Anthogyr Xpert Unit Connected Implantology Motor User Manual
- June 6, 2024
- Anthogyr
Table of Contents
- DESCRIPTION OF SYMBOLS
- AREA OF USE
- INDICATIONS
- SAFETY
- DESCRIPTION
- Technical Specifications
- CONNECTIONS – INSTALLATION
- COMMISSIONING
- USE
- PARAMETER MANAGEMENT
- CLEANING-DISINFECTION-STERILISATION
- CONDITIONS OF USE AND STORAGE
- WARRANTY
- MANUFACTURER’S RESPONSIBILITIES
- REGULATIONS
- ANOMALIES AND FAULTS
- FEASIBLE TROUBLESHOOTING OPERATIONS BY THE USER
- AGENCY – TECHNICAL INSPECTION OF SAFETY
- ELECTROMAGNETIC COMPATIBILITY
- REFERENCE
- Support
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Anthogyr Xpert Unit Connected Implantology Motor User Manual
DESCRIPTION OF SYMBOLS
- Alternating current.
- Type B equipment
- Warning : Adhere to the operating instructions.
- Protective earth.
- Manufacturer.
- Class IIa device in compliance with the European directive 93/42CEE.
- Thermodisinfectable device.
- Sterilisable at a specific temperature.
- Do not use if the packaging is damaged
- Protect from light.
- Permissible temperature limits.
- Fragile
- Store in a dry place
- Maximum permitted stack load 50 kg.
- Handle with care.
- Device expiry date.
- Device manufacturing lot number.
- Series number.
- Humidity limit.
- Atmospheric pressure limit.
- Commercial reference number of the device.
- Risk of getting crushed – Do not bring fingers in contact with the rotating parts when the pump is on.
- Transport in standing position; with the top facing the direction of the arrow.
- Sterile device
- Do not reuse, single-use device.
- Do not throw in the domestic dustbin.
- Do not tilt.
- Protection against foreign bodies > 1 mm.
- Protection against strong water spurts in all directions.
- Follow the operating instructions.
- Do not sterilise.
- Refer the instruction manual/brochure.
AREA OF USE
XPERT UNIT®, the implantology and dental surgery motor is exclusively meant for placing implants and for dental or bone surgeries (for example : tooth extractions, apicoectomy, bone modelling or bone smoothing). With regard to implant surgery procedures, it helps carry out surgery tasks while helping the practitioner to carry out drilling, tapping, and screwing.
The use of this appliance is reserved for certified dental health professionals who are competent and qualified, or for those who work under their supervision.
The Anthogyr product is exclusively intended to be used in dental medicine. Any misappropriate use is prohibited.
INDICATIONS
You have just acquired your new dental implantology motor, XPERT UNIT®. Anthogyr congratulates you for your choice and thanks you for the confidence that you have placed in it.
In order to get complete and sustainable benefit from this product, it is recommended that you read this user manual first and perform the recommended inspection and maintenance tasks. This Anthogyr product is used exclusively with the accessories provided by Anthogyr (micromotor, pedal, power supply cable). The use of any other accessory is prohibited.
This equipment is compatible with most of the surgical contra-angles and handpieces available on the market. However, the technical features of this product are usable only if it is combined with an Anthogyr Mont-Blanc contra- angle. It is strongly recommended to use the Anthogyr irrigation lines. The use of compatible and adaptable irrigation lines can create malfunctioning for which Anthogyr cannot be held responsible in any way.
It is essential to take all the precautions mentioned with the symbols E in order to use the equipment in the best condition and with total safety.
The manufacturer reserves the right to make changes or improvements in the equipment arising from new technical developments. This does not give retrospective right toward the equipment already installed.
To ensure that the Anthogyr product is operational at all times and to maintain it in good working condition, adhere to the maintenance, disinfection, and sterilisation instructions (see paragraph “10 – Cleaning, disinfecting, and sterilising”).
Only the Anthogyr technicians or those trained by Anthogyr are authorised to repair of the MD (check your equipment control parameters after maintenance operations; it is possible that it has been set to default values).
Copying and distributing these instructions requires prior written consent of the Anthogyr company.
The user is under obligation to:
- Use only products that are free of defects.
- Protect patients, third parties, and oneself from any danger.
- Avoid any contamination to spread through the product.
The user is responsible for checking the operating safety and the condition of the equipment before each use.
While using the product, it is essential to adhere to the national legal provisions and particularly:
- Provisions applicable to connecting and starting the MD.
- Provisions applicable to safety at work.
- Provisions for prevention of accidents.
SAFETY
WARNING SI GN
****Warning sign
In order to avoid any physical or material damage, this document divides the safety instructions into 3 levels of danger.
ATTENTION
Indicates a hazardous situation that may cause material damage or from
minor to moderate injuries.
WARNING
Indicates a hazardous situation that may cause serious/fatal injuries.
DANGER
Indicates maximum danger created by a situation that may immediately cause
serious/fatal injuries.
SAFETY INSTRUCTIONS
External interference
The equipment is certified to comply with the standards of electromagnetic compatibility in force.
ATTENTION
The user shall however ensure that no electromagnetic disturbance creates
further risk.
To avoid disturbing the proper functioning of the electromagnetic equipment, it is recommended to adhere to the stipulated distances among devices (“Electromagnetic compatibility”).
