UNITREND UT-P21 High Voltage Test Probe For Oscilloscope Instruction Manual
- June 5, 2024
- UNITREND
Table of Contents
UNITREND UT-P21 High Voltage Test Probe For Oscilloscope
Specification
Safety Instruction
This product is only for trained personnel or qualified engineer to use.
- Do not work alone when operating high voltage test.
- Check the test probe and cable is in good condition before use.
- Keep your hands, shoes and floor stay dry to avoid electric shock.
- Before connecting, disable the high voltage power supply. If you want to stop using, please turn off the power firstly.
Operational Approach
-
Please connect the ground wire to a well ground point.
-
Connect the BNC probe to the input end of the oscilloscope.
-
Set the demand position of the oscilloscope
Noted! Do not turn on the high voltage power supply until the above connection is completed. -
Turn on the oscilloscope to wait the condition is stable, and then to connect the test point and able the high voltage power supply.
Matters Needing Attention
- Do not measure the grounding cable before it is properly connected
- Under no circumstances shall the ground cable connected to a high-voltage power supply or the tip of the test probe be grounded.
- Before turning on the high voltage power supply make sure that no part of your body is in contact with the high voltage.
- When measuring DC1OKV or ACTKV, do not continue testing for more than 1 minute in order to avoid overheating and affect the accuracy of measurement. During this period, wait for cooling for about 5 minutes to continue measuring
- Before removing grounding, make sure the tip of the test prod is away from the high voltage supply.
Clean
- When cleaning the product, wipe it with a soft wet cloth. Do not immerse the product in water.