novopress ACO403 BT Press Gun ACO 403 with Batteries Instruction Manual
- June 6, 2024
- novopress
Table of Contents
- novopress ACO403 BT Press Gun ACO 403 with Batteries
- Proper use
- Basic safety information
- Definition of terms
- **Symbols on the device and information on the rating plate
- Scope of supply
- Technical Data
- Brief description of the press device (Fig.1)
- Start-up and operation
- Performing the pressing operation
- Re-pressing
- Cleaning, maintenance and repair
- Disposal
- Warranty and guarantee
- EU Declaration of Conformity
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
novopress ACO403 BT Press Gun ACO 403 with Batteries
Proper use
These press devices are intended exclusively for use with press collars manufactured by Novopress or approved by Novopress, in agreement with the system supplier. If you are uncertain whether your press device is suitable for the press tools described here, please contact the system supplier. The devices and the press collars are to be employed solely for the purpose of press-fitting pipes and fittings for which the relevant press collars have been designed. Any use beyond or different to that described here shall be regarded as improper use. All work with this tool that does not correspond to proper use may lead to damage to the press device, the accessories, or the pipes. Leaks and/or injury may result. Novopress is not liable for damage resulting from
- the use of unsuitable press tools or press tools from other manufacturers, or
- applications that do not conform to proper use.
Proper use also includes compliance with the Operating Manual, adherence to the inspection and maintenance conditions as well as compliance with the latest versions of all relevant safety regulations.
Basic safety information
The following pictograms are used to highlight sections of text. Please follow these instructions and act with particular caution in these cases. Pass all health and safety instructions on to other users and technicians.
WARNING!
This information indicates a potentially dangerous situation that could result
in serious injury or death.
CAUTION!
This information indicates a potentially dangerous situation that could result
in minor injuries and/or material damage.
Information!
This information is directly linked to the description of a function or an
operating sequence
Please read this Operating Manual carefully. The safety instructions it contains must be observed. Compliance with local safety regulations is essential!
WARNING!
Read all safety warnings and all instructions. Failure to follow the warnings
and instructions may result in electric shock, fire, and/or serious injury.
Therefore:
- Save all warnings and instructions for future reference.
WARNING!
Risk of injury from ejected fragments Incorrect or improper use or the use of
worn or damaged press tools and press devices carries a risk of injury from
ejected fragments.
Therefore:
- Press tools and press devices may only be used by trained personnel.
- Compliance with the specified service work and service intervals is mandatory.
- Check the press tools and press devices for cracks and other signs of wear before each use.
- Separate all press tools and press devices with material cracks or other signs of wear immediately and do not use them any longer.
- Only use press tools and press devices that are in perfect technical condition.
- Following incorrect use, do not use the press tool and press device any longer and have them inspected by an authorized workshop.
CAUTION!
Unskilled use may lead to a damaged or malfunctioning press tool and press
device.
Therefore:
- Replace worn press tools immediately and do not use them any longer.
- Use carrying cases for transport and storage and keep press tools and the press device in a dry room.
- Have damage checked immediately by an authorized workshop.
- Comply with the safety instructions regarding the cleaning and anti-corrosion agents used.
Information!
Consult the documents provided by the system supplier for handling and
assembly instructions for fittings and pipes.
Definition of terms
V | volt |
---|---|
A | ampere |
Hz | hertz |
W | watt |
kW | kilowatt |
g | gram |
kg | kilogram |
Bj | year of manufacture |
Ah | ampere hour |
--- | --- |
db(A) | decibel (sound pressure) |
bar | bar |
°C | degree Celsius |
kN | kilonewton |
a.c. / ∿ | Alternating current voltage |
d.c. / ⎓ | Direct current voltage |
F | force |
Ø | diameter |
--- | --- |
h | hour |
min | minute |
s | second |
m/s² | metre divided by square second (acceleration) |
Nr | number |
**Symbols on the device and information on the rating plate
**
Information on the rating plate
1 | Manufacturer logo incl. address |
---|---|
2 | Permissible uninterrupted operating time |
3 | Part No. |
4 | Serial No. |
5 | Year of manufacture mm.yy |
6 | Power consumption |
--- | --- |
7 | Country of manufacture |
8 | Permissible voltage range in volt |
9 | Nominal force |
10 | Type designation of device |
Explanation of possible pictograms on the rating plate
Scope of supply
The scope of supply for the standard equipment includes:
ACO403 BT
Press device incl. Operating Manual
Press collar in separate carrying case
Battery charger incl. operating manual
18 V d.c. rechargeable battery
Safety instructions
EU Declaration of Conformity
Carrying case
Other accessories may be included as additional extras. Please contact the system supplier for information on the supply scope.
