ARIES Filter work 931213 High Purity Water System Installation Guide
- June 5, 2024
- ARIES Filter work
Table of Contents
- HEALTH AND SAFETY
- SYSTEM OVERVIEW
- FLOW DIAGRAM
- INFLUENT WATER SPECIFICATIONS
- INSTALLATION
- DIMENSIONS
- SYSTEM CHECKLIST AND UNPACKING GUIDE
- MAIN CHECKLIST
- MOUNT INSTRUCTIONS
- SETUP
- INSTALLING THE CARTRIDGES
- INSTALLING THE UV BULB (OPTION)
- START-UP PROCEDURE
- DISPLAY AND KEYPAD FUNCTIONS
- MONITOR OPERATION AND PROGRAMMING
- MAINTENANCE
- SANITIZATION
- CONSUMABLES & SPARE PARTS
- TROUBLESHOOTING
- WARRANTY
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
- 2 Phillips Head Screw Driver
ARIES
High Purity Water System
Installation & Operation Manual
ARIES USER’S MANUAL
This manual contains details on the installation and operation of your Aries
Lab Water system. For optimum performance and safety please read and follow
these instructions carefully.
You have a quality, world-class water treatment system that will provide
ultrapure water for many years of trouble-free operation. Years of engineering
and experience went into the design and manufacture of this instrument which
will dispense water exceeding Type I water purity standards.
In order that we may keep you informed regarding your unit, please take a
moment to complete the information below that will assist with support should
it be needed.
Model Number:__
Serial Number:_
Contact:__
Cartridge #1:____
Cartridge #2:____
Cartridge #3:____
Options Installed:_| Date:____
Dealer:
Phone:
Part Number:__
Part Number:__
Part Number:__
---|---
ARIES HIGH PURITY WATER SYSTEM
HEALTH AND SAFETY
Installation and service of the equipment should be performed by authorized
and trained personnel. Safety measures and local guidelines must be followed.
Use good practices wearing personal protective equipment (PPE) during the
installation and service of the equipment.
Electricity – Aries Filterworks lab water equipment should be powered by a
grounded GFI electrical connection. Do not open the system unless the system
is unplugged. Before any installation or service is performed on the
equipment, make sure the power is isolated. To unplug the system, do not pull
directly on the cord but grip the plug and remove it from the outlet.
Ultraviolet Light – Aries Filterworks lab water systems may include an
ultraviolet light lamp in a stainless-steel housing. Make sure all power is
removed prior to accessing the side panel to perform any UV maintenance. Make
sure gloves are used when handling the UV bulb. (Gloves must be worn when
handling the UV bulb. Which protects the integrity of the bulb) Do not look
directly at the UV lamp while the system is operating.
Water Supply – Good piping practices should be followed by feeding water to
the lab equipment. Isolation valves, pressure gauges, and bleed valves are
recommended for the main water supply. Pressure should be regulated to meet
the minimum and maximum specifications stated for the equipment.
Sanitization – Make sure PPE is worn during any sanitization process. The
Aries system uses chlorine as a disinfectant. Chlorinated water will be
discharged during sampling and flushing sequences. Avoid breathing fumes and
ensure the area is sufficiently ventilated.
SYSTEM OVERVIEW
The Aries Lab water system delivers Type 1, 18 MO quality water on demand.
Water is purified in a staged process consisting of high-purity long exchange
resins to remove dissolved minerals and internal recirculation to maintain
purity. At discharge, a fina 0.2pm filter removes particulates and bacteria to
attain CLSI / ASTM Type 1 water specifications. Additional optional
technologies are incorporated depending on the model. The resistivity monitor
provides feedback on the final water quality.
The Aries system is designed to recirculate through multiple technologies to
maintain purity. A timer controller runs the system automatically for 30
minutes every 2 hours to maintain the integrity of the system. As the system
transitions from “Operating” mode to “Stand By” mode, the recirculation pump,
solenoid, resistivity display, and UV (if installed) will be automatically
turned off via the timer. A WAKE button on the side of the system will reset
the cycle timer back to the “Operating” condition.
