removing
the license from the scanner.
Charging
The battery can be charged in the charging dock, powered by a standard outlet. It can also be connected in the carry case for in-car charging.
LED sequence
One blue flashing | Charging, charge level below 33% |
---|---|
First LED solid blue A second LED lashing blue | Charging, charge level below |
66%
First and second LEDs are solid blue, the third LED flashing blue| Charging,
charge level below 100%
All LEDs solid blue – fully charged.| Fully charged
All LEDs flashing yellow| Temperature is too high (more than 45 Celsius, 113
Fahrenheit) for safe charging
All LEDs flashing blue| Temperature is too low (less than 0 Celsius, 32
Fahrenheit) for safe charging.
A fully discharged battery should be fully charged within 5.5 hours. While in
a good condition, the battery run time will be 5 hours (assuming 50% idle time
between scans).
Scanner will enter low power mode when idle. The Li-ion cells used in the
battery pack do not suffer from the ‘Memory effect’. For optimal battery
longevity, avoid letting the battery run completely down and recharge at
approximately 20 Celsius/68 Fahrenheit. Electrical connection to the power
pack is through the gold pads on the battery and scanner. Power pack output is
protected but care should be taken to prevent short-circuiting the gold
connector pins and pads.
Warning:
To reduce the risk of burns, fire, electric shock, or injury to persons an
appliance should never be left unattended when plugged into mains or 12 V
cigarette lighter.
Care & maintenance
Please note the following important points:
The unit is not sealed against jets of water. Water will penetrate your
scanner if it is hosed down or rinsed under a fast-flowing tap.
- The unit is NOT corrosion-proof and thus it should not be left wet. Take care to detach the external battery and ensure interconnect areas are dry also.
- The battery charger and charging carry case is not waterproof – avoid getting them wet.
Probe
While every attempt has been made to make the probe as rugged as possible the
crystal array remains vulnerable and should be protected from knocks. The
probe has a protective sheath covering the length of the cable subject to wear
and flex. Any damage to this sheath or the rest of the cable should be
repaired before further use to prevent moisture ingress that will rapidly
damage the probe. The probe should be checked regularly for such damage.
Cleaning
The scanner, battery, and probe can be wiped down with an antibacterial wet
wipe, reducing the need for drying. This prevents problems relating to
corrosion following wash-down.
The scanner and power pack can also be cleaned using a cloth, soap, and warm
water. Sponging down or even brief, shallow submersion is less aggressive than
rinsing or hosing down.
Warning: The scanner and power pack must not be stored damp – severe
corrosion will result. Thoroughly dry the scanner before storage and charging.
Take care to detach the external power pack and ensure interconnect areas are
dry also.
Gels
Use only gels recommended by Tecnovet. Do not store the probe in gel.
Storage
Clean and dry the probe head prior to storage in a dry location. Do not store
the transducer in gels or cleaning/disinfecting solutions. This will lead to
premature deterioration of the probe. Do not expose the transducer to
materials containing the following agents:
- Acetone
- Methanol
- Mineral oil
- Iodine
- Freon
- Industrial cleaners
- Materials containing perfumes (lotions, gels, etc.)
Warning:
Never sterilize the transducer with autoclave, ultraviolet, gamma radiation,
gas, steam, or heat sterilization techniques. Severe damage will result.
Specifications
Size| 19×8.3×6.2cm
7.5 x 3.25 x 2.4 in
---|---
Weight| 800g/1.76lbs
Mechanical| Rugged Glass-filled
Polypropylene casing with TPE over mould.
Scanner operating frequency bands: B1=2412MHz2462MHz and B2=5150MHz-5250MHz
Maximum radio-frequency power transmitted in the frequency bands: 16.1dBm for
B1 and 18dBm for B2.
User Output
Scanner wireless link to the compatible viewing device using Sound-Link over
standard WI-FI connection. Image and video storage Storage of images and cine
loops is limited only by available space on a smart device linked with the
scanner.
Probe
Broadband straight linear rectal
Active array length 64 mm
Frequency range 4.5 to 8.5 MHz
3 user selectable depths of scan: 6cm, 8cm, 12cm
128 element crystal array.
Power and charging
Removable Li-ion battery gives 5h runtime
Battery charge time 5.5h. Only use Tecnovet approved/supplied battery, 12V
vehicle power supply, battery charger, and DC power adapter.
ESG-CHARGER: Input: 12V, 2A / Output: 4.2V, 2A.
Tecnovet Vehicle Power Outlet/Cigarette Plug Cable
(Model: ACC-DC LEAD):12VDC, 2A Fused.
Only to be used when charging from a vehicle supply.
Only replace a fuse with 2A medium acting fuse rated to
125V, 10kA breaking capacity (F2A M125V).
DC Power Adapter (XP Power, Model: VEP24US12):
Input: 100-240VAC, 0.6A/Output: 12VDC, 2.0A
Battery (Model: ESG-BATT): 3.6Vdc, 6700mAhr. Scanner will enter low power mode
when idle.
Temperature range
Scanner: -10 Deg C to 45 Deg C .
Battery Charge: 0 Deg C to 35 Deg C *
Battery Charger has under and over-temperature
protection and LED status indication.
Warning:
If the equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired.
- Best charged at 25 Deg C, frequent charging at 35 Deg C / 95 F may shorten battery lifetime.
Warranty
Tecnovet warrants Tecnoscan BL against defects in materials for one year from
the date of purchase. Tecnovet does not warrant against normal wear and tear,
Smart devices compatibility
Android:
Minimum version: Kit Kat (Android V4.4) Minimum
RAM: 2Gb
Minimum Processor: Quad-Core 1.8GHz
Recommended WiFi: 802.11 a/ac (5GHz band)
Apple:
Minimum version: iOS 8
Minimum iPhone version: iPhone 5s Minimum iPad
version: iPad4
Hereby, Tecnovet SL, declares that the radio equipment type Tecnoscan BL is in
compliance with Directive 2014/53/EU. The full text of the EU declaration of
conformity is available at the following internet address: https://www.imv-
imaging.com/esg-doc/
This device complies with part 15 of the FCC Rules. Operation is subject to
the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation of the device.
If the user makes any modifications not expressly approved by the party
responsible for compliance it could void the user’s authority to operate the
equipment.
This device complies with Industry Canada’s license-exempt RSSs. Operation is
subject to the following two conditions: (1)This device may not cause
interference; and (2) This device must accept any interference, including
interference that may cause undesired operation of the device.
Contains IC4511-WL18DBMOD
Note: All Tecnovet products are for animal applications only.
The product shall not be used outside its scope as declared by the
manufacturer.
Tecnoscan BL: Instruction Manual