Respshop E122 Cylinder Flow Meter or Concentrator User Manual
- June 6, 2024
- Respshop
Table of Contents
- Respshop E122 Cylinder Flow Meter or Concentrator
- Troubleshooting Guide
- Check for blockages
- Cannot reach target flow
- Cannot reach target temperature
- Check operating conditions
- Check tube
- Condensation
- O2 too low
- O2 too high”Fig. 21
- Exxx
- APPENDIX A: TROUBLESHOOTING
- Support
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Respshop E122 Cylinder Flow Meter or Concentrator
Troubleshooting Guide
This Troubleshooting Guide is intended for all users, including
clinical/biomedical engineers and technical personnel, of the myAIRVO™ 2
humidifier. It applies to all myAIRVO 2 humidifiers from lot numbers 140910
and above. Refer to the myAIRVO Product Technical Manual and the myAIRVO 2
User Manual for additonal informations and detailed instructions of use.
If this troubleshooting guide does not resolve your issue, please contact your
local Fisher & Paykel Healthcare representative.
myAIRVO does not turn onA.
-
Press and hold the ON/OFF button for at least 2 seconds.
B. Is the myAIRVO 2 plugged into mains power?
C. Is the power cord securely inserted into the back of the myAIRVO 2?
D. Is the power cord damaged?- If yes, replace the damaged cord. See Section 5.1 of the myAIRVO Product Technical Manual for a 900PT410xx replacement power cord.
E. Connect the myAIRVO 2 into another power outlet.
F. Connect a different electrical device into the same power outlet. Turn on the device to confirm that the power outlet is working.
G. The myAIRVO 2 may be ‘on’ with a broken display.
Turn the myAIRVO 2 on without the heated breathing tube and check that the audible alarm activates.
- If yes, replace the damaged cord. See Section 5.1 of the myAIRVO Product Technical Manual for a 900PT410xx replacement power cord.
Power out (black screen)
The auditory alarm will sound for at least 120 seconds. The most likely
cause is a dislodged or disconnected power cord. A. Please follow the
instructions in Section 1.
Note: Press “audio pause” button to permanently silence the alarm ( The
device will not automatically restart.
Check water
FIG1
A. Is the water bag empty?
If yes, refill or replace the water bag and press the “mode” button ( reset
the alarm.
B. Is the water chamber empty?
-
* For HC360: Ensure the water level is below the indicated black line.
-
For MR290: If yes, replace the water chamber as it may be damaged. Contact your local Fisher &Paykel Healthcare representative about the faulty chamber.
Warning: The heater-plate and base of the water chamber may be hot.
C. For MR290: -
Open the vent cap near the water bag spike. This allows the pressure to equalize, letting the water flow into the water chamber.
-
Ensure that there are no kinks in the fluid line, preventing water from flowing into the chamber.
-
Check for leaks
Fig. 2 or “E122”
The most likely cause is a missing water chamber or the existing chamber has
not been pushed into place correctly.
WATER CHAMBER
A. Is the water chamber fitted correctly? Even if it appears to be:
- Remove the water chamber.
Push the chamber on firmly, until the finger guard “clicks” into placeFig. 3. Warning: The heater-plate and base of the water chamber may be hot. HEATED BREATHING TUBE
-
A. Is the heated breathing tube attached to the device correctly?Even if it appears to be:
-
Disconnect the heated breathing tube.
-
Check that the black O-ring is in placeFig.4.
If the O-ring is damaged or missing, replace with part 900PT408. -
-
Reconnect the heated breathing tube.
B. Confirm that the heated breathing tube is not visibly damaged.
PATIENT INTERFACE
A. Is the patient interface correctly fitted to the heated breathing tube?-
Even if it appears to be, disconnect and reconnect the patient interface. It should make a “click” sound when it is connected properly.
B. The unit may be in Junior mode, used an adult interface. -
Press and hold the “mode“ button ) for 5 seconds to change ( between Junior mode and Default mode.
Junior mode can be disabled in the advanced menu, see page 10.
Note: If the myAIRVO is in Junior mode and the 900PT500/501/500E Default tube is used with the OPT842/44/46/70 or RT013 interfaces, it may generate a “Check for leaks” alarm.
AIR FILTER & FILTER COVER
A. Is the air filter and filter cover (at the back of the device) correctly fitted, as per the User Manual?
-
-
Check for blockages
” Fig. 5 or “E121”
WATER CHAMBER AND NON-RETURN VALVE
A. Have the silicone flaps of the non-return valve, found inside the left-hand
chamber port, been displacedFig. 6?
-
* If yes, return them to the correct position using a non-sharp tool, such as a pair of non-sharp tweezersFig. 7.
