opera SR303 Relieve Overlay Mattress System Instruction Manual
Thank you for purchasing this anti-decubitus mattress replacement system. Please read these instructions carefully before setting up and using the device. Pay special attention to the warnings and other safety information. Use of genuine components is essential for optimal performance. If you do not fully understand all the instructions, safety precautions, and warnings, do not use this device. If you have questions, please contact Opera on 0333 222 8584.
Indications
This air alternating mattress system is designed for patients who endure pressure ulcer and potential patients who wish to reduce the likelihood of pressure ulcer. This device is intended to treat and prevent pressure ulcers by facilitating blood circulation and decreasing pressure of each tissues contact area. Always consult a physician or healthcare professional before using this mattress system. This anti-decubitus mattress system is mainly for low risk patient group.
Explanation of Symbols
Contradictions
Certain patient conditions are not suitable for using this type of device such as fracture of unstable vertebrae and illness of unstable vertebrae. Always consult a physician or health professional before using this device.
The use of this system does not replace the regular repositioning, monitoring and, nursing of the patient.
Safety Precautions
2.1 Proper Operation
To ensure proper operation, please inspect and verify that all parts are set up properly and are anchored securely to the bed system. Verify that the mattress replacement system does not interfere with the bed frame function. Do not place anything on top of the power unit. Make sure power cord set is underneath the bed frame and does not pose a hazard.
2.2 Use of Linens
It is recommended to limit bed linens to a single layer in order to allow moisture to escape efficiently through the coverlet. Breathable incontinent pads are recommended for use with this mattress system.
2.3 Flammability Hazard
Avoid using this device near open flames, lighters, or cigarettes.
Flammability hazard exists. This device draws air from the surrounding
environment. Thus, cigarette smoking may damage internal components.
2.4 Disinfection Between Patients
This system should be disinfected thoroughly between patients in order to avoid cross contamination.
2.5 Weight Capacity
Verify that the patient’s weight, accessories, and this mattress replacement system, do not exceed the bed frame’s manufacturer’s recommended weight capacity.
Warnings
3.1 Side Rails
Use this mattress with side rails that meet bed rail regulations (BS EN 60601-2-52:2010) to ensure that the gap between the side rail and the top of the mattress is not large enough to pose risk of head or neck entrapment. Failure to do so could result in serious patient injury or death. If applicable, adhere to facility or local guidelines regarding entrapment regulations.
3.2 Do Not Disassemble
Do not disassemble the power unit if you are not a qualified technician.
Please contact Opera on 0333 222 8584.
3.3 AP/APG Protection
This product is not AP/APG protected.
3.4 Periodic Repositioning
It is recommended that the patient be repositioned periodically while using this mattress.
3.5 Maintenance
If the equipment needs maintenance, contact Opera on 0333 222 8584 as soon as possible. For equipment that is no longer functional, make sure to follow national, state, and/or facility requirements for disposal of the unit.
3.6 Maximum Operating Temperature
The air mattress operates at a maximum temperature of 100 °F (37.7 °C).
3.7 Sharp Protrusions
Ensure that there are no protruding objects, sharp points, or bed springs under the mattress as these could puncture the air cells.
3.8 Disposal of Waste
This device must not be disposed of as normal household waste after the end of its service life. Contact your local waste recycling facility for further instructions.
System Package
4.1 Power Unit Package
Power Unit x 1
Power Cord x 1
Instructions For Use x 1
4.2 Mattress Package
Mattress replacement unit with coverlet x 1.
Power Unit Features
a. Power Switch
The green power LED will light when the main power rocker switch on the panel is turned on.
b. Lockable Patient Weight Knob
Patient Weight Knob is used to adjust the internal pressure of the mattress according to patient weight. Simply turn the knob to adjust the pressure setting if the mattress is too soft or firm to suit each patient’s needs. Caregivers should always perform a hand check by placing their hands underneath the patient’s pelvis area to check if there is sufficient air support to ensure that the patient is not bottoming out.
c. Couplers
Quick release female couplers are used to secure mattress air hoses to the power unit.
d. Strain Relief of Power Cord
Provided with a moulded-on anti-kink bushing held in place by integral slot in bottom enclosure.
e. Hanging Hooks
Hanging hooks are designed to hang the power unit on almost any foot board.
f. Air Filter and Cap
We recommend that the filter is kept clean to ensure optimal performance of the power unit.
g. Fuse
Fuse Holder
Mattress and Power Unit Installations
6.1 Secure Mattress Overlay
Place mattress overlay on top of the original mattress with the logo at the foot end. Secure the mattress using the elastic straps at four corners. Please verify all bed functions are working properly without interference before proceeding to the next step. Also, check the combined height of the Overlay and Mattress meets bed rail regulations (BS EN 60601-2-52:2010).
6.2 Secure Power Unit
Secure the power unit on to the foot board using hanging hooks.
