smeg BPW1260 Bedpan Washers User Guide

June 1, 2024
Smeg

smeg-LOGO

smeg BPW1260 Bedpan Washers

Specifications:

  • Cold Water Consumption (L): Min 0.5 ml/l, Max 2 ml/l
  • Hot Water Electricity Consumption (L): 28.5 (*39.5)
  • Total Descaler Consumption (ml): 22
  • Total Detergent Consumption (ml): Min 0.5 ml/l, Max 10 ml/l
  • Cycle Average Duration (min): 300

Product Usage Instructions

Program Selection:

  1. Refer to the brief table of programs to identify the specific program based on the type of load.
  2. Select the appropriate program number on the device interface.

Loading the Device:

  1. Load the urine bottles, bedpans, or other items according to the program selected.
  2. Ensure items are arranged properly for effective cleaning.

Detergent and Descaler Setup:

  1. Check the default detergent and descaler values for each program from the table.
  2. Adjust detergent and descaler dosages as needed, except for program no. 8 CLOSTRIDIUM.

Starting the Cycle:

  1. Close the device door securely.
  2. Press the start button to initiate the washing cycle.

Frequently Asked Questions (FAQ):

  1. Can I change the default detergent and descaler values for all programs?
    The detergent and descaler dosages can be adjusted for most programs except for program no. 8 CLOSTRIDIUM, where the default values cannot be changed.

  2. How do I know which program to select for different types of loads?
    Refer to the brief table of programs in the user manual to determine the suitable program based on the type of load, whether it’s urine bottles, bedpans, or other items.

BPW1260 – BPW4260 PROGRAM TABLE
BPW1260: Sanitary bedpan washer – Hospital Sector

WARNING
This document specifies the characteristics of the washing programs and constitutes an appendix to the device’s user manual; it must be consulted together with the manual for proper, safe use of the device. Programs may vary depending on the firmware version installed. If in doubt, have authorised staff check that the installed version is the same as the version to which the document refers, stated in the concluding table.

BRIEF TABLE OF PROGRAMS


| ****

NAME

| ****

TYPE OF LOAD

| ****

A0

| COLD WATER CONSUMP- TION (L)| HOT WATER CONSUMP- TION (L)| ELECTRICITY CONSUMP- TION (kWh)| TOTAL DESCALER CONSUMPTION (ml)| TOTAL DETERGENT CONSUMPTION (ml)

(when installed)

| CYCLE AVERAGE DURATION

(min)

---|---|---|---|---|---|---|---|---|---
MIN| | MAX| MIN| | MAX
0,5 ml/l| 2 ml/l| 0,5 ml/l| | 3 ml/l| | 10 ml/l
1| URINE BOTTLES| ONLY URINE

BOTTLES

| 300| 28,5| 22| 0,42| **14,5| …| 58| 5,5| | **33| | 110| 9


2

| NORMAL BEDPANS| URINE BOTTLES + BEDPAN| ****

300

| 28,5 (39,5)| 44 (55)| ****

0,43

| ****

**14,5

| ****

| ****

58

| 5,5 (*11)| ****

| **33 *(66)| **

| 110 (*220)| ****

11 (*14)


3

| INTENSIVE BEDPANS| URINE BOTTLES + BEDPAN| ****

300

| 28,5 (39,5)| 66 (88)| ****

0,44

| ****

**14,5

| ****

| ****

58

| 11 (*16,5)| ****

| **66 *(99)| **

| 220 (*330)| ****

15 (*18)

4| ECO URINE

BOTTLES

| ONLY URINE

BOTTLES

| 60| 24,8| 14,6| 0,37| **14,5| …| 58| **3,7| | 21,9| | 73| 8


5

| ECO NORMAL BEDPANS| URINE BOTTLES + BEDPAN| ****

60

| 24,8 (32,1)| 32,9 (40,2)| ****

0,40

| ****

**14,5

| ****

| ****

58

| **3,7 *(7,3)| **

| 21,9 (*43,8)| ****

| 73 (*146)| ****

9 (*11)


6

| ECO INTENSIVE BEDPANS| URINE BOTTLES + BEDPAN| ****

60

| 24,8 (32,1)| 47,5 (62,1)| ****

0,40

| ****

**14,5

| ****

| ****

58

| **7,3 *(11)| **

| 43,8 (*65,7)| ****

| 146 (*219)| ****

11 (*14)


