Bird 6150-ALM-02 Fiber-Fed NFPA Alarm Panel User Manual
- June 6, 2024
- Bird
Table of Contents
Bird 6150-ALM-02 Fiber-Fed NFPA Alarm Panel
©Copyright 2021 by Bird Technologies Inc.
Instruction Book Part Number 920-6150-ALM-02 Rev. A
Safety Precautions
The following are general safety precautions that are not necessarily related to any specific part or procedure and do not necessarily appear elsewhere in this publication. These precautions must be thoroughly understood and apply to all phases of operation and maintenance.
WARNING
-
Keep Away From Live Circuits
Operating Personnel must at all times observe general safety precautions. Do not replace components or make adjustments to the inside of the test equipment with the high voltage supply turned on. To avoid casualties, always remove power. -
Do Not Service Or Adjust Alone
Under no circumstances should any person reach into an enclosure for the purpose of service or adjustment of equipment except in the presence of someone who is capable of rendering aid. -
Safety Earth Ground
An uninterruptible earth safety ground must be supplied from the main power source to test instruments. Grounding one conductor of a two conductor power cable is not sufficient protection. Serious injury or death can occur if this grounding is not properly supplied. -
Resuscitation
Personnel working with or near high voltages should be familiar with modern methods of resuscitation. -
Remove Power
Observe general safety precautions. Do not open the instrument with the power applied.
Safety Symbols
WARNING
Warning notes call attention to a procedure, which if not correctly performed,
could result in personal injury.
CAUTION
Caution notes call attention to a procedure, which if not correctly performed,
could result in damage to the instrument.
Note: Calls attention to supplemental information.
Safety Statements
-
USAGE
ANY USE OF THIS INSTRUMENT IN A MANNER NOT SPECIFIED BY THE MANUFACTURER MAY IMPAIR THE INSTRUMENT’S SAFETY PROTECTION. -
USO
EL USO DE ESTE INSTRUMENTO DE MANERA NO ESPECIFICADA POR EL FABRICANTE, PUEDE ANULAR LA PROTECCIÓN DE SEGURIDAD DEL INSTRUMENTO. -
BENUTZUNG
WIRD DAS GERÄT AUF ANDERE WEISE VERWENDET ALS VOM HERSTELLER BESCHRIEBEN, KANN DIE GERÄTESICHERHEIT BEEINTRÄCHTIGT WERDEN. -
UTILISATION
TOUTE UTILISATION DE CET INSTRUMENT QUI N’EST PAS EXPLICITEMENT PRÉVUE PAR LE FABRICANT PEUT ENDOMMAGER LE DISPOSITIF DE PROTECTION DE L’INSTRUMENT. -
IMPIEGO
QUALORA QUESTO STRUMENTO VENISSE UTILIZZATO IN MODO DIVERSO DA COME SPECIFICATO DAL PRODUTTORE LA PROZIONE DI SICUREZZA POTREBBE VENIRNE COMPROMESSA.
About This Manual
This manual covers the installation, operating & maintenance instructions for: 6150-ALM-02
Changes to this Manual
We have made every effort to ensure this manual is accurate. If you discover
any errors, or if you have suggestions for improving this manual, please send
your comments to our Solon, Ohio factory. This manual may be periodically
updated. When inquiring about updates to this manual refer to the part number
and revision on the title page.
Introduction
The Bird NFPA Alarm Panel is an SBII+ Fiber-DAS alarm status annunciator that
provides a convenient means of monitoring your Fiber-DAS alarms from a
reasonably nearby location such as a security desk or main lobby desk located
in the same building as the Fiber-DAS system.
Figure 1 Alarm Panel
Monitored Alarms
The alarm panel is capable of monitoring the five alarms required for NFPA
compliance including:
Signal Booster Failure — This is a summed alarm that is active when any
number of fault conditions arise within the booster unit such as when an
overcurrent or high-temperature event occurs.
Loss of AC Power — Indicates that AC power to the booster unit has
failed.
Antenna Malfunction — Indicates there is a loss of RF cable integrity
between the signal booster unit and the remote antenna line sensor. In order
to detect an antenna malfunction, an antenna line sensor must be installed as
close to the Donor antenna as possible. This sensor works with the Bias-T
assembly inside the booster cabinet to verify the continuity of the antenna
line
Low Battery Condition — The source of this alarm is the battery backup
unit and it indicates that the battery backup voltage level has dropped
significantly.
Charger Failure — The source of this alarm is the battery backup unit and
it indicates failure of the battery charger. If the battery charger looses AC
power this alarm will be active.
Other Monitored Parameters
In addition to the NFPA alarms, the Alarm Panel displays the status following
parameters:
Power — when illuminated indicates the alarm panel is connected to/and
receiving power from a Head-End unit.