WARNING
The equipment is not designed to withstand shocks from an electric
defibrillator.
Electrical connection
- DANGER
- ELECTRICAL SHOCK / ELECTROCUTION
- Non-compliant electrical connection
- Compulsorily check that the power supply voltage is identical to the one mentioned on the control box.
- Check the power supply cable before every use.
- Never insert or pull the plug from the wall socket with wet hands.
WARNING
To avoid any risk of an electric shock, this equipment must only be
connected to a power supply network equipped with a protective earth.
Environment
- Do not tilt the device.
- Ensure that the cords do not obstruct free movement of people.
- Do not use the device near ionising radiation.
- The device must be connected to a wall socket close by that is easily accessible. The required power cords include a plug that can bear at least 125% of the rated current of the device.
- Use a maximum impedance network of 4Ω.
DANGER
- Do not use this product in premises with explosion hazards.
- Do not expose the equipment to water splashes or water mist.
- Do not put metallic objects into the equipment to avoid any risk of electric shock, short-circuit, or emission of dangerous substances.
DANGER
ELECTROCUTION
Accidental penetration of liquid in the box.
Before every use, check and ensure impermeability of the recipients of
liquids. In case any liquid penetrates the equipment, do not touch it and
immediately disconnect the power supply cable from the electricity network.
Make sure that the surface of the equipment is completely dry before
reconnecting the power supply cable.
Use of the device
ATTENTION
- Do not move the device while in use.
- We strongly recommend to limit the torque so as not to exceed the recommendations of the manufacturers for the instrumentation used.
- This is to limit the risk of injury and deterioration of instruments.
WARNING
Once the servicing work is completed, switch off the device. Unplug it from the electrical mains in the event of prolonged non-use
Storage conditions
- Recommended conditions : keep protected from moisture and at room temperature. Keep away from sunlight, sparks and other sources of ignition.
- Temperature : from -20 °C to +70 °C.
- Air humidity level : from 5% to 95%.
- Incompatible materials : acids and other solvents.
Peristaltic pump
ATTENTION
- Do not operate the pump if the irrigation line is blocked or closed by a clamp.
- Place the fluid to be pumped always above with respect to the pump, in order to ensure optimum operation of the latter.
- Always keep the rollers of the pump head and all the moving parts clean and free of contamination and debris.
ATTENTION
- Do not open the guard when the pump is rotating.
- Do not put your fingers in contact with the rotating parts when the pump is in operation.
DANGER
Never attempt any repair alone.
Micromotor and micrometer cable
ATTENTION
Refer to the usage instructions supplied with the micromotor and/or micromotor cable to familiarise yourself with all safety and usage aspects.
Hygiene and maintenance
DANGER
Sterilisable products are supplied in non-sterile condition and are required to be decontaminated and sterilised before the first use as well as immediately after each use.
ATTENTION
Do not use any spray directly on the MD to clean it; preferable to use wipes.
ATTENTION
Refer to the usage instructions supplied with the micromotor and/or micrometer cable to familiarise yourself with all maintenance aspects.
Malfunction
In the event of abnormal operation (flickering screen, noisy pump motor,
vibrating micromotor):
- Immediately stop using the device.
- Try to locate or eliminate the problem with the help of the descriptive document (paragraph ” Anomalies and faults”).
- If it is not possible to locate or eliminate the problem with the help of the descriptive document, switch off the device and call the Anthogyr After-Sales Service.
DANGER
Never attempt any repair alone.
DESCRIPTION
N°
|
Name
---|---
A
|
Wireless pedal
B
|
Cabinet
C
|
Irrigation line
D
|
Charging adapte
E
|
Power cord
F
|
Micromotor cord
G
|
Micromotor
H
|
Bracke
I
|
Micromotor protective casing
J
|
Front face of protection
K
|
Clips for irrigation line
L
|
Micromotor support
Rear face
Front face
N°
|
Name
---|---
1
|
Peristaltic pump
2
|
ON/OFF Button (I/O)
3
|
Power socket 220V / 110V
4
|
Touch screen
5
|
Connector for micromotor cable
Technical Specifications
Technical Specifications
Dimensions| Cabinet (w x d x h)| 243 x 222 x 102 mm
Pedal (l x d x h)| 280 x 170 x 70 mm
Micromotor (Ø x l)| Ø23 x 115 mm
Micromotor cable (l)| 2 m
Weight| Cabinet| 2.5 kg
Pedal| 1 kg
Micromotor| 110 g
Micromotor cable| 110 g
Electrical| Input power supply| from 100 to 240 V (50-60 Hz)
Power consumption| 200 to 300W
Input fuse| Ø5 x 20 – T5A 250 V
Classification| Cabinet| IP4X
Pedal| IPX6
Electrical classification| Class| I according to IEC 60601-1
Type| B
Classification Bluetooth| Group| 1
Class| B
FCC ID| QOQBGM111
IC| 5123A-BGM111
Micromotor| Rotation speed| 100 – 40000 rpm
Motor connection| Model with light : TYPE 1-short Model without light : TYPE
4-short
Peristaltic pump| Maximum flow rate at 100% with Anthogyr line| 120 ml/min
Pump pipe Øext| Ø7 mm
Pump pipe Øint| Ø4 mm
CE classification| Directive 93/42/CEE| Class IIa conforming to rule 8 for
cabinet / pedal / micromotor cable / micromotor class I conforming to rule 1
for the micromotor mount / clips
Display| N.cm| Torque (Newton-centimetre)
rpm| Speed (revolutions per minute)
CONNECTIONS – INSTALLATION
It is imperative to familiarise yourself with all the safety instructions
before installation and use of the device (see paragraph “2 – Safety”).
