Ai Mon Display and Monitoring Display Apps User Manual

June 6, 2024
Ai Mon

Ai Mon Display and Monitoring Display Apps

Features of App Menu Display

  1. Colored background indicates activation, while white (transparent) background indicates deactivation.
  2. For a more specific explanation of on-screen moving, see Help.
  3. The colors and shapes displayed on the main screen (real-time and cumulative) have the following meanings:

Alarm Functions Guide for Monitoring Information

Categories| Operating Criteria| Sound Type Frequency Pitch| Notes
---|---|---|---

App

|

Caution

| Heart Rate|

User Options

| Dingdong-Dingdong (double sound),

2x Dingdong Once

Slightly High Pitch

|
SpO2
Skin Temperature
Fall detection
Crying
Gateway| Band Battery| Less than 20%

of the battery

| Single Sound(Dingdong)

Once Low Pitch

| Gateway

alarm

Settings of five measurement information

  1. age-in-days and automatic settings
  2. The baby’s current age is the number of days calculated from the date of birth to the present. Depending on the baby’s current age, services are provided in two types (Mode1, Mode2).
    Info range| birth~365days| 366days~1,095days
    ---|---|---
    Heart rate| measuring
    SpO2| measuring| not measuring
    Skin temp| measuring
    Fall detection| measuring| not measuring
    Cry detection| measuring
  3. Settings of user range for heart rate, SpO2, skin temperature
  4. Depending on the mode, you can set the notification ranges for heart rate, SpO2 and skin temperature in the AIMON app. AIMON app(APP) : 《baby info settings》 / 〈range settings〉
  5. birth~365days
  6. Measured info Lower range setting Higher range setting
    Heart rate 65~75 bpm(70 bpm) 185~205 bpm(195 bpm)
    SpO2
    Skin temp 28~34.5 ℃(31 ℃) 35.5~40 ℃(36 ℃)
  7. 366days~1,095days
    Measured info| Lower range setting| Higher range setting
    ---|---|---
    Heart rate| 65~75 bpm(70 bpm)| 185~205 bpm(195 bpm)
    SpO2| –| –
    Skin temp| 28~34.5 ℃(31 ℃)| 35.5~40 ℃(36 ℃)

Specifications

Band & Gateway

Accessory band

Items Accessory band S Accessory band L
product / model Accessory band T1 / Accessory band T2 Accessory band T1 /

Accessory band T2
length| T1: approx.170mm, T2: approx.210mm| T1: approx.185mm, T2: approx. 250mm
weight| T1 : approx. 6g T2 : approx. 2.5g| T1 :approx. 7g T2 approx. :2.5g
material| tencel, cotton, nylon(velcro)| tencel, cotton, nylon (Velcro)
manufacturer| AIMON Co., Ltd.
COO| Republic of Korea
DOM| indicated separately

Charging Cradle

product/model AIMON SMART B Charger
apparent size approx. 33mm x 33mm x 8.5mm
weight approx. 6g
material PC
manufacturer AIMON Co., Ltd.
COO Republic of Korea
DOM indicated separately
  1. You can read the user manual before making a service inquiry and can troubleshoot any of the following symptoms.
  2. If the problem is not resolved, please contact (info@aimon.co.kr)
  3. If gateway reset is required
  4. An automatic connection is made when the connection between the gateway and registered WiFi (AP) is disconnected. If you want to connect manually because the gateway and registered WiFi (AP) are not automatically connected after tens of minutes have passed, reset the gateway.
  5. If the registered WiFi is not found after connecting to the gateway by Bluetooth pairing in the App or a new WiFi AP you want to connect to is not found (WiFi AP-2.4GHz must be nearby), reset the gateway. After resetting the gateway, try
  6. pairing and verify it by pressing the gateway alarm button for 7 seconds to restore the Bluetooth connection between the phone (App) and the gateway during use.
  7. After pairing is complete, the registered WiFi on your phone will be searched for or a new WiFi AP you are trying to connect to is searched for.
  8. If band reset is required
  9. An automatic connection should be made after the connection between the band and the gateway is disconnected, but if the automatic connection is not possible even they are close to each other, reset the band (you can check the connection through the gateway LED)
  10. Reset the band when it is not connected to the gateway for a long time even after the band is powered on. You can check the connection by LED blinking indicating the band connection among the gateway LEDs.
  11. After gateway reset
  12. or band reset
  13. check the communication status of the gateway and the band through the gateway LED and check the communication status through the paired phone (App).
  14. Among the LEDs of the gateway, you can check the interconnection and connection status by the LED blinking associated with the band or phone.
  15. Check if the communication connection between the gateway and the band is normal through the function of the phone (App) which is paired with the gateway. If the pairing between the gateway and the phone (App) is normal, you can check the communication status with the gateway on the paired phone (App) regardless of the cloud connection (//). Check your phone manufacturer’s OS patch. (Refer to 4. Information necessary for use of the application on user manual)
  16. Reinstall the apps in the following order.
  17. Delete the AIMON app → Install the latest patch provided by the mobile phone
    operating system and manufacturer → Use after reinstalling the AIMON app

