Jamr B02T Blood Pressure Monitor
A Special Thank You…
Thank you for choosing a B02T blood pressure instrument. We’re proud of the
care and quality that goes into the manufacture of each and every item that
bears our name. Only the finest materials are used to assure you of a timeless
instrument designed for optimum performance. You’ll quickly appreciate the
results, for you now own one of the finest digital blood pressure instruments
that money can buy. With proper care and maintenance, your B02T automatic
blood pressure monitor is sure to provide you with many years of dependable
service. Please read the following instructions and general information which
will prove helpful in allowing you to enjoy your B02T of Digital In hospitals
and physician’s offices throughout the world, where accuracy and dependability
are critical,B02T professional diagnostic products are the instruments of
choice.
Now you too can enjoy the benefits of B02T engineering and quality in the
home. This feature rich instrument was designed to simplify the measurement of
blood pressure and pulse rate at home and deliver consistent, dependable
results.Your B02T digital blood pressure monitor is a fully automatic digital
blood pressure measuring device for use on the upper arm. It enables very fast
and reliable measurement of the systolic and diastolic blood pressure as well
as the pulse by way of the oscillometric method. This device offers clinically
proven accuracy and has been designed to be user friendly. Read this booklet
thoroughly before attempting to use your new B02T Digital Blood Pressure
Monitor.
Thank you for your patronage. It is indeed our pleasure to serve you.
Sincerely,
Shenzhen Jamr Technology Co., Ltd.
1.Introduction and Intended Use
This manual is for B02T models. It is a fully automatic digital blood pressure
measuring device for use by adults on the upper arm at home or in your
doctor’s/nurse’s office. It enables very fast and reliable measurement of
systolic and diastolic blood pressure as well as pulse through the
oscillometric method. This device offers clinically proven accuracy and has
been designed to be user friendly.
Before using, please read this instruction manual carefully and then keep it
in a safe place. Please contact your doctor for further questions on the
subject of blood pressure and its measurement.
Warning: Not suitable for neonatal and infants.
Warning: Not suitable for people who cannot express their ideas correctly.
This device can not be used together with hf surgical equipment.
Remember…
- Only a health-care professional is qualified to interpret blood pressure
- This device is NOT intended to replace regular medical
- It is recommended that your physician review! your procedure for using this device.
- Blood pressure readings obtained by this device should be verified before prescribing or making adjustments to any medications used to control hypertension. Under no circumstances should YOU alter the dosages of any drugs prescribed by your doctor.
- This monitor is intended for use by adults only. Consult with a physician before using this instrument on a
- In cases of irregular heartbeat , measurements made with this instrument should only be evaluated after consultation with your doctor.
- %familiarize your self with the section titled Important Information on Blood pressure and its 4easurement. It contains important information on the dynamics of blood pressure readings and will help you to obtain the best results.
- Host products, including accessories, shall be processed in accordance with local regulations after reaching the life cycle.
NOTE!
- This device contains sensitive electronic components. Avoid strong electrical or electromagnetic fields in the direct vicinity of the device (e.g. mobile telephones, microwave ovens) during use. These can lead to erratic results.
- Do not attempt to service or repair this device yourself. Should a malfunction occur, refer to local distributor or the manufacturer.
Warning:
- Too frequent measurements can cause injury to the PATIENT due to blood flow interference.
- Don’t place the cuff over wound
- Pressurization of the CUFF can temporarily cause loss of function of simultaneously used monitoring ME EQUIPMENT on the same limb.
Contraindication
Use of this instrument on patients under dialysis therapy or on
anticoagulant, antiplatelets, or steroids could cause internal bleeding.
Warnings and Precautions
Warning: ** The device contains sensitive electronic components. Avoid
strong electrical or electromagnetic fields in the direct vicinity of the
device (e.g. mobile telephones, microwave ovens).These can lead to temporary
impairment of the measuring accuracy.
Warning: ** Do not use cuffs, AC adapters or batteries other than those
included with this product or replacement parts supplied by the manufacturer.