Status — Link and alarm status is displayed in the upper right corner of
the display, see Figure 2.
Figure 2 Status Indicator
Alarm Reporting
When an alarm condition is detected, the corresponding indicator will
change to RED. The status indicator in the top right of the display will
indicate the booster’s alarmed status, see Figure 3.
Link Status
Link status during normal operation is indicated by no alarm (green) or
alarms reported (red).
If communication is lost with the Head-End Unit, a disconnected icon (yellow)
is displayed. The Alarm panel will automatically attempt to reconnect with the
Head-End unit. If the Alarm Panel remains disconnected for approximately one
minute, the offline message is displayed indicating the connection could not
be reestablished, see Figure 3.
Cable Connections
Fiber-DAS Head-End Unit — The Alarm Panel connects to the Head-End unit
using a standard Ethernet cable with RJ-45 connectors. The Ethernet cable
provides operating voltage for the Alarm Panel and carries the Fiber-DAS
system status.
Facility Panel — Optionally, the NPFA alarm status may be extended from
the Alarm Panel (25-pin connector) to an existing facility alarm panel.
Figure 4 Alarm Panel Connections
Installing the Alarm Panel
The alarm panel is designed to be wall-mounted and has a built-in mounting
flange on the top and bottom of the enclosure. Each flange has two 5/16 screw
holes and all four holes should be used during installation to insure the box
is fastened firmly.
Place the panel in a convenient viewable location. The cable connections are
on the bottom of the enclosure, so the bottom should be accessible.
Figure 5 shows the location of the mounting brackets on the alarm panel
enclosure.
Figure 5 Alarm Panel Mounting
Booster – Alarm Panel Connection
The Alarm Panel is equipped with one RJ-45 Alarm input connector. This
connector is used to connect a Bird Fiber-DAS Head-End Unit to the alarm
panel. The RJ-45 connector is used for alarm panel power and alarm data. See
Figure 6.
Bird Boosters with an RJ-45 Alarm Panel output (data + POE) can connect
directly to the RJ-45 connector on the bottom of the alarm panel. A standard
Ethernet cable is used for the connection, the maximum length for this cable
is 300 feet (100 meters). The RJ-45 connection supplies power for the alarm
panel and interfaces the boosters NFPA alarms.
Figure 6 Alarm Input connections
Alarm Panel Output Connection
An alarm output connector is provided to allow the NFPA alarms to be extended
to a facility alarm panel. The Alarm Output connector is a standard D-sub
25-pin receptacle. See Figure 6.
One of the important considerations when installing the alarm panel wiring is
noise on the lines which may cause false alarm indications at the alarm panel.
Using good quality cable with sufficient shielding is important. When laying
the cable run avoid running it next to the building’s AC wiring or near strong
noise sources like motors.
If your installation requires the distance between the alarm panels to be
greater than several thousand feet consult with the factory before running the
signal cable. The wire gauge and shielding used for the signal lines will
determine the workable distance, thicker wire gives greater distance but can
be cost-prohibitive.
The alarm can be connected as normally open or normally closed circuits.
Normally closed is defined as a contact pair that is normally closed when the
booster is powered-up, operating normally, with no alarm conditions.
The pinout for the Alarm Connector are shown in Table 1 on page 6.
Table 1 Alarm 25-Pin Connector Output Pinout
Normal Operation
The LCD alarm panel connects to the head-end unit of the SBII+ Fiber-DAS and
will display alarms for the head end unit as well as each remote that is
connected to the head-end through fiber-optic cable.
When the Alarm Panel is connected to a head-end unit and the unit is powered
up, the alarm panel will display Power (green) and the alarm status of the
head-end unit.
When the alarm panel is powered-up and the head-end unit and remote units are
operating normally (no faults) the status lights will be green for the
monitored NFPA alarm parameters for each unit.
Alarm Condition
When a unit detects a malfunction for one of the monitored NFPA parameters,
the associated indicator will change to red. The status indicator will also
change to an alarm condition.
Figure 7 Alarm Panel Display
Communication Lost
If communication between the Fiber-DAS system and the Alarm panel is lost, see Figure 7, and does not reconnect, indicated by the offline message, troubleshoot the connection. See “ Troubleshooting” on page 8.
Troubleshooting
If Alarm panel is malfunctioning find the symptoms in the following paragraphs for the steps to isolate the cause.
Alarm Panel Display’s Offline Message
Recheck systems after each step.
-
Disconnect Ethernet cable from alarm panel, inspect connector, then reinsert.
If Ethernet connector is damaged, replace connector. -
Disconnect Ethernet cable from Head-end unit, inspect connector, then If Ethernet connector is damaged, replace connector.
