Drager 4544122 Polytron 5000
| WARNING
Strictly follow the Instructions for Use.
The user must fully understand and strictly observe the instructions. Use the
product only for the purposes specified in the Intended use section of this
document.
---|---
For your safety
General safety statements
- Before using this instrument, carefully read the Instructions for Use (IfU).
- Strictly follow the Instructions for Use. The user must fully understand and strictly observe the instructions. Use the instrument only for purposes and under the conditions specified in this document.
- Comply with all local and national laws, rules and regulations associated with this instrument.
- Only trained and competent personnel are permitted to inspect, repair and service the product as detailed in these Instructions for Use. Further maintenance work that is not detailed in these Instructions for Use must only be carried out by Dräger or personnel qualified by Dräger. Dräger recommends a Dräger service contract for all maintenance activities.
- Use only genuine Dräger spare parts and accessories, otherwise the proper functioning of the instrument may be impaired.
- The flameproof/explosion proof joints are not in accordance with the relevant minimum or maximum values of EN/IEC 6009-1. The joints are not intended to be reworked by the user.
- Do not dispose of the Instructions for Use. Ensure that they are retained and appropriately used by the instrument user.
- The measuring function of the gas detection transmitter for explosion protection, according to Annex II, clauses 1.5.5, 1.5.6 and 1.5.7 of Directive 94/9/EC is currently not covered.
- Substitution of components may impair Intrinsic Safety. Only if intrinsic safety is involved
Safe connection of electrical devices.
- Never connect this instrument to another electrical device before consulting the manufacturer or an expert.
Using the product in areas subject to explosion hazards:
- Instruments or components for use in explosion-hazard areas which have been tested and approved according to national, European or international Explosion Protection Regulations may only be used under the conditions specified in the approval and with consideration of the relevant legal regulations.
- The instruments or components may not be modified in any manner. The use of faulty or incomplete parts is forbidden.
The appropriate regulations must be observed at all times when carrying out repairs on these instruments or components.
Definition of alert icons
To reduce the risk of ignition of a flammable or explosive atmosphere, strictly adhere to the following Caution and Warning statements. The following alert icons are used in this document to provide and highlight areas of the associated text that require a greater awareness by the user. A definition of the meaning of each icon is as follows:
| DANGER
Indicates an imminently hazardous situation which, if not avoided, will result
in death or serious injury.
---|---
| WARNING
Indicates a potentially hazardous situation which, if not avoided, could
result in death or serious injury.
| CAUTION
Indicates a potentially hazardous situation which, if not avoided, could
result in physical injury, or damage to the product or environment. It may
also be used to alert against unsafe practices.
| NOTICE
Indicates additional information on how to use the instrument
Description
Product overview
The instrument is powered by 16 to 30 VDC. Gas concentrations, error messages,
and software menu choices are displayed on a 3 digit LCD display.
Access to the software is obtained by tapping the magnetic wand (part number
4544101) on the glass viewport at the appropriate indicators. In this way, the
instrument can be configured, calibrated and maintained non-intrusively, so
declassification of the area is not necessary for these procedures. The
Polytron 5000 is housed in either an epoxy powder-coated aluminum or stainless
steel 316 enclosure with 3/4″ NPT threaded ports located at the 3 and 9
o’clock positions.
Intended use
The Dräger Polytron 5000 is a transmitter for the detection of carbon
monoxide, hydrogen sulphide, oxygen, hydrogen, ammonia, chlorine, sulfur
dioxide and nitrogen dioxide in ambient air. The instrument is designed to be
installed in permanent locations, and depending on the version, it is approved
for use in certain specified hazardous, classified areas (see Section 6.1 on
Page 12). Polytron 5000 gas detectors are available in two versions, those
used for hazardous areas and others for general purpose only (see Section 6.1
on Page 12).
Caution: Not tested in oxygen enriched atmospheres (>21 % O2). High off-scale
readings may indicate an explosive concentration.
Hazardous area units
Polytron 5000, explosion-proof instrument for the monitoring of CO, H2S and
O2.
