SciCan HYDRIM ®
M2 G4 Quick Reference Guide
Operator’s Manual should be consulted before using this washer-disinfector.
+49 (0) 7561 98 343 – 0

BASIC OPERATION

  1. Power unit ON.

  2. Load Instruments.

  3. Close the door.

  4. Select cycle and press START.

Refilling the Water Softener Salt Reservoir

  1. Unscrew lid.
  2. Fill to the top (max. 1 kg / 2.2 lbs.)
  3.  Screw lid on tightly.
  4. Run a complete cycle, with no instruments.

IMPORTANT!
If the salt container is being filled for the first time, it MUST be filled with at least 1 litre of water before adding the salt.
Only tablet salt should be used in the softening unit. Fine grained salt such as table salt should NOT be used as it could clog the unit.
If salt was spilled while refilling, run a cleaning cycle.
The water softener salt is also available from SciCan.
Please contact your local dealer.

TOUCHSCREEN.

| Rinse and Hold Cycle
Prevents soil from drying on instruments.
Incomplete loads that will not be washed within the hour.
NOT a cleaning and disinfection cycle.
---|---
Heavy Duty Cycle with Disinfection
Heavily soiled instruments and cassettes.
| Door
Lock
---|---
Network
Connections

Water Softener/ Detergent Level/Door Status
| User/Technician Menu
---|---
USB
Enter
Back

Cleaning the Chamber

Pour 1.0L of vinegar or citric acid-based instrument neutralizer into the chamber and follow these steps:SciCan HYDRIM M2 G4 Automated Instrument
Washers - fig 4

Replacing the HIP ™ Cleaning Solution

Recommended precautionary measures with regard to safe handling of HIP solution:

  • Wear undamaged surgical gloves and eye protection.
  • After handling the product, remove and dispose of the gloves and wash hands prior to removing eye protection.

TROUBLESHOOTING

Refer to Operator’s Manual for a more comprehensive list. Call for service if suggested recommendations do not resolve the error message.

CF2 Chamber Filling
Water is not filling the chamber| Check that the water hoses are not kinked.
---|---
Check that water shut-off valves are opened.
Call for Service.
CF15 Water Reservoir Overflow
Overflow switch triggered| Restart the program.
Call for Service.
CF20 Condenser System
Unit is not able to cool down| Check that the water hoses are not kinked
Call for Service.
CF21 Dosing System
Detergent is not dispensing| Tighten detergent cap.
Check for leaks.
Call for Service.

Spare Parts and Accessories

CS-HIPL HIP Cleaning Solution (2 x 3.8L)
01-112594S Water Softener Salt
01-108305S Water Test Kit
01-114390S HEPA Filter

**** Terminal sterilization of instruments MUST follow processing in the HYDRIM M2 G4.

IMPORTANT!
For non-invasive medical devices, SciCan recommends terminal sterilization after processing in the SciCan HYDRIM M2 G4 according to local guidelines. For invasive medical devices, such as dental handpieces, terminal sterilization must be performed after processing in the SciCan HYDRIM M2 G4. FOR DETAILED INFORMATION PLEASE SEE PROVIDED OPERATOR’S MANUAL

MSDS available for download atwww.scican.com
96-113757 EU EN R2. HYDRO is a registered trademark and HIP is a trademark of SciCan Ltd.
Manufactured by SciCan Ltd. 1440 Don Mills Road, Toronto, Ontario, Canada, M3B 3P9.www.scican.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

SciCan User Manuals

Related Manuals