Options available with the Aries lab system are UV, recirculating dispensing
gun, secondary direct feed port, and wall mounting bracket.
FLOW DIAGRAM
**MODEL TYPES***|
---|---
ARS-102| Aries Lab Water System with 0.2 Micron Outlet Final Filter
ARS-105| Aries Lab Water System with 0.05 Micron Outlet Ultrafilter for
Endotoxin Removal
ARS-102-UV| Aries Lab Water System with installed UV and 0.2 Micron Outlet
Final Filter
ARS-105-UV| Aries Lab Water System with installed UV 0.05 Micron Outlet
Ultrafilter for Endotoxin Removal
- Submicron filter included. Start-Up Cartridges Sold Separately.
**MODEL OPTIONS**|
---|---
ARA-UV| Combination UV for TOC and Bacteria Reduction (installed)
ARA-DC| Dispensing Gun with Final Filter – Factory Installed option. A
recirculating gun is mounted on the side of the unit with approximately 5 feet
of span. The user manually dispenses water.
ARA-DF| Direct Feed – Factory installed option. Secondary Port on the side of
the system. Used to feed ancillary equipment such as analyzers and scientific
equipment. The internal pressure switch automatically keeps the system
operating during a water draw condition. A minimum flow of 0.1 Ipm is
required.
Note, that there is no final filter on this port. A downstream final filter
may be considered.
ARA-WB| Wall Mounting Bracket – Factory Installed option. Wall mounting
bracket supplied with additional support base plate to the main system.
UV was previously offered as an option with a separate part number.
**START UP CARTRIDGES***|
---|---
VPK-4010| RO Feed Start-Up Cartridges – (3) High Purity Series Mixed Bed
Cartridges
VPK-3805| Tap Water Feed / Service DI – (1) Organics Pretreatment & (2) Mixed
Bed Cartridges
- Sold Separately.
INFLUENT WATER SPECIFICATIONS
PARAMETER | REQUIREMENT |
---|---|
Source/Type | Reverse osmosis (RO) or Service deionization (SDI) is preferred. |
0.2pm particulate prefiltering is recommended for SDI and tap water feeds.
Conductivity| RO: < 20pS/cm
SDI: > 1 MO (Resistivity)
TOP (Using appropriate conditioning as necessary)| RO: < 50 ppb
SDI: < 200 ppb
Temperature| 20° F to 100°F (5° C to 38°C)
Pressure| 30 psig minimum/90 psig maximum (1.38 to 6.21 bar)
Fouling Index| Silt Density Index: < 3
Dissolved Carbon Dioxide| < 30 ppm
Free Chlorine| < .05 ppm
EFFLUENT WATER SPECIFICATIONS
PARAMETER | REQUIREMENT |
---|---|
Resistivity | > 18 MO-cm (Resistivity) |
Microorganism | < 10 CFU/mL |
< 1 CFU/mL (with 0.05 micron ultrafilter outlet)
TOC| < 10 ppb (with UV option & 50 ppb influent)
Chlorides| < 1 ppb
Sodium| < 1 ppb
Endotoxin| < 0.03 EU/mL (with 0.05 micron ultrafilter outlet)
CONNECTIONS
– Inlet – 3/8″ OD tubing – Outlet – 1/4 FNPT
– Direct Feed Port – 3/8″ CPC Female Bulkhead with CPC Male x 3/8″ OD Tube
Compression
INSTALLATION
It is recommended to review the pre-site requirements before proceeding with the installation. Read through the entire manual prior to the start of the installation. Review all paperwork of items that have been ordered and compare them to the items that have shipped. If any parts are missing, notify the factory or your sales contact.
DIMENSIONS
SYSTEM CHECKLIST AND UNPACKING GUIDE
Special care should be used for the shipping and delivery of these systems. Similar care should be considered when removing the system from the box. Items should be inspected and itemized upon removal to ensure they have not been damaged in transit. Immediately report any noticeable damage to the shipping carrier.