Technical Data
Device: | ACO403 BT |
---|---|
Rated voltage: | 18 V d.c. / 5,0 Ah / M18 (Milwaukee) |
Nominal force: | 120 kN |
Height: | 320 mm |
Length: | 650 mm |
Width: | 102 mm |
Net weight: | 12,8 kg (without battery) |
Max. noise level: | 89 db(A) 1) |
Noise pressure level: | 78 db(A) 1) |
Vibration value: | <2,5 m/s² 2) |
Type of protection: | IP20 |
Temperature range during operation: | -10°C to +50°C |
Wireless technology: | Bluetooth® 4.0 smart |
Frequency range: | 2402,0 MHz to 2480,0 MHz |
Maximum output power: | 1,8 dBm |
- Measurement uncertainty 3 db(A) 2) Measurement uncertainty 1,5 m/s²
Information! Wear hearing protection.
The stated vibration emission value was measured using a standard test
procedure and can be employed for the purpose of comparison with another
device. The stated vibration emission value can also be used to obtain an
initial estimate of interruptions during intermittent operation.
Caution The vibration emission value may differ from the stated value
while the device is actually in use, depending upon the manner in which the
device is used. It may be necessary to establish safety measures to protect
operating personnel, depending on the actual conditions of use (intermittent
operation.
Battery charger
For information on the function and operation of the battery charger and the
rechargeable batteries, please consult the battery charger operating
instructions.
Brief description of the press device (Fig.1)
The ACO403 BT press device comprises the press device (8) and a fixed adaptor jaw (9). The head with the adaptor jaw can be rotated through 180°. It is only possible to adjust the head before the pressing operation. The head cannot be rotated while the pressing operation is being performed. To carry out the pressing operation, hold the device tightly on handle (6) and on grip (7) on the adaptor jaw.
Automatic press cycle (Fig. 1)
The device features an automatic press cycle. This ensures that the pressing
operation is completed every time. For safety reasons, the automatic press
cycle only switches on when the pressing operation has started, i.e. from when
the fitting starts to be shaped. After the automatic press cycle has been
triggered, the green LED (2) goes out and the pressing operation runs
automatically. The pressing operation can then only be stopped by pressing and
holding the release button (5). After the pressing operation has been
completed, the motor automatically switches off and the green LED (2) lights
up again. If the start button (1) is released before the automatic press cycle
begins, the piston in the press device moves back to its starting position.
The pressing operation will not have been completed. The pressing operation
must be triggered again.
Release button (Fig. 1)
In the event of an emergency, the pressing operation can be stopped at any
time by pressing and holding release button (5). The piston in the press
device then returns to its starting position.
Wireless interface (Fig. 1)
The device ACO403 BT features a wireless interface. This wireless interface
together with the NovoCheck app can be used to connect the ACO403 BT to a
mobile terminal device, such as a smartphone, tablet, etc. (Android, IOS;
further information at www.novopress.de). The
following functions are available:
- Reading the press device
- Checking the press device
- Making settings
- Generating construction site reports.
To establish a connection, the press device must be operational and the
NovoCheck app installed on the terminal device.
The press device can now be connected with the NovoCheck app. (For more
information, go to www.novopress.de) When the
connection to the wireless interface was established successfully, the blue
LED (10) lights up. The blue LED (10) goes out if no connection has been
established until the first pressing operation is performed. A new attempt to
establish the connection can only be made if the press device has been
switched off in the meantime. It may be necessary to pull the rechargeable
battery for this. If you do not wish to establish a connection, press the
start button (1) once again and perform the pressing operation.
Battery
Removing the battery (Fig. 2)
Push in (1) the two release buttons (A) and then remove the battery (2).
Inserting the battery (Fig. 3)
Slide the battery into the device as illustrated until it clicks into place.
Battery level indicator (Fig. 2)
The battery level is displayed when button (B) is pressed. The number of LEDs lit up denotes the battery level. If the battery is inserted in the press device when the battery level is checked, the last pressing operation must have been completed at least 1 minute prior to this. Otherwise, the display will be inaccurate.
Information!
The ACO403 BT press device must only be operated with 18 V rechargeable
lithium-ion batteries. The rechargeable 18 V battery must only be used in the
appropriate press devices.
