Jamr Technology M523 Ultrasonic Fetal Doppler User Manual
- June 13, 2024
- Jamr Technology
Table of Contents
- Jamr Technology M523 Ultrasonic Fetal Doppler
- INTRODUCTION
- Packing List
- COMPONENTS OF THE DEVICE
- BATTERY INSTALLATION
- CARE AND MAINTENANCE
- TROUBLESHOOTING
- TECHNICAL SPECIFICATIONS
- Ultrasound Intensity and Safety
- FCC Statement
- WARRANTY
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Jamr Technology M523 Ultrasonic Fetal Doppler
INTRODUCTION
- Thank you for purchasing the Jamr Ultrasonic Fetal Doppler. Please read this manual carefully before using the device and save the manual to inspect at any time.
- The ultrasonic fetal Doppler is a hand-held, battery-powered Doppler fetal heart rate detector. It is intended to detect and display fetal heart rate and is used by health care professionals in hospitals, clinics, communities, and homes for singleton pregnancies after 16 weeks gestation.
Contraindications
None.
The following symbols appear in the manual or device, understand what these
symbol means help you to use the device correctly.
Symbol
|
Description
| ISO/IEC
symbol number and registration
date
|
Symbol
|
Description
|
ISO/IEC symbol
number and registration date
---|---|---|---|---|---
WARNING
|
Indicates a potentially hazardous situation which, if not avoided, could result
in death or serious injury.
|
N/A
|
CAUTION
|
Indicates a potentially
a hazardous situation which, if not avoided, may result in
minor or moderate injury to the user or patient, or cause damage
to the equipment or other property.
|
ISO 7000-0434A 2004-01- 15
|
indicates a carrier that contains unique device identifier information
|
N/A, conform to ISO
15223- 1-2021
|
| Do not dispose of this product
as unsorted municipal waste.
Collection of
such waste separately for special
treatment is necessary.
|
N/A, conform to EN 50419:2022
|
Refer to the instruction manual/booklet
|
N/A, ISO 7010-M002
|
IP22
| Degree of protection against
harmful ingress of particulate matter and
water.
|
N/A, conform to IEC 60529
---|---|---|---|---|---
|
Manufacturer
| ISO 7000-3082
2011- 10-02
| | Solids:
Date of manufacture
| ISO 7000-2497
2004-01- 15
|
Serial number
| ISO 7000-2498
2004-01- 15
| | Type BF applied part| N/A, conform to IEC 60417-5333
|
Indicate the item is a medical device
|
N/A, conform to ISO
15223- 1-2021
|
| Caution: Federal law restricts this device to sale by or on the
order of a physician.
|
N/A
| This device has not been tested for use in an MR environment and should not be used exposed to MR
environments. Keep it outside the MRI scanner room.
|
N/A, conform to Testing and Labeling
MedicalDevices for Safety in the Magnetic Resonance (MR)
Environment
|
|
Humidity limitation
|
ISO 7000-2620
2004-01- 15
|
Temperature limit
|
ISO 7000-0632
2014-06-04
|
|
Atmospheric pressure limitation
|
ISO 7000-2621
2004-01- 15
|
This way up
This is the correct upright position of the distribution packages for transport and/or storage.
|
ISO 7000, No.0623
|
|
Fragile, handle with care
|
ISO 7000-0621
2014-06-04
|
Keep away from sunlight
|
ISO 7000-0624
2014-06-04
|
|
Keep dry
|
ISO 7000-0626
2014-06-04
---|---|---|---|---|---
| Stacking limit by| |
| |
number| | |
Maximum number of identical transport packages/items which may be
stacked on the bottom package,
|
ISO 7000, No.2403
|
The device complies with part 15 of the FCC rules
|
N/A
where “n” is the limiting number.| | |
Packing List
No. | Name | Quality |
---|---|---|
1 | Main unit | 1 |
2 | User manual | 1 |
3 | Bluetooth User manual | 1 |
4 | AA battery | 2 |
SAFETY INFORMATION
Indicates a potentially hazardous situation, if not avoided, which could result in death or serious injury.
- WARNING: The device and the APP are for use by healthcare professionals only. The device is WARNING not intended for treatment.
- WARNING: The device is not explosion-proof and cannot be used in the presence of flammable anesthetics.
- WARNING: Do not throw batteries in the fire as this may cause them to explode.
