FORENSICS DETECTORS FD-90A-O2 Single Gas Detector User Manual
- June 3, 2024
- FORENSICS DETECTORS
Table of Contents
90A-O2 Single Gas Detector
User Manual
WARNING
- KEEP THE DETECTOR AWAY FROM ELECTROMAGNETIC & MAGNETIC INTERFERENCES (i.e. PHONES & MAGNETS)
- STORE DETECTOR WITHIN SPECIFICATIONS
- IF UNWELL, SEEK CLEAN AIR & MEDICAL HELP.
- DO NOT OPEN THE UNIT
- KEEP AWAY FROM DUST & PARTICULATE
- NEVER EXPOSE TO EXHAUST GAS or CONCENTRATED VAPORS, HARSH CHEMICALS OR EXTREMELY HIGH CONCENTRATION LEVELS AS IT MAY POISON THE SENSOR
- READ & FOLLOW INSTRUCTIONS
- TO ENSURE ACCURACY, CALIBRATE THE DEVICE AT AT LEAST EVERY 6 MONTHS
- BUMP TEST WHEN FIRST UNPACKING TO CONFIRM DETECTOR OPERATION
- BUMP TEST BEFORE USE TO CONFIRM DETECTOR OPERATION
INTRODUCTION
You have purchased the SINGLE GAS DETECTOR by FORENSICS DETECTORS™. The detector has a temperature indicator, time, alarm functionality, adjustable alarms and calibration friendly for industrial, business, home or R&D.
OPERATION
ON/OFF: Press POWER button for 5 seconds. After self check, normal
operation begins and the gas level is shown.
MENU MODE: Quickly press POWER button to enter the main MENU SELECTION.
Use UP and DOWN buttons to make your selection, then press POWER button to
select.
MENU OPTIONS
Gas Zero: Expose to ZERO air for 2 minutes using certified gas or fresh
air (only for O2 Detectors, expose to pure N2). Maintain a flow of about
0.5L/min when using gas bottles and use the sensor cap provided to deliver the
gas to the detector. Then press Save to register the Zero reading.
Gas Calib: Enter passcode 8888. Enter CAL gas concentration – usually mid
point of detection range or lowest alarm level. Expose to CAL air for 2
minutes using certified gas. For O2 detectors, simply expose the detector to
fresh air that has 20.9% of O2. Maintain a flow of about 0.5L/min when using
gas bottles and use the sensor cap provided to deliver the gas to the
detector.
Set Time: Follow screen instructions. Move cursor and adjust time digits
as desired.
Record: Time-stamped history of alarm activation.
LA Set: Low Alarm point setting. Follow screen instructions. Move the
cursor and adjust alarm levels.
HA Set: High Alarm point setting. Follow screen instructions. Move the
cursor and adjust alarm levels.
Unit Set: Allows the user to select between ppm or mg/m3.
Note: For O2, the unit is measured in %vol
ESC: Exit menu options and back to normal operation.
BATTERY CHARGING
The product has a built-in lithium battery and is charged via micro-USB. When the battery mark on the screen is full, charging is completed. To Operate the Detector whilst charging, plug the charging cable whilst the detector is ON. DO NOT charge in dangerous test locations to avoid fire or explosions.
SPECIFICATIONS
Sensor: Electrochemical Sensor
Sensor Life: 2-3 years (comes with calibration certificate)
Detection Range: see Table 1
Error: <±5% F.S. of detection range (see Table 1)
Recovery/Response Time: < 30 seconds
Storage / Operating Temperature: 14°F – 122°F
Storage / Operating Humidity: <95%RH
Battery: DC3.7V Li-Ion battery 1500mAh
Dimension/Weight: 4.3×2.3×1.7 inches & 5.4oz
Rating: ATEX certified Ex ib IIB T3 Gb. IP65 certified.
Charging Time: 3 hours, Operating Time: >24 hours
Contact Us
WEB: www.forensicsdetectors.com
Email:sarah@forensicsdetectors.com
SINGLE GAS DETECTOR (PROFESSIONAL SERIES)
FORENSICS DETECTORS™
Table 1: Professional Series Model FD-90A gas detectors offered by FORENSICS DETECTORS™
Gas | Range | Low Alarm | High Alarm |
---|---|---|---|
H2 | 0-1000ppm | 35ppm | 250ppm |
H2S | 0-100ppm | 10ppm | 15ppm |
CO | 1000ppm | 50ppm | 200ppm |
CO2 | 0-50,000ppm | 1000ppm | 2000ppm |
C2H4O | 0-20ppm | 10ppm | 15ppm |
O2 | 0-30% | 19.50% | 23.50% |
NH3 | 0-100ppm | 25ppm | 50ppm |
Cl2 | 0-20ppm | 5ppm | 10ppm |
O3 | 0-20ppm | 5ppm | 10ppm |
SO2 | 0-20ppm | 2ppm | 5ppm |
PH3 | 0-20ppm | 0.3ppm | 5ppm |
NO | 0-250ppm | 20ppm | 50ppm |
NO2 | 0-20ppm | 5ppm | 10ppm |
HCN | 0-500ppm | 10ppm | 20ppm |
HCl | 0-50ppm | 10ppm | 20ppm |
CH2O | 0-10ppm | 2ppm | 5ppm |
VOC | 0-100ppm | 20ppm | 50ppm |
What is CALIBRATION?
Your detector comes already calibrated, and ready to use. Turn it ON and GO.
However, calibration is an important function to be performed to ensure your
gas detector operates accurately (EVERY 6 MONTHS). Accuracy and Calibration
drift can happen over time because of the chemical degradation of sensors and
the natural drift in electronic components. There are two parts to the
calibration, ZERO Calibration, and SPAN Calibration.
ZERO CALIBRATION: Ensures a good baseline to ZERO target gas exposure.
This ensures the detector reads a true ZERO. For example, for CO detectors,
this is performed in the fresh air, with NO carbon monoxide present.
SPAN CALIBRATION: Ensures accurate gas concentration reading (i.e. ensure
that the display reading in ppm is accurate and true). For example, an OSHA
safety officer using a CO detector used in the field would want to calibrate
to a concentration of 50ppm, since ambient CO is usually in the lower range.
The span calibration gas concentration chosen is best chosen to represent the
concentration that the sensor typically is exposed to, as to ensure maximum
accuracy for daily application usage.
What is Bump Testing?
Bump testing is to expose the gas detector to a small amount of “blast” of target gas to ensure the detector operates and alarms as programmed. The function of this test is to verify detection operation and build user confidence, particularly in hazardous and critical user applications. Recommended bumping test when first purchased and unpack the detector and weekly thereafter.
GAS SAMPLING PUMP
Not required but recommended for continuous monitoring or gas sampling in
isolated areas (hard to get to areas) such as sewers, tanks, shafts, etc…
Sold Separately. Amazon.com
Product ASIN: B0859JDCNB
ROBUST WORK DESIGN
The professional series detector is a robustly designed unit. It arrives with
a sling, calibration cap, and cable charger with an attached metal belt clip.
The detector conforms to a variety of qualifications: CE ATEX certified Ex
ib IIB T3 Gb IP65 certified
Product Tested, Calibrated, QA/QC in California, USA
Product Packaged in California, USA
Product Made in China