SciCan HYDRIM C61W G4 Instrument Washer User Guide
- June 5, 2024
- SciCan
Table of Contents
SciCan HYDRIM ®
C61W G4 Quick Reference Guide
The operator’s Manual should be consulted before using this washer.
1-800-572-1211
BASIC OPERATION
-
Power on the unit.
The button is inside the kick plate. -
Load Instruments and close the door.
-
Select cycle.
-
Press START.
Refilling the Water Softener Salt Reservoir
- Unscrew lid.
- Fill to the top (max. 0.5 kg / 1.1 lbs.)
- Screw lid on tightly.
- Run a complete cycle with no instruments.
IMPORTANT!
If the salt container is being filled for the first time, it MUST be filled
with at least 1 liter of water before adding the salt.
Only tablet salt should be used in the softening unit. Fine-grained salt such
as table salt should NOT be used as it could clog the unit.
The water softener salt is also available from SciCan. Please contact your
local dealer.
TOUCHSCREEN
| Rinse and Hold Cycle
Prevents soil from drying on instruments.
Incomplete loads that will not be washed within the hour.
NOT a cleaning cycle.
---|---
Regular Cycle
Moderately soiled loose instruments.
Heavy Duty Cycle
Heavily soiled instruments and cassettes.
| Door
Lock
---|---
Network
Connection
Water Softener/ Detergent Level/Door Status
| User/Technical
Menu
---|---
USB
Enter
Back
Cleaning the Chamber
Pour 0.5L of vinegar or citric acid-based instrument neutralizer into the chamber and follow these steps:
IMPORTANT! Do a Chamber Cleaning cycle every 25 cycles. Failure to clean the chamber can promote corrosion.
Replacing the HIP ™ Cleaning Solution
Recommended precautionary measures with regard to safe handling of HIP solutions:
- Wear undamaged surgical gloves and eye protection.
- After handling the product, remove and dispose of the gloves and wash hands prior to removing eye protection.
TROUBLESHOOTING
Refer to Operator’s Manual for a more comprehensive list. Call for service if suggested recommendations do not resolve the error message.
CF2 Chamber filling failure | Check that the water hoses are not kinked. |
---|
Check that water shut-off valves are opened.
Call for Service.
CF23 Top wash arm low RPM
Top wash arm speed too low.| Check for obstructions to the rotation of top
wash arm.
Check that holes in wash arms are not plugged.
CF24 Bottom wash arm low RPM
Bottom wash arm speed too low.| Check for obstructions to the rotation of the
bottom wash arm.
Check that holes in wash arms are not plugged.
Spare Parts and Accessories
CS-HIPC-U | HIP- Ultra Cleaning Solution (8 x 750 mL) |
---|---|
01-112594S | Water Softener Salt |
01-108305S | Water Test Kits |
01-114390S | HEPA Filters |
WTI-SK | Wash Test Indicator – Starter Kit |
WTI-RK | Wash Test Indicator – Refill Kit |
**** Terminal sterilization of instruments MUST follow processing in the HYDRIM C61W G4.
www.scican.com
MSDS available for download atwww.scican.com
HYDRIM is a registered trademark and HIP is a trademark of SciCan Ltd.
Manufactured by SciCan Ltd. 1440 Don Mills Road, Toronto, Ontario, Canada, M3B
3P9.95-115868 NA EML R2
Documents / Resources
|
SciCan HYDRIM C61W G4 Instrument
Washer
[pdf] User Guide
HYDRIM C61W G4, Instrument Washer, HYDRIM C61W G4 Instrument Washer
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>