MAIN CHECKLIST
System
-
ARS-102 – Main System with 0.2-micron capsule filter
Or -
ARS-102-UV – Main System with UV module installed. Bulb shipped loose. 0.2-micron filter included.
-
ARS-105 – Main System with 0.05-micron ultrafilter
-
ARS-105-UV – Main System with UV module installed. Bulb shipped loose. 0.05-micron ultrafilter include
Options
- Wall Mounting Bracket Plate (ARA-WB).
- Installed Dispensing Gun (ARA-DG). Dispensing Gun Tubing connected to side system with bracket. 0.2-micron filter (PF-00-6402) included with ARA-DG
- Direct Feed Port (ARA-DF). Side port with CPC 3/8″ Female panel mount
Included Accessories
- Final Filter as offered with the specific model
- Outlet angle valve with Teflon tape
- Certificate of Conformance
- Hose nipple adapter for ARS-105 ultrafilter
- UV bulb for UV option
- A final filter for the ARA-DG dispensing gun option
- Quick connect CPC 3/8″ Male x 3/8″ Compression fitting for direct feed option
Recommended Items needed for Installation
- (2) 3/8″ Quick Connection Isolation Valve for upstream isolation and depressurization (not included)
- 0-100 psi pressure gauge (300-grade stainless steel) for diagnostics (not included)
- Tube Cutter
-
2 Phillips Head Screw Driver
- Teflon Tape
- Basic Hand tools (adjustable wrench, pliers, socket wrench, etc.)
- Tie-wraps
Optional Items
- Drill for Wall Mount Level for Wall Mount
- Measuring Tape
- Small electronics size flat screwdriver (Electronics diagnostics)
- 6 feet of 3/8″ Hose for commissioning
INSTALLATION & OPERATION MANUAL
MOUNT INSTRUCTIONS
The Wall mount must be capable of supporting a minimum weight of 50 lbs (22.7
kg). The wall mount bracket has holes on 16″ centers to be secured to studs.
If studs are not available, a plywood sheet mounted to the wall is
recommended. Make sure holes are level.
Mount bracket with small slots in tabs pointing up.
SETUP
Select a location for the unit which is near the water supply and electrical
supply connections. If the water supply does not already have an isolation
valve for servicing the Aries Lab System, install one ahead of the unit
connection.
Install the right-angle valve into the front of the unit outlet.
Do not install the final capsule filter into the outlet valve until the final
step of a startup.
Using the tube fittings provided with the unit, connect the unit inlet to the
water supply.
(DO NOT open the supply isolation valve at this time.)
INSTALLING THE CARTRIDGES
there are three (3) cartridge positions. The water flows from left to right.
Any pretreatment filters must be installed in position #1 (furthest from the
display) and a polishing filter must be installed in position #3 (closest to
the display).
Locate the water purifying cartridges selected for your system. Unwrap the
filters, remove the shipping plugs and affix the labels.
Place the large opening at the bottom of the filter onto the plunger. Push the
cartridge down and tilt back until the filter is aligned with the top post.
Allow the filter to slide up over the upper fitting and gently rotate left and
right to insure a good seal with the top and bottom 0-rings.
Repeat the above procedure for filters #2 and #3.
Filters should always be installed in the 1,2,3 order and removed in revere –
3,2,1.
INSTALLING THE UV BULB (OPTION)
WARNING – ELECTRICAL SHOCK HAZARD. Make sure all power is removed from the
system.
Remove the top cover by removing the Phillips head screws that attach it.
Gently lift the top cover up and set it aside. The UV chamber will be visible
at the top of the lab unit.
Unclasp the clamps that hold the stainless-steel body of the UV light. Gently
lift the body up about 2 inches to allow clearance for bulb removal.
Unscrew the blue-collar which is threaded into the stainless-steel body.
Remove 0-ring gasket and stainless-steel washer.
Hold the UV lamp by the plastic ends ONLY. To avoid oils and fingerprints, it
is recommended to use gloves when handling the UV bulb. DO NOT touch the
glass. If the lamp should become oily or dirty, clean it with a clean soft
tissue or cloth wetted with isopropyl alcohol.
