Mini Implant
Instruction Manual
Product description
Anthogyr Mini Implant is a single-component implant with an Optiloc®
prosthetic connection.
The Anthogyr Mini Implant system includes instruments and prosthetic and
surgical parts. Anthogyr Mini Implants are implants with a sandblasted BCP
bone anchoring surface and a DLC (a-C:H) coated connection. The Optiloc®
matrix system is a prefab connetor for the retention of removable restorations
on Anthogyr Mini Implants. The Optiloc® matrix is made of a housing and
interchangeable plastic (PEEK) retention inserts with different colourcoded
retention values or extraction forces. These instructions for use are valid
for the following dental implants:
– Anthogyr Mini Implants
Materials:
Titanium-6aluminium-4vanadium alloy:
Chemical components | Composition, % (mass/mass) |
---|---|
Aluminium | 5.50 to 6.50 |
Vanadium | 3.50 to 4.50 |
Other (Fe, O, C, N, H) | Total max 0.53 |
Titanium | Balance |
Intended use
Anthogyr Mini Implants are intended for the replacement of missing tooth roots and allows the stabilisation of removable full dentures.
Indications
Anthogyr Mini Implants are indicated for the stabilisation of complete
removable dentures in the mandible and/or maxilla with a total edentulousness.
They allow immediate loading in an appropriate clinical situation (good
primary stability).
Clinical benefits
- Stabilisation of removable prosthesis
- Restoration of masticatory function
- Aesthetic restoration
Patient type and intended user
Anthogyr Mini Implants are intended for use with partially or totally edentulous adult patients who do not present any of the conditions mentionned in the “contraindications”section. Anthogyr Mini Implants must be used by a surgeon trained in dental implantology.
Contraindications
Anthogyr Mini Implants are contraindicated in following cases:
- Implant divergence greater than 40° between two implants.
- D4 bones .
- Allergy or hypersensitivity to chemical components in the materials used and mentioned in the “Product description” section.
- Absolute contraindications: serious diseases (tumours, heart disease, etc.), metabolism disorders, uncompensated haematologic diseases, drug addition, alcoholism, psychosis, functional disorders, xerostomia, immune deficiency, leukocyte disorder, local or systemic treatments (steroid, anticoagulant, chemotherapy or radiation therapy, etc.).
- Relative contraindications: bruxism, occlusal stress, parafunction, unfavourable bone anatomy, pregnancy, growth not finished, insufficient oral hygiene, smoking lack of motivation or cooperation, irradiated bone, uncontrolled periodontal disease, oral infections or inflammations.
- Localised contraindications: Excessive resorption and/or insufficient bone quality, local radicular residues.
Warning
Implant surgery is a complex dental procedure.
Incorrect techniques can cause implant failure and/or loss of bone support.
Appropriate training and qualification and a good knowledge of surgical
techniques with Anthogyr products are required. Anthogyr offers specific
training.
Caution/precautions
Clinical use:
Single-use devices: do not reuse or re-sterilise. Risk of contamination and
risk of alteration of the functional surfaces. It is important to perform a
pre-clinical assessment and treatment plan that takes into account the
anatomical constraints of the future restoration.
Do not use an implant after the expiry date indicated on the packaging
Safety information regarding magnetic resonance imaging (MRI):
The safety and compatibility of Anthogyr devices that remain in the patient’s
body have not been evaluated in the magnetic resonance (MR) environment. They
have not been tested for heat build-up, migration or artefacts in MR
environments. The safety of Anthogyr devices in an MR environment is unknown.
Performing an MRI examination on a patient wearing such a device may result in
injuries.
Unwanted side effects
The clinical outcome of dental treatment is influenced by multiple factors. The following residual risks and possible side effects are related to the use of Anthogyr Mini Implant and may lead to additional dental treatment at the dental practice: swelling, haematoma, risk of swallowing/inhaling small parts during the procedure, phonetic problems, pain, hypersensitivity/ allergic reactions, inadequate function or device failure (mobility, loss of integrity), aesthetic problems, gingival injuries, damage to existing dentition, bone damage, loss of marginal bone, hyperplasia, exfoliation, dysesthesia, postsurgical paraesthesia, perforation of sinus or alveolar plates, systemic or local infection (including peri-implantitis, periodontitis, gingivitis, fistula), local or systemic infections including bacterial endocarditis.
Compatibility information
Anthogyr Mini Implants must be used in combination with the corresponding
Optiloc® system components and Optiloc® compatible overdentures.
Only use original parts with the Optiloc® connection to place Anthogyr Mini
Implants. If unsuitable ancillary equipment is used: risk of injury, damage
and malfunction of the implant, risk of damage to the ancillary equipment. For
more information, please refer to the manuals listed in the “Further
information” section.