Warning: ** Do not use the batteries and the AC adapter to provide
power at the same time.
Warning: This system may fail to yield specified measurement accuracy if
operated or stored in temperature or humidity conditions outside the limits
stated in the specifications section of this manual.
Warning: ** The separate ac adapter which is intended to connect USB
interface of Blood Pressure Monitor has not been evaluated according to IEC
60601-1. The safety of the product shall be reappraised when it power supply
by a separate ac adapter.
Warning: ** Remove the battery if the ME EQUIPMENT is not likely to be
used for some time.
Warning: ** The user must check that the equipment functions safely and
see that it is in proper working condition before being used.
Warning: ** No modification of this equipment is allowed.
Warning: ** The device is not suitable for use in the presence of
flammable anesthetic mixtures with air or with oxygen or nitrous oxide.
Warning: ** This equipment shall not be serviced or maintained while in
use with the patient.
Warning: ** The patient is an intended operator, the functions of
monitoring blood pressure and pulse rate can be safely used by patient. The
routine clean and changing batteries can be performed by the patient.
Warning: **** Use of power adapters
- Adapter: input 100-240v, 50/60hz output DC 5V 1A
- Do not to position the device to make it difficult to operate the disconnection device while using adaptor.
- Do not be prone to water leakage, high temperature, moisture, direct sunlight and more or more corrosive gas And Do not use this product in the above environment.
Caution: ** To avoid any possibility of accidental strangulation, keep
this unit away from children and do not drape tubing around your neck.
Caution: ** To avoid damaging the device, keep this unit away from
children and pets.
Caution: ** The standard material used for the bladder and tubing is
latex-free.
Attention: ** Self-measurement means control, not diagnosis or
treatment. Unusual values must always be discussed with your doctor. Under no
circumstances should you alter the dosages of any drugs prescribed by your
doctor.
Attention: ** The pulse display is not suitable for checking the
frequency of heart pacemakers!
Attention: ** In cases of irregular heartbeat, measurements made with
this instrument should only be evaluated after consultation with your doctor.
Note: To obtain the greatest accuracy from your blood pressure
instrument, it is recommended that the instrument be used within the specified
temperature and the relative humidity, please see the Technical
Specifications.
Note: The cuff is treated as the applied part. The user should contact
the manufacturer for assistance, if needed, in setting up, using or
maintaining the device.
Important Information on Blood Pressure and its Measurement
How does high or low blood pressure arise?
Your level of blood pressure is determined in the circulatory center of the
brain and adjusts to a variety of situations through feedback from the nervous
system. To adjust blood pressure, the strength and speed of the heart (Pulse),
as well as the width of circulatory blood vessels is altered.Blood vessel
width is controlled by fine muscles in the blood vessel walls.
Your level of arterial blood pressure changes periodically during heart
activity: During the “blood ejection” (Systole) the value is highest (systolic
blood pressure value). At the end of the heart’s “rest period” (Diastole)
pressure is lowest (diastolic blood pressure value).
Blood pressure values must lie within certain normal ranges in order to
prevent particular diseases.
Which values are normal?
Please refer to the diagram below(Picture-01)
There are six grids in the display of device. Please refer to the picture-01-01. Different grids represent different interval scales of WHO.
Blood pressure is very high if your diastolic pressure is above 90 mmHg and/or your systolic blood pressure is over 160 mmHg, while at rest. In this case, please consult your physician immediately. Long-term values at this level endanger your health due to continual damage to the blood vessels in your body. If your systolic blood pressure values are between 140 mmHg and 159 mmHg and/or the diastolic blood pressure values between 90 mmHg and 99 mmHg, consult your physician. Regular self-checks are necessary. If you have blood pressure values that are too low, (i.e., systolic values under 105 mmHg and/or diastolic values under 60 mmHg), consult your physician. Even with normal blood pressure values, a regular self-check with your blood pressure monitor is recommended. You can detect possible changes in your values early and react appropriately. If you are undergoing medical treatment to control your blood pressure, keep a record of values along with time of day and date. Show these values to your physician. Never use the results of your measurements to independently alter the drug doses prescribed by your physician.