Non-hazardous area units
Polytron 5000, general purpose instrument for the monitoring of NH3, Cl2, SO2,
NO2, NO, EO and H2O2 (see Section 6 on Page 12, and Section 8 on Page 13 for
details).
Intended operating area and operating conditions
Hazardous areas classified by zones:
The device is intended to be used only in hazardous areas classified zone 1 or
zone 2, within a temperature range as marked on the device, where gases of
explosion groups IIA, IIB or IIC and temperature class T4 or T6 (depending on
the maximum ambient temperature) or dusts of groups IIIA, IIIB or IIIC may be
present.
Hazardous areas classified by divisions:
The device is intended to be used only in hazardous areas Class I&II, Div. 1
or Div. 2, within a temperature range as marked on the device, where gases or
dusts of groups A, B, C, D or E, F, G and temperature class T4 or T6
(depending on the maximum ambient temperature) may be present.
Operation
Principle of operation
Dräger electrochemical sensors are electrochemical measuring transducers for measuring the partial pressure of gases under atmospheric conditions. The ambient air being monitored diffuses through a membrane into the liquid electrolyte in the sensor. The electrolyte contains a measuring electrode, a counter-electrode and a reference electrode. An electronic potentiostatic circuit ensures a constant electrical voltage between the measuring electrode and reference electrode. Voltage, electrolyte and electrode material are selected to suit the gas being monitored so that it is transformed electrochemically on the measuring electrode and a current flows through the sensor. This current is proportional to the gas concentration. At the same time, oxygen from the ambient air reacts at the counter-electrode electrochemically. The current flowing through the sensor is amplified electronically, digitized and corrected for several parameters (e.g. the ambient temperature). The resulting measured value is given as an analog, 4-20 mA signal.
Design principle
- Meter
- Potentiostat
- Direct current supply
- Measured gas
- Membrane
- Measuring electrode
- Electrolyte
- Reference electrode
- Counter electrode
- Temperature sensor
\Installation
- The flameproof/explosion proof enclosure provides three ¾” NPT openings, which can be used for field wiring, direct attachment of a sensor or wiring of a remote sensor.
- Secondary circuit intended to be supplied from an isolating source (N/A for relay circuits).
- The optional increased safety terminal box provides four 20 mm openings, which can be used for field wiring or wiring of a remote sensor. The permissible cable diameter range is 7 to 12 mm.
- When installed at locations exceeding ambient temperatures of 55 °C, use only appropriate wiring, specified for at least 25 °C above the maximum ambient temperature.
- Strip wire insulation by 5-7 mm.
- Connect the wires as indicated in wiring figure (also showing grounding conductor terminal).
Mount the instrument at an appropriate height for the gas to be detected,
taking into account the density of the gas, air flow patterns in the room,
personnel, and other environmental considerations. For most toxic gases, where
personnel protection is the main goal, mounting the Polytron 5000 at an
average breathing height of about 5 feet (1.5 meter) is common practice.
Responsibility for correct placement of the Polytron 5000 rests exclusively
with the end user; if in doubt about placement, consult with Dräger
application engineers. Polytron 5000 is a two-wire instrument powered by 16 to
30 V DC. Wiring terminals are located on the bottom of the enclosure; pull out
the bezel by grasping the notches on either side of the display with your
fingers and pulling up.
1 Grasp notches with fingers and pull up.
- Connect the wires as shown below.
Secondary circuit intended to be supplied from an isolating source.
Tightening torque
** Part| TQ Lb. In.| TQ Nm**
---|---|---
lid| min. 266| min. 30
sensors| min. 266| min. 30
plugs| min. 266| min. 30
conduit hubs| min. 443| min. 50
Tightening torque and wire size for field wiring terminals
Electronic
TQ Lb. In.| Wire Size AWG| Wire Size mm 2
All field wiring terminals 4.4 – 7.0
(0.5 – 0.8 Nm)| 24 – 12| 0.2 – 2.5
- In case of optional increased safety terminal box, securely screw the transmitter onto the e-box using 4 screws with a tightening torque of 8 Nm.