-
Inspect Ethernet cable for damage.
If Ethernet cable is damaged, repair or replace the cable. -
Cycle power on the Head-End unit.
If Offline message is still displayed, contact Customer Service.
No Alarm Panel Display, Power LED On
- Perform the steps in Alarm Panel Display’s Offline Message.
If symptoms persist, contact Customer Service.
No Alarm Panel Display, Power LED Off.
Recheck systems after each step.
-
Verify power is applied to the Head-End unit.
If power to Head-End is off, determine case and restore power to the Head-End -
Disconnect Ethernet cable from alarm panel, inspect connector, then reinsert.
If Ethernet connector is damaged, replace connector. -
Disconnect Ethernet cable from Head-end unit, inspect connector, then reinsert.
If Ethernet connector is damaged, replace connector. -
Inspect Ethernet cable for damage.
If Ethernet cable is damaged, repair or replace the cable. 5. If symptoms persist, contact Customer Service.
Customer ServiceAny maintenance or service procedure beyond the scope of those
in this chapter should be referred to a qualified service center.
If the unit needs to be returned for any reason, request an Return Material
Authorization (RMA) through the Bird Technologies website. All instruments
returned must be shipped prepaid and to the attention of the RMA number.
Bird Service Center
30303 Aurora Road
Cleveland (Solon), Ohio 44139-2794 Fax: 440-248-5426
E-mail: bsc@birdrf.com
For the location of the Sales Office nearest you, visit our Web site at:
http://www.birdrf.com
Specifications
Specification | Value |
---|---|
Alarms Monitored | Signal Booster Failure,Loss of AC Power, Antenna |
Malfunction, Low Battery Condition, Charger Failure.
Display| 7” LCD
Alarm Indications
Virtual LED| Red
Alarm Input
Alarm Input| Ethernet
Alarm Input Connector| RJ-45
Alarm Output
Alarm Output Connector| D-sub 25-pin
Alarm Types| Normally Open,
Normally Closed
Max Allowable Voltage| 30 VDC
Max Current| 1 A
Wire Size| 18 to 26 AWG
Enclosure Rating| NEMA 4
Operating Temperature| 32 °F to 122 °F (0 °C to 50 °C)
Enclosure Dimensions| 8.5” x 8.7” x 5.2”
(21.6 cm x 22.1 cm x 13.2 cm) H x W x D
Enclosure Weight| 7 lbs (3.18 kg)
Limited Warranty
All products manufactured by Seller are warranted to be free from defects in material and workmanship for a period of one (1) year, unless otherwise specified, from date of shipment and to conform to applicable specifications, drawings, blueprints and/or samples. Seller’s sole obligation under these warranties shall be to issue credit, repair or replace any item or part thereof which is proved to be other than as warranted; no allowance shall be made for any labor charges of Buyer for replacement of parts, adjustment or repairs, or any other work, unless such charges are authorized in advance by Seller.
If Seller’s products are claimed to be defective in material or workmanship or not to conform to specifications, drawings, blueprints and/or samples, Seller shall, upon prompt notice thereof, either examine the products where they are located or issue shipping instructions for return to Seller (transportation- charges prepaid by Buyer). In the event any of our products are proved to be other than as warranted, transportation costs (cheapest way) to and from Seller’s plant, will be borne by Seller and reimbursement or credit will be made for amounts so expended by Buyer. Every such claim for breach of these warranties shall be deemed to be waived by Buyer unless made in writing within ten (10) days from the date of discovery of the defect.
The above warranties shall not extend to any products or parts thereof which have been subjected to any misuse or neglect, damaged by accident, rendered defective by reason of improper installation or by the performance of repairs or alterations outside of our plant, and shall not apply to any goods or parts thereof furnished by Buyer or acquired from others at Buyer’s request and/or to Buyer’s specifications. Routine (regularly required) calibration is not covered under this limited warranty. In addition, Seller’s warranties do not extend to the failure of tubes, transistors, fuses and batteries, or to other equipment and parts manufactured by others except to the extent of the original manufacturer’s warranty to Seller.
The obligations under the foregoing warranties are limited to the precise terms thereof. These warranties provide exclusive remedies, expressly in lieu of all other remedies including claims for special or consequential damages. SELLER NEITHER MAKES NOR ASSUMES ANY OTHER WARRANTY WHATSOEVER, WHETHER EXPRESS, STATUTORY, OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY AND FITNESS, AND NO PERSON IS AUTHORIZED TO ASSUME FOR SELLER ANY OBLIGATION OR LIABILITY NOT STRICTLY IN ACCORDANCE WITH THE FOREGOING.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>