Feature of the press device when the battery is empty:
When a pressing operation begins, a check is made to determine whether the
battery charge is sufficient to complete the pressing operation. If this is
not the case the device does not start. The green LED (2) flashes. If the
green LED (2) flashes following a pressing operation, the pressing operation
was carried out correctly. The battery must be recharged prior to the next
pressing operation.
Energy-saving mode
If the press device is not used over an extended period, the device switches off. No LED is lit up any longer. Press the start button (1) briefly to activate the press device.
Start-up and operation
CAUTION!
Fingers risk being crushed if the device is operated without a press collar.
(Fig. 4a) If the press device is not placed on a press collar, fingers could
be placed in the at-risk zone. Fingers could be crushed when the device is
started.
Therefore
-
* Do not operate the device without a press collar.
- Keep your fingers away from the at-risk zone
Information!
The ACO403 BT press device is not suitable for continuous operation. A short
break of at least 15 minutes must be taken after approx. 15-20 pressing
operations in succession in order to allow the device to cool down.
Meaning of the LED display
LED display | Status/cause | Measure |
---|---|---|
All LEDs off. | The device is switched off. | Briefly press the start button (1) |
(see section 7.5).
Green LED (2) lights up.| On standby.|
Green LED (2) off (pressing operation in progress).| Automatic press cycle is
on, device ends the pressing operation automatically.|
Green LED (2) flashes.| Insufficient battery charge.| Charge or replace
battery.
Blue LED (10) flashes.| The press device is ready for connection.| The press
device can be connected with the NovoCheck app.
Blue LED (10) lights up.| The press device is connected with the NovoCheck
app.|
Red LED (3) flashes.| Device not within the temperature range| The device only
works within a temperature range of -10 to +50°C. If the red LED is flashing,
the device must be moved into a warmer or colder area. The device can be
warmed to operating temperature by means of dry runs.
If the device is too warm it must be cooled down by means of a short break.
Red LED (3) lights up.| Device fault.| Press the start button (1). If this is unsuccessful, the press device is defective.
Send the device to a specialist workshop. NOTE!
The pressing operation may not have been completed; please check and repeat if required.
Red (3) and green (2) LED flash alternately.| Service interval reached.| Take
the press device to be serviced within the next 100 pressing operations.
Red (3) and green LED
(2) flash slowly and simultaneously.
| Service interval reached. It is possible to perform another 500 pressing
operations before the device is blocked automatically.| Take the press device
to the maintenance service within the next 500 pressing operations.
Red (3) and green LED
(2) flash quickly and simultaneously.
| The maximum number of pressing operations before maintenance is reached, the device is blocked.| Take the press device to the maintenance service.
Starting position of the press collar (Fig. 5)
Information!
To ensure that the press collars function correctly, the sliding segments must
be free to move. The sliding segments are always pressed back automatically
into the correct starting position by springs.
Therefore:
-
* Ensure that marking lines (M) on shells (C) are aligned with the edges of the sliding segments (D) when the press collar is in starting position. (Fig. 5)
- If this is not the case, have the press collars repaired.
The press collar must be positioned in the way shown (oval shape) before it is placed on a fitting. After the press collar has been opened and removed from a fitting, the press collar must again be positioned in the way shown.
Closing the collar (Fig. 6 + 7)
- By pressing the outer end of the shell gently, the shell moves inwards. (Fig. 6)
CAUTION!
There is a risk that the press collar will break if it is positioned
incorrectly If the press collar is positioned incorrectly on the fitting the
press collar may become damaged.
Therefore:
-
* Line (A) must be at the same height of the edge (B). (Fig. 7)
- If this is not the case, remove the pressed collar again and reposition it.
- The line (A) must be at the same height of the edge (B) when the press collar is positioned correctly. (Fig. 7)
Opening the collar (Fig. 8 + 9)
To open the press collar, the two outer shells can be swivelled away from the
pipe, either individually or together, as described below. First try the
process with one hand before you try it with both hands together. If you just
open one side first and then the other you must hold the press collar tightly
when opening the second side so that it does not fall off.
- To open the press collar the following 2 steps must be carried out. (Fig. 8)
- Press lever (15) at the position marked in the direction of the pipe and hold it there.
- Grip shell (16) at the position marked from below and swivel it away from the pipe.
- Once a certain point has been reached the shell automatically moves to its final position and makes an audible click.
- By pulling the two outer shells (16) apart, open the press collar far enough to remove it from the press fitting. (Fig. 9)
Performing the pressing operation
WARNING!
Risk of injury from ejected fragments Incorrect or improper use of press tools
or the use of worn or damaged press tools carries a risk of injury from
ejected fragments.
Therefore
- Ensure that the nominal size of the press fitting matches the nominal size of the press tool.
