GASLAB PRO TS-202 TecPen MAP Oxygen & CO2 Headspace Analyzer User Manual

June 5, 2024
GASLAB PRO

GASLAB PRO TS-202 TecPen MAP Oxygen & CO2 Headspace

Analyzer

Perform a measurement

  1. Acclimatize the device to the surrounding temperature (1-2 hours).
  2. Press ON/OFF to switch on the TecPen.
  3. The PWR LED will light up green.
  4. The TecPen will indicate the battery status and oxygen level with ****%.
  5. The TecPen will take 2 minutes to warm up.
  6. Connect the gas inlet of the TecPen to the provided filter and connect the sampling needle onto the filter using the Luer lock.
  7. Optional: Connect the tubing to the TecPen. Put the filter onto the free end of the tubing. Then connect the needle to the filter.
  8. Take a septum and place it on the package. Make sure that that the septum is placed on a place where enough gas is available to make the test.
  9. Insert the needle into the gas-filled space to be measured.
  10. Press START immediately after piercing the sample.
  11. The “MEAS” will appear on the display. The oxygen value will change continuously during measurement. Depending on your model, the oxygen level will display in %O2 or ppm.
  12. Once the measurement stops, the “MEAS” disappears from the display leaving the measured concentration. If the value is **** the O2 concentration was greater than the calibrated range.
  13. Press the SAVE button to save the measured value. The saving process is indicated by “LOG” on the display. The value is stored in the internal memory.
  14. Press ON/OFF again to turn the TecPen off.

For additional settings and features refer to the sections in this manual.

Introduction

The TecPen Handheld MAP Sensor is designed to provide fast, accurate measurements of gas concentrations in closed or pierce-able containers. It uses an optical gas sensor with OLED display as well as USB connectivity. This makes it an excellent tool for:

  • Food packaging in protective gas atmosphere (MAP packaging)
  • Headspace of vials, cell breeding vessels, infusion containers, syringes
  • Micro-bioreactor systems

Package Contents

Quantity Item
1 TecPen
1 Case

2

2

| Particulate Filters

Sampling needle 25/0.8

1| USB Cable
1| USB Protection

Cover

1| USB Flash Drive
1| Instruction Manual
12| Septum
1| Flexible Extension
1| Calibration Certificate

Layout & Controls

  1. START Button: starts and stops measurement.
  2. OLED Display
  3. PWR (Power) LED – Lights up when the device is switched on
  4. BT LED – Lights up when charging and turned on
  5. SAVE Button – saves measurement to memory
  6. ON/OFF Button – press briefly: switch on device, hold down: switch off
  7. USB Port – for charging or connecting to a PC to down data logs
  8. Sampling Gas Inlet
  9. Gas Outlet

Operation

IMPORTANT! If the device is taken to a room with a significantly different ambient temperature, an acclimatization period of 1-2 hours is required.

Before You Begin

  • The TecPen can be charged with any 5V USB cable and adapter. Full charging takes place in under 2 hours.
  • If the device is taken to a room with a significantly different ambient temperature, an acclimatization period of 1-2 hours is required.
  • Before taking a measurement, a filter should be attached to the TecPen to prevent contamination of the gas chamber. To do this, place the tubing on the gas inlet’s male Luer lock fitting (8) and fasten the filter to the other end. Attach the sampling needle to the front of the filter.

Note : All connections are made with Luer lock fittings. These must be securely engaged for proper installation and seal.

Turning the TecPen On

Press the ON/OFF button to turn on the unit. The TecPen shows the battery status and a value of *****. This is a placeholder and not an actual measurement result.

  1. After turning it on let the TecPen warm up for 2 minutes to insure accuracy.
  2. The TecPen can only be switched on when it is not connected to the charging cable.
  3. If no buttons are pressed within five minutes, the TecPen will switch itself off. The unit can also be switched off by pressing the ON/OFF button for at least two seconds.