Further information
- If your values are mostly normal under resting conditions but exceptionally high under conditions of physical or psychological stress, it is possible that you are suffering from so-called “labile hypertension.” Consult your doctor.
- Correctly measured diastolic blood pressure values above Z20mm;g require immediate medical treatment
What can be done if regular high or low values are obtained?
- Consult your doctor.
- Increased blood pressure values (various forms of hypertension) are associated with considerable health risks over time. Arterial blood vessels in your body are endangered due to constriction caused by deposits in the vessel walls (Arteriosclerosis). A deficient supply of blood to important organs (heart, brain, muscles) can result from arteriosclerosis. Furthermore, the heart will become structurally damaged with increased blood pressure values.
- There are many different causes of high blood pressure. We differentiate between the common primary (essential) hypertension, and secondary hypertension. The latter group can be ascribed to specific organ malfunctions. Please consult your doctor for information about the possible origins of your own increased blood pressure values.
- There are measures which you can take to reduce and even prevent high blood pressure.
Components of your blood pressure monitor
Measuring unit
.
The symbols on the LCD display
- Systolic blood pressure
- Diastolic blood pressure
- Irregular heartbeat symbol
- Pulse unit symbol
- Memory symbol
- Date/Time display
- USER A
- USER B
- Movement error symbol
- Mute symbol
- Average value symbol
- Battery low symbol
- Cuff self-checking function
- Pulse display
- Heartbeat symbol (Flashes during measurement)
- WHO symbol
Features of Model B02T
- Talking function
- Double users: 2 x 120 sets memory
- Cuff self-checking function
- Irregular heartbeat checking
- Average value function
- Low battery display
- WHO function
- Auto power-off
- External power adapter support
- Volume adjustment
- Date/time display
- Bluetooth function
- 1/3 Average Measurements(choose the switch between 1 and 3 on the right side of the device, 1 means the measurement of one time. And 3 means device will automatically shows the average result of 3 times measurements )
Note: Arm circumference should be measured with a measuring tape in the middle of the relaxed upper arm. Do not force cuff connection into the opening. Make sure the cuff connection is not pushed into the AC adapter port.
Using your Monitor for the First Time
Activating the pre-installed batteries Battery Installation
Use only 1.5V “AAA” alkaline batteries with this device.
- Press the hook on the bottom of the battery cover and lift the cover off in the direction of the arrow.
- Install 4 “AAA” size batteries so the + (positive) and (negative) polarities match the polarities of the battery compartment, replace the battery cover. Make sure that the battery cover is securely in position.
Battery replacement
Low Battery Indicator
- When the Low Battery Indicator appears on the display, turn the monitor off and remove all the batteries. Replace with 4 new batteries at the same time. Long-life alkaline batteries are recommended.
- To prevent the damage of monitor from leaked battery fluid, please take out of battery if the monitor unused in a long time(generally more than 3 months). If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Contact a physician immediately.
- Attached battery is only for testing the function of the monitor, Long-life alkaline batteries are recommended.
- Dispose of the device, components and optional accessories according to applicable local regulations. Unlawful disposal may cause environmental pollution
- Battery is dangerous stuff, do not mix it with other
System Settings
After you load the battery or connect power for the monitor.
-
Setting the User:
Press the SET button and then you can set the A/B user by shifting the A/B user button. -
Setting the Year/Month, Date/Time/Volume:
Long press the SET button for more than 3s, and then you can start to set.
Setting the Year:
Initial year is 2019, when the year display is flashing, press the MEM
button, the year will increase by I year each, hold the MEM button and it will
increase continuously 1 by 1, until 2049, and then rollover to 2019, once the
year SET is OK, press SET button to confirm.
Setting Month/Date:
Initial Month/Date is 1/01, when the Month display is flashing, press the MEM
button, the month will increase by 1, press SET button to confirm, and do in
the same way to set the date. Press SET button to confirm.