Refer to the usage instructions supplied with the micromotor and/or
micrometer cable to familiarise yourself with all safety, maintenance and
usage aspects.
MAINS CONNECTION
**Use only the power cable supplied by the manufacturer.
Do not forget to verify that the supply voltage is exactly same as the one
mentioned on the control cabinet and the power switch is in the Position
0**.
CONNECTION / DISCONNECTION OF THE MICROMOTOR
Align the red dot of the micromotor socket to the red dot of the connector on the front face of the cabinet and clipper:
To disconnect, remove the safety mechanism in the following way:
Pull the mobile part of the connection in the direction of arrow.
****Do not disconnect the micromotor cable from the console directly by
pulling the cable!
CONNECTION / DISCONNECTION OF A CONTRA-ANGLE ON THE MICROMOTOR
Engage the contra-angle on the motor nose.
Rotate the contra-angle on the motor nose until it is positioned correctly.
On the contra-angles opening, a small ‘snap’ confirms that the positioning is
correct.
ATTENTION
The LED micromotor (Ref XP311L) is not compatible with LED contra-angles.
POSITIONING OF THE BRACKET AND THE BAG
Orient the bracket plate with the plate at the rear of the cabinet.
The positioned bracket must not turn by itself.
Suspend the bag of physiological liquid on the bracket.
PLACEMENT OF THE IRRIGATION LINE IN THE PERISTALTIC PUMP
Part to be inserted in the pump
ATTENTION
- Do not operate the pump, if the irrigation line is blocked or closed by a clamp.
- Place the fluid to be pumped always above with respect to the pump, in order to ensure optimum operation of the latter.
- Keep the rollers of the pump head and all the moving parts always clean and free of contamination and debris.
- Do not open the guard when the pump is rotating.
PLACEMENT OF IRRIGATION CLIPS
Place the irrigation clips on the micrometer cable at regular intervals
(between 15-20 cm).
Position the irrigation line along the micrometer cable by inserting it into
the clips.
COMMISSIONING
POWER-UP
Set the power switch to position I
At startup, all the pictograms on the screen light up and the device emits a beep.
- The code Er02002 at startup means there is an anomaly with the date and time. The date and time synchronise automatically with the mobile app when you are connected.
WIRELESS PEDAL CONNECTION
At the 1st use
At unit startup, the pictogram display means that the pedal is on standby.
Press on the pedal lever to quit the standby mode.
- As soon as the pictogram lights up, the pedal is ready to be used.
- If the pictogram remains lit up, proceed to pair the pedal with the unit.
When using daily:
At unit startup, after a few seconds, the pictogram goes out and the
pictogram lights up .
If the pictogram remains lit after the unit startup, press on the pedal lever to quit the standby mode.
- As soon as the pictogram lights up, the pedal is ready to be used.
- If the pictogram remains lit up, proceed to pair the pedal with the unit.
How to pair the pedal with the unit
-
Press on the pedal lever.
-
Press the 3 green buttons on the screen at the same time.
-
The logo flashes until the pairing is complete (time to achieve pairing = about 30 sec)
-
Press the 2 end buttons on the pedal.
-
A beep at the console confirms the pairing : the logo disappears and the logo is displayed.
- If after all these actions, the pictogram remains lit, see the section “Anomalies and errors”
How to place the pedal in extended standby.
If you don’t plan to use your motor for an extended period of time (> 30 days), or you wish to transport it by air, placing the pedal in extended standby mode is recommended.
To do so, just press on the 3 pedal buttons simultaneously.
The persistent display of the symbol on the screen confirms that the pedal is on standby.
To leave standby mode, just press on the pedal lever.
USE
DEFINITION OF THE DISPLAY
- the coloured square indicates the current program
Pump
0%
|
25%
|
50%
|
75%
|
100%
| | | |
Battery
0%
|
25%
|
50%
|
75%
|
100%
| | | | __
PARAMETERS OF PROGRAMS (FACTORY SETTING)
To select a program on the screen, press on the pictogram of the selected program. For navigation to the pedal between the implantology sequence programs.
Speed (rpm)
| 1200| 20| 15| 40 000| 2000| 40| 30|
15 000
---|---|---|---|---|---|---|---|---
Torque (N.cm)
| 80| 80| 80| 80|
80
|
80
| 80|
80
Direction of rotation
|
right
|
right
|
right
|
right
|
left
|
left
|
left
|
left
Pump
| 100%| 100%| 0%| 100%| 0%| 0%| 0%| 0%
Display of the torque
|
NO
| YES| YES| NO| NO| YES| YES|
NO
Display of the speed
|
YES
| YES| YES| YES| YES| YES| YES|
YES
Beep
|
/
| /| /| /| YES| YES| YES| YES
At the startup, the program “drilling” is active. The program parameters can be changed by following the instructions given below.
- The range of setting from 5 to 80 N-cm. These values are valid only for the contra-angles of the MontBlanc
® Implant series (10400X-XL / 16400X-XL). With these contra-angles, the accuracy of torque is ± 10% for the range from 5 to 80 N.cm.