Installing the sensor
The Polytron 5000 sensor connector is keyed and can only be installed one way. Use only DrägerSensors in this instrument. To install the sensor, see the following figure for details:
- Sensor housing
- Sensor
- Set screw
- Sensor cover
1. Loosen the set screw in the stainless steel sensor cover.
2. Unscrew the stainless steel sensor cover at the end of the sensor housing.
3. Insert the ribbon cable connector into the socket in the DrägerSensor®1.
4. Push the DrägerSensor into the sensor housing.
5. Screw the stainless steel sensor cover back on.
6. Tighten the set screw to secure the sensor cover.
7. Allow the sensor to warm-up according to the sensor data sheets.
- Always test a newly-installed sensor with target gas to verify proper operation.
- When installed at locations exceeding ambient temperatures of 55 °C, use only appropriate wiring specified for at least 25 °C above the maximum ambient temperature.
1) DrägerSensor® is a registered trademark from Dräger
Replacing the sensor
- Loosen the set screw in the sensor cover.
- Unscrew the stainless steel sensor cover at the end of the sensor housing.
- Pull the ribbon cable out of the sensor connector and remove the old sensor.
- Insert the ribbon cable connector into the socket of the replacement DrägerSensor
- Push the DrägerSensor into the sensor housing.
- Screw the stainless steel sensor cover back on.
- Tighten the set screw to secure the sensor cover.
- Always test a newly-installed sensor with target gas to verify proper operation.
Menu
In the standard operating mode, the gas concentration of the target gas will
be displayed. To access the software menu, tap the magnetic wand once against
the glass viewport above the Down arrow. The display shows the first menu
item, Zero Adj. The displays flashes -0-‘,
Adj’, and then shows the
target gas concentration.
Menu navigation
Tap the magnetic wand over the Up and Down arrows to scroll through the menu selections. If the magnetic wand is held over the switch for > 0.5 second, this will be considered multiple taps and the menu will scroll. When you reach the last item Password Adj, the menu will bottom-out, and you will have to use the Up arrow to scroll back up through the menu. The active menu item as well as its current value or status will flash on the display as it scrolls.
Changing parameter values/status
To enter a new value, or change a status, tap OK with the magnetic wand when the desired menu item is displayed. The current value or status will flash to indicate a change to data entry mode. The Up and Down arrows allow you to adjust the value of a numerical parameter or to toggle between preset choices. Once the display shows the desired value or choice, tap OK to validate the new parameter. This will take you back to the menu, where you can scroll to another menu item, if desired.
Exiting the Menu
To get back into the standard measurement mode, just scroll to the gas concentration menu item at the top of the menu. The actual gas concentration will be displayed.
Menu items
Sensor lock
Upon power up, the Polytron 5000 checks to see if the correct sensor is
installed in order to prevent accidental installation of the wrong sensor
type.
If the correct sensor type is installed, the software goes directly to
measurement mode, displaying the gas concentration (see Section 3.6.2 on Page
9). If the installed sensor differs from the sensor installed previously, the
message ‘sns’, ‘loc’, ‘on’ (sensor lock on) will flash. You then
have two choices: install a sensor of the correct type, or verify that you are
purposely changing to a different sensor.
To verify the switch to a different sensor, hold the magnetic wand over OK.
Theflashing message will change to ‘sns’, ‘loc’, ‘5’, then to ‘sns’,
‘loc’, ‘4’, counting down from 5 to 1. You must hold the magnetic wand on OK
during the entire countdown. If you remove the magnet at any time during the
countdown, ‘sns’, ‘loc’, ‘on’ will begin flashing again. After counting
down to ‘sns’, ‘loc’, ‘1’, the instrument will switch to the measurement
mode, accepting the new sensor type.
Gas conc
Displays the current value of the concentration of the target gas in ppm. This field is read-only, and cannot be modified by the operator.