7

| MULTIPURPOSE RACK| BUCKETS + VARIOUS

CONTAINERS

| ****

60

| ****

39,5

| ****

44

| ****

0,34

| ****

**14,5

| ****

| ****

58

| ****

5,5

| ****

| ****

**33

| ****

| ****

110

| ****

11


8

| ****

CLOSTRIDIUM

| URINE BOTTLES + BEDPAN| ****

600

| 28,5 (39,5)| 77 (99)| ****

0,42

| ****

**14,5

| ****

| ****

58

| ****

N/A

| **253 *(286)| **

20 (*23)

Note

  1. the value in brackets refers to consumption with device fitted with the support for washing two bedpans (ref. “BPW-2B”, code 905423). The difference in electricity consumption betweenthe two scenarios is negligible.
  2. the descaler and detergent values in bold type refer to the standard default setup. Dosages can be set depending on requirements (except for program no.8 CLOSTRIDIUM, where thedefault value cannot be changed).

ABBREVIATIONS and TERMS USED

CODE DESCRIPTION
CW Cold water
HW Hot Water
P1 Peristaltic dispensing pump P1, for detergent – optional.

When the washing program indicates P1: if the optional accessory is installed, the relative

detergent is dispensed; otherwise the device performs the phase without chemical additives.

P2| Peristaltic dispensing pump P2, for acid.
Pi(n)| Refers to: dosage of n ml/litre of the product dispensed by pump Pi. [P1 (3) indicates dosage of alkaline detergent with concentration of 3ml/litre]
A 0| Thermal disinfection parameter, see specific point.
IdProg| Unique identifier of the program, used in the internal memory of the electronic circuit board and in the printout (if the printer is installed).
Ph| Phase. “Ph 1” refers to Phase 1 of the washing cycle.
TD| Thermal disinfection phase, performed using steam: the spray systems are not activated and instead the steam generated by the device heats the load to a high temperature to

reach the target value of parameter A0.

T_target| Target temperature during the final steam removal phase.
R| Suffix applied to a specific phase (e.g. P1-R) to indicate that the phase is performed when washing of Bedpan 2 (on the right) is enabled and is actually a repetition of the previous phase, performed for Bedpan 1 (left).

N.B.: when bedpan 2 is not placed in the device, the phase is bypassed.

Steam removal| At the end of the thermal disinfection process, the device removes the steam from the chamber:

–          By fan-assisted convection: if the specific system is installed.

–          By natural convection: otherwise.

The process ends when the target temperature in the chamber (T_target), generally set around 65 °C, is reached.

N.B.: this process is intended to remove the steam and not for washing and disinfection

purposes; it is not marked as a specific phase on the display.

Pause| A controlled period of time between one phase and the next.

Pauses only occur in specific cycles and are included in the parameters in the table.

ABBREVIATIONS REFERRING TO THE TYPE OF SPRAYING AND WASHING PERFORMED

CODE DESCRIPTION
U.B. Urine Bottles. Washing process for this type of sanitary aids.
BP.1 Bedpan 1, left. Washing process for this type of load.
BP.2 Bedpan 2, right. Washing process for this type of load, only

active if the specific circuit is installed.
U.S.| Upper Spraying.
M.W.| Multipurpose wash , ”multipurpose” rack washing for various types of support
ALL| All the spraying systems installed in the device are activated simultaneously.

N.B.: the different types of spraying can also be combined. For example “M.W. + U.B.” indicates that both spraying systems, M.W. and U.B. as identified above, are activated.

PARAMETER A0

Parameter A0 – referred to below as a specific characteristic of each program – is used in the EN ISO 15883-1 product standard to assign a numerical value to the thermal disinfection efficacy achieved.

A0 is calculated by performing an addition, as follows:

Where:

z = 10 °C| T = temperature in °C| t = time in seconds of the interval considered
---|---|---

For calculation of the parameter, the device considers:

  1. The final, Thermal disinfection phase.
  2. The moments during which the temperature is above 75 °C, adding them up in accordance with the formula.

The A0 parameter values in the table refer to the standard default values. Any changes to the parameter can be made by an operator with “Superuser” credentials, using the relative password.

RECOMMENDED CHEMICAL PRODUCTS

P1 – Detergent:

  • Smeg DETER-BPW
  • Neodisher Mediclean Forte

P2 – Acid neutralising agents:

  • Smeg DEKAL 100

The chemical concentration values in the table refer to the standard default values. Any changes to these values can be made by an operator with “Superuser” credentials, using the relative password.

Programs were tested in accordance with the relevant performance and safety standards, with reference to the chemical agents mentioned in this document and recommended by the authorised staff.
Programs were tested with positive outcome both with and without the optional pump P1.