Taking a Measurement

The sampling needle is used to puncture the packaging to be measured. After inserting the sampling needle briefly press the START button to take a measurement. The pump will draw a sample until it reads the lowest stable group of measurements. If it cannot read a stable measurement the pump will automatically shut off after 10 seconds. To try another measurement, press the START button again.

Saving a Measurement

Press the SAVE button. The measured value currently on the display is saved in the TecPen’s internal memory.

Turning the TecPen Off

Switch the unit off by pressing the ON/OFF button for at least two seconds.

Menu Display

The TecPen has multiple sub-screens for setup, calibration and troubleshooting information. To access these screens.

  1. Press and hold the SAVE button until the Current Recipe screen is shown.
  2. Press the SAVE button to cycle through the remaining screens.

It is not possible to cycle through the screens in reverse order. See the following Menu option page to learn more about each screen.

Main screen| | battery level measured value

Temporary: Memory/Log Indicator

Indicator for current measurement

---|---|---
Current  Recipe File| | Consecutive / internal

number product name, EAN code

Date  &Time| | 24 hour  clock used for data logging
Measurement Data

Display

| | Amplitude, sensor life, temperature
Device Information| | Device serial number,

device state (status) code.

0 is normal

Zero point calibration| | Select option YES / NO NOTE: 100% Nitrogen only!

Menu Options

In addition to the main screen the TecPen has 6 menu options screens that can be accessed by pressing and holding the SAVE button.

Change Data File (Recipes)

  1. Hold the SAVE button for 2 seconds to access the menu.
  2. A list of preset product recipe/configurations will appear on the display.
  3. Tap the SAVE button to scroll down or START button to scroll up to select a recipe.
  4. Press the SAVE button for 2 seconds to confirm and save the setting.

See the chapter on “Customizing the Output Data (Recipes)” to create your own custom recipes.

Set Date and Time

  1. Press and hold the SAVE button for two seconds. Repeat this step until the date and time display appears.
  2. Press the START button for two seconds.
  3. The year will be underlined.
  4. To scroll down or decrease number press SAVE
  5. To scroll up or increase number press START
  6. After setting the year, press START for two seconds.
  7. The year is saved when the month is underlined.
  8. Repeat steps 4-7 until everything is set.
  9. Press SAVE for 2 seconds to save the settings.

View Measurement Data Display

  1. Hold the SAVE button for 2 seconds to access the menu.
  2. Press the SAVE button to switch to the View Data Display screen.
  3. Press the SAVE button for 2 seconds to exit the menu.

View Device Status

  1. Hold the SAVE button for 2 seconds to access the menu.
  2. Press the SAVE button to switch to view the device status (state) code.
  3. Press the SAVE button for 2 seconds to exit the menu.

Zero Point Calibration

The TecPen sensor is calibrated once at the factory and does not require recalibration. However, a zero point adjustment can be carried out to insure the most accurate reported oxygen values over the life of the sensor. See the chapter on “Zero Point Calibration” for specific instructions.

Saving Data

Default Operation

After each measurement is completed and the pump stops running, the currently displayed measured value can be stored in the internal memory of the TecPen by briefly pressing the SAVE button. By default, a new data file is created each day using the date as the file name and the measured oxygen level and a time stamp as data.

Output Data File Format

The TecPen outputs a comma-delimited text file (.txt) with a header row that can be imported into any spreadsheet program. The data file format is as follows:

Data Label Example
Date 2019-03-15 (YYYY-MM-DD format)
Time 14:33:30 (hours:minutes:seconds 24 hr. clock)
Product_Number 123456 (user defined)
Product_Name NAME56789ABCD (user defined)
Product_EAN 0123456789ABC (user defined)
Product_Oxygen_Value 0.53472 (range 0-5%)
Amplitude 14016 (signal strength for error checking)
Tau 64.610248 (signal life for error checking)
Temperature[°C] 25.187500 (Centigrade only)
Battery_Level[%] 86 (range 0-100%)
Device Status 0=Normal (see device status codes)

Exporting Saved Data Files

  1. Press the ON/OFF button to turn on the unit.
  2.  Connect the TecPen to a PC using the supplied USB cable.
  3. The TecPen appears on your PC like a USB drive with one or more saved data files.
  4. Copy the files from the TecPen using your PC’s file manager.
  5. Disconnect the TecPen properly before removing the USB device from your PC

USB Data Safety

Before unplugging the USB cable, the TecPen MUST be ejected properly like any other standard USB device to avoid damaging the TecPen’s internal file system.