Password
The use of a password is optional with the Polytron 5000. A password consists of a 3-digit number from 000 to 999; a value of 000 disables password protection and allows anyone to access the software interface. The instrument is delivered with the password set to 000. If a password has been set, it must be entered to gain access to the software menu. Tap the magnetic wand over the Down arrow until the `PAS’ menu item is displayed. The 3-digit LCD will then show `000′, with the first zero on the left blinking. Use the Up and Down arrows to increment or decrement this digit, then tap OK. The second (middle) digit will blink, and the correct value should be set using the Up and Down arrows as before. Repeat the process for the third digit. Tap OK when the full password is displayed. If the displayed value matches the set password, you will gain access to the rest of the menu. If an incorrect password is entered, the instrument will return to the measurement mode.
Password adj
Use the Down arrow to scroll to the last menu item, Password Adjust. The
display will show the message PAS’,
ADJ’, `000′. If a password other
than 000 has already been entered, that number will appear in place of `000′.
Tapping the magnetic wand over OK causes the 3-digit display to flash. Tap
the Up or Down arrows to scroll the value to the desired new password. The
display will stop scrolling at a maximum value of 999 or a minimum of 000; it
will not roll over. Tap OK when the desired password is displayed to accept
this as the new password. Once a password is set, you will have to enter it to
gain access to the menu.
Zero adj
Allows you to adjust the zero reference point of the sensor when no target gas is present, such as during calibration (see Section 4.1 on Page 10).
| NOTICE
Zero Adjust is not active when an oxygen sensor is installed.
---|---
Span adj
Allows you to adjust the displayed gas concentration to match the known concentration of an applied calibration gas. For example, if a 100 ppm calibration gas is applied to the sensor, the Span Adj value should be adjusted to 100 once the sensor reading has stabilized (see Section 4.1 on Page 10).
FSD adj
Permits adjustment of the Full Scale Deflection, or range, of the instrument. It establishes the upper limit value of the 4-20 mA output of the sensor. For example, if the FSD is set to 100 ppm, then 0 ppm of the target gas will produce a 4 mA output, while 100 ppm will produce a 20 mA signal.
A1 adj
This menu item is not active for the Polytron 5000.
A2 adj
This menu item is not active for the Polytron 5000.
A1 lat
This menu item is not active for the Polytron 5000.
A2 lat
This menu item is not active for the Polytron 5000.
A1 acn
This menu item is not active for the Polytron 5000.
A2 acn
This menu item is not active for the Polytron 5000.
Cal sig
Calibration signal is the signal that is transmitted by the 4-20 mA output anytime you access the software menu. It is user-selectable. The two possibilities are:
- a steady 3 mA signal
- an oscillating 3-5 mA signal with a frequency of 1 Hz
The default value is a steady 3 mA signal.
Output and display variations
The following table shows the status of the 4-20 mA output, and the LCD display for various conditions.
Condition | Analog Output | LCD Display |
---|---|---|
warming-up | actual signal | actual signal |
after warm-up | actual signal | actual signal |
after power outage | actual signal1 | actual signal |
new sensor installed | actual signal | actual signal |
sensor removed | fault | “pls con snr” |
bad/unsupported sensor | fault | “snr err” |
hardware fault | fault | “flt” |
microprocessor lock-up | fault | frozen display |
in menu | maintenance signal | menu function |
in cal modes | maintenance signal | cal function displays |
1) latest calibration data is restored upon power-up
Maintenance
Prior to calibration, each sensor must be allowed to warm up for the length of time specified in the sensor data sheet. During this time the sensor is capable of detecting the target gas, but its performance will deviate from specifications. Please see sensor data sheet for details.
Calibration procedure for toxic gases
For oxygen sensor see Section 4.2 on Page 11. Calibration of this instrument
must be performed at regular intervals as detailed in the sensor data sheet.
- Regulator
- Calibration gas cylinder
- Calibration adapter
Zero calibration
- Attach the pressure regulator to the nitrogen (N2) or Zero Air calibration gas cylinder.
- Fit the calibration adapter tightly to the end of the sensor.
- Turn the gas on.
| NOTICE
Allow the gas to flow for at least three minutes before proceeding.
---|--- - Scroll through software menu to Zero Adj and tap OK. Thecurrent zero value will be displayed.
- Wait for the zero to stabilize.