The use of other chemical products may cause unsatisfactory washing results or lead to damage for the device or the surrounding environment and harm to users. The manufacturer declines all liability for uses other than those specified in the documentation accompanying the product.

WASHING PROGRAMS

INTRODUCTORY NOTES
Below, each program is subdivided into its constituent phases, explaining, by means of abbreviations:

  • the type and quantity of water taken in,
  • The type of chemical agent dispensed,
  • The spraying mode used in the phase, since the device has different spraying circuits depending on the type of load (urine bottles, bedpans or other items).

The descriptive name of the program, as provided below in full (e.g. “URINE BOTTLES”), is used in this document to clarify the intended type of load, and is saved in the device’s internal memory but does not appear on the display.
The type of load which can be processed depends on the washing support used and the device’s actual equipment. Consult authorised after-sales service staff for information about the supports compatible with the device.
“Preparation” is the initial phase which prepares the device for the washing cycle as such, in which it takes in cold water.

SERVICE PROGRAMS

Programs PRDIS, FILL, PRSR1, RESET and TEST [Also saved in the memory with IdProg. 800, 801, 802, 803 and 804] are service programs marked with the relative icon (as shown in the illustration) and cannot be used to process the load.


NAME

| ****

TYPE OF LOAD

| ****

A 0

| COLD WATER CONSUMPTION (L)| HOT WATER CONSUMPTION (L)| ELECTRICITY CONSUMPTION

(kWh)

| AVERAGE DESCALER CONSUMPTION

(ml)

| AVERAGE DETERGENT CONSUMPTION

(ml)

| CYCLE AVERAGE DURATION

(min)

---|---|---|---|---|---|---|---|---


PRDIS

| URINE BOTTLES

+ BEDPAN

| ****

600

| ****

24,8

| ****

14,6

| ****

0,45

| ****

23,5

| ****

| ****

9

FILL| –| –| 11| –| –| 5,5| –| 2
PRSR1| –| –| –| 29,3| –| 110| 110| 7
RESET| –| –| –| –| –| –| –| –
TEST| –| 6000| 28,5| 33 (44)| 0,77| 108| 99 (198)| 24 (*27)

PROGRAM 01 – URINE BOTTLES

On display: Pr 01 – IdProg. 701
Intended purpose: washing and thermal disinfection of Urine Bottles.

Phase

No.

| Phase Type| Water Type| Water Intake

[Litres]

| Additive| Washing Type / Target
---|---|---|---|---|---
Ph.1| Preparation| CW| 6.5| P2(0,5)| –
Ph.2| Prewash| CW| 11| –| U.B. [Urine Bottles]
Ph.3| Wash| HW| 11| P1(3)| U.B. [Urine Bottles]
Ph.4| Rinse| HW| 11| P2(0,5)| ALL
Ph.5| Thermal

disinfection

| CW| 11| P2(0,5)| TD , A0 = 300
–| Steam removal| –| –| –| T_target = 55 °C

PROGRAM 02 – NORMAL BEDPANS

On display: Pr 02 – IdProg. 702
Intended purpose: washing and thermal disinfection of sanitary bedpans and urine bottles.

Phase

No.

| Phase Type| Water Type| Water Intake

[Litres]

| Additive

(ml/litre)

| Washing Type / Target
---|---|---|---|---|---
Ph.1| Preparation| CW| 6.5| P2(0,5)| –
Ph.2| Prewash| CW| 11| –| BP.1 [Bedpan 1, left]
Ph.2- R| Prewash| CW| 11| –| BP.2 [Bedpan 2, right]
Ph.3| Prewash| HW| 11| –| U.B. [Urine Bottles]
Ph.4| Wash| HW| 11| P1(3)| BP.1 [Bedpan 1, left]
Ph.4- R| Wash| HW| 11| P1(3)| BP.2 [Bedpan 2, right]
Ph.5| Rinse| HW| 11| –| U.S. [Upper Spraying]
Ph.6| Rinse| HW| 11| P2(0,5)| ALL
Ph.7| Thermal

disinfection

| CW| 11| P2(0,5)| TD , A0 = 300
–| Steam removal| –| –| –| T_target = 55 °C

PROGRAM 03 – INTENSIVE BEDPANS

  • On display: Pr 03 – IdProg. 703
  • Intended purpose: washing and thermal disinfection of sanitary bedpans and urine bottles – intensive washing validated under 15883.

