EXTECH 445815 Humidity Alert II Remote Probe Hygro Thermometer
Introduction
Congratulations on your purchase of Extech’s Humidity Alert lI Remote Probe Hygro-Thermometer. Humidity alarms warn when the %RH exceeds a pre-set HI or LO limit. The remote probe conveniently mounts on the meter or extends for measurements in ducts or remote locations. It also features Humidity and Temperature adjustments. The unit can be wall-mounted or placed on a flat surface using the fold-out stand. This professional meter, with proper care, will provide years of safe reliable service.
Meter Description
- temperature-humidity Sensor
- %RH Display
- Dew Point Display
- Temperature Display
- %RH LED alarm
- Max/Min push button
- Reset push button
- Sensor cable storage
- Tilt Stand
- Humidity calibration adjust
- Temperature units switch (°F/PC)
- Temperature calibration adjust
- Wall hanger
- Battery compartment
- Audible Alam ONIOff switch
Cautions
This device is not a toy and must not reach children’s hands. It contains
hazardous objects as well as small parts that the children could swallow. In
case a child swallows any of them, please contact a physician immediately Do
not leave batteries and packing material lying around unattended; they can be
dangerous for children if they use them as toys
In case the device is going to be unused for an extended period of time,
remove the batteries to prevent them from training Expired or damaged
batteries can cause cauterization on contact with the skin. Always, therefore,
use suitable hand gloves in such cases. See that the batteries are not short-
circuited. Do not throw batteries into the fire.
Copyright 2005-2015 FLIR Systems, Inc.
All rights reserved including the right of reproduction in whole or in part in
any form
ISO-9001 Certified www.extech.com
Operation
Preparation for use
Open the battery compartment by pushing the cover on the rear of the
instrument downward as indicated by the arrow. Remove the battery insulating
strip and the protective film cover on the LCD display.
Temperature Units
Select F or °C units using the rear selection switch.
Audible alarm
Select or Xusing the rear selection switch to turm the audible alarm on or
off.
Maximum/ Minimum display
- Press the MAXMIN button. The highest value measured since the Reset button was last pressed appears on the display.
- Press the MAXIMIN button again to display the lowest value measured since the Reset button was last pressed.
- Press the MAXMIN button again to return to normal display operation.
Resetting (clearing) the MAX/MIN memory
- Press the MAX/MIN button to enter the MAX/MIN mode.
- Press and hold the RESET button for 1 second to clear the memory and start recording new max/min values.
Setting the %RH Alarm Limits.
- Simultaneously Press and HOLD the MAX/MIN and RESET button until “HI” and the stored HI %RH limit appears flashing.
- Press the RESET button to set the Hi %RH limit. Hold the button in for rapid adjustment.
- Press the Max/Min button. “LO” and the stored LO %RH limit will appear flashing
- Press the RESET button to set the LO %RH limit.
- Press the Max/Min button to turn on or off the alarm. The alarm iconE will appear in the display when the alarm is active.
- Press and hold the RESET button to save the setings and exit the mode.
Alert
If %RH exceeds the HI limit or goes below the LO limit the alert will be
activated. The alert consist of two flashing ‘ALERT” icons in the upper
display, a flashing red light (for 60 seconds) and an audible buzzer (for 60
seconds)
Calibration
RH Calibration
- Insert the meter’s sensor into a humidity chamber. The reference should be 85% RH at 21°C.
- Check the reading after 1 hour.
- Adjust the RH calibration pot in % turn increments, waiting for the display to update after each adjustment until the reading is within the accuracy specification.
Temperature Calibration
- Place the sensor in a stabilized environment of approximately 21°C.
- Check the reading after 1 hour.
- Adjust the temperature calibration pot in % turn increments, waiting for the display to update after each adjustment until the reading is within the accuracy specification.
Battery Replacement
Open the battery compartment by sliding the battery cover on the rear of the
instrument downward as indicated by the arrow. Replace the 1.5V AAA’ batteries
and replace the cover.
You, as the end-user, are legally bound (Battery ordinance) to return all used
batteries and accumulators; disposal with the household garbage is prohibited!
You can hand over your used batteries/accumulators, gratuitously, at the
collection points for our branches in your community or wherever batteries
/accumulators are sold!
Disposal: Follow the valid legal stipulations in respect of the disposal
of the device at the end of its lifecycle
Specifications
Note: If subjected to an electrostatic discharge the meter may malfunction and require the user to perform a reset.