- Trim the stabilized value to zero on the display using the Up and Down arrows.
- Accept the value by tapping OK with the magnetic wand.
- Turn off the gas flow and remove the calibration adapter from the sensor
| NOTICE
Ambient air can be used to zero the sensor instead of nitrogen or Zero Air if
the area is known to be free of the target gas or any gas to which the sensor
may be cross-sensitive (as listed on the sensor data sheet). In this case, no
cylinder or calibration adapter is needed for the zero calibration.
---|---
Span calibration
- Attach the pressure regulator to the calibration gas cylinder.
- Fit the calibration adapter tightly to the end of the sensor.
- Turn the gas flow on.
| NOTICE
Allow the gas to flow for at least three minutes before proceeding
---|--- - Scroll through software menu to Span Adj and tap OK. The span value will be displayed.
- Wait for the span value to stabilize.
- Trim the stabilized value to the calibration gas concentration using the Up and Down arrows.
- Accept the value by tapping OK with the magnetic wand.
- Turn off the gas flow and remove the calibration adapter from the sensor.
Calibration procedure for oxygen
Zero calibration
Zero calibration is not required for the oxygen version of the Polytron 5000.
Span calibration
- The span calibration of an oxygen sensor requires oxygen calibration gas or ambient air. Normally a cylinder containing 20.9 % O2 is used to match the atmospheric concentration.
- Fit the calibration adapter tightly to the end of the sensor if calibrating with a cylinder containing oxygen. Alternatively, if you want to calibrate the span using ambient air, leave the adapter off and skip the next two steps.
- Attach the pressure regulator to the calibration gas cylinder containing 20.9 % O2.
- Turn the gas flow on.
| NOTICE
Allow the gas to flow for at least three minutes before proceeding.
---|--- - Scroll through software menu to Span Adj.
- Wait for the signal to stabilize.
- Trim the stabilized value to the calibration gas concentration (normally 20.9 %) using the Up and Down arrows.
- Accept the value by tapping OK with the magnetic wand.
- Turn off the gas flow and remove the calibration adapter from the sensor.
Error messages
Error Code | Condition | Solution |
---|---|---|
pls con snsr | no sensor is connected | connect sensor |
sensor connection is bad | check to ensure sensor is seated in connector | |
snr err | unsupported sensor is connected or sensor EEPROM data is corrupted |
install supported sensor
flt| hardware fault| replace electronics
AFE err| wrong software version installed in analog front end (AFE)| install
most recent AFE software
defective AFE| replace electronics
AFE out| AFE microcontroller is out of its socket or not installed| check
microcontroller
defective AFE| replace electronics
Disposal of electrical and electronic equipment
EC-wide regulations for the disposal of electrical and electronic equipment, which have been defined in the EU Directive 2002/96/EC and in national laws, have been effective since August 2005 and apply to this device. Common household appliances can be disposed of using special collecting and recycling facilities. However, as this device has not been registered for household usage, it must not be disposed of through these means. The device can be returned to your national Dräger Sales Organization for disposal. Please do not hesitate to contact Dräger if you have any further questions on this issue.
Technical data
Approvals
See printout of approval label.
Signal transmission to central control unit
Analog | Transmission by 2-core shielded cable |
---|---|
Measurement current | 4 to 20 mA |
Transmitter fault | <2 mA |
Maintenance signal | 4 ± 1 mA, 1 Hz modulation or steady 3 mA signal |
(user selectable)
Voltage of power supply
Operating voltage | 16 to 30 VDC |
---|---|
Operating current | 2.0 to 21 mA @ 24 VDC |
Maximum input current | 21 mA @ 24 VDC |
Connector accepts 24 to 12 AWG wire (0.2 to 2.5 mm2)
Physical specifications
Enclosure | NEMA 4x IP 65, 66, 67 |
---|---|
Size (approx.) | L x W x D, approx. 9.5” x 5.8” x 5.1” |
(240 x 150 x 130 mm)
Weight (approx.)| aluminum version:
5.9 lbs (2.7 kg)
stainless steel version:
9.7 lbs (4.4 kg)
Environmental parameters
Temperature | –40 to +150 °F, (–40 to +65 °C) |
---|---|
Pressure | 20.7 to 38.4 in. of Hg (700 to 1300 mbar) |
Humidity | 5 to 95 %RH, non-condensing |
Maximum Air Velocity | < 19.5 ft/s ( < 6 m/s) |
Ambient influences
See sensor data sheets.