6.3 Connect Hose Couplers
Firmly connect the air hose couplers to the couplers on the power unit.
6.4 Power Up Unit
Plug the power unit into an electrical outlet, and turn on the main power rocker switch on the control panel.
6.5 Set Patient Weight
Turn the lockable patient weight knob on the power unit’s control panel to maximum setting to shorten the initial inflation time if needed.
6.6 Wait for Inflation
Wait approximately 30 minutes for the mattress to inflate fully before allowing the patient to lie down on the mattress.
Program Settings
7.1 Adjusting Firmness
Place the patient in the centre of the mattress. Adjust the mattress’ internal pressure according to the patient weight by using the lockable patient weight knob on the control panel of the power unit. If the mattress is too soft or too firm, increase or decrease the mattress’ internal pressure one increment at a time and wait for the system to stabilise before making another change. Continue this process until comfort is achieved.
7.2 Perform Hand Check
Caregivers should always perform a hand check by placing their hands underneath the patient’s pelvis area to check if there is sufficient air support to ensure the patient is not bottoming out.
Tucking the bed sheet in too tightly may reduce the effectiveness of the system.
Patient Transfer and Transport
8.1 Transfer
It is recommended to have the mattress system fully inflated during the transfer process. Make sure the bed is secured before proceeding.
8.2 Transport
In the event of a patient transport, two options are available.
- Detach the mattress’ air hose couplers from the power unit’s quick release couplers and connect the two air hose couplers together to retain air in the mattress. The mattress will stay inflated for approximately 2 hours, depending on the patient’s weight.
- Unplug the power unit’s power cord from the wall outlet and the mattress should stay inflated for approximately 40 – 50 minutes. This mattress system also has a safety foam base underneath the air cells to support the patient for a short period of time in case of deflation. To resume normal operation, please follow the instructions beginning in Section 6.1.
Emergency CPR Deflation
In the case of emergency, pull on the air hose couplers from the control unit and turn off the power unit by pressing the power/mute button on the control panel at the same time for emergency deflation. The air will discharge from the mattress with the patient’s own weight. To resume normal operation, simply re-insert the air hose couplers, press the power/ mute button again and reset the patient weight. See instructions in Section 6.1.
Cleaning Instruction
The air mattress and power unit must be cleaned thoroughly between patients to avoid cross contamination. The following is a suggested guideline. Be sure to follow local infection control policies.
Regular cleaning can be performed at bedside with a disinfectant followed by drying with a clean dry cloth. Use only mild detergents to clean the coverlet and the mattress. Any appropriate non-phenolic cleaning agent may be used for heavy soiling from urine, blood or other bodily fluids. Please ensure that the air mattress and coverlet are completely dry before letting the patient lie on the surface again.
Machine wash warm water at maximum 71 °C.
Do not use electric or tumble dryers. Do not iron.
Always unplug the power unit before cleaning. Routine cleaning of power unit can be done by wiping down with damp cloth using disinfectant or mild detergent. Never spray liquids directly on the unit itself.
Routine Maintenance and Storage
Remove air filter from the rear panel of the power unit by opening up the filter cap. Inspect the filter for dirt or dust, and clean it with mild soap and water. Reinsert the dried air filter after cleaning and ensure that the cap is secure. If a replacement is needed, contact Opera on 0333 222 8584.
Only disinfect and dry systems should be stored. Disconnect the air hoses from the power unit. Roll up the mattress starting from the head end and working down toward the foot end. Use the straps to secure it and store in clean plastic or other storage bag. Store in a cool, dark place.
Troubleshooting
Return for Service
Service and repair must be performed by Opera authorised technicians or representatives. Please contact Opera on 0333 222 8584.
Warranty
We warrant the product to be free from defects from the date of purchase.
Please inspect all accessories when you purchase our product. If there is any damage or missing accessories when you receive the product, please ask for a replacement from Opera on 0333 222 8584 within three days of purchase.
The warranty period for the products are according to the regulations in your country, the minimum period is 2 years from date of purchase for the power unit and 2 years for the mattress and the coverlet. The warranty coverage of any product is contingent up on its purchase from Opera.
Warranty coverage will not be extended to any product on which the production lot number has been removed or defaced on which repair has been attempted by any person or agency not authorized by our company or if in the sole opinion of our company that the system shows evidence of tampering, abnormal or unreasonable abuse, negligence, accident or operation without regard for the restrictions specified in the instructions which accompany the system. This warranty does not cover normal maintenance such as cleaning, adjustment, lubrication, and updating of equipment or parts. If the damage is a result from improper operation, a reasonable service fee and part cost will be charged.
The warranty stated above is the only warranty made and is in lieu of all other warranties whether expressed or implied, including any warranty of merchantability or fitness for a particular reason, we will not be liable for consequential or incidental damages of any kind.
Product Specifications