Setting Time :
When the hour display is flashing, press the MEM button, the hour will
increase by 1, press SET button to confirm, and do in the same way to set
the minute. Press SET button to confirm.
Setting Volume:
When display with SP is flashing, press MEM button to switch volume 1, volume
2, volume 3 or OFF. Press SET button to confirm.
Record Delete:
When you checking the memory data, long press MEM button to delete existing
user measurement data.
Note: You can’t delete all measurement record from the monitor storage at one time, if you decide to delete the all record, please keep the record in another way, in case you need it some days later. Take the battery out won’t lead to a record missing.
Cuff tube connections
Insert the cuff tube into the opening on the left side of the monitor indicated by the drawing of a cuff.
Measurement Procedure
Note: You should always be seated and calm before and during measurement.
Before measurement:
- Avoid eating and smoking as !ell as all forms of exertion directly before measurement. These factors influence the measurement result. Find time to relax by sitting in an armchair in a quiet atmosphere for about ten minutes before taking a measurement.
- Remove any garment that fits closely to your upper
- Always measure on the same arm normally left .
- Always compare measurements taken at the same time of day, since blood pressure changes during the course of the day, as much as 20-40 mmHg.
Common sources of error:
Note: Comparable blood pressure measurements always require the same conditions!
- Conditions should always be quiet.
- All efforts by the user to support the arm can increase blood pressure. 4ake sure you are in a comfortable, relaxed position and do not flex any of the muscles in the measurement arm during the measurement. Use a cushion for support if necessary.
- If the arm artery lies considerably lower or higher than the heart, an erroneously high or low blood pressure will be measured! Each 25-30cm difference in height between your heart and the cuff results in a measurement error of 10 mmHg!
- Cuffs that are too narrow! or too short result in false measurement values. Selecting the correct cuff is extremely important. Cuff size is dependent upon the circumference of the arm (measured in the center). The permissible range is printed on the cuff.
Cuff works Under the pressure range 0-300MMHG The wide range rigid cuff is : 8.7” – 15.7” (22 – 40 cm)
Note: Only use approved cuffs!
- A loose cuff or a sideways protruding air pocket causes false measurement
- With repeated measurements, blood accumulates in the arm, !hitch can lead to false results. Consecutive blood pressure measurements should be repeated after a 1 minute pause or after your arm has been held up in order to allow the accumulated blood to flow If you decide to take your Averaging Mode measurement again, be sure to wait at least one minute beforehand.
Fitting the Cuff
Please refer to picture-06
- The cuff is preformed for easier use. Remove tight or bulky clothing from your upper arm.
- Wrap the cuff around your upper left arm. The rubber tube should be on the inside of your arm extending downward to your hand. Make certain the cuff lies approximately 1/2” to 3/4” (2 to 3 cm) above the elbow. Important! The on the edge of the cuff (Artery Mark) must lie over the artery which runs down the inner side of the arm.
- To secure the cuff, wrap it around your arm and press the hook and loop closure together.
- There should be little free space between your arm and the cuff. You should be able to fit 2 fingers between your arm and the cuff. Cuffs that don’t fit properly result in false measurement values. Measure your arm circumference if you are not sure of proper fit.
- Lay your arm on a table (palm upward) so the cuff is at the same height as your heart. Make sure the tube is not kinked.
- Remain seated quietly for at least two minutes before you begin the measurement.
Measure Procedure
Refer to picture-07
The monitor is designed to take measurements and store the measurement values in memory for two people using User ID A and User ID B.
- Sit comfortably in a chair with your feet flat on the
- Select your User ID (A or B).
Stretch your arm forward on the desk and keep relaxing, make sure the palm of hand is upturned. Make sure arm is in correct position, to avoid body movement. Sit still and do not talk or move during the measurement. After the cuff has been appropriately positioned on the arm and connected to the blood pressure monitor, the measurement can begin:
-
Operate via the App on smart phone with Bluetooth
Install the App from Google play store or Apple app store. Open Bluetooth on smart phone, and then Turn on the App, the home interface will show up as picture App 01. Please refer the below steps tell how to remote control on the App:-
Complete My Profile
Click the SET button on the top left corner as picture App 02, then select My Profile button to edit and save user information (male/female, name, age, height and weight) as picture App 03. -
Setting Language
After setting My Profile, back to select and save the BM monitor Language as picture App 04.
