KONBEST KB-9690K Rechargeable &Portable Iniermttent Pneumatic Compression User Manual

June 5, 2024
KONBEST

KONBEST KB-9690K Rechargeable &Portable Iniermttent Pneumatic

Compression User Manual
\

ACCESSORIES

  1. 2pcs 1eg cuffs with connecting hose
  2. Main Unit
  3. Charger
  4. Handbag
  5. Arm sleeve (OPTIONAL)
  6. Waist rap(OPTIONAL)

LED TOUCH CONTROL PANEL

PUT ON THE CUFFS

QUICK START GUIDE

Before using please check the battery power by pressing power button- to try starting up the pump; if it could not start up,it means the pump is out of battery,please use charger to charge for it.The indicator light on adapter will be lighted red color.The color wiell become to greencolor when full charging and battery indicator light show full.

  1. .Securely cornect air hose to cuffs.
  2.  Securely connect the main plug of air hose to pu.
  3.  Put on the cuffs comfortably.

QUICK START GUID

  1. Press the power button – to rnake the pump on standby mode.
    Note:The power button is at the front side of the pump. Press 0 can turn off the pulmp.

  2.  Set the desired time by choosing button (10~90mins)
    Note:Default is 30mins .One-time short push to reduce/increase 5mins. When time runs out, the pumg shuts off automatically.

  3. Set the desired pressure level by pushing button (30- 260mmg)
    Note: Default is 80mmHg. One-time short push to reduce/increase 5mmHg. The purp stops work below 80mmHg.

  4. Select desired massage mode by pushing any of the
    buttons.
    Note: Default is mode A. Only under the state of pause, the mode could be switched to the desired mode

  5. Selsct desired massage area by pushing the buttons.
    Note: Default is all chambers light on.Only under the state of pause,the area could be turned off/on to tha desired position and relevant light . off/on.

  6. Prsss the ON/PADSE button. 10.Note : After all the Settings are completed,press the start key to run formally.
    Classic Relax Mode:routing ①→②→③-④→③,chamber④inflates, then defates,next inflates,then deflates,same way till chamber④,then repeat.①②③④⑤⑥⑦⑧⑨

MASSAGE MIODES FOURTYPES

B:
Classic Relax Mode:routing ①→②→③-④→③,chamber④inflates, then defates,next inflates,then deflates,same way till chamber④,then repeat.①②③④⑤⑥⑦⑧⑨
Sequential Compression Mode: routing ①-①②-①②③-①②③④-①②③④⑤ →①②③④⑤⑥,chamber ①inflates,then hold the pressure,②starts inflating,
①② hold pressure at the same time,then③start inflating,same way till after ⑥ Chamber①②③④⑤⑥deflates together.Then repeat.

C

Pulse Mode:routing ①-①②-①②③-②③④→③④⑤-④⑤⑥.Chamber infla- tes then deflates for 3 times,then hold the pressure;next② holds the pressure after in flating then defiates for 3 times,while 1still keeps pressure;③holds the pressureafter infating then defates or 3 times,while ①②still keeps pressure;④holds pressure after inflating then deflates for 3 times,while ②③ still keeps pressure but ①deflates,working
as this way.

D

Note: Under the mode D, for the FIRST CYCLE, it works as mode B to warm
up; from 2nd cycles on, it works as PULSE mode D from 2nd cydles.

In this mode, all chambers inflate together,then holds pressure,and then together , then the cycle repeats.
In this mode, all chambers inflate together,then holds pressure,and then deflates

After Use

Just needs to disconnect the hose from the machine by pulling the main end out. The cuffs connected with hose are always kept and used as a whole;
When use, it just need insert the other end of the hose the machine had better not disconnect the hose from the cuffs frequently; Abuse may cause damage to the pins or hose.
Note: When travel with the product, it could disconnect the hosefrom the cuffs to save space. Notice: No pulling the hose. Disconnect the hose from cuff by holding pins to pull up.

PUMPSP ECIFICATION
Weight: 1 .6KG size: 21138.6CM Power Consumption:49W Charger Input:AC100v~ 240v Charger Output: DC12.6V;4A .

After using, when storing the pump,be careful to pull the hose plug of from the pump by pressing clips at the double sides of the hose plug, to avoid breaking the plug.
.Avoid removing the hoses on the boots all the time. When you are traveling with your boots,you coulddis- assemble to optimize the space in the bag, but i t is not necessary . Avoid tearing the hoses when you disassem- ble by holding the pins to pull up.

GENTLE REM INDER

CLEANING AND MAINTENANCE
Turn off pump before cleaning.
Don’t wash the machine and cuffs.
Wipe with a dry or damp cloth.
Keep in a clean and dry place.
Protect the equipment from direct sunlight or heat,such as gas stove.

WARNINGS

Do not unzip the cuffs when the unit is on. Iways turn the power off before unzipping the cuff.
Do not get the pump wet.
Make sure the pump is on a flat surface.
When using the machine, do not move with the cuffs.
The equipment cannot be repaired and maintained during use. Please consult your physician before use.This product is not intended to diagnose,treat, or cure any diseases.