SETTING THE VALUES OF SPEED / TORQUE / FLOW RATE
Setting the rotation speed and the maximum torque
- Pressing on the parameter to be changed (speed or torque).
- The parameter starts flashing and the buttons +/- appear.
- Adjust the value of the parameter with the buttons + or –.
- Press on the parameter to validate. The parameter stops flashing.
- The parameter automatically stops flashing after 5 seconds and the set value is saved.
NOTE 1 : Value of the parameter is saved permanently for the program.
NOTE 2 : You cannot operate the pedal during the setting of the
parameters.
Setting the irrigation flow rate
For setting the irrigation flow rate, press on the pictogram .
Each press increases the flow rate by 25% (display of an additional drop).
At 100%, an additional press makes the irrigation flow rate 0%.
To cut off the irrigation and reset to a given flow rate, use the ON / OFF
function of the pedal.
DISPLAY OF THE LAST TORQUE REACHED
The XPERT UNIT® motor allows you to validate the tightening torques to be saved. These torques will be exported with the mobile app XPERT UNIT® and used on the web portal.
Displaying and saving the tightening torques
-
In tightening mode, (button lit up) when the motor has stopped, the torque displayed is the maximum torque programmed (Screen A).
-
When the motor is rotating, the torque displayed on the screen is the instantaneous torque (Screen B).
-
When the motor stops, the last torque remains displayed for 30 seconds, before the display returns to the maximum torque programmed (Screen A).
- To re-display the last torque, long press on the torque shown on the set point screen (Screen A).
-
The last torque obtained is displayed with the active patient No. (Screen C).
-
When long pressing on the value of the torque, the value of the torque is saved (a confirmation beep is heard).
-
To return to the setpoint screen without saving the torque value with the patient number press .
Patient numbering
To make it easier to identify the saved torques, you can enter the patient
number corresponding to the saved torques. This information will appear in the
data exported with the mobile app.
To edit the patient number, press on the PXX area. This area will start
flashing and the + and – buttons will appear.
Once adjusted, press on the PXX area to validate or wait 3 seconds without
doing anything.
To return to the set point screen without saving the torque value with the patient number, press
You may also edit the patient number after the set point screen (Screen A) by
pressing if this is already lit.
To return to the set point screen without saving the torque value with the patient number, press .
IMPORTANT : EACH TIME THE PEDAL IS PRESSED, THE LAST TORQUE OBTAINED IS
ERASED.
IMPORTANT : EACH TIME THE EQUIPMENT IS TURNED ON, THE PATIENT NUMBER GOES
BACK TO P01.
USE OF THE PEDAL
Description of pressing different buttons and lever.
Pressing lever (D) | Variation in the rotation speed of the micromotor |
---|---|
Pushing lever (D) | Illumination of the light for 3 seconds |
Short pressing on the left button (A) | Irrigation ON / OFF (the flow rate set |
on the screen is preserved)
Long pressing on the left button (A)| Startup of the pump at the maximum flow
rate without the micromotor rotating (rinsing function)
Short pressing on the middle button (B)| Activation / deactivation of the
reverse mode
Long pressing on the middle button (B)| Torque reduction in 5 increments
Short pressing on the right button (C)| Moving from one sequence to the next
Long pressing on the right button (C)| Torque increase increased in 5
increments
BACKUP POWER FOR THE WIRELESS PEDAL
__ This pictogram symbolizes the charge level of the battery of the pedal.
__ When the battery charge level has reached 25%, at least one day of use of the pedal will remain. The device beeps and the pictogram flashes, indicating the need to recharge soon (see instructions below).
__When charge level of the battery reaches 0%, the device will beep every 30 seconds and it is not possible to use the motor. Recharge according to the following instructions.
RECHARGING THE PEDAL BATTERY
Connect the charging adapter to the pedal through the connector located at the rear of the pedal. Connect the adapter to the mains to power it.
__On screen, this logo is animated to indicate that the charging is in progress.
__When this logo is displayed, the battery is fully charged.
- The motor can be used in a normal manner during the charging of the pedal battery.
STOPPING THE MOTOR
At the end of the dental procedure :
- Set the switch to position 0.
- Remove the tool fixed to the contra-angle.
- Remove the bottle or the bag of physiological liquid from the bracket.
- Remove the drill of the irrigation line from the bottle or the bag of physiological liquid.
- Remove the clips for irrigation lines.
- Disconnect the contra-angle from the micromotor.
- Lay the micromotor on the micromotor support.
MEDICAL DEVICE LIFECYCLE
If used in a proper manner, the micromotor and its cable have a lifecycle
corresponding to 250 cleaning and sterilisation cycles; the pedal has a
lifecycle corresponding to 5 years of use.
However, these indications are not a warranty because wear may appear
prematurely, depending on how the device is maintained (cleaning and
sterilisation).
PARAMETER MANAGEMENT
ACTIVATION / DEACTIVATION OF AUDIO WARNINGS
List of audio warnings
Description
|
Can be deactivated
---|---
Weak battery|
NO
Operating in reverse|
NO
Programmed maximum torque reached|
YES
Validation of sending data to a connected device|
YES
Pairing of the pedal to the console|
NO
Complete integrity at the startup of the console|
NO
CLEANING-DISINFECTION-STERILISATION
DANGER
Sterilisable products are supplied in non-sterile condition and are required to be decontaminated and sterilised before the first use as well as immediately after each use.
