RESMED VPAP ST With iVAPS BiLevel Machine User Guide

June 5, 2024
ResMed

VPAP™ ST with iVAPS
NONINVASIVE VENTILATOR
Information Guide 

Please read the entire Information and Welcome Guides before using your device.

Ind ications for use

The VPAP ST is indicated to provide noninvasive ventilation for patients weighing more than 13 kg or more than 30 kg in iVAPS mode with respiratory insufficiency or obstructive sleep apnoea (OSA). The VPAP ST is intended for home and hospital use.

Contraindications
Positive airway pressure therapy may be contraindicated in some patients with the following preexisting conditions:

  • severe bullous lung disease
  • pneumothorax or pneumomediastinum
  • pathologically low blood pressure, particularly if associated with intravascular volume depletion
  • dehydration
  • cerebrospinal fluid leak, recent cranial surgery, or trauma.

Adverse effects

Patients should report unusual chest pain, severe headache, or increased breathlessness to their prescribing physician. An acute upper respiratory tract infection may require temporary discontinuation of treatment.
The following side effects may arise during the course of therapy with the device:

  • drying of the nose, mouth, or throat
  • nosebleed
  • bloating
  • ear or sinus discomfort
  •  eye irritation
  • skin rashes.

Troubleshooting

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the device enclosure.

Problem/Possible cause Solution

No display
Power is not connected.| Ensure the power cord is connected and the power outlet (if available) is on.
The DC plug is partially inserted into the back of the device or inserted too slowly.| Fully insert the DC plug.

Insufficient air delivered from the device

Ramp time is in use. Wait for air pressure to build up or change ramp time.
The air filter is dirty. Replace the air filter.
Air tubing is not connected properly. Check air tubing.
Air tubing is blocked, pinched, or punctured. Unblock or free the air tubing.

Check the air tubing for punctures.
Mask and headgear are not positioned correctly.| Adjust the position of the mask and headgear.
Incorrect air tubing selected| If you are using the SlimLine, Standard or 3 m air tubing ensure that you have the correct air tubing selected via the menu.
The non-vented mask is used.| Only use a vented mask.
Mask vents might be blocked.| Check if you have sufficient venting. Unblock mask vents if necessary.
EPAP may be set too low.| Talk to your clinician about your settings.

Device does not start when you breathe into the mask
The breath is not deep enough to trigger SmartStart/Stop.
Take a deep breath in and out through the mask.

Problem/Possible cause Solution

SmartStart/Stop is disabled
because Leak Alert is enabled.| Press Start/Stop to start therapy.
SmartStart/Stop is disabled because Confirm Stop is enabled.| Press Start/Stop to start therapy.
SmartStart/Stop is disabled.| Talk to your clinician about enabling the SmartStart/Stop feature.
There is an excessive leak| Adjust the position of the mask and headgear. Connect the air tubing firmly at both ends.

The device does not stop when you remove your mask

SmartStart/Stop is disabled because Leak Alert is enabled| Press Start/Stop to stop therapy.
---|---
SmartStart/Stop is disabled because Confirm Stop is enabled.| A message appears on the screen. To stop therapy, select Yes and press the Push Dial.
SmartStart/Stop is disabled.| Talk to your clinician about enabling the SmartStart/Stop feature.

SmartStart/Stop is enabled but the device does not stop automatically when you remove your mask

An incompatible mask system is being used.| Only use equipment recommended by ResMed.
---|---
Incorrect mask setting being used.| Check the selected mask type in the Setup menu. Change it if necessary.
The patient is using a nasal pillow mask with a set pressure of less than 6 cm H2O.| Disable SmartStart/Stop.
The patient is using a pediatric mask with a set pressure less
than 8 cm H2O.| Disable SmartStart/Stop.

Pressure rises inappropriately

Talking, coughing or breathing in an unusual manner.| Avoid talking with a nasal mask on, and breathe as normally as
possible.
---|---
The Mask cushion is buzzing against the skin.| Adjust the headgear.
Cushion seated incorrectly causing the excessive leak.| Adjust headgear or re- fit cushion.

Displays message: High-temperature fault, refer to the user manual

Problem/Possible cause Solution
The device has been left in a hot environment. Allow cooling before re-use.

Disconnect the power cord and then reconnect it to restart the device.
The air filter is blocked.| Replace your air filter. Disconnect the power cord and then
reconnect it to restart the device.
Air tubing is blocked.| Check your air tubing and remove any blockages. Disconnect the power cord and then reconnect it to restart the device.
The humidity level setting is too high, resulting in an accumulation of water in the air tubing.| Turn the humidity level setting down and empty the water from the air tubing.