- WARNING: Do not attempt to recharge normal dry-cell batteries, they may leak and may cause a fire or even explode.
- WARNING: Don’t touch the signal input or output connector and the patient simultaneously.
- WARNING: Do not use the device during the process of surgical equipment (including high-frequency surgical equipment) and MRI examinations to avoid jeopardizing physician or patient safety.
- WARNING: Don’t dismantle or fix the device by yourself, or use not authorized parts and accessories, otherwise it may result in a detection error or damage to the device.
- WARNING: If the electrolyte in the battery gets into your eye, please immediately. flush with a great deal of clearwater, or it will result in blindness or other harmful danger.
- WARNING: Contraindication for women with multiple fetuses
- WARNING: Device not for use in water
- WARNING: Device interference (e.g., performing ultrasound imaging or Doppler flow measurements together with ultrasound fetal monitoring may cause false FHR readings, and the trace recording may deteriorate.)
- WARNING: We recommend that exposure to ultrasound should be kept as low as reasonably achievable. This is considered to be good practice and should be observed at all times.
- WARNING: Precaution to perform the ultrasound procedure prudently using the principle of ALARA
- CAUTION: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient, or cause damage to the equipment or other property.
- CAUTION: Operate the device only as intended, do not use it for any other purpose. Remove batteries if the device is not likely to be used for 3 months or longer time.
- CAUTION: Don’t use a cell phone for calling near the device, or it may influence the normal function of the device.
- CAUTION: Dispose of device, component, and optional accessories according to applicable local regulations. Unlawful disposal may cause environmental pollution.
- CAUTION: Don’t subject the device to strong shocks like dropping it on the floor, otherwise it may lead to the measurement value of the product being incorrect.
- CAUTION: DO NOT disassemble or attempt to repair this device or other components by yourself, this may cause an inaccurate reading.
- CAUTION: DO NOT drop or subject this device to strong shocks or vibrations.
- CAUTION: DO NOT insert batteries with their polarities incorrectly aligned.
- CAUTION: Remove batteries if this device will not be used for a long period of time.
- CAUTION: DO NOT use batteries after their expiration date.
COMPONENTS OF THE DEVICE
The following components are included with the package, make sure all the components are complete when you open your new purchase.
The | Symbols on the LCD display |
---|---|
No. | Explanation |
1
| Fetal heartbeat symbol| | 1. The fetal heartbeat symbol indicates fetal heartbeat information. It flashes with a fetal heartbeat.
2. When the fetal heart rate is between 110 and 160 bpm, the green one flashes.When the fetal heart rate is out of this
range, the red one flashes.
2
| Fetal heart signal quality|
| Fetal heart signal quality is divided into three grades: average, good, very good
3
|
Fetal heartbeat value
| | display range is 60 to 240 when the fetal heart rate is not
between 60 and 240, it will display “ “.
4
|
Voice
| |
Voice level: The Voice level and fetal heart rate value display use the same area, when there is a voice control operation, display voice level and then recover to display the current values of fetal heart rate. The voice level shows the range of 0 to 6.
0 represents mute.
5
|
Battery power symbol
| |
indicates full battery;
indicates approximately 70% remaining battery; indicates approximately 40% remaining battery; indicates approximately 10% remaining battery.
When the battery level is critically low, the low battery indicator will flash 10 times before automatically shutting down. (Please replace the battery promptly when this occurs.)
---|---|---|---
6
|
Bluetooth symbol
| | In standby mode, Bluetooth is awaiting connection, and the Bluetooth symbol will flash. Once the connection is successful, the Bluetooth symbol will remain steady.
7
|
Probe symbol
| |
Indicates that the probe is on.
BATTERY INSTALLATION
CAUTION: Use only 1.5V ”AA” alkaline batteries with the device.
Follow the device’s instructions to open the battery cover. Insert two AA-size batteries into the battery compartment, ensuring that the positive and negative terminals of the batteries align with the corresponding markings inside the compartment. Place the batteries flat and then close the battery cover.
- Low Battery Indicator
- When appears on the display, turn the device off and remove all the batteries. Replace with new 2 “AA” size batteries at the same time.
- CAUTION: Don’t replace other model AA alkaline batteries, otherwise may cause breakdown.
- CAUTION: Do not mix and use new and old batteries, otherwise may cause leakage, fever, or fracture of the battery to damage the product.