Note: If the Non-return valve is damaged or missing, replace with part 900PT911. Upon replacement, ensure the spine is sitting verticallyFig. 7. If placed horizontally, this may cause the bottom flap to open due to gravityFig. 6b. This may cause both “Check for leaks” and “Check for blockages” warnings.
B. Is the water chamber overfilled above the black line?-
For HC360: Ensure the water level is below the indicated black line.
-
For MR290: If yes, replace the water chamber as it may be damaged. Contact your local Fisher &Paykel Healthcare representative about the faulty chamber.
HEATED BREATHING TUBE
A. Is the heated breathing tube visibly blocked or kinkedFig. 8?
PATIENT INTERFACE AND myAIRVO MODE
A. Is the patient interface visibly blocked or kinkedFig. 9?
B. The unit may be in Default mode with a junior cannula. -
Press and hold the “mode” button ) for 5 seconds to change ( between Default mode and Junior mode.
Note: If the myAIRVO is in Default mode and the 900PT531 Junior tube is used with the OPT316 and OPT318 cannula interfaces it may generate a “Check for blockages” alarm.
AIR FILTER
A. Is the air filter significantly discolored/dirty? -
Replace with part 900PT913.
Note: A promptFig. 10 for filter change will occur once the myAIRVO 2 has counted 1,000 hours of use. Choose “Now” or “Later”Fig. 11 by using the “up” or “down” buttons and press the “mode” button () to confirm.
Selecting “Now” will zero the counter. Selecting “Later” will activate the prompt at the start of next use.
B. Is there a foreign object blocking the air filter or filter holder?
CONDENSATION
Please see Section 10.
ALTITUDE
A. The myAIRVO 2 is designed to operate at an altitude below 2,000 meters.
-
Cannot reach target flow
Fig. 12A.
- Press the “mode” button ( ) to continue normal operation at a lower (maximum
achievable) flow rate.
B. Is the target flow setting too high for the patient interface?
• Check the swing tag/User Manual for the appropriate flow range for each patient interface.
Note: If the myAIRVO 2 cannot reach the target flow setting, it will automatically select a maximum achievable flow rate and prompt the user to press the “mode” button ( ) to confirm.
C. Follow steps in Section 5 — “check for blockages”.
D. Is the altitude above 2,000 m?
The myAIRVO 2 is designed to operate at an altitude below 2,000 meters.
Cannot reach target temperature
The most likely cause is operating the myAIRVO 2 at a high flow rate in a cold
room. Consider decreasing the target flow setting.
A. Press “mode” button ( ) to continue.
Note: The humidity level may be compromised.
B. Is the ambient room temperature below 18 °C (64 °F)?
-
* If yes, proactive management of condensation may be required. See Section 10 on prevention and management of condensation.
Check operating conditions
This alarm may be caused by a sudden change in ambient room temperature, e.g.
storing the unit in a cold place, then using it in a warm place.
A. Is the ambient room temperature less than 10 °C (50 °F) or greater than 30
°C (86 °F)?
B. Leave the unit running for 30 minutes.
Switch the unit off, then restart.
Check tube
Fig. 15 or “E38
A. Is the heated breathing tube attached correctly?
-
* Even if it appears to be, unplug and reconnect the heated breathing tube.
B. Is the heated breathing tube visibly damaged?
- Check the electrical pins and the tube itself.
C. Try using a new heated breathing tube.
- Check the electrical pins and the tube itself.
Condensation
PREVENTION OF EXCESSIVE CONDENSATION
A. Is the myAIRVO 2 being used in ambient conditions between 18 – 28 °C (64 –
82 °F)?
-
* If the room is less than 18 °C (64 °F), condensation is more likely to occur.
B. Is there a local source of cooling acting on the heated breathing tube?
-
A fan to cool the patient,
-
An air-conditioning unit, vent or an open window?
-
Are you able to remove or minimize these sources of cooling, e.g. redirect the fan, cooling the patient, away from the heated breathing tube?
CONDENSATION MANAGEMENT
A. Implement a system to check the heated breathing tube for condensate regularly.
B. Is the myAIRVO 2 placed below head heightFig. 16? -
This will allow condensate to drain towards the water chamber, away from the patient.
C. If condensation is present, drain it back into the water chamberFig. 17: -
Disconnect the patient interface from the heated breathing tube.
-
Drain the tube by lifting the patient end of the tube, allowing the condensate to run into the water chamber.
-
At higher target flow rates, it may be necessary to first reduce the target flow rate to 30 L/min or below, to ensure the condensate drains into the water chamber.
D. If condensate persists, consider turning the target temperature down. -
A lower target temperature will decrease the humidity output of the myAIRVO 2, decreasing the level of condensation.
Note: The temperature and humidity level delivered to the patient will also be reduced.
-
O2 too low
The measured oxygen level has fallen below the allowed limit.
Ensure the oxygen source matches that of the oxygen setting – see “Oxygen
Input Settings” on page 11.