-
App01 App02 App03
* **Connect Bluetooth
** After the setting is finished, click the HOME button on the top right corner as picture App 01, then select Bluetooth Device button. A moment later, the device will be connected as picture App 05 and App 06.
App04 App05 App06
* **Start the measurement
** Once the Bluetooth is connected, click the “START” button on the App to
start the testing as picture App 06. When measuring is done, the assessment
will be voiced out and a interface about checking the result will show up as
picture App 07 and App 08.
App07
App08
* **Checking the measurement
** In the frame of home interface, it will show the last measurement as
picture App 09. You can check the measurements with detail by clicking the
History button as picture App 10 and app 11
-
Operate on the device
-
Press the START/STOP button. The pump begins to inflate the cuff. In the display, the increasing cuff pressure is continually displayed.
-
After automatically reaching an individual pressure, the pump stops and the pressure slowly falls. The cuff pressure is displayed during the measurement.
-
When the device has detected your pulse, the heart symbol in the display begins to blink.
-
When the measurement has been concluded, the measured systolic and diastolic blood pressure values, as well as the pulse will be displayed.
-
The appearance of this symbol signifies that an irregular heartbeat was detected. This indicator is only a caution. It is important that you be relaxed, remain still and do not talk during measurements.
-
The measurement results are displayed until you switch the device off. If no button is pressed for 60 seconds, the device switches off automatically.
-
Cuff self-checking symbol
The cuff correct symbol will be displayed if the cuff position is correct, otherwise the wrong symbol will be displayed. Please check again the cuff if the wrong symbols displayed. -
Movement error symbol
The Movement Error Symbols displayed if you move your body during the measurement. Please remove the cuff, and wait 2-3 minutes. Reapply the cuff and take another measurement.
-
NOTE:
Patient Position:
- Comfortably seated
- Legs uncrossed
- Feet flat on the floor
- Back and arm supported
- Middle of the CUFF at the level of the right atrium of the heart
Recommended Use Methods
- Recommendation that the PATIENT relax as much as possible and not talk during the measurement PROCEDURE
- Recommendation that 5 min should elapse before the first reading is taken 3.Any reading can be affected by the measurement site, the position of the PATIENT, exercise, or the PATIENT’S physiologic condition
- Performance of the AUTOMATED SPHYGMOMANOMETER can be affected by extremes of temperature, humidity and altitude
- To stop the inflation or measurement, push the START/STOP The monitor will stop inflating, start deflating, and will turn off.
- After the monitor has detected your blood pressure and pulse rate, the cuff automatically deflates. Your blood pressure and pulse rate are displayed.
- The monitor will automatically turn off after one minute.
Irregular Heartbeat Detector
This symbol – indicates that certain pulse irregularities were detected during the measurement. In this case, the result may deviate from your normal basal blood pressure – repeat the measurement. Information for the doctor on frequent appearance of the Irregular Heartbeat Symbol. This instrument is an oscillometric blood pressure monitor device that also analyzes pulse frequency during measurement. The instrument is clinically tested.
If pulse irregularities occur during measurement, the irregular heartbeat symbol is displayed after the measurement. If the symbol appears more frequently (e.g. several times per week on measurements performed daily) or if it suddenly appears more often than usual, we recommend the patient to seek medical advice. The instrument does not replace a cardiac examination, but serves to detect pulse irregularities at an early stage.
Error Indicates
The following symbol will appear on the display when measuring abnormal
SYMBOL | CAUSE | CORRECTION |
---|---|---|
No display appears | Weak battery or improper placement | Replace both batteries |
with new ones. Check the battery installation for proper placement of the
battery polarities.
Er 1| Sensor abnormal| Check if the pump is working or not. If it is working,
then the problem is sensor abnormal. Please send it to the local distributor.