- CAUTION: Turn the device off before replacing the batteries.
- CAUTION: When the batteries are replaced or taken out, you need to reset the date and time.
- CAUTION: To prevent damage to the device from leaking battery fluid, please take out of the battery if the device has been unused for a long time (generally more than 3 months). If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Contact a physician immediately.
- CAUTION: Dispose of the useless batteries according to local environmental protection regulations.
Powering On and Off
- Insert one end of the probe cable into the Type-C port at the rear of the main unit and the other end into the Type-C port at the rear of the handle, ensuring a proper connection between the main unit and the probe handle.
- To power on the device from the off state, press and hold the power button for approximately 2 seconds. During the screen initialization, a full-screen display will appear for 1 second (the heart-shaped symbol alternates between red and green on the full-screen display, with each color displayed for 0.5 seconds), and then the device will enter the “0” detection screen.
- In the initial state, the voice defaults to level 3. Regardless of the voice level adjusted during usage, the device will remember the voice level from before it was powered off.
- Press the side voice + and – keys briefly to adjust the device’s speaker volume, which ranges from 0 to 6 levels, with 0 representing mute.
- To power off the device, press and hold the power button for approximately 2 seconds. A short press does not initiate any operational commands.
- To prolong battery life, if there is no operation and no signal input within 3 minutes (the display remains at “0”), the fetal heart monitor will automatically power off.
- Fetal Heart Signal Quality Indicator The fetal Doppler will display three levels based on the recognized signal strength: Level one, level two, and level three represent weak fetal heart signal strength, moderate fetal heart signal strength, and strong fetal heart signal strength, respectively. See the illustration below for reference:
Before detection
Check whether the device is in good condition and whether there is obvious
damage that may affect the safety of pregnant women or the performance of the
device; if any damage is found, please stop using it immediately and replace
the damaged parts.
Detection Procedure
Follow the steps below for fetal heart rate detection:
- The pregnant woman lying on a bed;
- Apply an appropriate amount of coupling to the acoustic surface of the ultrasound fetal Doppler;
- Turn on the power switch;
- Touch the pregnant woman’s abdomen to determine the position of the fetus;
- Place the device on the pregnant woman’s abdomen and move or tilt the device slowly within the range of the fetus until hearing clear and rhythmic fetal heart sounds and seeing a stable fetal heart rate value;
- After the detection, turn off the device first, and then use a clean soft cloth or paper to wipe off the remaining coupling on the surface of the pregnant woman’s abdomen and the ultrasound head of the device.
CAUTION: Do not mistake maternal heart rate as fetal heart rate.
CAUTION: Do not take gloves to operate the buttons. If there is water,
couplant, and other substances on the surface of the fingers, please clean
them first, otherwise the test results will be affected.
Note: We recommend you use the couplant manufactured by Hony Medical Co., Ltd., the number of 510(k) is K221999. You can order this couplant from the manufacturer. Manufacturer’s product website:http://www.yafho.com/product/222.html.
Look for the fetal heart rate
Since the fetus moves at any time in the abdomen, the position of the fetal
heart is not fixed. You can refer to the uterine fundus height of the
following different pregnancy cycles to determine the approximate range of the
fetus:
- At the 12th week of pregnancy, the height of the uterine fundus is 2-3 transverse fingers above the pubic symphysis (about 2-3cm).
- At the 16th week of pregnancy, the height of the fundus is between the navel and the pubic bone.
- At the 20th week of pregnancy, the height of the fundus of the uterus is 1 finger below the navel (about 1cm).
- At the 24th week of pregnancy, the height of the fundus of the uterus is 1 finger above the navel (about 1cm).
- At the 28th week of pregnancy, the height of the fundus of the uterus is 3 fingers above the navel (about 3cm).
- At the 32nd week of pregnancy, the height of the fundus of the uterus is between the navel and the xiphoid process.
- At the 36th week of pregnancy, the height of the fundus of the uterus is 2 fingers below the xiphoid process (about 2cm) .
See the position picture of the fetal heart below:
CARE AND MAINTENANCE
To keep your ultrasonic fetal doppler in the best condition and protect the unit from damage, follow the directions listed below:
Maintenance
- Before each use, the device must be checked whether has obvious damage that could affect the safety of pregnant women or the performance of the device. If damage is found, make necessary repairs or replacements.