Displays message: Check ResMed 30/90W Power Supply Unit and fully insert the connector

The DC plug is partially inserted into the back of the device or inserted too slowly.| Fully insert the DC plug.
---|---
A non-ResMed power supply unit is connected to the device.| Remove the power supply unit and replace it with a ResMed power
supply unit.
The power supply unit is being covered by bedding.| Make sure that the power supply unit is free from bedding, clothes
or other objects that could cover it.

Displays message: No tube, please check your tube is connected

Flow is high because air tubing is not connected properly.
Note: The tube disconnection check may not operate when an antibacterial filter is used.| Connect the air tubing firmly at both ends.
---|---

Displays message: Tube blocked, please check your tube

Air tubing is blocked.| Check your air tubing and remove any blockages. Disconnect the power cord and then reconnect it to restart the device.
---|---

Displays message: High leak, please check system setup and all connections

There is an excessive leak. Note: If Leak Alert is enabled, an audible alert is activated and a high leak message is displayed.| Adjust the position of the mask and headgear. Connect the air tubing firmly at both ends
---|---

The following message is displayed on the LCD after you try to update settings or copy data to the SD card: Card error, please remove the SD card and contact service, provider

SD card is not inserted correctly you may have removed the
SD card before settings were copied to the device.| Ensure that the SD card is inserted correctly.
Reinsert the SD card and wait for the Home screen or the “Settings updated successfully, press any key” message to appear on the LCD.
Note: This message only appears once. If you re-insert the SD card after you have updated your settings, the message will not be redisplayed.
---|---

The following message is NOT displayed on the LCD after you try to update the settings using the SD card : Settings updated successfully, press any key

The settings were not updated| Contact your clinician/service provider immediately.
---|---

General technical specifications

Power supply| 90W power supply unit
Input range: 100-240V, 50-60Hz, 115V, 400Hz nominal for aircraft use Typical power consumption: 70W (80VA)
Maximum power consumption: 110W (120VA)
---|---
Power supply| 30W power supply unit
Input range: 100-240V, 50-60Hz, 115V, 400Hz nominal for aircraft use Typical power consumption: 20W (40VA)
Maximum power consumption: 36W (75VA)
Power supply| 90W DC/DC converter
Nominal inputs: 12V, 24V
Typical power consumption: 70W Maximum power consumption: 110W
Environmental conditions| Operating temperature: +5°C to +35°C
Note: The airflow for breathing produced by this therapy device can be higher than the temperature of the room. Under extreme ambient temperature conditions (40°C) the device remains safe.
Operating humidity: 10 to 95% non-condensing
Operating altitude: Sea level to 2,591 m; air pressure range 1013 hPa to 738 hPa Storage and transport temperature: -20°C to +60°C Storage and transport humidity: 10 to 95% non-condensing
Aircraft use| ResMed confirms that the device/s meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21, category M) for all phases of air travel.
Electromagnetic compatibility| The product complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-2, for residential, commercial, and light industry environments.
It is recommended that mobile communication devices are kept at least 1 m away from the device.
Information regarding the electromagnetic emissions and immunity of this ResMed device can be found on www.resmed.com, on the Products page under Service and Support. Click on the PDF file for your language.
IEC 60601-1 classification| Class II (double insulation), Type BF, Ingress protection IP21

VPAP ST technical specifications

Mode pressure ranges

| CPAP mode
Set Pressure: 4-20 cm H2O
S, ST, T and PAC modes
IPAP: 4-25 cm H2O; EPAP: 2-25 cm H2O iVAPS mode
PS: 0-23 cm H2O; EPAP: 2-25 cm H2O
---|---
Maximum single fault pressure| Maximum single fault steady-state pressure: 30 cm H20—if pressure exceeded for > 6 sec; 40 cm H20—if pressure exceeded for >1 sec
Physical| Nominal dimensions (L x W x H): 153 mm x 140 mm x 86 mm Weight: 835 g
Housing construction: Flame retardant engineering thermoplastic Air outlet: 22 mm conical air outlet (complies with ISO 5356-1:2004)
Air filter| Hypoallergenic air filter: Acrylic and polypropylene fibers in a polypropylene carrier
Standard air filter: Polyester non-woven fiber