Type of implant | Type of connection | Compatible components |
---|---|---|
Anthogyr Mini implant | Optiloc® | Optiloc® |
Compatible instruments
Optiloc® screwing instruments included in the kit.
Cleaning and disinfection
Anthogyr Mini Implants are supplied sterile (GAMMA sterilisation) and are intended for single use. Do not clean or sterilise the implants. Cleaning, disinfection and sterilisation can compromise the essential material and design features of the implants and result in device failure.
Sterilisation
Anthogyr dental implants are supplied sterile. Check that the entire packaging
of the device is undamaged before opening. Implant with a damaged packaging
must not be used. It is recommended to have a replacement implant readily
available for use. The intact blister pack protects the sterilised implant
against any external influence and, if stored properly, guarantees sterility
until the expiry date. The blister pack must not be opened before implant use.
When removing the implant from the sterile packaging, asepsis rules must be
followed.
Anthogyr declines all responsibility for resterilised implants, regardless of
who carried out the re-sterilisation or the method used. Under no
circumstances should a previously used or unsterile implant be implanted. If
the original packaging is damaged, Anthogyr will not accept the return of the
content.
Protocol for use
Refer to the brochures listed in the “Further information” section for
detailed step-by-step instructions.
Step 1 : Preoperative planning
The length of the implant, as well as the number of implants to be used and
their positioning, must be determined in advance, taking into account the
patient’s anatomy and the oral environment.
For this purpose, use the X-ray templates available for the range.
A sufficient number of implants must be used to provide support and distribute
the support loads:
→ A minimum of 4 mini implants Ø2.6 must be placed for mandibular
restorations.
→ A minimum de 6 mini implants Ø2.6 must be placed for maxillary restorations.
Step 2 : Implant bed preparation
→ Mark the site with a pointer drill.
→ Each site should be prepared using a progressive sequence of drilling
diameters in accordance the bone density. Make sure never to exceed the depth
of the planned drilling: presence of depth marks on each rotating instrument
or use of drills with a stop or a contraangle fitted with a stop system.
Please refer to the “step-by-step” guidelines listed in the “Further
information” section.
Avoid overheating the bone when drilling and tightening the implant to reduce
the risk of bone loss during the osteointegration phase. The risk of
overheating the bone can be reduced by using irrigation and by controlling the
torque.
Step 3 : Removing the implant from the packaging
The Anthogyr Mini Implant is packaged in a cardboard box with sterile
packaging consisting of a sealed blister pack and a capped tube.
They are delivered mounted in the cap which serves as an implant holder.
→ Remove the blister from the cardboard box outside of the sterile field.
→ Open the seal without touching the inside of the blister.
→ Let the capped tube gently fall on the sterile field.
Warning: Do not use an Anthogyr Mini Implant if it is detached from the
cap after opening the blister.
Warning: All handling should be done so as to avoid direct contact with
the exterior surface of the implant. When handling the implant, be very
careful not to drop it in the patient’s mouth. Make sure that the implant does
not touch the tube.
Step 4 : Inserting the implant
PRE-SCREWING THE IMPLANT :
Pre-screwing of the implant should be done manually using the packaging cap
until more torque is required.
PLACEMENT USING A CONTRA-ANGLE:
→ Adjust the output speed of the contra-angle to the recommended speed of 15
rpm.
→ Using the contra-angle and the Optiloc® adapter, screw the implant into the
channel to the desired depth. The implant must be placed so that the Optiloc®
connector is above the gum. Warning: Check the tightening torque frequently to
make sure it does not exceed 45 N.cm. Untighten and retighen to reduce the
screw pressure if needed. MANUAL PLACEMENT: Using the ratchet wrench
(available in the kit):
→ Assemble the Optiloc® key adapter and the surgical ratchet wrench
→ Screw the implant into the channel to the desired depth. The implant must be
placed so that the Optiloc® connector is above the gum.
Warning: When inserting the implant with the surgical ratchet wrench
control of the tightening torque is not possible. However, the verification of
the torque using a surgical dynamometric wrench to tighen the implant is
possible. A torque too high can damage the connection. Untighten and retighen
to reduce the screw pressure if needed.
Step 5 : Fabrication of the prosthesis
The prosthesis must be fabricated according to the prosthetic techniques of
removable restoration on mini implants. For immediate loading, a minimum
insertion torque of 35 N.cm is recommended. Abnormal loading of the implant
may result in implant fracture.
For more information on the prosthetic procedure, please refer to the “Further
Information” section.
Step 6 : Maintenance
MAINTENANCE BY THE CLINICIAN:
During periodic visits, please inspect the condition of each secondary part of
the denture retention system. Always use plastic instruments during the visit.
The proper seating of the prosthesis on the mucosa should be checked at least
once a year and, if necessary, the prosthesis should be relined to avoid
abnormal stress on the implant structure. MAINTENANCE BY THE PATIENT:
(please pass on the following information to the patient):
The patient’s attention should be drawn to the need for regular oral hygiene
as well as the need to thoroughly clean the Optiloc® connection of the Mini
Implants every day.