First, slide the 0-ring gasket of the male, pronged end of the UV bulb until
it seats. Second, place the washer on top of the gasket. Third, place the
blue-collar on top of the gasket.
Connect the female wire connection to the UV male prong until it is fully
seated.
Install the bulb into the stainless-steel chamber. Gently thread the blue-
collar until hand tight.
Gently push UV body back into the bracket. Squeeze clamps until they click
firmly about the UV body.
Replace the top cover and turn the unit on.
START-UP PROCEDURE
Plug in the unit. The electrical outlet should be properly grounded and fused.
It is recommended that the system run on its own dedicated electrical circuit.
Open the outlet valve and place a container underneath.
Slowly turn on the water supply to the unit and correct any leaks at the
supply connections.
Turn the unit on at the power switch and allow water to flow. Water will be
visible filling the cartridges up-flow from the left to the right. Allow the
air to bleed from the system through the outlet valve into the open container.
When all air has been purged from the system close the outlet valve.
Install the final capsule filter on the outlet valve. Open the outlet valve
and bleed off any residual air in the filter by opening the thumb screw on the
side of the filter. Close the thumb screw when water is present and close the
outlet valve.
Check connections and cartridges for leaks.
“INITIAL POLISHING” OF NEW CARTRIDGES
The cartridge may require up to 24 hours to rinse up to 18 Megohms.
4 liters of water should be run through the filters as an initial purge of the
system.
We recommend a sample be collected in an appropriate sterile sample container
and tested for water purity. Periodic laboratory water testing to confirm the
effectiveness of the treatment process is the owners’ responsibility.
DISPLAY AND KEYPAD FUNCTIONS
PRIMARY BUTTONS
Wake – Wakes the unit up from the Standby mode and initiates the
30-minute polishing cycle.
( NOTE: The Wake key must be pressed longer than 3 seconds or the unit
will not wake up.)
Power- Turns on the power, and start the timer controller. Polishing –
Green LED is on when the unit is running, Solenoid, pump, and meter are on.
Stand By – Amber LED is on when the unit is in Stand By mode. Solenoid,
pump, and meter are powered down.
RESISTIVITY MONITOR
MONITOR OPERATION AND PROGRAMMING
TO CHECK WATER TEMPERATURE
Press and Hold TEMP
– The water temperature is displayed while the button is pressed.
– The monitor returns to normal operation when released.
– The displayed temperature can be toggled from C to F by repeated pressing.
TO CHECK SET-POINT
Press and Release SET-POINT
-LO- will be displayed momentarily, followed by the currently stored
value. The Low set-point value is the point where the relay will energize.
-HI- will then be displayed followed by the High set-point value. The
High set-point value is the point where the relay will de-energize.
– The monitor will automatically return to normal operation.
TO ADJUST SET-POINT
Press and Hold SET-POINT then also press ADJUST Release both buttons
-LO- will be displayed momentarily, followed by the current Low set-point value with the LCD’s highest digit flashing.
– To skip this digit, Press and Release ADJUST again OR, Press SET-
POINT repeatedly until the desired number is achieved.
Then Press and Release ADJUST
– The next lowest digit will begin flashing. – Press SET-POINT to adjust
the digit – Press ADJUST to Skip.
Continue this sequence until all the digits are set and press ADJUST
-HI- will be displayed momentarily, followed by the current High set-point value with the LCD’s highest digit flashing.
– Press SET-POINT to adjust the digit
– Press ADJUST to Skip.
Continue this sequence until all the digits are set and press.
– The Monitor will automatically return to its normal operating mode.
MAINTENANCE
This unit was designed and manufactured to have the ability to produce the purest water that can be obtained for laboratory usage. It purifies water that meets or exceeds the specifications recommended by CLSI, and ASTM, for Type I Reagent Grade Water when it is operated and maintained according to these instructions.
- The ultraviolet lamp (if installed) must be changed once each year.
