BASIC OPERATION
TOUCHSCREEN
MAINTENANCE SCHEDULE
Weekly
| Clean the chamber and trays.
---|---
Drain water reservoir.
Disinfect external surfaces.
Inspect and clean the chamber filter
(or replace if necessary).
Monthly
or
160 Cycles
| Inspect the water reservoir filter
(clean or replace).
---|---
Clean the external fill water tank.
Every
6 Months
or
1,000 Cycles
| Perform all monthly cleaning tasks.
---|---
Replace bacteriological air filter.
Replace drawer seal.
WARM-UP MODE
- STAND-BY LOW – For low to average use: 70°C / 158°F
- STAND-BY HIGH – For high use: 120°C / 248°F
- STAND-BY OFF (default) – For infrequent use. The wait time will be longer (up to 12 minutes from a cold start)
INSTRUCTIONAL VIDEOS
LED COLOUR
LOCK ICON
- Drawer is locked; the chamber is still under pressure
- Drawer remains locked; not safe to open
- Drawer is safe to open
- Drawer remains unlocked; unsafe to start cycle
Spare Parts and Accessories
Spare Parts and Accessories
01-116292S| Air Filter
01-116293S| Drawer Seal, 6L
01-116294S
__
| Tray, 6L
01-116296S| Exhaust Tube
01-116297S| Extended Exhaust Tube
01-116298S| Direct-to-Drain
01-116299S| Water Filling
01-116300S| Waste Bottle
01-116301S| USB Memory Stick
01-116302S| Water Reservoir Cap
SCI-BDSK134V| Bowie-Dick Test Kit
97902001| Helix Test Kit with 100 Chemical Indicators
97902002| Helix Text Kit with 400 Chemical Indicators
01-110282S| Power Cord EU 16A/250V
01-110285S| Power Cord UK IND plug 16A/250V
01-116496| StatMat Silicone Mat STATIM B
01-116497S| Reservoir Chamber Filter
01-116498S| Pouch Rack Kit
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>