Note: If the TecPen is unplugged from the USB port without first selecting the “Safely Remove Hardware and Eject Media” option, all data  still on the device may be lost.

Customizing the Output Data (Recipe) File

The Product_Number, Product_Name and Product_EAN fields can be customized so that several TecPens can be used on the same QC line, and all will have different names in their data files to assist in QA reporting. These “recipes” can also be used if the TecPen is moved to different QC departments throughout the day. To get started, you need to modify a file named Rezepte.csv loaded on in TechPen’s internal memory.  The file can be downloaded via USB, modified and uploaded, or you can create a new Rezepte.csv file in any spreadsheet program that can save a .csv file. The required format is:

| A| B| C
---|---|---|---
1| 100ABC| Scanner1| 12345
2| 200DEF| Scanner2| 67890

The 3 alpha-numeric fields (do not use commas) in the table will replace the Product_Number, Product_Name and Product_EAN fields in your output data files. By default, the first row will always be loaded when the TecPen is turned on. If you want to select a different recipe, you must do so from the menu options after you turn the TecPen on and before you start collecting data.

Zero Point Adjustment

The TecPen uses 100% 5.0 pure nitrogen gas for zero adjustment. To calibrate the TecPen, it must be first be purged of any remaining gas from previous use by running the pump for several seconds until the device is filled with pure nitrogen. Follow the steps below:

  1. Connect the TecPen to pure nitrogen at 1L/min flow and perform 2 measurements to insure the device is completely devoid of oxygen.
  2. Press and hold the START button, then press and hold the SAVE button until the calibration screen appears.
  3. A screen will appear with the text “Recalibration 0%?” followed by “NO”.
  4. To exit recalibration, press and hold the SAVE button. You will read “Nothing Changed.”
  5. To begin recalibration, press the START or SAVE button momentarily to switch the display to “YES”.
  6. Once “YES” is shown calibration has started. Wait 5 minutes. Press and hold the Save button until the screen shows “recalibrated.” You will be returned to the main screen and the oxygen level should now show 0.00%. If not, set the zero point again.

A video with these instructions is available beginning at Part 1 for setup, then skip to Part 3 (2:27).
https://www.youtube.com/watch?v=FYy08FOLNYc

IMPORTANT!
Zero point adjustment may only be carried out with class 5.0 or purer nitrogen or the oxygen measurement results will not be accurate.After zero point adjustment the original factory values are overwritten. If the adjustment has not been carried out correctly (impure nitrogen, fresh air calibration, etc.) the original factory values can be restored by following the instructions on this video beginning at Part 2 (0:48). https://www.youtube.com/watch?v=FYy08FOLNYc

Maintenance

Cleaning

  • The unit can be cleaned externally with isopropanol and a fine, soft cloth. Never immerse the device in water or other liquids.
  • Do not clean the unit with ultrasonic cleaner.
  • Do not fill with solvents, alcohol, or cleaning agents, they may damage the device.

Battery

  • LiPo batteries are used for the TecPen. The duration per battery charge depends on how and under what circumstances the devices are used. Charge the battery using the supplied USB cable. Once the battery is fully charged, remove the battery from the charging cable.
  • Do not charge the battery longer than necessary (overnight). This may cause the charging capacity to decrease more quickly, the battery to overheat or become defective.
  • Should the charging capacity decrease unusually quickly within one year or with a maximum of 300 charge cycles (whichever is reached first) due to a production error, please contact GasLab.

ISO Certification

The instrument must be returned annually to the manufacturer for recalibration in accordance with ISO:9000 certification.

Device Status Codes

The device state is indicated by a status code that can be read from the menu. These codes are useful for troubleshooting.