Sound DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871:1996

| Pressure level (CPAP mode)
With SlimLine air 26 dBA with uncertainty of 2 dBA as measured according to tubing: to EN ISO 17510-1:2009
With Standard air 27 dBA with uncertainty of 2 dBA as measured according to tubing: to EN ISO 17510-1:2009
With either SlimLine 28 dBA with uncertainty of 2 dBA as measured according to or Standard air to EN ISO 17510-1:2009 tubing and H5i:
Power level (CPAP mode)
With SlimLine air 34 dBA with uncertainty of 2 dBA as measured according to tubing: to EN ISO 17510-1:2009
With Standard air 35 dBA with uncertainty of 2 dBA as measured according to tubing: to EN ISO 17510-1:2009
With either SlimLine 36 dBA with uncertainty of 2 dBA as measured according to or Standard air to EN ISO 17510-1:2009 tubing and H5i:
Supplemental oxygen| Recommended maximum supplemental oxygen flow: 15 L/min (CPAP, S, ST, T, PAC); 4 L/min (iVAPS mode)

Air tubing technical specification

Air tubing Material Length Inner diameter
ClimateLine heated air tubing Flexible plastic and electrical components 2
m 15 mm
ClimateLinemAx heated air tubing Flexible plastic and electrical components
1.9 m 19 mm
SlimLine air tubing_ Flexible plastic 1.8 m 15 mm
Standard air tubing Flexible plastic 2 m 19 mm
3 m air tubing Flexible plastic 3 m 19 mm
Heated air tubing temperature cut-out: 41°C

Notes:

  • The manufacturer reserves the right to change these specifications without notice.
  • The temperature and relative humidity settings displayed for Climate Control have not measured values.
  • Check with your clinician/service provider before using the SlimLine air tubing with devices other than the S9 or H5i.
  • The electrical connector end of the heated air tubing is only compatible with the H5i air outlet and should not be fitted to the device or mask.
  • When using the SlimLine or ClimateLine above 20 cm H2O, the device’s optimum performance may not be reached if used with an antibacterial filter. The device performance must be checked prior to prescribing the SlimLine for use with an antibacterial filter.
  • The ClimateLine or ClimateLine MAX is designed only for use with the H5i.

Humidifier performance

The following settings have been tested at 22°C ambient temperature:

Mask pressure RH output % Nominal system output AHa , BTPSb
cm     H2O Setting 3 Setting 6
3 90 90 100
10 95 95 100
20 95 95 100
25 10 100 100

a. AH – Absolute Humidity in mg/L.
b. BTPS – Body Temperature Pressure Saturated.

Pneumatic flow pathRESMED VPAP ST With iVAPS BiLevel
Machine

  1. Flow sensor
  2. Blower
  3. Pressure sensor
  4. Mask
  5. Air tubing
  6. H5i
  7.  Device
  8. Inlet filter

Flow (maximum) at set pressures
The following are measured at the end of the specified air tubing:

Pressure, cm H2O| VPAP ST and
Standard, L/min| VPAP ST, H5i and
Standard, L/min| VPAP ST and
SlimLine, L/min| VPAP ST, H5i and
ClimateLine, L/min
---|---|---|---|---
4| 200| 170| 195| 170
8| 200| 170| 190| 170
12| 200| 170| 184| 170
16| 200| 170| 175| 170
20| 190| 170| 168| 161
25| 180| 161| 144| 125

Displayed values

Value Range Display resolution
A pressure sensor at the air outlet
Mask pressure  2-25 cm H2O 0.1 cm H2O
Flow derived values
Leak 0 0–200 L/min 1 L/min
Tidal volume 0–4000 mL 1 mL
Respiratory rate 0–50 BPM 1 BPM
Minute ventilation 0–30 L/min 0.1 L/min
Ti 0.1  0.1–4.0 sec 0.1 sec
I:E ratio 1:50–2:1 0.1
Value Accuracy
--- ---
Pressure measurement
Mask pressure ±0.5 cm H2O (+4% of measured value)
Leak ±12 L/min or 20% of reading, whichever is greater, at 0 to 60 L/min
Flow measurements 60 L/min
Tidal volume.c ±20%
Respiratory rate,c ±1BPM

Minute ventilation
b, c| ±20%

a. Results are expressed at ATPD (Ambient Temperature and Pressure, Dry).
b. Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL or minute ventilation <3 L/min.
c. Measurement accuracy verified as per EN ISO 10651-6:2009 for Home Care Ventilatory Support Devices (Figure 101 and Table 101) using nominal ResMed mask vent flows.
Pressure accuracy
Maximum static pressure variation at 10 cm H2O according to EN ISO 17510-1:2009

| Standard air tubing| SlimLine air tubing
---|---|---
Without H5i|  9.89 cm H2O to 9.97 cm H2O| 9.76 cm H2O to 9.87 cm H2O
With H5i| 9.82 cm H2O to 9.98 cm H2O| 9.78 cm H2O to 9.88 cm H2O