- Make sure that there is no dirt, swarf etc. between the press tool and the press fitting.
- Do not hold the press tool askew on the press fitting.
- If a press tool is used incorrectly, do not use it any longer and have it inspected by an authorised workshop.
WARNING!
Risk of injury from ejected fragments Material fatigue may cause the press
tool to break. Press tools are wearing parts. Frequent pressing will result in
material fatigue. Press tools worn or damaged in any other way as the result
of material fatigue can break even if used correctly; there is then a risk of
injury from ejected fragments. The risk of breakage is further increased in
the event of improper use.
Therefore
- Check the press tools for cracks and other signs of wear before each use.
- Discard all press tools with material cracks or other signs of wear immediately and do not use them any longer.
- Only use press tools that are in perfect technical condition.
CAUTION!
The danger of crushing! There is a risk of fingers and hands becoming crushed.
Therefore
- Keep all body parts and foreign objects away from the jaw levers on the fixed adaptor jaw when initiating the pressing operation.
- Do not hold the jaw levers of the adaptor jaw with your hands during the pressing operation. Use the grip (7). (Fig. 1)
CAUTION!
Keep fingers away from the at-risk zone when the adaptor jaw is closed. (Fig.
4a) Fingers may get crushed.
Therefore
- Leave the adaptor jaw open. (Fig. 4b)
Information!
To ensure that the press collars function correctly, the sliding segments must
be free to move. The sliding segments are always pressed back automatically
into the correct starting position by springs.
Therefore:
-
* Ensure that marking lines (M) on shells (C) are aligned with the edges of the sliding segments (D) when the press collar is in starting position. (Fig. 5)
- If this is not the case, have the press collars repaired.
- Check that the nominal size of the press-fitting matches the nominal size of the pressed collar.
Information!
Comply with the system supplier’s instructions when positioning the press
collar on the press fitting.
CAUTION!
There is a risk that the press collar will break if it is positioned
incorrectly If the press collar is positioned incorrectly on the fitting the
press collar may become damaged.
Therefore:
-
* The line (A) must be at the same height of the edge (B). (Fig. 7)
- If this is not the case, remove the press collar again and reposition it
- Swivel the outer shells of the press collar outwards as far as possible and then position the press collar on the press fitting.
- Close the press collar. (Fig. 6)
- Open the adaptor jaw fully using the grip (7).
WARNING!
Risk of injury from ejected fragments If the adaptor jaw is positioned on the
press collar incorrectly, the adaptor jaw and the press collar may break.
Therefore
-
* Claws (21) of the adaptor jaw must grip pins (22) of the press collar. (Fig. 11, 12 + 13)
- The markings (L/Fig. 11 + 12) on the claws (21) of the adaptor jaw must be fully covered when the press collar has been placed (Fig. 11+ 12)..
- Position the adaptor jaw on the press collar as follows.
- Push claws (21) of the adaptor jaw as far as possible into grooves (24) in the press collar. (Fig. 10)
- Close the adaptor jaw using the grip (7).
- Check that the markings (L/Fig. 12) on the claws (21) of the adaptor jaw are fully covered. If a marking is partially visible, the adaptor jaw must be placed again.
Information!
Ensure there is no gap between the segments of the press collar when the
pressing operation is complete.
Information!
If the press device is switched off, briefly press the start button (1) to
activate the press device. Press the start button again to start the pressing
operation.
- In order to start the process, press the start button and hold it down.
- After the pressing operation has been completed, open the adaptor jaw using the grip and detach it from the press collar.
- Open the press collar and remove it.
Re-pressing
Re-pressing is always necessary if a pressing operation has not been fully completed. Check the system supplier’s instructions to ascertain whether re- pressing is permitted. If it is permitted, please note the following differences from the initial pressing operation that must be complied with
WARNING!
Risk of injury from breaking claws on the adaptor jaw and damaged pins on the
press collar Holding the press device incorrectly and changing its position
before and during repressing means that the claws can no longer grip the pins
properly. The tips of the claws may press directly on the pins, causing the
claws and pins to break and fragments to fly off.
Therefore:
-
* After positioning on the press collar, press the jaw levers on the adaptor jaw together as far as possible using the grip. (Fig. 14)
- Make sure that the claws of the adaptor jaw always grip the press collar pins throughout the re-pressing operation and that the markings (L) on the claws (21) are fully covered. (Fig. 11 + 12)
Cleaning, maintenance and repair
CAUTION!
Risk of injury during cleaning or repair work from inadvertently pressing the
On switch.