- The resin cartridges must be replaced when the purity meter indicates the water purity is below your requirements or every 6 months.
- The exterior surfaces may be wiped clean with a mild detergent and a soft cloth to keep the unit clean and new looking.
REPLACING THE FILTER CARTRIDGES AND FINAL FILTER
The cartridges are expendable items that have a finite capacity to deionize
and polish the water. When the digital display indicates the purity is less
than your minimum requirements, the DI resin cartridges have exhausted their
capacity to perform. The cartridge should be replaced every 6 months as a
matter of routine maintenance. The final filter should be changed every time
the cartridges are replaced.
- Turn the unit off using the power switch.
- Close feed water valve to the system.
- Place a container under the outlet valve to drain water from the system, which releases pressure. Open the outlet valve and leave it open until water stops draining from the system.
- Before removing DI cartridges, wrap the bottom of each plunger with paper towels or a towel to absorb the small water that will drain from the bottom of the filter once it is disengaged from the plunging apparatus. Quickly place the cartridge filled with water into a local basin or sink. Covering the bottom cartridge with your hand will help minimize the excess water.
- Remove the three cartridges starting with Cartridge #3. Push down on the cartridge gently and tilt forward. Quickly remove the filter and discard it in the( trash. There will be some residual water inside the filter. Water will come out of the outlet valve (still open) when the cartridge is depressed.
- Remove cartridges #2 and #1 in this order using the same procedure above. When the plunger is depressed some water will come out of the upper fitting next to the filter.
- Dry the unit and replace the cartridges as described in Section 5.2, Installing the Cartridge.
- The used cartridges may be disposed of in the same manner as any water filter cartridge.
- Replace the Final Filter with a new one each time the filter cartridges are replaced.
SANITIZATION
-
Use Sanitization Kit HPA-010.
-
Turn off the system and disconnect the power plug.
-
Open dispense valve.
-
Remove cartridges.
-
Remove 0.2-micron final filter.
-
Insert sanitization cartridges in the following order:
CARTRIDGE 1 – SANITIZER, VP-17-0001
CARTRIDGE 2 – BLANK, VP-17-0000
CARTRIDGE 3 – BLANK, VP-17-0000 -
Insert the power plug into an electrical outlet and turn on the system.
-
Draw off about two liters of water from the system, then close the dispense valve.
-
Allow the system to remain in recirculation for 45 to 60 minutes. *Note: The unit must remain in the Polishing Mode while sanitizing. If the Aries HP water system was manufactured after the year 2005, the “WAKE” button will need to be pressed once within the first 30 minutes of the sanitization interval so that the water remains in recirculation. Pressing the “WAKE” button interrupts system standby to start the 30-minute run cycle.
-
After 15 minutes, open the dispense valve draws off approximately 250 ml, and use test strips to verify minimum chlorine residual of 5 ppm.
-
After 45 to 60 minutes, dispense about 20 liters to drain.
-
After flushing the system, open the dispense valve draw off 250 ml, and use test strips to ensure the chlorine is removed.
-
Turn off the system and disconnect the power plug.
-
Open dispense valve.
-
Remove cartridges.