A. Adjust the level of oxygen from the oxygen source as necessary, i.e.
increase the oxygen flow rate through the oxygen flow meter.
B. Is the oxygen source (wall/cylinder flow meter or concentrator) turned on?
C. Is the oxygen source empty or faulty?
D. Is the “AIRVO 2 oxygen inlet kit”Fig. 19 installed correctly, as per the
instructions included with part 900PT422 and confirmed that there are no kinks
in the “AIRVO 2 oxygen inlet kit” oxygen tubing?
E. Is the oxygen source tubing correctly and securely fitted to the myAIRVO 2?
F. Is the minimum oxygen limit set to 25%?
-
* A prompt will appear with an option to change this lower limit to 21 %. Select “Yes” or “No” by using the “Up“ and “Down” buttons. Press the
“mode” button ( ) to confirm selectionFig. 20.
See Section 2 – Advanced Settings in the myAIRVO Product Technical Manual to change this lower oxygen limit.
G. Ensure the oxygen source matches the oxygen input setting.- For oxygen concentrators: The “Oxygen Input Setting” should be 95 %.
- For 100 % oxygen sources: The “Oxygen Input Setting” should be 100 %. See page 11 for details on “Oxygen Input Settings”
H. Allow the device to sufficiently warm up; rapid changes in temperature can affect the sensor.
O2 too high”Fig. 21
The measured oxygen level has risen above the allowed limit.
Ensure the oxygen source matches that of the oxygen setting – see “Oxygen
Input Settings” on page 11.
A. Adjust the level of oxygen from the oxygen source as necessary, i.e.
decrease the oxygen flow rate through the oxygen flow meter.
See Section 2 – Advanced Settings in the myAIRVO Product Technical to change
this lower oxygen limit.
B. Ensure the oxygen source matches the oxygen input setting.
-
* * For oxygen concentrators: The “Oxygen Input Setting” should be 95 %.
- For 100 % oxygen sources: The “Oxygen Input Setting” should be 100 %. See page 11 for details on “Oxygen Input Settings”
Exxx
A. Follow the instructions in Appendix A if a fault with an error code is displayed on the myAIRVO screen.
APPENDIX A: TROUBLESHOOTING
The following pages provide troubleshooting advice for fault/error / “E” codes that may appear during use of the myAIRVO 2.
Support
For more information please contact your local Fisher & Paykel Healthcare representative
Manufacturer
Fisher & Paykel Healthcare Ltd 15 Maurice Paykel Place
East Tamaki, Auckland 2013
PO Box 14 348, Panmure Auckland 1741
New Zealand
Tel: +64 9 574 0100
Fax: +64 9 574 0158
Email: info@fphcare.com Web:
www.fphcare.com
-
Australia
Fisher & Paykel Healthcare Pty Limited 36-40 New Street,
PO Box 167
Ringwood, Melbourne
Victoria 3134, Australia
Tel: +61 3 9879 5022 Fax: +61 3 9879 5232 -
Austria
Tel: 0800 29 31 23 Fax: 0800 29 31 22
Benelux
Tel: +31 40 216 3555 Fax: +31 40 216 3554 -
China
Tel: +86 20 3205 3486 Fax: +86 20 3205 2132 -
France
Tel: +33 1 6446 5201 Fax: +33 1 6446 5221 -
Germany
Tel: +49 7181 98599 0 Fax: +49 7181 98599 66 -
India
Tel: +91 80 4284 4000 Fax: +91 80 4123 6044 -
Irish Republic
Tel: 1800 409 011 -
Italy
Tel: +39 06 7839 2939 Fax: +39 06 7814 7709 -
Spain
Tel: +34 902 013 346
Fax: +34 902 013 379 -
Sweden
Tel: +46 8 564 76 680
Fax: +46 8 36 63 10 -
Switzerland
Tel: 0800 83 47 63
Fax: 0800 83 47 54 -
Taiwan
Tel: +886 2 8751 1739
Fax: +886 2 8751 5625 -
Turkey
Fisher Paykel Sağlık Ürünleri Ticaret Limited Şirketi,
Alinteri Bulvari 1161/1 Sokak No. 12-14, P.O. Box 06371 Ostim,Ankara, Turkey
Tel: +90 312 354 34 12
Fax: +90 312 354 31 01 -
UK
Fisher & Paykel Healthcare Ltd
Unit 16, Cordwallis Park
Clivemont Road, Maidenhead Berkshire SL6 7BU, UK
Tel: +44 1628 626 136
Fax: +44 1628 626 146 -
USA/Canada
Tel: +1 800 446 3908 or +1 949 453 4000
Fax: +1 949 453 4001
REF 185048390 REV A © 2014-11 Fisher & Paykel Healthcare Limited
www.fphcare.com
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>