Therefore:
- Comply with the safety instructions and always remove the battery before performing cleaning, servicing or repair work.
Service addresses
Novopress GmbH Pressen und Presswerkzeuge & Co. KG
Welserstr. 7
41468 Neuss
Germany
You can find the addresses of authorized specialist workshops by contacting
Novopress or at www.novopress.de.
Maintenance intervals
Maintenance and repair work may only be carried out by Novopress or the
authorised NOVOPRESS specialist workshops
Information!
The next maintenance service is indicated on the press device and the press
collar. This maintenance service must be performed at least once a year. If
more than 2,500 pressing operations are performed in a year, maintenance
service must be performed after these 2,500 operations at the latest. An LED
indicates that the maintenance interval has been reached 100 pressing
operations in advance (see Section 8.1 Meaning of the LED display). The red
and green LED flash slowly and simultaneously to indicate that 500 pressing
operations are left before reaching the maximum number of operations. The
device must be taken to maintenance service within these 500 pressing
operations, otherwise it automatically blocks itself after these 500 pressing
operations.
Regularly (before use, at the start of the working day) and when soiled (Fig. 13)
- Check the press collar, adaptor jaw and device for visible defects such as damage, material cracks and other signs of wear. If defects are found, do not use the device any more. Replace it or send it to an authorised specialist workshop.
- Clean the press collar with compressed air.
- Lubricate the adaptor jaw pins.
- Spray the entire adaptor jaw with lubricant.
- Remove deposits from pressing contour (K).
- Clean press collar pressing contour (K).
- Completely coat pressing contour (K) with lubricant.
- Check that sliding segments (D) and shells (C) can move freely.
Recommended lubricant:
- enclosed oil
- oil containing MoS2
After every 50 pressing operations:
- Spray joints (G) of the press collars with lubricant.
- Spray lubricant between sliding segments (D) and shells (C).
Six-monthly inspection
Have the device inspected by a qualified electrician or a Novopress workshop
Annually to get a 1-year extended warranty (see section 13 Warranty and
guarantee)
Have the press device inspected and serviced at Novopress or a Novopress
workshop?
Disposal
CAUTION!
Groundwater hazard The device contains hydraulic fluid. Hydraulic fluids pose
a hazard to groundwater. Uncontrolled drainage or improper disposal is
punishable by law.
Therefore:
-
* Dispose of the device in an environmentally responsible manner.
The instructions in the operating instructions provided with the battery charger must be complied with when disposing of rechargeable batteries and the charger. The product must not be disposed of as normal household waste. Novopress recommends that disposal be carried out by authorised specialist companies. Alternatively, waste devices can be returned directly to Novopress (or a specialist workshop) for proper disposal.
Warranty and guarantee
Novopress provides the full statutory warranty of 24 months for its press
devices and tools. The
warranty period always commences on the date of delivery, which must be proven
by means of the
sales documentation in case of doubt.
Within the warranty period, the warranty covers the repair of any damage or
malfunction of these tools
that is attributable to material or production faults.
The following are not covered by the warranty:
- Damage caused by improper use or inadequate servicing.
- Damage caused by the use of products not approved by Novopress for use with its press devices.
- Damage caused by pressing unsuitable pipes or fittings.
For wearing parts, the Novopress warranty lasts only for the intended service life. Service work and parts provided in response to warranty claims do not have to be paid for. However, all shipping costs shall be borne by the user. Claims can only be accepted if the device is delivered to Novopress or a Novopress workshop in an assembled state. Repair or replacement of the device under warranty does not result in an extension of the warranty period. Repair or replacement can only be performed using as-new components, the function of which corresponds to that of the old components. All parts that are faulty and consequently replaced are the property of the manufacturer.
Extended warranty on expiry of the statutory warranty period
We voluntarily provide the following extended warranties beyond the
statutory warranty period 1-year extended warranty for new press devices,
adaptor jaws and press collar Novopress provides an extended warranty of one
year for all new press devices, adaptor jaws and press collars that have been
serviced at least every 12 months by Novopress or by an authorised Novopress
workshop. There is a charge for this service work. For details of costs,
please contact Novopress or an authorised Novopress workshop.
Repair warranty
Novopress provides the following repair and spare parts warranty beyond the
warranty period for new products that undergo repair after this period:
- 6-month warranty for replaced spare parts.
- 12-month warranty for replaced assemblies.
- 12-month warranty for replaced press devices.
EU Declaration of Conformity
EU declaration of conformity and list of maintained directives and applicable standards see attached document EU declaration of conformity.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>