Phase

No.

| Phase Type| Water Type| Water Intake

[Litres]

| Additive

(ml/litre)

| Washing Type / Target
---|---|---|---|---|---
Ph.1| Preparation| CW| 6.5| P2(0,5)| –
Ph.2| Prewash| CW| 11| –| BP.1 [Bedpan 1, left]
Ph.2- R| Prewash| CW| 11| –| BP.2 [Bedpan 2, right]
Ph.3| Wash| HW| 11| P1(3)| U.B. [Urine Bottles]
Ph.4| Wash| HW| 11| P1(3)| BP.1 [Bedpan 1, left]
Ph.4- R| Wash| HW| 11| P1(3)| BP.2 [Bedpan 2, right]
Ph.5| Rinse| HW| 11| –| U.S. [Upper Spraying]
Ph.6| Rinse| HW| 11| –| BP.1 [Bedpan 1, left]
Ph.6- R| Rinse| HW| 11| –| BP.2

[Bedpan 2, right]
Ph.7| Rinse| HW| 11| –| U.B. [Urine Bottles]
Ph.8| Rinse| HW| 11| P2(0,5)| ALL
Ph.9| Thermal

disinfection

| CW| 11| P2(0,5)| TD , A0 = 300
–| Steam removal| –| –| –| T_target = 55 °C

PROGRAM 04 – ECO URINE BOTTLES

On display: Pr 04 – IdProg. 704
Intended purpose: washing and thermal disinfection of Urine Bottles – Eco: water and energy saving.

Phase

No.

| Phase Type| Water Type| Water Intake

[Litres]

| Additive| Washing Type / Target
---|---|---|---|---|---
Ph.1| Preparation| CW| 6.5| P2(0,5)| –
Ph.2| Prewash| CW| 7.3| –| U.B. [Urine Bottles]
Ph.3| Wash| HW| 7.3| P1(0,5)| U.B. [Urine Bottles]
Ph.4| Rinse| HW| 7.3| P2(0,5)| ALL
Ph.5| Thermal

disinfection

| CW| 11| P2(0,5)| TD , A0 = 60
–| Steam removal| –| –| –| T_target = 55 C

PROGRAM 05 – ECO NORMAL BEDPANS

On display: Pr 05 – IdProg. 705
Intended purpose: washing and thermal disinfection of sanitary bedpans and urine bottles – Eco: water and energy saving.

Phase

No.

| Phase Type| Water Type| Water Intake

[Litres]

| Additive

(ml/litre)

| Washing Type / Target
---|---|---|---|---|---
Ph.1| Preparation| CW| 6.5| P2(0,5)| –
Ph.2| Prewash| CW| 7.3| –| BP.1 [Bedpan 1, left]
Ph.2- R| Prewash| CW| 7.3| –| BP.2 [Bedpan 2, right]
Ph.3| Prewash| HW| 7.3| –| U.B. [Urine Bottles]
Ph.4| Wash| HW| 7.3| P1(0,5)| BP.1 [Bedpan 1, left]
Ph.4- R| Wash| HW| 7.3| P1(0,5)| BP.2 [Bedpan 2, right]
Ph.5| Rinse| HW| 7.3| –| U.S. [Upper Spraying]
Ph.6| Rinse| HW| 7.3| P2(0,5)| ALL
Ph.7| Thermal

disinfection

| CW| 11| P2(0,5)| TD , A0 = 60
–| Steam removal| –| –| –| T_target = 55 °C

PROGRAM 06 – ECO INTENSIVE BEDPANS

On display: Pr 06 – IdProg. 706

Intended purpose: washing and thermal disinfection of sanitary bedpans and urine bottles – Eco: water and energy saving.

Phase

No.

| Phase Type| Water Type| Water Intake

[Litres]

| Additive

(ml/litre)

| Washing Type / Target
---|---|---|---|---|---
Ph.1| Preparation| CW| 6.5| P2(0,5)| –
Ph.2| Prewash| CW| 7.3| –| BP.1 [Bedpan 1, left]
Ph.2-R| Prewash| CW| 7.3| –| BP.2 [Bedpan 2, right]
Ph.3| Wash| HW| 7.3| P1(3)| U.B. [Urine Bottles]
Ph.4| Wash| HW| 7.3| P1(3)| BP.1 [Bedpan 1, left]
Ph.4- R| Wash| HW| 7.3| P1(3)| BP.2 [Bedpan 2, right]
Ph.5| Rinse| HW| 7.3| –| U.S. [Upper Spraying]
Ph.6| Rinse| HW| 7.3| –| BP.1 [Bedpan 1, left]
Ph.6- R| Rinse| HW| 7.3| –| BP.2 [Bedpan 2, right]
Ph.7| Rinse| HW| 7.3| –| U.B. [Urine Bottles]
Ph.8| Rinse| HW| 7.3| P2(0,5)| ALL
Ph.9| Thermal