  1. Normal operation
  2. Not specified
  3. UART buffer overrun
  4. Temperature sensor error
  5. Not specified
  6. Internal SD card is not recognized

IMPORTANT

  • If status codes 1, 2, 4 or 5 appear, switch off the device, wait a few moments and then switch it back on.
  • If the same status code appears after switching it back on, contact the manufacturer immediately
    Note: If the TecPen displays status code 3 during operation turn off the device and contact GasLab immediately.

Troubleshooting Guide

| Problem| Solution
---|---|---
1| The Display measurement shows

**** after a sample

| The oxygen level read is higher than 5% oxygen. The TecPen will not measure oxygen in fresh air (21%). The over range value is saved when logged
2| The PWR lights do not light up after switching on| Does the display show the main screen?

The LED light may be broken, but unless there is an error on the LCD in most cases the device is not charged.

3| The Display shows BATT| Battery is uncharged.
4| The Display remains black after switching it on| Charge the device and try to switch it on again

Did you disconnect your device properly from the

Computer the last time?

Yes: The battery might be empty. No/ I do not know: Contact GasLab.

6| The device is connected to the power supply, but the BT light does not light up| If the device is switched off while connected to a power outlet, the BT light does not light up while the device is charging.
7| The device is connected to the

Computer, but I cannot find it

| You must switch on the device before connecting to the computer. If the device is switched on but you cannot find it, check for available  COM ports.
8| The measured value is not saved| Did you press the SAVE button? Did you observe the

LOG sign on the display?  If yes, contact GasLab.

9| The log file shows a wrong date

The log file shows a wrong time

| Check the date  and time on the device is set.  To change date/time see the manual.
10| The measured value varies between two measurements| Give the TecPen time to acclimatize between different rooms or temperatures.
11| The device is not measuring| Example: Can you hear the pump?

Yes: Did you follow operation procedure? Yes: Contact GasLab.

Is the display showing MEAS? No: Start the measurement.

Yes: Check screen 4 if there is a value of amplitude higher than 0. If yes, contact GasLab.com

Specifications

TS-System TecPen MAP O2

Measurement range

Resolution

| 0-5%
Range

0-0.5%

0.5-2.5%

2.5-5%

| Accuracy

± 2% Mev**

±3% Mv*

±5% Mv*

0.001%
Response time at 25°C/ 77°F| <150ms
Pump flow max.| 67 mL
Temp. range Min./Max| -10°C/ +60°C
14°F/140°F
Medium| Gas
Power supply| 5V USB and LiPo battery
Rechargeable battery lifetime| > 3 hours per charge
Data Interface| USB
Temperature compensation| 10-30°C
50-86 °F
Display| OLED Display
Cleaning housing| no organic solvents. 40% EtOH
Parts touching sample| St.1.4404/ PTFE/ Glas
Case| aluminum anodized
Protection| IP65
Full service interval by Manufacturer (includes necessary retrofitting)| 3 years
Life time of sensor spot  (protective atmosphere; 0- 40°C)| 24 months
Sensor cap replacement interval| 2 years
Recommended zero-point adjustment interval by customer| 6 months
Warranty| 1 Year ex works

  • **Mev = measured end value
  • *Mv = measured value
    TecSense sensors are CE, FCC, IC, KC and MIC marked, meet the requirements of ASTM F271408, and the European and Canadian standards for food safety. The company is DIN ISO 9001/2015 certified.

Support

The quickest way to obtain technical support is via email. Please include a clear, concise definition of the problem and any relevant troubleshooting information or steps taken so far, so we can duplicate the problem and quickly respond to your inquiry.

Warranty

The sensor comes with a 1-year warranty starting from the date it was shipped to the buyer. The TecPen sensor cap comes with a 90 day warranty. For more information visit our website: https://www.gaslab.com/pages/terms- conditions

Contact Us

If the troubleshooting guide above does not help you solving your problem or for more information, please contact us using the information below. [email protected] , 386-872-7668 (M-F 9:00am–5:00pmEST) www.gaslab.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>