Use disinfectant products in accordance with the instructions given by their
manufacturers.
The MDs that can be sterilised can withstand a temperature between 134 °C and
137 °C for 3 or 18 minutes following the laws of the country in force.
Remove the irrigation line and throw it in a safety container meant for soiled
medical items (single-use irrigation line).
ATTENTION
Do not use any spray directly on the MD to clean it; preferable to use wipes.
ATTENTION
Refer to additional instructions on the micromotor and micromotor cable for more information.
OVERVIEW OF MAINTENANCE
Single-use
|
The rmodisinfectable ****
|
Disinfectable by wipes and nonsterilisable
|
Disinfectable by wipes and sterilisable (noimmersion)
|
Disinfectable by immersion(ultrasound) and sterilisable
---|---|---|---|---
- Bags of physiological liquid
- Irrigation lines
- Screen Protection
- Micromotor protection casing
|
- **** Micromotor
- Contra-angle
|
- **** Cabinet
- Pedal
| ****
- **** Micromotor cable
- Micromotor
|
- Bracket
- Motor support
- Clips for fixing the irrigation line
- Contra-angles of the MontBlanc®
CLEANING-DISINFECTION-STERILISATION CYCLE
Only for the contra-angles
CONDITIONS OF USE AND STORAGE
Conditions of use and operation
Temperature
|
+10°C to +40°C (50°F to 104°F)
Relative humidity
|
15 to 80 %
Atmospheric pressure
|
70 to 106 kPa
Conditions of use and storage
Temperature
|
-20°C to +70°C (-4°F to 158°F)
Relative humidity
|
5 to 95 %
Atmospheric pressure|
70 to 106 kPa
WARRANTY
WARRANTY 24 months
- This device is guaranteed for parts and labour against any manufacturing defect for a period of 24 months from the date of invoice.
- This warranty does not apply to wearing parts.
- Any modification or addition to the product without the express consent of the company Anthogyr will nullify this warranty.
- The warranty will lapse in case of non-observance of the technical instructions provided with all our devices.
- Anthogyr cannot be held responsible for damages and their consequences resulting or possibly resulting from normal wear and tear, incorrect use, cleaning or maintenance, non-observance of the instructions regarding use or connection, scaling or corrosion, impurities in the water supply system, or chemical or electrical influences that are unusual or non-compliant with the user manual, maintenance and assembly instructions of Anthogyr and other instructions of manufacturers.
- The transport expenses for return of goods for repair at Anthogyr are to be borne by the customer, even if the repair is done under warranty.
- The warranty covers shipping costs for return of the items to the customer.
- In order to consider warranty claims, please keep a copy of the invoice or a copy of the delivery note along with the device.
- The MDs must be returned in cleaned and sterilised condition in a closed bag with proof of passage to the autoclave, the authentic coloured indicator, with a letter stating the reason for return.
- The renewal of spare parts is assured for 7 years after the commercial stoppage of the product.
MANUFACTURER’S RESPONSIBILITIES
The present instruction does not substitute instructions of peripheral products.
Anthogyr reserves the right to modify the device and/or the user manual without prior notice.
The following points, if not followed, will annul the responsibility of Anthogyr :
- All the recommendations of Anthogyr must be respected during installation (voltage supply, electromagnetic environment, etc.)
- The device must be used in an electrical installation conforming to the regulations in force.
- The use of this device must be restricted to the operation for which it has been designed.
- No modification, extension and setting as an afterthought, should be made by a third party or any other person.
- Opening and repairs can be carried out only by persons trained by Anthogyr or persons who have been authorised.
- Only original spare parts of Anthogyr must be used.
- The product must be used in accordance with the assembly, service and maintenance instructions.
- All the instructions contained in this document must be followed.
REGULATIONS
According to the standards ISO11498 and ISO7494, the device comes under the category of dental treatment devices.
This MD is classified as IIa according to European directive 93/42/EC.
This equipment is manufactured in conformity with the currently applicable
standard pursuant to IEC60601-1.
This equipment has been designed and manufactured according to an ISO 13485 quality assurance system.
As an electrical and electronic equipment, disposal of the product must be carried out as per a specialized collection, removal, and recycling or destruction chain (particularly in the European market, with reference to the directive no. 2002/96/CE dated 23/01/2003).
When the product reaches its end of life, we recommend that you contact your dental equipment reseller, in order to know how to proceed.
The box, wireless pedal, micromotor, and the micromotor cable include components that must be treated as non residential waste.
ATTENTION
It is prohibited to throw the pedal into a domestic bin.
Put the pedal in a special container meant for the treatment of e-waste.
ANOMALIES AND FAULTS
This part is intended to help the user in case of problems.
In the event of abnormal operation (flickering screen, noisy pump motor, vibrating micromotor), immediately stop using the device and call the Anthogyr After-Sales Service. If it is not possible to locate or remove a problem with the help of this descriptive document, or if the malfunction signals still remain active, switch off the device and call the Anthogyr After-Sales Service. Don’t attempt to disassemble the control cabinet, the motor or the control pedal.
DANGER
Never attempt to carry out repairs alone.