Er 2| Monitor could not detect pulse wave or cannot calculate the blood
pressure data| Check if the air releasing is too slow or not. If it is too
slow, please check if there is any dust in the tube plug of the cuff and the
cuff port in the device. If yes, please clean and start the measurement again.
If no, please send the device back to the local distributor.
Er 3| Measurement result is abnormal (SYS≦45mmHg, DIA≦24mmHg)| Occasionally-
measure for one more time/ Always – send it to local distributor
Er 4| Too loose cuff or air leakage (Cannot inflate to 30mmHg within 15s)| Tie
the cuff correctly and make sure the air plug is properly inserted in the unit
Er 5| The air tube is crimped| Correct it and make the measurement again
Er 6| The sensor is sensing great fluctuation in the pressure| Please keep
quiet and don’t move
Er 7| The pressure that the sensor sensing is over the limit| Please send back
to the local distributor
Er 8| The demarcation is
incorrect or the
device has not been
demarcated| Please send back to the local distributor
Trouble Remover
SYMBOL DESCRIPTIONS
The following symbols may appear in this manual, on the Digital Blood Pressure Monitor B02T, or on it’s accessories. Some of the symbols represent standards and compliances associated with the Digital Blood Pressure Monitor B02T and its use.
Memory
At the end of a measurement, this monitor automatically stores each result with date and time. Each unit stores 120 sets measurements for 2 users, totally 240 sets(User A and B).
Viewing the stored values
With the unit off, press the 4emory button. The display first shows A,
then shows an average of all measurements stored in the unit. Please note:
Measurements for each user are averaged and stored separately. Be certain that
you are viewing the measurements for the correct user. Pressing the Memory
button again displays the previous value. To view a particular stored memory,
press and hold the Memory button to scroll to that stored reading.
Discontinuing a Measurement
If it is necessary to interrupt a blood pressure measurement for any reason (e.g the patient feels unwell), the Start/Stop button can be pressed at any time. The device then immediately lowers the cuff pressure automatically.
Battery Change Indicator
Batteries discharged– replacements required
When the batteries are discharged, the battery symbol will flash as soon as
the instrument is switched on. You cannot take any further measurements and
must replace the batteries
The battery compartment is located on the back side of the unit.
- Remove cover from the bottom plate, as illustrated below picture-08
- Insert the batteries (4 x size AAA ). Always use AAA long life batteries or alkaline 1.5v batteries.
- The memory retains all values although date and time must be reset – the year number therefore flashes automatically after the batteries are replaced.
- To set date and time, follow the procedure described in Section 4.2.
Which batteries and which procedure?
Use four new, long life 1.5V AAA batteries. Do not use batteries beyond their expiration date. If the monitor is not going to be used for a prolonged period the batteries should be removed.
Using rechargeable batteries
You can also operate this instrument using rechargeable batteries.
- Only use Nimh; reusable batteries!
- If the battery symbol the batteries must be removed and recharged! They must not remain inside the instrument, as they may become damaged through total discharge even when switched off. The batteries must NOT be discharged in the blood pressure monitor! If you do not intend to use the instrument for a week or more, always remove the rechargeable batteries!
- Recharge these batteries using an external charger and follow! manufacturer’s instructions carefully.
Using the AC Adapter
You may also operate this monitor using the AC adapter (output 5V DC/1A with
Micro USB plug).
Use only the approved AC adapter to avoid damaging the unit.
- Ensure that the AC adapter and cable are not
- Plug the adapter cable into the AC adapter port on the right side of the blood pressure monitor.
- Plug the adapter into your electrical outlet. When the AC adapter is connected, no battery current is consumed.
Note : No power is taken from the batteries while the AC adapter is connected to the monitor. If electrical power is interrupted,(e.g., by accidental removal of the AC adapter from the outlet) the monitor must be reset by removing the plug from the socket and reinserting the AC adapter connection.
Care and Maintenance
Wash hands after each time measurement.
If one device is used by different patients, wash hands before and after each
use.