- The accuracy of the fetal heart rate is controlled by the device and cannot be adjusted by oneself.
- If the fetal heart rate result is not reliable, use other methods such as using a stethoscope to verify, or contact your local agent or manufacturer for assistance.
- The device is a precision instrument and must be handled with care. The excess couplant on the ultrasonic probe must be wiped off after the instrument is used.
Cleaning
- We recommend thorough cleaning of the device.
- Monitor/Main Unit should be wiped down with a clean paper towel, soft cloth, or damp cloth with clean, warm water after each use. If a damp cloth or wipe is used, the monitor should be dried with a clean dry paper towel or cloth. The fetal monitor should never be put into water or other liquids for cleaning.
- Visual inspection for residual soil and an additional repeat of previous cleaning steps in the event that residual soil is visible on the device.
Probe
- Ensure that any acoustic coupling gel is completely wiped off the probe.
- Use a moistened soft cloth or wipe to remove any remaining contaminants that remain on the probe.
- Do not re-use cloths or wipes.
- Soap, detergents, or enzymatic cleaners should be used in accordance with the manufacturer’s instructions.
- Use a lint-free soft and clean dry cloth or wipe to thoroughly dry the probe.
Disinfection
Re-usable fetal dopplers should be cleaned and disinfected after each use. Re-
usable fetal dopplers can be disinfected by being wiped down with a low-level
disinfectant such as Ethyl or isopropyl alcohol (70-90%).
Storage
- Keep your device in the storage case when not in use.
- Store your device in a clean, safe location.
- Do not store your device:
- If your device is wet.
- In locations exposed to extreme temperatures, humidity, direct sunlight, dust or corrosive vapors such as bleach.
- In locations exposed to vibrations or shocks.
Disposal
- Dispose of the device in accordance with the regulations applicable at the place of operation.
- If you have any queries, please refer to the local authorities responsible for waste disposal.
NOTES
- Handing of battery and wastes method, please act according to the native law to proceed to handle.
- To protect the environment, dispose of empty batteries at your retail store or at appropriate collection sites according to national or local regulations.
BLUETOOTH CONNECTION
- Install the App from the Google play store or Apple app store.
- Open Bluetooth on your smartphone, and then Turn on the App.
- Bluetooth pairing for the first time
- Press the power button to turn on the device and the Bluetooth symbol will flash, then operate Bluetooth pairing according to the settings on the APP, The Bluetooth symbol will stop flashing after the connection is successful.
- Bluetooth paired successfully
- Bluetooth will be automatically searched and connected when it is powered on.
- For details, please refer to the Bluetooth APP user manual
- Note: Only healthcare professionals can use the APP.
TROUBLESHOOTING
Due to operation or other problems, you may not be able to achieve the expected results, please refer to the following troubleshooting measures to solve.
SYMBOL | CAUSE | CORRECTION |
---|
Can’t turn on or
turn off
immediately after turning on
| The battery is low| Replacement battery
Switch on and off the device without following the instructions| Long press
the switch button to turn on and off
Device malfunction| Contact customer service to resolve
After power on, the LED shows, but|
The probe did not start working
| The device needs to touch the skin probe to start working
there is no sound from the
loudspeaker
| the volume is minimal and no sound| Voice adjustment
---|---|---
Device malfunction
|
Contact customer service to resolve
Fetal heart rate jumps when detecting fetal
heart sounds
| Some strong interference sources, such as high-frequency devices,
mobile phones, etc
| Avoid using it in environments with
strong electromagnetic interference sources
Pregnant women experience fetal movement and the position of the fetal heart changes| Readjust the position of the
ultrasound fetal Doppler to find the best position
Device and belly friction
misidentification
| Find the best fetal heart position and hear stable fetal heart sounds
When used, the sensitivity is low and the noise is too large
|
Some strong sources of interference
| Avoid using it in environments with
strong electromagnetic interference sources
No ultrasonic couplant applied| Apply ultrasonic couplant
The device was not placed in the optimal fetal heart position
| Adjust the position and angle of the
ultrasound fetal Doppler on the belly of pregnant women