ERROR CODES AND MEANINGS
It is possible to remove the error message from the screen by pressing on it. In the event of a major problem, the cabinet restarts. If the error message reappears, please contact the Anthogyr After-Sales Service.
Error code
|
Description of the fault
| Possible solutions
---|---|---
Er 02 001|
- Torques as well as the system parameters (setting, error file) are not saved any longer
- Check the different parameters of the sequences (speed, torque, flow rate)
| The assembly can still be used.
Contact the Anthogyr After-Sales Service for more information.
Er 02 002|
- The date and time are incorrect
- The date and time can be updated using the mobile application.
- Attention : if the date and time are not correct, the retrieval of information using the mobile applica- tion will be inaccurate.
Er 05 001|
- The motor is still usable, but the pump runs the risk of not operating.
Er 02 004|
- The screen encounters operating problems
- The micromotor/pedal/micromotor assembly can still be used safely with the pedal to complete the ongoing care.
| If the problem per- sists, do not use the motor / pedal / micro- motor any longer.
Contact the Anthogyr After-Sales Service for more infor- mation.
Er 02 005|
- The touch bar encounters operating problems
- The micromotor/pedal/micromotor assembly can still be used safely with the pedal to complete the ongoing care.
Er 02 006
Er 02 007
|
- Update the cabinet software using the mobile application
Er 04 006|
- Overheating of the micromotor.
- Let the micromotor be idle for 30 minutes to 1 hour so that it cools.
Error code | Description of the fault | Possible solutions |
---|---|---|
Er 02 005 – Er 02 006 | The electronics encounters a hardware or software | |
failure | Do not use the micro- **motor/pedal/ micromotor any longer |
Return to the Antho- gyr After-Sales Ser- vice**
is mandatory
Er 02 007 – Er 04 001| The electronics encounters a hardware or software
failure
Er 04 002 – Er 04 003| The electronics encounters a hardware or software
failure
Er 04 004 – Er 04 005 Er 04 006 – Er 04 007| The electronics encounters a
hardware or software failure
Er 04 008 – Er 04 009| The electronics encounters a hardware or software
failure
Er 04 010 – Er 04 011| The electronics encounters a hardware or software
failure
Er 04 012 – Er 05 001| The electronics encounters a hardware or software
failure
If it is not possible to locate or solve the problem with the help of different descriptive documents, switch off the device and call the Anthogyr After-Sales Service.
Depending on the error code, it is also possible to use the mobile application to send information to the Anthogyr After-Sales Service.
MALFUNCTION
Malfunction
|
Probable causes
|
Possible solutions
---|---|---
**** The micromotor does not rotate|
- The micromotor is not connected to the cabinet
- The micromotor is jammed
- The pedal is not detected and no longer allows to interact with the cabinet.
|
- Ascertain that the cable is connected to the cabinet
- Mount a contra-angle with a tool of large diameter and try to rotate it
- See pedal problems below.
The pedal is not detected by the cabinet|
- The pedal is not (or longer) paired with the cabinet / the pedal is of
- The battery is completely discharged
|
- Proceed to the pairing of the pedal (see section 7.2.3)
- Turn on the pedal (see section 7.2.1)
- Put the pedal for recharging
The battery logo indicates that recharg- ing is not taking place (no animation)|
- The battery does not get charged
- The charger is defective or assembled incur- rectly
|
- **** Contact the After-Sales Service
After putting the pedal battery for charg- ing, the battery logo is always flashing and the level has remained at 25%|
- **** The battery does not get charged
|
- Contact the After-Sales Service
No spray or low flow rate, or leakage problem|
- The flow rate of the pump is 0% or too low
- The irrigation line has not been placed correctly
- The irrigation line is defective or has been pinched
|
- Adjust the flow rate of the pump on the screen
- Check the positioning of the line in the pump.
- Check the mounting direction of the line in the pump
- Change the irrigation line
- Use the Anthogyr irrigation line
**** The tool of the contra-angle does not turn|
- Bad connection of the tool in the contra-angle
- The contra-angle has been damaged
- The motor is jammed
|
- Check the tool and its placement
- Change of contra-angle
- Check the integrity of the motor, if the cable is con- nected properly to the cabinet
The screen does not light up|
- The power to the device has been switched off
- The fuses have blown
- The power socket is defective
- The differential circuit breaker of the electrical installation is OFF
- The power cord is defective
- The electronic card and/or the screen are HS
|
- Ascertain that the cabinet is attached properly to the mains and/or that the (I/O) button is on position I
- Replace the fuses (see §16)
- Change the wall socket
- Reset the differential circuit breaker of the electrical installation
- Replace the power cord
- Contact the After-Sales Service
The use of “compatible” or “adaptable” lines may generate malfunctions. Prefer the use of Anthogyr lines. (Ref. 11052-10 or 11052-5)
If it is not possible to locate or eliminate the problem with the help of above different descriptive documents, switch off the device and call the Anthogyr After-Sales Service.
Anthogyr After-Sales Service
Tel. : +33 (0)4 50 58 50 53
Fax. : +33 (0)4 50 93 78 60
contact@anthogyr.com
FEASIBLE TROUBLESHOOTING OPERATIONS BY THE USER
Checking and replacement of fuses.
- Switch off the cabinet with the ON / OFF button (I/O).
- Unplug the power cord from the wall socket.
- Unplug the power cord from the cabinet.