Healing phase
The healing period required for osseointegration varies considerably and
depends on the individual patient and treatment.
It is the sole responsibility of the surgeon to decide when the implant can be
loaded. If temporary components are used during the healing phase, they must
be placed in sub-occlusion.
Further information
For more information on the use of Anthogyr products, please contact your
local Anthogyr sales representative or contact Anthogyr customer service or
visit ifu.anthogyr.com and
www.anthogyr.com.
For more specific information on Anthogyr Mini Implants, please refer to:
- Anthogyr Mini Implants User Guide (MIO_NOT)
- Mini Implant System Catalog (C210)
- Mini Implant System Leaflet (C211)
Subject to the availability of the European Medical Device Database (EUDAMED), the summary of safety and clinical performance characteristics (SSCP) is available at https://ec.europa.eu/tools/eudamed.
Storage
Store these products in a clean, dry area, at room temperature. Improper storage may compromise the essential characteristics of the materials and design, which may lead to device failure.
Waste treatment
Waste resulting from the intervention (packaging, part extracted, etc.) must be handled as medical waste under the responsibility of the user.
Patient information
Patients must accept regular medical follow-ups and should consult their
doctor in the event of any unexpected change in the performance of the
prosthetic reconstitution. Patients must be informed of the need to ensure
regular oral hygiene.
Patient must be advised to remain cautious for the first few weeks after
surgery.
Traceability information is available to patients via the detachable labels on
the device.
Notes
The practitioner must have the necessary knowledge to practice dental
implantology and must be familiar with the handling instructions for Anthogyr
products as described in this document in order to use Anthogyr products
safely and in accordance with their instructions for use. Anthogyr products
must be used in accordance with the manufacturer’s instructions for use. The
dental surgeon is solely responsible for the proper use of Anthogyr products
in accordance with their instructions for use and to determine whether the
product is suitable for the individual patient’s situation.
Anthogyr products are part of a complete range and must be used in combination
with the corresponding original components and instruments distributed by
Anthogyr, its parent company and any affiliates or subsidiaries of the parent
company (“Straumann”), unless otherwise specified in these instructions for
use. The use of third-party products not distributed by Anthogyr voids any
warranty or other obligation, express or implied, of Anthogyr.
Any product-related issues must be reported to the local Anthogyr organisation
together with the product in question. In the event of a serious incident, the
user must file a report with the local Anthogyr organisation and the
appropriate competent authority in accordance with local regulations. Anthogyr
also offers an online complaint service in the countries concerned.
Validity
The publication of this document supersedes and replaces all previous versions. Anthogyr all rights reserved. Anthogyr® and/or other trademarks and logos of Anthogyr® mentioned herein are trademarks or registered trademarks of Anthogyr.
Availability
Some components of the Anthogyr implant system are unavailable in certain countries.
Symbols
The following table describes the symbols that may be printed on the packaging label. Please refer to the label on the packaging for the applicable product symbols.
Symbol
| Description of symbol|
Source of symbol
---|---|---
| Manufacturer| ISO 15223-1
| Date of manufacture| ISO 15223-1
| Catalog number| ISO 15223-1
| Batch code| ISO 15223-1
| Serial number| ISO 15223-1
| Consult the instructions for use or consult the digital instructions for
use| ISO 15223-1
| Medical Device| PR 15223-1 (2020)
| Unique identifier of the device| PR 15223-1 (2020)
| CE marking – compliance with current regulations| Directive 93/42/CEE —— MDR
(EU) 2017/745
| GOST-R logo for certification of the Russian Federation|
| FDA certification logo| 21 CFR 801.109(b) (1)
| Use-by date| ISO 15223-1
| Single sterile barrier system| ISO 7000
| Single sterile barrier system with protective packaging inside| ISO 7000
| Sterilized using irradiation| ISO 15223-1
| Do not resterilise| ISO 15223-1
| Non-sterile| ISO 15223-1
| Suitable for sterilisation in a steam steriliser (autoclave) at the
specified temperature (135°C)| ISO 7000 – 2868
| Non suitable for sterilisation in a steam steriliser (autoclave) at the
specified temperature (135°C)| Anthogyr
| Do not use if packaging is damaged or consult instructions for use| ISO
15223-1
| Keep away from sunlight| ISO 15223-1
| Do not reuse| ISO 15223-1
| Warning| ISO 15223-1
| Contains hazardous substances| ISO 7000
Anthogyr
2237 Av. André Lasquin
74700 Sallanches – France
Phone : +33(0)4 50 58 02 37
www.anthogyr.com
Email : contact@anthogyr.com
Validity Date : 2021-03
REF : 063IMPL-MIO_NOT