Device malfunction| Contact customer service to resolve
TECHNICAL SPECIFICATIONS
Model | M523 |
---|---|
Intended patient | Pregnant woman |
Contacting location | Abdomen surface |
Operation Principle | Continuous wave |
fetal heart rate range
| ****
60bpm-240bpm
Accuracy| ±2bpm
Ultrasound frequency| 3.0MHz
Display screen| LED
Unit| bpm
Sensitivity| ≥90dB
Ultrasound working frequency| 3.0MHz ± 10%
Power supply| d.c.3.0V AA battery
Lifetime| 3 years
Degree of Ingress Protection| IP22
Anti-electric shock degree| Type BF
Anti-electric shock type| Internally powered equipment
Operating Environment
| Temperature: 5℃ to 40℃
Humidity: 10% to 80% RH
Atmospheric pressure: 80 Kpa~105 Kpa
Storage Environment
| Temperature: -20℃ to 55℃
Humidity: ≤93% RH
Atmospheric pressure: 80 Kpa~105 Kpa
Weight| 169.2g
Size| Main Unit:138.6× 59.6× 29.8mm
Probe Handle:132.1× 31.7× 31.7mm Probe Cable:0.8m
Conformance standard
| IEC 60601- 1, IEC 60601- 1-2, IEC60601- 1- 11, IEC
60601-2-37
Ultrasound Intensity and Safety
Ultrasound in Medicine
The use of diagnostic ultrasound has proved to be a valuable tool in medical
practice. Given its known benefits for non-invasive investigations and medical
diagnosis, including investigation of the human fetus, the question of
clinical safety with regards to ultrasound intensity arises. There is no easy
answer to the question of safety surrounding the use of diagnostic ultrasound
equipment. Application of the ALARA (As Low As Reasonably Achievable)
principle serves as a rule-of-thumb that will help you to get reasonable
results with the lowest possible ultrasonic output. The American Institute of
Ultrasound in Medicine (AIUM) states that given its track record of over 25
years of use and no confirmed biological effects on patients or instrument
operators, the benefits of the prudent use of diagnostic ultrasound clearly
outweigh any risks.
Ultrasound Safety and the ALARA Principle
Ultrasound waves dissipate energy in the form of heat and can therefore cause
tissue warming. Although this effect is extremely low with Doppler, it is
important to know how to control and limit patient exposure. Major governing
bodies in ultrasound have issued statements to the effect that there are no
known adverse effects from the use of diagnostic ultrasound, however, exposure
levels should always be limited to As Low As Reasonably Achievable (the ALARA
principle).
Explanation of MI/TI MI (Mechanical Index)
Cavitations will be generated when ultrasound wave passes through and contacts
tissues, resulting in instantaneous local overheating. This phenomenon is
determined by acoustic pressure, spectrum, focus, transmission mode, and
factors such as states and properties of the tissue and boundary. This
mechanical bioeffect is a threshold phenomenon that occurs when a certain
level of ultrasound output is exceeded. The threshold is related to the type
of tissue. Although no confirmed adverse mechanical effects on patients or
mammals caused by exposure at intensities typical of present diagnostic
ultrasound instruments have ever been reported, the threshold for cavitation
is still undetermined. Generally speaking, the higher the acoustic pressure,
the greater the potential for mechanical bioeffects; the lower the acoustic
frequency, the greater the potential for mechanical bioeffects. The AIUM and
NEMA formulate mechanical index (MI) in order to indicate the potential for
mechanical effects. The MI is defined as the ratio of the peak-rarefactional
acoustic pressure (should be calculated by tissue acoustic attenuation
coefficient 0.3 dB/cm/MHz) to the acoustic frequency.
MI = Pr, α
fawf ×CMI
CMI = 1 (MPa / MHz)
TI (Thermal Index)
Heating of tissues is caused by absorption of ultrasound when the ultrasound
energy is applied. The temperature rise is determined by the acoustic
intensity, exposed area and thermo physical properties of the tissue. In order
to indicate the potential for temperature rise caused by thermal effects, the
AIUM and NEMA formulate thermal index (TI). It is defined as the ratio of the
total acoustic power to the acoustic power required to raise the tissue
temperature by 1ºC (1.8°F). According to different thermo physical properties
of the tissue, TI is divided into three kinds: TIS, TIB and TIC. TIS (Soft
Tissue Thermal Index): It provides an estimate of potential temperature rise
in soft or similar tissues. TIB (Bone Thermal Index): It provides an estimate
of potential temperature rise when the ultrasound beam passes through soft
tissue and a focal region is in the immediate vicinity of bone. TIC (Cranial
Bone Thermal Index): It provides an estimate of potential temperature rise in
the cranial bones or superficial bones.
Measurement Uncertainties
The uncertainties in the measurements were predominantly systematic in origin;
the random uncertainties were negligible in comparison. The overall systematic
uncertainties were determined as follows:
- Hydrophone Sensitivity: ±12percent for intensity, ±6 percent for pressure. Based on the hydrophone calibration report by ONDA. The uncertainty was determined within ±1 dB in frequency range 1-15 MHz.
- Digitizer: ±0.3 percent for intensity. ±0.15 percent for pressure. Based on the stated accuracy of the 8-bit resolution of the Agilent DSO6012 Digital Oscilloscope and the signal-to-noise ratio of the measurement.
- Temperature: ±2.4 percent for intensity uncertainty, ±1.2 percent for pressure uncertainty. Based on the temperature variation of the water bath of ±1ºC (1.8°F).
- Spatial Averaging: ±3.5 percent for intensity, ±1.75percent for pressure.
- Non-linear Distortion: N/A. No effects of nonlinear propagation were observed. Since all the above error sources are independent, they may be added on an RMS basis, giving a total uncertainty of ±12.73 percent for all intensity values reported, ±6.37 percent for all the pressure values, ±12.6 percent for the Mechanical Index, uncertainty of ±12.73% percent for power, ±0.15 percent for center frequency, ±6.87%for the MI.
Prudent Use Statement
Although no confirmed bioeffects on patients caused by exposure from present
diagnostic ultrasound equipment have ever been reported, the potential exists
that such bioeffects may be identified in the future. Therefore, the
ultrasound should be used prudently. High levels of acoustic output and long
exposure time should be avoided while acquiring necessary clinical
information.
Reference for Acoustic Output and Safety
- “Bioeffects and Safety of Diagnostic Ultrasound” issued by AIUM in 1993
- “Medical Ultrasound Safety” issued by AIUM in 1994
- “Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment, Revision 3” issued by AIUM/NEMA in 2004
- “Standard for real-time display of thermal and mechanical acoustic output indices on diagnostic ultrasound equipment, Revision 2” issued by AIUM/NEMA in 2004
- “Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers” issued in 2008.
- “Medical electrical equipment—Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment” issued by IEC in 2007.
Acoustic Output Reporting Table for Track 1 Acoustic output reporting table
for IEC60601-2-37(IEC 60601-2-37 (ed.2), am1 for use in conjunction with
IEC60601-1 (ed.3), am1 with Corr1 and Corr2, table 201.103)
Transducer Model: M523, Operating Mode: CW mode
Acoustic Output Reporting Table for Track 1 (non-autoscanning Mode)
- System Model: M523 SN: 20211224P612000001|
- Transducer Model: /
- Nominal Frequency:3.0MHz
- Operating Mode: CW Mode
Acoustic Output Reporting Table for Track 1 (non-autoscanning Mode)
- System Model: M523 SN: 20211224P612000002
- Transducer Model: /
- Nominal Frequency:3.0MHz
- Operating Mode: CW Mode
Acoustic Output Reporting Table for Track 1 (non-autoscanning Mode)
- System Model: M523 SN: 20211224P612000003
- Transducer Model: /
- Nominal Frequency:3.0MHz
- Operating Mode: CW Mode
Cybersecurity related information
- Bluetooth devices use the serial port and Bluetooth gatt layer encapsulated profile port for communication with the app, the serial port is the use of baud rate of one megabit, and the app docking profile is part of the public standard 32bit public protocol, and 128bit private protocol combination of the profile
- Instructions on how to maintain the device cybersecurity and how to update the firmware/software for the device.
- Software update can only be done by way of serial port burn-in and Jieli private protocol for over-the-air upgrade, does not support ota upgrade on the app
- Instructions for any security actions that the user or user facility are expected to take/implement to ensure secure use of the device.
- The device does not involve any security issues, and there is no docking with the gateway, so it only does the work of verification of the transmitted data
- Inform users of any cloud/networking/connection dependencies and provide instructions for users if the dependencies are unavailable.
- APP establishes communication with the device through Bluetooth and receives the data transmitted by the device; APP can store the data in the cloud server through the Internet to facilitate the user’s view of the history;
- Bluetooth connection status, users use APP to connect with the device via Bluetooth, after successful connection, during the measurement process, once APP detects the disconnection of Bluetooth connected with the device, it will give a reminder and try to reconnect the Bluetooth of the device, users can also choose to reconnect the Bluetooth device via manual way;
- network connection status, after the user has registered and logged in successfully through the APP, in the process of using the APP, once the APP detects that the current network connection is abnormal, it will give a reminder and try to reconnect the network;
- Description of cybersecurity events that can be detected by the device and a description of how users will be informed of such events.
- For some common network security events, APP will try to circumvent and remind. For example, anti-interface hijacking, in key interfaces such as login window or user privacy input, APP detects whether the foremost active application is itself, and gives a reminder if it is not its own system to avoid malware stealing user privacy information.
- Instructions for what a user should do if a cybersecurity event is detected or suspected
- In the User Agreement, “VII. Rights and Obligations” gives some suggestions and instructions, such as: the user is responsible for the safety of the registered account information and account password, the user needs to take legal responsibility for the registered account and the actions under the password. The user agrees not to disclose the account and password information to others under any circumstances, and to notify JAMR immediately if the user suspects that others are using the user’s account.
FCC Statement
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
- This device may not cause harmful interference, and
- this device must accept any interference received, including interference that may cause undesired operation.
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However,
there is no guarant ee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and
on, the
user is encouraged to try to correct the interference by one or more of the
following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
Caution: Any changes or modifications not expressly approved by the party
responsible for compliance could void the user’s authority to operate the
equipment.
This equipment complies with FCC radiation exposure limits set forth for an
uncontrolled environm ent. This transmitter must not be co‐located or
operating in conjunction with any other antenna or transmitter.
EMC DECLARATION
The ME EQUIPMENT or ME SYSTEM with following ESSENTIAL PERFORMANCE is intended
used in home and healthcare environment facility environment.
ESSENTIAL PERFORMANCE
Device-Specific Function| Device-Specific Pass/Fail Criteria|
Detection/Testing Method
---|---|---
Fetal heart rate range| 60bpm-240bpm| ****
IEC 60601-2-37:2007/AMD1:2015
Accuracy| ±2bpm
If Essential Performance is lost or degraded due to electromagnetic disturbances, this may result in loss of product function, please read below important information before to avoid possible electromagnetic disturbances.
Warning:Don’t near active HF surgical equipment and the RF shielded room
of an ME system for magnetic resonance imaging, where the intensity of EM
disturbances is high.
Warning: Use of this equipment adjacent to or stacked with other
equipment should be avoided because it could result in improper operation. If
such use is necessary, this equipment and the other equipment should be
observed to verify that they are operating normally.
Warning: Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of this equipment could result in
increased electromagnetic emissions or decreased electromagnetic immunity of
this equipment and result in improper operation.”
Warning:Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the ultrasonic fetal doppler, model: M523 ,
including cables specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL
PERFORMANCE with regard to electromagnetic disturbances for the excepted
service life.
Guidance and manufacturer’s declaration -electromagnetic emissions and
Immunity.
Table 1
Guidance and manufacturer’s declaration – electromagnetic emissions
The “ Ultrasonic Fetal Doppler P612” is intended for use in the electromagnetic environment specified below. The user of the “Ultrasonic Fetal Doppler” should ensure that it is used in
such an environment.
Emissions test| Compliance| Electromagnetic environment – guidance
RF emissions CISPR 11
| ****
Group 1
| The P612 uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11| ****
Class B
| The P612 is suitable for use in all establishments, including domestic
establishments and those directly
Harmonic
emissions
IEC 61000-3-2
| ****
Not application
| connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
---|---|---
Voltage| |
fluctuations/ flicker emissions IEC 61000-3-3| ****
Not application
|
Table2
Guidance and manufacturer’s declaration – electromagnetic immunity
The “ Ultrasonic Fetal Doppler P612” is intended for use in the
electromagnetic environment specified below. The user of the “Ultrasonic
Fetal Doppler” should ensure that it is used in such an environment.
Immunity test| IEC 60601
test level
| Compliance level| Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
| ****
±8 kV contact
±2,4,8, 15 kV
air±6KV indirect
| ****
±8 kV contact
±2,4,8,15 kV air
±6KV indirect
| Floors should be wood, concrete or ceramic tile. If floors are covered with Synthetic material, the relative humidity should be at
least 30 %.
Electrical fast transient/bur st
IEC
61000-4-4
| Signal
ports: ±1 kV Input a.c. power
ports: ±2 kV
Input d.c. power
ports: ±2 kV
| ****
Not application
| ****
Mains power quality should be that of a typical commercial or hospital
environment.
Surge IEC
61000-4-5
| Line to line:
±1kV,
Line to ground:
±2 kV
| ****
Not application
| Mains power quality should be that of a typical commercial or hospital
environment.
Voltage dips,
short interruptions and
voltage
variations on power supply
input lines IEC
61000-4- 11
| <5% U T
(>95 % dip in
U T)
for 0,5 cycle 40% U T
(60 % dip in U T) for 5 cycles 70% U T
(30 % dip in U T)
for 25 cycles
<5% U T
(>95 % dip in
U T)
for 5 sec
| ****
Not application
| ****
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the “ Ultrasonic Fetal Doppler P612” requires
continued operation during power mains interruptions, it is recommended that the “Ultrasonic Fetal Doppler” be
powered from an uninterruptible power supply or a battery.
Power frequency (50/60 Hz)
magnetic field
IEC
61000-4-8
| ****
30 A/m
50Hz/60Hz
| ****
30 A/m
50Hz/60Hz
| ****
Power frequency magnetic fields should be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
NOTE U T is the a.c. mains voltage prior to application of the test level.
Table3
Guidance and manufacturer’s declaration – electromagnetic immunity
The “ Ultrasonic Fetal Doppler P612” is intended for use in the electromagnetic environment specified below. The user of the “Ultrasonic Fetal Doppler” should ensure that it is used in
such an environment.
Immunity test| IEC60601
Test level
| Complianc e level| ****
Electromagnetic environment – guidance
| ****
3 Vrms
150 kHz
to 80 MHz
3 V/m
80 MHz
to 2,7 GHz
| | Portable and mobile RF communications equipment should be used no closer
to any part of the
| | “ Ultrasonic Fetal Doppler P612 ” , including cables,
| | than the recommended separation distance calculated from the equation
applicable to the frequency of the
| | transmitter.
Conduc| | Recommended separation distance
ted RF IEC 61000-
4-6
| ****
3 V
| d=1.2√P
d=1.2√P 80MHz to 800MHz d=2.3 √P 800MHz to 2.5 GHz
Radiate
d RF IEC
61000-
| ****
10V/m
| where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey, a should
be less than the compliance level in each frequency range’s
Interference may occur in the vicinity of equipment marked with the following symbol:
4-3| |
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
a mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the “ Ultrasonic Fetal Doppler P612” is used exceeds the applicable RF compliance level above, the Ultrasonic Fetal Doppler P612 should be observed to verify
b normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the “ Ultrasonic Fetal Doppler P612 ” . Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [ V 1] V/m.
Table4
Recommended separation distances between portable and mobile RF communications equipment and the Ultrasonic Fetal Doppler P612
The “ Ultrasonic Fetal Doppler P612” is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of the Ultrasonic Fetal Doppler can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the “ Ultrasonic Fetal Doppler” as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of
transmitter
W
| Separation distance according to frequency of transmitter (m)
150 kHz to 80
MHz
| 80 MHz to 800
MHz
| 800 MHz to 2,5
GHz
0,01| 0.12| 0.12| 0.23
0,1| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
WARRANTY
- Your Jamr Ultrasonic Fetal Doppler is guaranteed for 2 years against manufacturers’ defects for the original purchaser only, from date of purchase. The warranty does not apply to damage caused by improper handling, accidents, professional use, not following the operating instructions or alterations made to the instrument by third parties.
- Warranty only applies to the instrument.
- There are no user serviceable parts inside. Batteries or damage from old batteries is not covered by the warranty.
- Note: According to international standards, your monitor should be checked for accuracy every year.
CONTACT INFORMATION
- Ultrasonic Fetal Doppler is manufactured by:
- Shenzhen Jamr Technology Co., Ltd.
- A101-301, D101-201, Jamr Science & Technology Park, No. 2 Guiyuan Road,
- Guixiang Community, Guanlan Street, Longhua District, Shenzhen 518100, PEOPLE’S
- REPUBLIC OF CHINA
- Website: www.jiemeirui.com
- Telephone: 0086 755 85292057