disinfection

| CW| 11| P2(0,5)| TD , A0 = 60
–| Steam removal| –| –| –| T_target = 55 °C

PROGRAM 07 – MULTIPURPOSE RACK

On display: Pr 07 – IdProg. 707
Intended purpose: washing and thermal disinfection of other non-critical, non-invasive aids and containers similar to bedpans.
Optional accessory R-UNIV (multipurpose rack) must be installed in order to use this program.

Phase

No.

| Phase Type| Water Type| Water Intake

[Litres]

| Additive

(ml/litre)

| Washing Type / Target
---|---|---|---|---|---
Ph.1| Preparation| CW| 6.5| P2(0,5)| –
Ph.2| Prewash| CW| 11| –| M.W. [Multipurpose Wash]
Ph.3| Prewash| HW| 11| –| U.S. [Upper Spraying]
Ph.4| Wash| HW| 11| –| M.W. [Multipurpose Wash]
Ph.5| Rinse| HW| 11| P1(3)| U.S. [Upper Spraying]
Ph.6| Rinse| HW| 11| –| M.W. + U.S. [Multipurpose Wash + Upper Spraying]
Ph.7| Thermal

disinfection

| CW| 11| P2(0,5)| TD , A0 = 300
–| Steam removal| –| –| P2(0,5)| T_target = 55 °C

PROGRAM 08 – CLOSTRIDIUM

On display: Pr 08 – IdProg. 708
Intended purpose: washing and thermal disinfection of sanitary bedpans and urine bottles, with special treatment for removal of “Clostridium Difficile” bacteria.

In order to use this program, P1 must have been installed and enabled within the program (by the engineer).

Phase

No.

| Phase Type| Water Type| Water Intake

[Litres]

| Additive

(ml/litre)

| Washing Type /

Target

---|---|---|---|---|---
Ph.1| Preparation| CW| 6.5| P2(0,5)| –
Ph.2| Prewash| CW| 11| –| BP.1 [Bedpan 1, left]
Ph.2- R| Prewash| CW| 11| –| BP.2 [Bedpan 2, right]
Ph.3| Prewash| HW| 11| –| U.B. [Urine Bottles]
Ph.4| Wash| HW| 11| P1(3)| BP.1 [Bedpan 1, left]
Ph.4- R| Wash| HW| 11| P1(3)| BP.2 [Bedpan 2, right]
Ph.5| Wash| HW| 11| P1(10)| BP.1 + U.B [Bedpan 1 + Urine Bottles]
–| Wash| HW| 11| P1(10)| BP.2 + U.B [Bedpan 2 + Urine Bottles]
–| –| –| –| –| [ + Pause 60 sec. ]
Ph.6| Rinse| HW| 11| –| U.S. [Upper Spraying]
Ph.7| Rinse| HW| 11| –| BP.1 [Bedpan 1, left]
Ph.7-R| Rinse| HW| 11| –| BP.2 [Bedpan 2, right]
Ph.8| Rinse| HW| 11| P2(0,5)| ALL
Ph.9| Thermal

disinfection

| CW| 11| P2(0,5)| TD , A0 = 6000
–| Steam removal| –| –| –| T_target = 55 °C

CUSTOM PROGRAMS

On display: PrC 01, PrC02 … – IdProg. 751, …, 760
Intended purpose: Washing and thermal disinfection of sanitary bedpans, urine bottles or other non-critical, non-invasive aids, such as containers similar to bedpans – programs customisable by “Superuser” user.
The fixed phases of every Custom program which cannot be edited by the user are listed in the table below.

Phase

No.

| Phase Type| Water Type| Water Intake

[Litres]

| Additive| Washing Type / Target
---|---|---|---|---|---
Ph.1| Preparation| CW| 6,5| P2(0,5)|
Ph.2|  |  |  |  |
Ph.3|  |  |  |  |
|  |  |  |  |
|  |  |  |  |
Ph.15| Rinse| HW| 11| P2(0,5)| ALL
Ph.16| Thermal

disinfection

| CW| 11| P2(0,5)| TD, A0 = 60
–| Steam removal| –| –| –| T_target = 55°C

The other, intermediate phases can be activated and customised using the parameters shown in the WD-Connect Multi app.
These are the programs which can be modified to meet specific requirements. Modifications can be made using the Smeg WD-CONNECT MULTI software or by authorised engineers on request. For minimum installation requirements, contact your authorised dealer.

