pokit POK-PRO Portable Oscilloscope Multimeter User Manual
- June 5, 2024
- pokit
Table of Contents
- Warranty and Limitation of Liability
- Introduction
- Contacting Pokit
- Installing the App
- Switch Explanation
- Safety Information
- Fuse replacement
- Technical Specifications
- Recharging
- Taking Measurements
- Accessories Ratings
- Cleaning and Maintenance
- Troubleshooting
- Getting Started with the App
- Appendix A
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Pokit Pro
User Manual
MODEL: POK-PRO
Pokit ® Innovations
Warranty and Limitation of Liability
This Pokit product will be free from defects in materials and workmanship under normal use as described in the published product documentation for 12 months from the date of original purchase (Limited Warranty). This Limited Warranty is only valid and enforceable if you have purchased the product directly from Pokit Innovations or our authorised resellers and distributors. It does not apply where you purchase the product from an unauthorised reseller.
The following link contains our complete Limited Warranty and Limitation of Liability:
www.pokitinnovations.com/warranty-policy/
TO THE EXTENT PERMITTED BY LAW, POKIT INNOVATIONS PTY LTD AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE OR ANY SIMILAR STANDARD IMPOSED BY APPLICABLE LEGISLATION AND POKIT INNOVATIONS PTY LTD’S RESPONSIBILITY TO REPAIR, REPLACE, OR OFFER A REFUND FOR DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO ITS CUSTOMERS UNDER THIS WARRANTY
To obtain service for your Pokit product, please contact us.
Introduction
The Pokit Pro (MODEL: POK-PRO) is a battery powered multimeter, oscilloscope
and logger.
This device meets CAT III IEC61010-1 Edition 3 Standards. IEC61010-1 defines
four measurement categories (I to IV) based on the potential danger
from transients. CAT III devices are designed to protect against transients
when measuring equipment permanently installed in a building. Examples of this
kind of equipment are distribution boards, circuit breakers, and mains socket
outlets.
Contacting Pokit
Website: www.pokitinnovations.com
Email: support@pokitmeter.com
Address: Pokit Innovations Suite 2.2, 56 Delhi Rd, North Ryde, NSW, 2113
Australia
Installing the App
The Pokit Pro requires a Bluetooth connection to a smartphone running the
official Pokit App.
No measurements can be taken without the app.
- Download the Pokit App from the Apple App Store or Google Play Store.
- For information about device compatibility visit: pokitmeter.com/devices.
- Ensure the device is charged and press the button once to power on.
- Open the app and see your devices listed.
- Connect to your Pokit and start measuring.
Switch Explanation
Warning: Personal Injury or damage to the Pokit Pro can occur if you
attempt to take a measurement with the switch in an incorrect position.
Ensure the switch is in the correct position before taking a measurement by
adjusting the switch and selecting the desired measurement mode in the
app. Measurement modes will not be selectable if the switch is in the wrong
position.
Measurement Mode
Voltage:
Select this mode for voltage measurements.
|
---|---
Resistance / Low Current / Capacitance / Diodes:
Select this mode for resistance measurements, current measurements below
300 mA, diode measurement, external temperature probe use and capacitance
measurement.
|
---|---
High Current:
Select this mode for high current measurements. Warning: Restrict use to
10A max current for a maximum of 30 seconds every 15 minutes.
|
---|---
Safety Information
- Use the device only as specified in this manual. Do not exceed the measurement limits defined in the specifications.
- Do not use the device if it appears damaged, or if it is not operating properly. Inspect the probe lead, button, switch, USB port and unit for signs of wear before use.
- The black probe lead is double insulated with a white inner insulation layer. If the white layer is visible then stop using the device immediately.
- Always use proper switch position and measurement mode for measurements (Refer to Section 5).
- Turn off circuit power before connecting the device to take a current measurement. Remember to connect the device in series with the circuit when measuring current.
- Do not apply more than the rated voltage (as marked on the device) between terminals, or between any terminal and earth ground.
- Use caution with voltages above 30V ac rms, 42V ac peak, or 60V dc. These voltages pose a shock hazard.
- Disconnect circuit power and discharge all high-voltage capacitors before testing resistance, continuity, diodes, or capacitance.
- Do not use the device in an environment where explosive gas or vapor may be present.
- Keep your fingers behind the finger guards at all times when measuring with the probes.
- When measuring mains voltages, only use accessories that are rated with measurement category III or IV ratings for voltages that meet or exceed the rated voltage of the device and are compatible with Pokit Pro.
- Comply with local and national safety requirements when working in hazardous locations.
- Use proper protective equipment, as required by local or national authorities, when working in hazardous areas.
- Avoid working alone.
- Do not use device if internal water damage is suspected.
- Use only the specified replacement fuse or the protection may be impaired and the warranty voided.
- Verify the device’s operation by measuring a known voltage prior to use.
Symbol | Description |
---|---|
Caution Consult manual for more information. | |
Equipment protected throughout by double insulation or reinforced |
insulation.
| Fuse required – See section 7 of manual for details.
Fuse replacement
Only use the appropriate fuses for Pokit Pro:
Fuse for High Current: 10A, 600V AC/DC FAST 10kA Fuse (PN: 0ADAC9100-BE)
Fuse for Low Current: 0.5A, 600V AC/DC FAST 10kA Fuse (PN: 0ADAC0500-BE)
Before replacing the fuse ensure the device is removed from any external
circuits.
1. Lift external rubber fuse cover.
|
2. Remove internal fuse cover screw and remove internal fuse
cover.
---|---
3. Replace blown fuse.
|
4. Ensure replacement fuse matches the correct ratings as specified on the device.
Warning: Ensure all fuse covers are securely in place prior to operation.
Technical Specifications
Maximum voltage between any terminal and earth ground: 600V Surge Protection 6
kV peak per IEC 61010-1 600V CAT III, Pollution Degree 2
Fuse for High Current: 10A, 600V AC/DC FAST 10kA Fuse (PN: 0ADAC9100-BE)
Fuse for Low Current: 0.5A, 600V AC/DC FAST 10kA Fuse (PN: 0ADAC0500-BE)
Safety Compliance IEC 61010-1:2010+A1:2016 (Ed 3.0)
Safety Compliance IEC 61010-2-033:2012 (Ed 1.0)
Maximum Operational Altitude: 2000m
Characteristic | Specification |
---|---|
Voltage | 1mV to 600V AC(True RMS)/DC ±1% |
Current | 1µA to 10A* AC(True RMS)/DC ±1% |
Resistance | 100mΩ to 1MΩ ±1%, 1MΩ to 5MΩ ±5% |
Temperature | 0 to 60˚C ±1˚C, 32 to 140˚F ±1.8˚F |
Capacitance | 1nF to 1000µF ±2% |
Diode | 0V-1.5V |
Continuity | 100ms, Internal Buzzer |
Logging | 128kB, 3hr/sample @ 6 months |
Logged Parameters | Voltage, Current, Temperature |
Logging Sample Rate | 100ms to 18hr |
Input Impedance (DC) | 10MΩAll measurements are ±3 readings |
Operating Temperature | -10 to 40˚C, 14 to 104˚F |
Battery Charging Temperature | 10 to 45˚C, 50 to 113˚F |
Max Sample Rate | 1M Samples/sec |
BLE Range | 10m |
*Do not measure Currents above 6A for more than 30 seconds every 15 Minutes.
Full specifications are available in the app.
EMC Compliance: CE, FCC, RCM, MiC
- Changes or modifications not expressly approved by the party responsible for compliance could void your authority to operate the equipment.
- This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses,
and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
Supplier’s Declaration of Conformity
Trade Name: | Pokit Innovations Pty. Ltd. |
---|---|
Model Number: | POK-PRO |
Responsible Part—U.S. Contact Infomation
Company:| Pokit Innovations Pty. Ltd.
Street Address:| Suite 2.01, 56 Delhi Rd
City, State:| Macquarie Park, NSW
Zip Code:| 2113
Telephone number:| +61 1300 611 388
Internet contact infomation:|
www.pokitinnovations.com
FCC Compliance Statement:
This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference recieved,
including interference that may cause undesired operation.
Recharging
Ensure the device is removed from any external circuits before recharging the
device.
Do not perform measurements when the device is charging.
Included is a USB C cable which can be used to charge your device from any
compliant USB-A port. The device should fully recharge in 3 hours or less.
Taking Measurements
As a safety precaution when working on mains circuits, if possible, isolate
circuits before taking measurements. Always wear appropriate PPE safety gear
when measuring live circuits.
Pokit Pro is Rated for CAT II 600V circuits when no accessories are fitted. To
measure CAT III circuits, probe shrouds must be added to the probe tips in
order to safeguard from short circuits as per IEC61010. Some of the
accessories are also rated to CAT III and may also be used to measure CAT III
circuits (Refer to Section 11).
Refer to Appendix A for more info regarding measurement categories.
Voltage
When measuring voltage, make sure the switch is set to voltage measurement. To
test the Pokit Pro devices, take a measurement on a known live circuit to
ensure the device is operating correctly. Check (or replace) the 0.5A fuse if
0V is measured.
Warning: Stop using the device immediately if the fuse is OK and 0V is
measured.
Current
When measuring Low Current (<300mA), make sure the switch is set to Low Current measurement.
When measuring High Current (>300mA), make sure the switch is set to High Current measurement.
For Currents above 6A, do not exceed the 30sec/15min rating of the high
current mode. After 30 seconds of measurement wait 15 minutes before taking
your next reading. The unit may become hot if measurements are taken for
longer periods of time.
Isolate the circuit to be measured before connecting and disconnecting the
Pokit Pro. Do Not connect to a live circuit as this may cause arcing to occur
between the Pokit Pro test probes and the circuit contacts.
Resistance/Continuity
When measuring Resistance, make sure the switch is set to Resistance
measurement.
Warning: Do not connect to Live Mains Voltages in Resistance mode.
Capacitance
When measuring Capacitance, make sure the switch is set to Capacitance
measurement.
Warning: Do not connect to Live Mains Voltages in Capacitance mode.
Diodes
When measuring Diodes, make sure the switch is set to Diode measurement.
Do not connect to Live Mains Voltages in Diode measurement mode.
See app for detailed measurement instructions.
Accessories Ratings
WARNING The applicable measurement category and voltage rating for the Pokit Pro and any attached accessories is always the lower of the two. Ensure both the accessory and the Pokit Pro are adequately rated before measurement.
Cleaning and Maintenance
Dirt or moisture on the probe pins can affect readings. Wipe the device with a damp cloth and mild detergent. Do not use abrasives or solvents.
Troubleshooting
If none of the below-troubleshooting tips resolves your issue, visit the Pokit Innovations website for more troubleshooting guides or contact our support team at support@pokitmeter.com.
No BLE Connection
If the Pokit Pro does not appear in the app, then try the following:
- Press the button once to wake the Pokit Pro up if it hasn’t been used for a while.
- Move closer to the Pokit Pro (max rated BLE range is 10m).
- Plug the Pokit Pro into a USB-compliant charger; the RGB LED should blink green to indicate a charging state.
- Try disconnecting other Bluetooth devices from your phone.
- Refresh the list of connectable devices.
- Restart the app.
- Restart the phone.
- Make sure Bluetooth is enabled on your phone and the Pokit Pro app has its Bluetooth permissions set correctly.
Poor BLE Connection
If the Pokit Pro keeps disconnecting while in use, try the following:
- Move closer to the Pokit Pro (max rated BLE range is 10m).
- Try disconnecting other Bluetooth devices from your phone.
- Restart the app.
- Restart the phone.
Ambient electrical noise can also cause poor connection.
Voltage Measurement
If the Pokit Pro is not recording a Voltage measurement, try the following:
- Make sure the switch is correctly indexed in the Voltage measurement position (refer to section 10.1).
- Make sure that you have selected the correct measurement mode for the signal i.e., AC for AC Signals such as switchboards, outlets, and household appliances and DC for DC Signals such as computer equipment, batteries, and logic circuits. If you are unsure, use oscilloscope mode in DC Coupling to determine the signal type.
- If you have multiple Pokit devices, ensure that you are connected to the correct Pokit Pro. You can flash the RGB indicator by changing devices or by pressing ‘locate’ in the app.
- Check the fuse is OK by disconnecting from the circuit you are measuring. Move the switch into Low Current measurement mode, select continuity in the Pokit Pro App and short the probe leads together. If there is no continuity, replace the 500mA fuse (Refer to
section 7).
If Voltage measurement is not working correctly after troubleshooting, an internal fault may exist, stop using the device immediately.
Low Current Measurement
If the Pokit Pro is not recording Low Current measurement, try the following:
- Make sure the switch is correctly indexed in the Low Current measurement position (refer to section 10.2).
- Make sure that you have selected the correct measurement mode for the signal i.e AC for AC Signals such as switchboards, outlets and household appliances and DC for DC Signals such as computer equipment, batteries and logic circuits. If you are unsure, use oscilloscope mode in DC Coupling to determine the signal type.
- If you have multiple Pokit Pro’s, ensure that you are connected to the correct Pokit Pro. You can flash the RGB indicator by changing devices or by pressing locate in the app.
- Check the fuse is OK by disconnecting from the circuit you are measuring. Move the switch into Low Current measurement mode, select continuity in the Pokit Pro App and short the probe leads together. If there is no continuity, replace the 500mA fuse (Refer to section 7).
If Low Current measurement is not working correctly after troubleshooting an internal fault may exist. Stop using the device immediately.
High Current Measurement
If the Pokit Pro is not recording High Current measurement , try the following:
- Make sure the switch is correctly indexed in the High Current measurement position (refer to section 10.2).
- Make sure that you have selected the correct measurement mode for the signal i.e., AC for AC signals such as switchboards, outlets and household appliances and DC for DC Signals such as computer equipment, batteries and logic circuits. If you are unsure, use oscilloscope mode in DC Coupling to determine the signal type.
- If you have multiple Pokit devices, ensure that you are connected to the correct Pokit Pro. You can flash the RGB indicator by changing devices or by pressing locate in the app.
- Replace the 10A Fuse (Refer to section 7).
If High Current measurement is not working correctly after troubleshooting, an internal fault may exist. Stop using the device immediately.
Other Measurements
If the Pokit Pro is not recording Other measurements, try the following:
- Make sure the switch is correctly indexed in the Low Current measurement position (refer to section 10.2).
- If you have multiple Pokit Pro’s, ensure that you are connected to the correct Pokit Pro. You can flash the RGB indicator by changing devices or by pressing locate in the app.
- Check the fuse is OK by disconnecting from the circuit you are measuring. Move the switch into Low Current measurement mode, select continuity in the Pokit Pro App and short the probe leads together. If there is no continuity, replace the 500mA fuse (Refer to section 7).
Getting Started with the App
To get started with the app, navigate to the Apple App store or Google Play
store (respective of your device manufacturer) and download the official Pokit
App that is compatible with Pokit Pro.
When you start the App, you will receive a prompt requesting you to connect
your Pokit devices. Once you select “Connect your devices,” you will receive
another prompt requesting permissions for the app to connect to Bluetooth
and/or to allow access to your device’s location or likewise. In order to use
the app, you will need to enable these permissions. Pokit Innovations does not
store or transmit any personal location data however, the
software modules we use need these permissions to operate correctly.
Connect your device by pressing the connect device button. Using the app,
select your device by scrolling down the list of devices until you find the
one you’re looking for. When you select the device, you will be prompted with
important safety information relating to the use of the Pokit devices. Make
sure you read this carefully before using your device. Once you have connected
to a device, press locate. The RGB indicator LED should blink and the buzzer
should sound. If it doesn’t, you have likely connected to a different device.
Press measure to start taking measurements with your Pokit device.
Appendix A
Measurement Overvoltage Categories
OVERVOLTAGE CATEGORY l is not used within the context of IEC 61010. It is
intended for equipment used to connect to a mains supply for which
precautionary measures have been taken to substantially and reliably reduce
transient overvoltages to a level where they cannot cause a hazard.
OVERVOLTAGE CATEGORY ll is intended for equipment for which power is supplied
from the building wiring. It applies to both plug-connected equipment and to
permanently connected equipment.
OVERVOLTAGE CATEGORY lll is intended for equipment that forms part of a
building wiring installation. Such equipment includes socket outlets, fuse
panels, and some mains installation control equipment.
OVERVOLTAGE CATEGORY lV is for equipment installed at or near the origin of
the electrical supply to a building, between the building entrance and the
mains distribution board. Such equipment may include electricity tariff meters
and primary overcurrent protection devices.
Pokit ® Innovations
References
- Pokit Innovations | Small Wireless Multimeter | Portable Oscilloscope | Best Bluetooth Digital Multimeter
- Warranty Policy - Pokit Innovations
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>