Maximum dynamic pressure varies according to EN ISO 17510-1:2009

 Pressure (cm H20) 10 BPM 15 BPM 20 BPM

VPAP ST and Standard air tubing without H5i / VPAP ST and Standard air tubing with H5i
4| 0.18 / 0.18| 0.30 / 0.30| 0.51 /0.51
| 0.21 /0.20| 0.26 / 0.24| 0.38 / 0.36
12| 0.21 /0.20| 0.26 / 0.23|  0.34 / 0.31
16| 0.22 / 0.21| 0.27 / 0.26| 0.36 / 0.33
20| 0.23 / 0.22| 0.26 / 0.28| 0.38 / 0.35
25| 0.30 / 0.31| 0.54 / 0.50| 0.74 / 0.71
Pressure (cm H20)| 10 BPM| 15 BPM| 20 BPM
VPAP ST and SlimLine air tubing without H5i / VPAP ST and SlimLine air tubing with H5i
4| 0.22 / 0.20| 0.28 / 0.29| 0.47 / 0.53
8| 0.23 / 0.19| 0.32 / 0.29| 0.41 / 0.42
12| 0.22 / 0.21| 0.35 / 0.29| 0.41 /0.45
16| 0.22 / 0.23| 0.41 / 0.33| 0.44 / 0.50
20| 0.24 / 0.27| 0.37 / 0.34| 0.48 / 0.50
25| 0.31 /0.31| 0.50 / 0.54| 0.78 / 0.84

Symbols
The following symbols may appear on your product or packaging.

** Caution;
Read instructions before use;
IP21 Protection against insertion of fingers and against vertically dripping water;
Type BF equipment;
Class II equipment;
Start/Stop;
Manufacturer;
European Authorised Representative;
RoHS ; European RoHS;
LOT Batch code;
REF Catalogue number;
SN Serial number;
Direct current;
Lock/unlock;
China pollution control logo 1;
China pollution control logo 2;
IP20** Not drip proof;
Keep dry;

Environmental information
This device should be disposed of separately, not as unsorted municipal waste. To dispose of your device, you should use the appropriate collection, reuse, and recycling systems available in your region. The use of this collection, reuse, and recycling system is designed to reduce pressure on natural resources and prevent hazardous substances from damaging the environment. If you need information on these disposal systems, please contact your local waste administration. The crossed-bin symbol invites you to use these disposal systems. If you require information on the collection and disposal of your ResMed device please contact your ResMed office, or local distributor or go to www.resmed.com/environment.
Servicing
The VPAP ST device is intended to provide safe and reliable operation when operated in accordance with the instructions provided by ResMed. ResMed recommends that the VPAP ST device be inspected and serviced by an authorized ResMed Service Centre if there is any sign of wear or concern with device function. Otherwise, service and inspection of the devices generally should not be required during the five-year design life of the device.
Limited warranty
ResMed Pty Ltd (hereafter ‘ResMed’) warrants that your ResMed product shall be free from defects in material and workmanship from the date of purchase for the period specified below.

Product Warranty period

Mask systems (including mask frame, cushion, headgear, and tubing)—excluding single-use devices
Accessories—excluding single-use devices
Flex-type finger pulse sensors
Humidifier water tubs| 90 days
Batteries for use in ResMed internal and external battery systems| 6 months
Clip-type finger pulse sensors
CPAP and bilevel device data modules
Oximeters and CPAP and bilevel device oximeter adapters
Humidifiers and humidifier cleanable water tubs
Titration control devices| 1 year
CPAP, bilevel, and ventilation devices (including external power supply units)
Battery accessories
Portable diagnostic/screening devices| 2 years

This warranty is only available to the initial consumer. It is not transferable.
If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components.
This Limited Warranty does not cover: a) any damage caused as a result of improper use, abuse, modification, or alteration of the product; b) repairs carried out by any service organization that has not been expressly authorized by ResMed to perform such repairs; c) any damage or contamination due to cigarette, pipe, cigar or other smoke; and d) any damage caused by water being spilled on or into an electronic device.
Warranty is void on products sold, or resold, outside the region of original purchase.
Warranty claims on defective products must be made by the initial consumer at the point of purchase.
This warranty replaces all other expressed or implied warranties, including any implied warranty of merchantability or fitness for a particular purpose. Some regions or states do not allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you.
ResMed shall not be responsible for any incidental or consequential damages claimed to have resulted from the sale, installation or use of any ResMed product. Some regions or states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you.
This warranty gives you specific legal rights, and you may also have other rights which vary from region to region. For further information on your warranty rights, contact your local ResMed dealer or ResMed office.