- Do not expose the device to either extreme temperatures, humidity, dust or direct sunlight.
- The cuff contains a sensitive air-tight bubble. Handle this cuff carefully and avoid all types of stress through twisting or buckling.
- Clean the device with a soft, dry cloth. Do not use gas, thinners or similar solvents. Spots on the cuff can be removed carefully with a damp cloth and soapsuds. The cuff with bladder must not be washed in a dishwasher, clothes washer, or submerged in water.
- Handle the tube carefully. Do not pull on it. Do not allow the tubing to kink and keep it away from sharp edges.
- Do not drop the monitor or treat it roughly in any way. Avoid strong
- Never open the monitor! This invalidates the manufacturer’s
- Batteries and electronic instruments must be disposed of in accordance with the locally applicable regulations, not with domestic waste.
Accuracy test
Sensitive measuring devices must be checked for accuracy from time to time. We recommend a periodical inspection of your unit by an authorized dealer every 1 year. Please turn to local distributor or the manufacturer
Warranty
Your blood pressure monitor is guaranteed for 1 year against manufacturers’
defects for the original purchaser only, from date of purchase. The warranty
does not apply to damage caused by improper handling, accidents, professional
use, not following the operating instructions or alterations made to the
instrument by third parties. Warranty only applies to the instrument. All
accessories including the cuff are guaranteed for one year, USB charging cable
is not included. There are no user serviceable parts inside. Batteries or
damage from old batteries is not covered by the warranty.
Note: According to international standards, your monitor should be
checked for accuracy every year.
Certifications
Device standard:
This device is manufactured to meet the European blood pressure monitors:
~NZ060-Z • ~NZ060-3 • I~C 8060Z-2-30 • ISO8Z060-Z • I~C6060Z-Z-ZZ • IEC60601-1
Electromagnetic compatibility:
Device fulfills the stipulations of the International standard IEC60601-1-2
Technical Specifications
Model: B02T
Wight: 267g (Batteries and AC adapter are not included)
Display: 5885mm【3.35”x2.28”】 LCD
Digital Display Size: 118 (W) x 110 (L) x 52(H) mm【 .6 W x .33 x2.0 ; 】
Accessories: 1×Main Device, 1×Cuff, 1×Users manual, 1×Warranty card Operating
Conditions:
Temperature: 5℃ to 40℃;Humidity: 15% to 93% RH; Storage And Shipping
Conditions: Temperature: -25℃ to 70℃; ;umidity:s 93% R;; Atmospheric pressure
range:70kPa~106kPa
Measuring method: Oscillometric Pressure sensor: Resistive Measuring range:
0-280mmHg Pulse: 40 to 199 per minute
Cuff pressure display range:<300mmHg
Memory: Automatically stores the last 120 measurements for 2 users (total 240)
Measuring resolution: 1 mmHg
Accuracy: Pressure within ± 3 mmHg / pulse ± 5 % of the reading Power source:
a) 4*AAA batteries, 1.5 V b) AC adapter INPUT:100-240VAC 50/60HZ OUTPUT:5V DC
1A Accessories: Wide range rigid cuff 8.7” – 15.7” (22 – 40 cm)
Automatically power off : 60 seconds
Users: Adult Expected service life of the device and accessories: 5 years
Technical alterations reserved!
EMC Declaration
- This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment.
- Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.
- Caution: This unit has been thoroughly tested and inspected to assure proper performance and operation!
- Caution: this machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be used
FCC Statement
This device complies with part 15 of the FCC rules. Operation is subject to
the following two conditions: (1) this device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
Caution: Changes or modifications not expressly approved by the party
responsible for compliance could void the user’s authority to operate
the equipment.
NOTE: This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates
uses and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to
radio communications. However, there is no guarantee that interference will
not occur in a particular installation. If this equipment does cause
harmful interference to radio or television reception, which can be determined
by turning the equipment off and on, the user is encouraged
to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help important announcement
Radiation Exposure Statement
The device has been evaluated to meet general RF exposure requirement. The
device can be used in portable exposure condition without restriction.