- Remove the fuse box located between the (I/O) switch and the power socket with a flat screw driver.
- Take out the 2 fuses.
- Check the status of the fuses and, if necessary, change them (SPT Ø5×20-T5A-250 V with a breaking capacity of 1500A)
- Put back the 2 fuses in the box.
- Keep the box in place again
DANGER
Don’t attempt to disassemble the control cabinet, the motor or the control
pedal.
AGENCY – TECHNICAL INSPECTION OF SAFETY
The XPERT UNIT® motor for implantology and dental surgery must be subjected to technical inspection by a regular agency at least once in 3 years (unless the legislations of countries in force are different).
The technical inspection of safety must be carried out by an agency authorised by Anthogyr.
- Visual inspection of all the external parts
- Measure the leakage current of the MD (cabinet + pedal)
- Measurement of the leakage current at the patient side
- Measurement of electrostatic discharge (ESD)
- Inspection of the motor operation
- Inspection of the pedal operation
- Inspection of the pump operation
- Inspection of the screen operation
- Internal visual inspection, if external damages, signs of overheating, abnormal noise or failures are observed.
Anthogyr After-Sales Service
Tel : +33 (0)4 50 58 50 53
Fax : +33 (0)4 50 93 78 60
contact@anthogyr.com
ELECTROMAGNETIC COMPATIBILITY
- This product requires special precautions to be taken regarding electromagnetic compatibility. It must be installed and commissioned according to the instructions given in the paragraph “Connections/installation”.
- Certain types of mobile telecommunication devices, such as mobile phones, are likely to interfere with this product. Therefore, it is not recommended to use them during the use of the MD.
- The separation distances recommended in this paragraph must therefore be followed.
- This product must not be used near or on another device. If this cannot be avoided, it is necessary to check prior to use its proper operation under the conditions of use.
- The use of accessories other than those specified or sold by Anthogyr as replacement parts, can result in increased radiation or decreased immunity of this product.
Directives and declaration of the manufacturer – Electromagnetic radiations
The XPERT UNIT® motor is intended to be used in the environment specified below.
It is advisable for customer or user of the XPERT UNIT® motor to ensure that it is used properly in such an environment.
Radiation test| References of the standards|
Compliance
| | Speaker por t : Group 1 Class B 10 m
| |
- 30MHz – 230MHz = 30dBµV/m
| |
- 230MHz – 1GHz = 37dBµV/m
| | Input power : 240Vac / 50Hz
RF radiations| CISPR 11 Group1 Class B| Limits : Group 1 Class B 0.15MHz
to 0.5MHz
| |
- 66 to 56dBµV QP / 56 to 46dBµV AV
| | 0.5MHz to 5MHz
| |
- 56dBµV QP / 46dBµV AV
| | 0.15MHz to 0.5MHz
| |
- 60dBµV QP / 50dBµV AV
Harmonic radiations| IEC 61000-3-2| 220Vac 50Hz access : Limit Class A
Voltage fluctuation / flickers| IEC 61000-3-3| 230Vac 50Hz access : PST <1|
PLT > 0.65
The motor XPERT UNIT® uses RF power only for its internal functions.
Therefore, its RF radiations are very low and are not likely to cause any
interference in nearby electronic equipment.
XPERT UNIT® motor is suitable for use in all premises, including domestic
premises and those directly connected to the low-voltage public electricity
supply network feeding the buildings for domestic use.
Directives and declaration of the manufacturer – Electromagnetic immunity
The XPERT UNIT® motor is intended to be used in the environment specified
below.
It is advisable for customer or user of the XPERT UNIT® motor to ensure that
it is used properly in such an environment
Radiation test
|
References of the standards
|
Compliance
| Electromagnetic environment – directives
Off-peak period of power, short interruptions and voltage fluctuations in
electrical power supply input lines| IEC 61000-4-11| AC power
supply : 240Vac @50Hz
- 0% UΤ : 0.5 cycle at 0 °, 45 °, 90 °, 135 °, 180 °, 225 °, 270 °, 315 ° (criterion A)
- 0% UT : 1 cycle – Single phase at 0 ° (criterion A)
- 70% UT : 25 cycles – Single phase at 0 ° (criterion A)
- 0% UT : 250 cycles (criterion C)
| It is advisable that quality of the electricity supply network is that of a
typical commercial or hospital environment.
If user of the XPERT UNIT® motor requires continu- ous operation during
interruptions in the electricity supply network, it is recommended to power
the
XPERT UNIT® motor from an uninterrupted power supply or a battery.
Conducted RF disturbances| IEC 61000-4-6| Effective 3V from 150kHz to 80 MHz
(criterion A)| It is advisable that portable and mobile RF communi- cation
devices should not be used near any part of the XPERT UNIT® motor, including
cables, than the rec- ommended separation distance calculated from the
equation applicable to the frequency of the transmit- ter. Recommended
separation distance : d = 1.2 √P
Electrostatic discharges| IEC 61000-4-2| ±8kV at the contact
±2kV, ±4kV, ±8kV, ±15kV in the air (criterion B)
| It is advisable that the floor is made of wood, concrete or ceramic tiles.
If floors are covered with synthetic materials, then the relative humidity
should be at least 30%.
