SKC 232-01 Single Polypropylene Fitted Bags
The 232, 236, 252, and 262 Series sample bags feature a single polypropylene fi tt ing that is used for both fi lling the bag and removing the sample for analysis. The fi tt ing contains a syringe port with septum and a hose connection and acts as a valve to open or close the hose connection.
Guidelines for Bag Sampling
- Ensure that the bag material and fi tt ings are appropriate for the compounds to be sampled.
- Use only PTFE tubing to connect the sample bag to the pump to prevent sample loss by adsorption on tubing walls.
- Before using, flush the bag thoroughly with purifi ed air or nitrogen.
- Analyze the sample within 24 to 48 hours. Long-term storage of air-contaminant mixtures in bags is not recommended.
- Do not ship sample bags by air unless the cargo cabin is pressurized.
- Inflate bag properly (see Proper Bag Inflation for details).
- Due to the physical properties of FlexFoil® bag material, minor crimps and creases may appear on the bags during SKC manufacturing, quality testing, and user handling. These marks only affect the aesthetics of the bag and not its performance. FlexFoil bags exhibiting minor crimps and creases are not subject to replacement or refund as functionality is not affected.
Notes and Cautions
- Maximum Operating Temperature Do not exceed these specifi cations:
- 236 Series Bags: 140 F (60 C) due to the SamplePro® FlexFilm bag material temperature rating
- Do not use sample bags to sample compounds with boiling points > 249.8 F (121 C).
- SKC sample bags are designed for sampling air at atmospheric pressure only. Att empting to pressurize the bag can result in bag rupture and sample loss. Do not ship bag samples by air freight in non-pressurized cargo cabin. Bags can burst under such conditions.
- All federal and state packaging and transporting regulations apply.
- SKC bag fitt ing valves are extremely durable but are not intended for use as handles or hanging devices. This type of handling may damage the seal, causing leakage. This is considered to be misuse and will void the SKC warranty.
- Failure to follow warnings and cautions voids any warranty.
- Sample bags are designed for single use only.
Operation
- Flush the bag at least 3 times with purifi ed air or nitrogen before use.
- To fill a bag, connect PTFE tubing from the exhaust port of an air sample pump to the hose connection on the bag (the side stem protruding from the polypropylene fi tting).
- To open the valve, hold the side stem and turn the entire upper portion of the fi tt ing (including the brown or white septum cap and the white septum body to which it is attached) counterclockwise one revolution. Turn on the pump and sample (see Proper Bag Inflation).
- When sampling is complete, turn off the pump. To close the valve, hold the side stem and turn the entire upper portion of the fi tt ing (including the brown or white septum cap and the white septum body to which it is attached) clockwise until it is snug.
- To withdraw samples using a needle and syringe, carefully insert the needle into the syringe port in the center of the brown or white septum cap and pierce the septum.
Proper Bag Inflation
Avoid filling any bag more than 80% of its maximum volume. At the end of the
sampling period, turn off the pump and close the valve on the bag.
Filling Bags Using Negative Pressure
The SKC Vac-U-Chamber is a rigid air sample box that allows the direct filling
of an air sample bag using negative pressure provided by most personal air
sample pumps. When using the Vac-U-Chamber, the air sample enters the bag
directly, without passing through the pump. This eliminates the risk of
contaminating the pump or sample. The Vac-U-Chamber has rigid walls that will
not collapse under vacuum conditions, thus eliminating possible errors. All
surfaces in contact with samples are constructed of inert materials.
- Small Vac-U-Chamber (supplied without pump) with polypropylene fittings; suitable for use with 1-liter bags ………………………………………………………..Cat. No. 231-940
- Large Vac-U-Chamber (supplied without pump) with stainless steel fittings; suitable for use with 8-liter bags ………………………………………………………..Cat. No. 231-939
- Extra-large Vac-U-Chamber (supplied without pump) with stainless steel fittings; suitable for use with 10-liter bags ………………………………………………………Cat. No. 231-944
- Jumbo Vac-U-Chamber (supplied without pump) with stainless steel fittings; suitable for use with 25-liter bags ………………………………………………………Cat. No. 231-946
SKC sample bags are available in Tedlar®, Standard FlexFoil®, and FlexFoil
PLUS with fittings made of stainless steel, polypropylene, or PTFE in sizes
ranging from 0.5 to 100 liters. SKC also manufactures custom sample bags.
Contact SKC at 724-941-9701 or
skcorder@skcinc.com or go to
www.skcinc.com/media/documents/85177.pdf.
For sample bag stability reports, visit
www.skcinc.com/media/documents/1805.pdf.
Accessories
- PTFE Tubing, fi ts over bag fi tting and pump exhaust, 3/16-inch ID, 1/4-inch OD ………………………………………………… 10 feet ….Cat. No. 231-9-23
- PTFE Tubing, fi ts inside bag fi tting, 1/16-inch ID, 1/8-inch OD ………………………………………………… 10 feet ….Cat. No. 231-9-21
- PTFE Tubing, fi ts Vac-U-Chamber sample inlet port, 1/4-inch ID, 5/16-inch OD …………………………………………………. 10 feet …..Cat. No. 231-937 50 feet …..Cat. No. 231-924
skcinc.com Form 3781
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which
provides SKC’s sole liability and the buyer’s exclusive remedy. To view the
complete SKC Limited Warranty and Return Policy, go to
skcinc.com/warranty
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>