WARNINGS

  • Read the entire manual before using the device.
  • Use the device only as directed by your physician or healthcare provider.
  • Use the device only for the intended use as described in this manual. The advice contained in this manual should not supersede instructions given by the prescribing physician.
  • If you notice any unexplained changes in the performance of the device, if it is making unusual or harsh sounds, if the device or the power supply is dropped or mishandled, if water is spilled into the enclosure, or if the enclosure is broken, discontinue use and contact your ResMed Service Center.
  • Beware of electrocution. Do not immerse the device, humidifier, power supply, or power cord in water. In the event of a spill, disconnect the device from the power supply and let the parts dry. Always unplug the device before cleaning and make sure that all parts are dry before plugging in the device.
  • Explosion hazard—do not use in the vicinity of flammable anesthetics.
  • Make sure the power cord and plug are in good condition and the equipment is not damaged.
  • Keep the power cord away from hot surfaces.
  • The device should only be used with masks (and connectors 1) recommended by ResMed, or by a physician or respiratory therapist. A mask should not be used unless the device is turned on. Once the mask is fitted, ensure that the device is blowing air. The vent hole or holes associated with the mask should never be blocked.

Explanation: The device is intended to be used with special masks (or connectors) which have vent holes to allow a continuous flow of air out of the mask. When the device is turned on and functioning properly, new air from the device flushes the exhaled air out through the mask vent holes. However, when the device is not operating, insufficient fresh air will be provided through the mask, and the exhaled air may be rebreathed. Rebreathing of exhaled air for longer than several minutes can, in some circumstances, lead to suffocation. This applies to most models of CPAP or bilevel devices.

  • Oxygen supports combustion. Oxygen must not be used while smoking or in the presence of an open flame.
  • Always ensure that the device is turned on and airflow is generated before the oxygen supply is turned on. Always turn the oxygen supply off before the device is turned off so that unused oxygen does not accumulate within the device enclosure and creates a risk of fire.
  • Do not leave long lengths of air tubing around the top of your bed. It could twist around your head or neck while you are sleeping.
  • Do not use electrically conductive or antistatic air tubings.
  • Do not use the air tubing if there are any visible signs of damage.
  • Only ResMed air tubing and accessories should be used with the device. A different type of air tubing or accessory may alter the pressure you actually receive, reducing the effectiveness of the treatment.
  • Only use the ResMed 90W or 30W power supply unit. Use the 90W power supply unit to power the system comprising the device, H5i, air tubing, DC/DC converter, and battery pack. The 30W power supply unit is designed to power the device only and is recommended for traveling.
  • Only ResMed products are designed to be connected to the module connector port. Connecting other devices could damage the device.
  • Blocking the air tubing and/or air inlet of the device while in operation could lead to overheating of the device.

  CAUTIONS

  • Do not open the device enclosure. There are no user-serviceable parts inside. Repairs and servicing should only be performed by an authorized ResMed service agent.
  • Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturizing or antibacterial soaps, or scented oils to clean the device, humidifier, or air tubing. These solutions may cause damage and reduce the life of these products.
  • Incorrect system setup may result in incorrect mask pressure reading. Ensure the system is correctly set up.
  • Be careful not to place the device where it can be bumped or where someone is likely to trip over the power cord.
  • Make sure that the area around the device is dry and clean and clear of bedding, clothes or other objects that could block the air inlet or cover the power supply unit.
  • Ensure that the device is protected against water if used outdoors. Enclose the device in the S9 travel bag for transport.

Manufacturer: ResMed Pty Ltd
Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia.
See www.resmed.com for other ResMed locations worldwide.
For patent information, see www.resmed.com/ip.
S9, H5i, ClimateLine, SlimLine,
SmartStart and VPAP are trademarks of ResMed. S9, ClimateLine,
SlimLine, SmartStart and VPAP are registered in U.S.
Patent and Trademark Office.
© 2017 ResMed. 368904-eng/2 2017-10
0123
Global leaders in sleep and respiratory medicine