Default values
Polytron 5000 | FSD (Full Scale Deflection, user adjustable) |
---|---|
CO | 50 / 100 / 200 / 300 / 500 ppm |
H2S zero-bias | 10 / 20 / 50 / 100 ppm |
H2S LC | 10 / 20 / 50 / 100 ppm |
H2S HC | 100 / 200 / 300 / 500 ppm |
O2 | 25 vol% |
NH3 LC 1 | 50 / 100 ppm |
NH3 HC 1 | 300 / 500 ppm |
Cl2 1 | 5 / 10 / 20 / 50 ppm |
SO2 1 | 5 / 10 / 20 / 50 / 100 ppm |
NO 2 | 50 / 100 ppm |
NO2 1 | 5 / 10 / 20 / 100 ppm |
H2 | 500 ppm |
H2O2 LC 1 | 5 / 10 / 20 / 50 / 100 ppm |
EO 2 | 20 / 50 / 100 ppm |
Resolution:
if FSD > 50 the resolution is 1
if FSD < 50 the resolution is 0.1
- Only to be used with Polytron 5000 General Purpose part number 4544165.
- Notice: Gases which react with moisture will not repeatedly pass through the wire mesh, if the transmitter is installed in a nonclimate controlled environment. If it can be ensured, that condensation on the wire mesh will not occur; all sensors can be used in the explosion proof versions 4544120, 4544122. Otherwise, these sensors should only be used with Polytron 5000 General Purpose 4544165.
Order list
Description | Part Number |
---|---|
Polytron 5000, DS d A | 4544120 |
Polytron 5000, DS d S | 4544122 |
Polytron 5000, DS d A General Purpose | 4544165 |
Sensors
Description | Part Number |
---|---|
DrägerSensor CO | 6809605 |
DrägerSensor H2S | 6810435 |
DrägerSensor H2S LC | 6809610 |
DrägerSensor H2S HC | 6809710 |
DrägerSensor O2 LS | 6809630 |
DrägerSensor H2 | 6809685 |
DrägerSensor NH3 LC | 68096801 |
DrägerSensor NH3 HC | 68096451 |
DrägerSensor Cl2 | 68096651 |
DrägerSensor SO2 | 68096601 |
DrägerSensor NO | 68096252 |
DrägerSensor NO2 | 68096551 |
DrägerSensor H2O2 LC | 68097051 |
DrägerSensor EO | 68107402 |
- Only to be used with Polytron 5000 General Purpose part number 4544165.
- Notice: Gases which react with moisture will not repeatedly pass through the wire mesh, if the transmitter is installed in a nonclimate controlled environment. If it can be ensured, that condensation on the wire mesh will not occur; all sensors can be used in the explosion proof versions 4544120, 4544122. Otherwise, these sensors should only be used with Polytron 5000 General Purpose 4544165.
Accessories
Description | Part Number |
---|---|
Pipe Mount Kit | 4544198 |
Magnet with Key Chain | 4544101 |
Splash Guard | 6812510 |
Calibration Adapter PE | 4509314 |
Calibration Adapter PE, Europe | 6806978 |
Calibration Adapter Viton | 6810536 |
Calibration Kit (includes carrying case, calibration adapter 4509314, 500
cc/min cylinder regulator 4557020 and 100 % N2 103 L @ 1000 PSI nitrogen /
zero gas cylinder)| 4594620
Calibration Gases| on request
Spare parts
Description | Part Number |
---|---|
Bezel Polytron 5000 | 4544184 |
PCB Main Polytron 5000 | 4544185 |
PCB Filter Polytron 5000 | 4544252 |
Sensor Housing Complete | 4544189 |
Sensor Cap | 4544190 |