Warnings, safety information and Tips

AIMaON band

The AIMON band is not a medical device. It is not intended for use as a medical device or to replace a medical device. does not and is not intended to diagnose, cure, treat, alleviate or prevent ny disease or health condition or investigate, replace or modify anatomy or any physiological process.

  1. · Wear it with the method recommended in the manual, and if it is detected as being worn, heart rate, oxygen saturation(SpO2), skin temperature, fall detection and cry detection are measured.
  2. The measurement signal can be affected by the baby’s development status, skin brightness, and circulating blood volume in the skin, and there are differences between babies.
  3. All pulse and oximetry sensors are motion-sensitive and cannot calculate accurate readings when moving. This is due to the technical characteristics of pulse and oximetry.
  4. During wearing the strap, if strap wearing is not like as recommended in the manual, loose, or if there is a lot of separation between the recommended ankle area of the band and the skin, the measured value of oxygen saturation may not be accurate.
  5. Skin temperature is a technology that measures the temperature of the skin, so you can check the warmth of your baby.
  6. Regarding fall detection, a very large activity of the baby can be detected as a fall, and it may not be detected if the band is at off-body or too loose, or the band is turned off.
  7. There is a possibility that the high-frequency noise generated in the living environment is recognized as crying. (It is recommended to use in a quiet place close to the baby.)
  8. In the AIMON band, the measured values are passed through the gateway and delivered through the cloud server. Data reception may be delayed depending on the communication environment.

Accessary band

  • The accessory band of the AIMON band is made of natural materials and medical TPE to prevent irritation on the baby’s skin (it may vary depending on the baby’s skin sensitivity).

  • The accessory band of the AIMON Band is a consumable that wears out naturally over time and can be purchased separately.

  • Pulling or pushing the accessory band hard can damage it.
    The warranty period for the consumable accessory band is 3 months.

  • For babies 12 months and older, it is recommended to wear it inside the ankle while sleeping.

FCC and IC Compliance Statement

FCC

This device complies with part 15 of the FCC rules.(Part 15.19(3)) Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.” This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

  1. Reorient or relocate the receiving antenna.
  2. Increase the separation between the equipment and receiver.
  3. Connect the equipment into an outlet on a circuit different from that which the receiver is connected to.
  4. Consult the dealer or an experienced radio/TV technician for help

IC

This device complies with RSS-247 of the Industry Canada Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference

Caution

Please use ONLY the supplied power adapter. Failure to do so may void our limited warranty. Any changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate this equipment. The gateway complies with RF exposure requirements set forth for an uncontrolled environment. The gateway should be installed and operated with minimum distance 20 cm between the radiator & your body.

Product Warranty Statement

  1. Product
  2. Purchase route / Date
  3. Order Number
  4. Purchaser
  5. Warranty Period

We guarantee as follows

  • Free service is available if the product fails within one year of purchase or within the warranty period (within 3 months for expendables). The warranty period will be halved if ordinary products are used for sales purposes.
  • You may be charged for some fixed cost even within the warranty period or may not be able to use the warranty service IF:
    • Product fails due to impact, negligence, carelessness, or water-damaged by user
    • Arbitrarily disassembled or modified
    • Product fails due to natural disasters such as fire, earthquake, or flood etc.
      Everything else will be compensated in accordance with the Consumer Damage Compensation Regulations.