Fast transient bursts| IEC 61000-4-4| AC power supply : ±2kV (100Hz) / 240Vac
@50Hz DC power supply : ±2kV (100Hz)
Line input / output : ±1kV (100hz) (CriterionB)
| It is advisable that quality of the electricity supply network is that of a
typical commercial or hospital environment.
Magnetic field at the electrical power supply network fre- quency (50/60 Hz)|
IEC 61000-4-8| Level : 30A/m (50Hz) (Criterion A)| It is advisable that the
magnetic fields at the electri- cal power supply network frequency have
character- istic levels of a representative location in a typical commercial
or hospital environment.
Radiated RF disturbances| IEC 61000-4-3| 3V/m from 80MHz to 2.7GHz (Criterion
A)| Recommended separation distance : w d=0.4√P from 80 MHz to 800 MHz w
d=0.7√P from 800 MHz to 2.5 GHz
where P is the characteristic of maximum output power of the transmitter in
watts (W), according to the manufacturer of the transmitter and d is the
recommended separation distance in metres (m). It is advisable that the field
intensities of fixed RF transmitters, determined by an electromagnetic
investigation on site a, are below the level of compli- ance in each range of
frequencies b.
Interferences may occur close to the device markedwith the following symbol :
H
Surge voltage| IEC 61000-4-5|
- AC power supply : 240Vac @50Hz
- ±0.5kV, ±1kV, ±2kV between phase and earth
- ±0.5kV, ±1kV, between phases DC power supply :
- ±0.5kV, ±1kV, ±2kV between phase and earth
- ±0.5kV, ±1kV, between phases
Line input / outpu t : ±2kV (criterion B)
| It is advisable that quality of the electricity supply network is that of a
typical commercial or hospital environment.
Recommended separation distances between portable and mobile RF communication
devices and the XPERT UNIT® motor
XPERT UNIT® motor is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. Customer or user of the XPERT UNIT® motor can contribute to prevent electromagnetic interferences by maintaining a minimum distance between the portable and mobile RF communication devices (transmitters) and the XPERT UNIT® motor as recommended below, depending on the maximum transmission power.
Rated maximum output power of the transmitter in W
|
Separation distance depending on the transmitter frequency (m)
From 150kHz to 80MHz d= 1.2√P|
From 80MHz to 800MHz d= 0.4√P
| From 800MHz to 2.7GHz d= 0.7√P
0.01| 0.12| 0.04| 0.07
0.1| 0.38| 0.13| 0.22
1| 1.20| 0.40| 0.70
10| 3.80| 1.30| 2.20
100| 12.00| 4.00| 7.00
For transmitters for which the rated maximum output power has not been given above, the recommended separation distance “d” in metres (m) can be estimated by using the equation applicable to the transmitter frequency, where ‘P’ is the characteristic of maximum transmission power of the transmitter in watts (W), according to the manufacturer.
NOTE 1 : UT is the AC voltage of the network before application of the
test level.
NOTE 2 : At 80 MHz and at 800 MHz, the higher frequency range applies.
NOTE 3 : These directives may not apply in all situations.
Electromagnetic propagation is affected by the absorption and the reflection
by structures, objects and persons.
- a. The field intensities of fixed transmitters, such as the base stations for the radio phones (cellular / wireless) and land mobile radios, amateur radio, AM and FM broadcasting and TV broadcasting, cannot be provided theoretically with accuracy. To evaluate the electromagnetic environment due to fixed RF transmitters, it is advisable to consider an electromagnetic investigation on site. If intensity of the field, measured at the location where the XPERT UNIT ® motor is used, exceeds the applicable RF compliance level as above, it is advisable to monitor the XPERT UNIT ® motor to verify that its operation is normal. If abnormal performances are observed, then additional measures may be necessary, such as reorientation or relocation of the XPERT UNIT ® motor.
- b. Over the frequency range from 150 kHz to 80 MHz, it is advisable that field strengths are less than 3 V/m.
REFERENCE
Reference
|
Description
---|---
XP000
|
XPERT UNIT motor without light
XP000L
|
XPERT UNIT motor with light
XP000L-P
|
XPERT UNIT motor with light
XP010
|
XPERT UNIT motor without light + 10400X contra-angle
XP010L
|
XPERT UNIT motor with light + 10400X contra-angle
XP020
|
XPERT UNIT motor without light + 16400X contra-angle
XP020L
|
XPERT UNIT motor with light + 16400X contra-angle
XP411
|
Motor support
XP412
|
Clips for hanging of the irrigation line (x 10)
XP311L
|
LED Micromotor
XP311
|
Non-LED Micromotor
XP312
|
Micromotor cable
XP313
|
Micromotor XPERT UNIT non LED + cable
XP313L
|
Micromotor XPERT UNIT LED + cable
XP216
|
AC/DC charging adapter 8W-5V
(INPUT :100-240VAC;0.4A;50-60Hz / OUTPUT : 5.0VDC;1.6A)
XP415
|
Bracket
11053
|
Screen protection film
11054
|
Micromotor protection casing
11052
|
Irrigation line for bag or bottle
16400XL
|
Implantology contra-angle light XS
16400X
|
Implantology contra-angle XS
Support
Anthogyr
2 237, Avenue André Lasquin
74700 Sallanches – France
Tél. +33 (0)4 50 58 02 37
Fax +33 (0)4 50 93 78 60
www.anthogyr.com
Photos credits : Anthogyr
All rights reserved
Not contractual photos