-
Please refer to the “Installing the Cartridge” and “Start-Up Procedure” sections
CONSUMABLES & SPARE PARTS
PART NUMBER | MODEL DESCRIPTION |
---|---|
1200014 | Outlet Angle Valve |
1200102 | Resistivity Calibration Cell Block |
ARA-WB | Wall Mounting Bracket |
HPA-003 | UV Bulb for Combination UV System |
HPA-008 | 220 Volt Operation Step Down Converter |
HPA-010 | Sanitization Kit (Includes Blank Filters, Sanitization Cartridge, |
Gloves & Test Strips)
HPA-011| 0-ring Repair Kit for Upper Outlet and Plunger (3-sets)
HPA-013| Lower Fitting Repair Kit (Includes 3-sets of the plunger, cup,
springs, o-rings, and fittings)
HPA-014| Upper Repair Kit (Includes 3-sets of upper fittings)
PF-00-6402| 0.2-micron capsule filter – 1/4″ MNPT x 1/4″ hose barb
PF-00-6505-HN| 0.05-micron ultrafilter – 1/4″ MNPT x 1/4″ hose barb
PF-00-7101| POU 0.2-micron Endotoxin Removal Capsule Filter – 1/4″ MNPT x 1/4″
hose barb
PX-11-0001| Sanitization Cartridge Only
VP-17-0001| Single Sanitization Cartridge
VPK-3805| VP Series Cartridge Kit for Tap Water Feed (1-Organics Pretreatment
& 2 Mixed Beds)
VPK-4010| VP Series Cartridge Kit for RO Feed (3 Mixed Beds)
TROUBLESHOOTING
SYMPTOM | CAUSE | REMEDY |
---|---|---|
The unit will not WAKE up | The wake button was pushed too quirky. | The WAKE |
button must be pushed for at least 3 seconds to initiate the polishing mode.
The display s off.| No power to the unit or unit in Stand By mode.| Check the
unit is plugged into a live outlet. Check the unit’s fuse – replace it if it
is blown.
Press the Wake switch to initiate the polish code.
Unit blows fuses.| An electrical short exists in the unit.| Return the unit to
the factory for repairs or
have the unit serviced by an authorized dealer.
The unit will not dispense water.| Supply of water s not on. Bad Solenoid.
A filter cartridge is loaded with particles
and plugged.
No flow| Confirm that the supply water is turned on.
Have the unit serviced by an authorized dealer or returned to factory.
Replace the filter cartridge.
The capsule filter is bound with air. Open bleed valve.
The display indicates that %neter purity is the below-set point.| The supply
water is improperly treated. The resin cartridges are not seated. Resin is
exhausted.| Test the supply water TDS and add additional treatment.
Open cartridge housing and reseat cartridge. Install fresh cartridges.
Resin cartridges exhaust quickly.| The supply water is improperly treated.|
Test the supply water TDS. Add additional pretreatment to obtain higher
quality.
Submicron filter plugs rapidly.| The supply water has colloids and is
improperly treated.| Perform a Silt Density Test. If it is higher than 1 add
additional pretreatment.
WARRANTY
Systems and options have a one-year warranty from the date of purchase
accompanied by system registration. If any component fails to function
normally within the first year, return the unit to the factory prepaid. If in
the opinion of the factory, failure was due to defects in materials or
workmanship, repair or replacement will be made without charge. A reasonable
service charge will be made for diagnosis or repairs not covered by this
warranty. Aries Filterworks / ResinTech, Inc. assumes no other responsibility
or liability.
This warranty does not cover any product which has been subject to misuse,
neglect, accident, or used in violation of operating instructions nor does it
extend to any units altered or repaired for warranty defects by anyone other
than Aries Filterworks / ResinTech, Inc.
IN NO EVENT SHALL ARIES FILTEWORKS / RESINTECH, INC. BE LIABLE FOR ANY
INCIDENTAL, CONSEQUENTIAL, SPECIAL, INDIRECT, PUNITIVE, OR EXEMPLARY DAMAGES
OR LOST PROFITS FROM ANY BREACH OF THIS WARRANTY OR OTHERWISE.
Aries Filterworks / ResinTech, Inc., reserves the right to make changes in
design, specifications, and prices without notice.
ARIES WATER PURIFICATION MAINTENANCE LOG
Serial Number: | Date Installed: |
---|---|
Asset # or ID: | Location: |
DATE | MAINTENANCE PERFORMED | BY | COMMENTS |
---|---|---|---|
Copyright Notice and Trademarks
© 2020 ResinTech, Inc.
This manual or parts thereof may not be reproduced in any form without the
express written permission of Aries Filterworks, Inc. Aries Filterworks® is a
registered trademark of ResinTech, Inc. Camden, NJ USA. For more information,
please visit our website:
www.ariesfilterworks.com
Copyright © 2019 Aries FilterWorks. All rights reserved.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>