Always keep a record of the various phases and characteristics of the new custom programs.

Note

  1. construction of a custom washing program requires specific knowledge both of the load treatment process and of the parameters of the device being used. This function is password-protected.
  2. always proceed in compliance with the regulations in force in the place of installation: a custom program used for processing the load must be validated by the user in accordance with the relevant regulations and standards. Consult the authorised technical service for further information.
  3. For the thermal disinfection phase, it is possible to edit the A0 value.

SERVICE PROGRAMS

The following Programs, identified as Service, must not be considered as washing programs.

PrDis – SELF-DISINFECTION

On display: PrDis – IdProg. 800
Intended purpose: thermal disinfection of the chamber, without load.

Phase

No.

| Phase Type| Water Type| Water Intake

[Litres]

| Additive

(ml/litre)

| Washing Type / Target
---|---|---|---|---|---
Ph.1| Preparation| CW| 6.5| P2(0,5)| –
Ph.2| Service| CW| 7.3| –| ALL
Ph.3| Service| HW| 7.3| P2(2)| ALL
Ph.4| Service| HW| 7.3| –| BP.1 [Bedpan 1, left]

ALL

Ph.5| Thermal

disinfection

| CW| 11| P2(0,5)| TD , A0 = 600
–| Steam removal| –| –| –| T_target = 55 °C

Fill – PREPARATION

On display: Fill – IdProg. 801
Intended purpose: filling of the steam generator, to be used after a procedure which requires emptying of the generator (e.g. replacement of the heating element).

Phase

No.

| Phase Type| Water Type| Water Intake

[Litres]

| Additive

(ml/litre)

| Washing Type / Target
---|---|---|---|---|---
Ph.1| Preparation| CW| 11| P2(0,5)| –

PrSr1 – SERVICE PERISTALTIC PUMP

On display: PrSr1 – IdProg. 802
Intended use: priming of the additive circuits (e.g. after replacement of a jerry can which has run out).

Phase

No.

| Phase Type| Water Type| Water Intake

[Litres]

| Additive

(ml/litre)

| Washing Type / Target
---|---|---|---|---|---
Ph.1| Wash| HW| 11| P1(10)| ALL
Ph.2| Rinse| HW| 11| P2(10)| ALL
Ph.3| Rinse| HW| 7.3| –| ALL

ReSet – RESET

On display: ReSet – IdProg. 803
Intended purpose: emptying of the tank.

Phase

No.

| Phase Type| Water Type| Water Intake

[Litres]

| Additive

(ml/litre)

| Washing Type / Target
---|---|---|---|---|---
Ph.1| –| –| –| –| ALL

TeSt – TESTING

On display: TeSt – IdProg. 804
Intended purpose: test program after an installation/procedure by the engineer.

Phase

No.

| Phase Type| Water Type| Water Intake

[Litres]

| Additive

(ml/litre)

| Washing Type / Target
---|---|---|---|---|---
Ph.1| Preparation| CW| 6.5| P2(0,5)| –
Ph.2| Service| CW| 11| –| U.B. [Urine Bottles]
Ph.3| Service| HW| 11| P1(9)| BP.1 [Bedpan 1, left]
Ph.3- R| Service| HW| 11| P1(9)| BP.2 [Bedpan 2, right]
Ph. 4| Service| HW| 11| P2(9)| M.W. [Multipurpose Wash]
Ph. 5| Service| HW| 11| –| U.S. [Upper Spraying]
Ph.6| Thermal

disinfection

| CW| 11| P2(0,5)| TD , A0 = 6000
–| Steam removal| –| –| –| T_target = 55 C

195920397

|  |  |  |
---|---|---|---|---
 |  |  |

01 – EN

|

Program set code 16 812 0040 02

|

02/10/2023

| Initial summary table updated and phase data in detailed table corrected.

Custom program section added.

Service Programs: names updated.

00 – EN

|

Program set code 16 812 0040 00

|

15/02/2023

|

First issue.

Doc.| Rev.| Firmware version| Doc. Date| Update notes

Smeg S.p.A.
Via Leonardo da Vinci, 4
42016 Guastalla – Reggio Emilia – Italy
www.smeg-instruments.com

Italy

International Customers
Please contact your Local Smeg Distributor or write an email to service.instruments@smeg.it

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals