ROHDE SCHWARZ RT-Z2T Probe Interface Adapter Instruction Manual

June 3, 2024
ROHDE SCHWARZ

ROHDE LOGO R &S® RT-Z2T
Probe Interface Adapter
Instruction ManualROHDE SCHWARZ RT Z2T Probe Interface
Adapter

Product Description

The R&S RT-Z2T probe interface adapter is a product designed to connect a series of active Tektronix oscilloscope probes with TekProbe BNC™ level II interface to every R&S oscilloscope that is equipped with the Rohde & Schwarz probe interface.
The adapter provides power to operate the probe. The R&S oscilloscope is adjusted to the correct settings by selecting the probe from a list of predefined probes in the probe menu of the oscilloscope.
1.1 Supported Tektronix Probes
The following probes are supported by the R&S RT-Z2T adapter:

  • Single-ended probes: P6205, P6243, P6245, P6241, P6249
  • Current probes: TCP202
  • Differential probes: P6246, P6247, P6248, P6250, P6251
  • High-voltage differential probes: P5205, P5210
  • Electro-optical probes: P6701B, P6703B, P6711, P6713

1.2 Usage Information
The R&S RT-Z2T probe adapter is an add-on probe adapter, indented to be used in laboratory or test and measurement areas. These areas are used for analysis, testing, and servicing and where equipment is operated by trained personnel.
CAUTION
Risk of injury

Use the accessory carefully to prevent electric shock, personal injury, or damage. Read and observe the “Basic Safety Instructions” that are delivered as a printed brochure.
Also, read and observe the safety instructions of the oscilloscope to which the accessory is connected.
1.3 Deliveries
The R&S RT-Z2T delivery package contains the following items:

  • R&S RT-Z2T probe interface adapter
  • Manual
  • “Basic Safety Instructions” brochure

1.4 Inspecting the Contents

  • Inspect the package for damage.
    Keep a damaged package and the cushioning material until the contents have been checked for completeness and the device has been tested.
    If the packaging material shows any signs of stress, notify the carrier and your Rohde & Schwarz service center. Keep the package and cushioning material for inspection.

  • Inspect the product.
    If there is any damage or defect, or if the R&S RT-Z2T does not operate properly, notify your Rohde & Schwarz service center.

Device Description

The R&S probe interface of the R&S RT-Z2T probe interface adapter is attached to a Rohde & Schwarz oscilloscope.

ROHDE SCHWARZ RT Z2T Probe Interface Adapter - Device
Description Figure 2-1: R &S probe interface of the R&S RT-Z2T

1 = Release knob
2 = Rohde & Schwarz probe interface with 7 mm (276 mil) coaxial connector and 6 pogo pins
NOTICE
Risk of interface damage

Connect the adapter only to a base unit that has a Rohde & Schwarz probe interface. Do not connect the adapter to a commercially available BNC jack, as this connection can damage the interface irreparably. From the outside, the 7mm coaxial connector of the R&S probe interface looks like a standard BNC connector. However, it is constructed differently and has a significantly higher frequency limit.
The Tektronix active probe is connected to the adapter’s TekProbe BNC™ level II interface. This interface also provides the power for the probe.

ROHDE SCHWARZ RT Z2T Probe Interface Adapter -
OperationFigure 2-2: TekProbe BNC™ level II interface of the R &S RT-Z2T

Putting into Operation

The R&S RT-Z2T has been designed to withstand a moderate amount of physical and electrical stress. Treat the adapter with care.
NOTICE
Risk of adapter and instrument damage due to excess power
Voltages above the specified limits of the R&S RT-Z2T probe interface adapter or the connected R&S oscilloscope can damage the adapter or the oscilloscope. Do not exceed the lesser of the maximum input voltages specified for the adapter or the oscilloscope.
If the oscilloscope is set to 50 Ω, input coupling, the maximum input voltage is usually determined by the oscilloscope. If the oscilloscope is set to 1 MΩ input coupling, the maximum input voltage is usually determined by the adapter.
NOTICE
Damage caused by electrostatic discharge
To avoid damage to electronic components, protect the operational site against electrostatic discharge (ESD).

ROHDE SCHWARZ RT Z2T Probe Interface Adapter - Damage
caused

To protect the operational site, use one or both of the following measures:

  • Wrist strap with cord to ground connection
  • Conductive floor mat and heel strap combination

3.1 Connecting the Adapter to an R &S Oscilloscope
The R&S RT-Z2T has been designed for use with R&S oscilloscopes that feature an R&S probe interface.
A user manual is delivered with the adapter, which describes the usage of the adapter in detail.

ROHDE SCHWARZ RT Z2T Probe Interface Adapter - Tektronix
Probe

► Connect the adapter (1) to the Rohde & Schwarz probe interface of the base unit (2). The adapter snaps in when connected properly to the port.
► To disconnect, press the release button (3). Pull the adapter away from the front panel of the base unit.
3.2 Connecting the Tektronix Probe to the Adapter
To connect a Tektronix probe to the probe interface of the R&S RT-Z2T adapter:

  1. Turn the lever of the probe to the left side in the open position.
  2. Connect the BNC connector of the probe to the BNC socket of the R&S RTZ2T adapter.
  3. Lock the probe by pushing the lever to the right into the closed position.
  4. Adjust the oscilloscope settings by choosing the correct predefined probe or by entering the values for external attenuation, input coupling, and display unit manually.

For a description of the usage of a Tektronix probe, refer to its user manual.

Maintenance and Service

4.1 Service Strategy
Like all Rohde & Schwarz products, Rohde & Schwarz probes and adapters are of high quality and require only minimum service and repair. However, if service or calibration is needed, contact your Rohde & Schwarz service center. Return a defective product to the Rohde & Schwarz service center for diagnosis and exchange.
You can return the R&S RT-Z2T probe interface adapter for calibration. The service personnel carries out the required tests.
4.2 Contacting customer support
Technical support – where and when you need it
For quick, expert help with any Rohde & Schwarz product, contact our customer support center. A team of highly qualified engineers provides support and works with you to find a solution to your query on any aspect of the operation, programming, or applications of Rohde & Schwarz products.
Contact information
Contact our customer support center at www.rohde- schwarz.com/support , or follow this QR code:

ROHDE SCHWARZ RT Z2T Probe Interface Adapter - QR Code http://www.rohde-schwarz.com/support
Figure 4-1: QR code to the Rohde & Schwarz support page

4.3 Returning for servicing
Use the original packaging to return your R&S RT-Z2T to your Rohde & Schwarz service center. A list of all service centers is available at: www.services .rohde-schwarz.com
If you cannot use the original packaging, consider the following:

  1. Use a sufficiently sized box.
  2. Protect the product from damage and moisture (e.g. with bubble wrap).
  3. Use some kind of protective material (e.g. crumpled newspaper) to stabilize the product inside the box.
  4. Seal the box with tape.
  5. Address the package to your nearest Rohde & Schwarz service center.

4.4 Cleaning

  1. Discharge any static electricity from your hands before cleaning the facing surfaces of the sensor head.
    Thus, you ensure that no high voltage caused by static electricity is applied to the product. Application of a high voltage can damage the internal Hall elements or circuitry. You can attract static electricity to your hands by touching a nearby metal object.

  2. Clean the outside of the product using a soft cloth moistened with either distilled water or isopropyl alcohol. Keep in mind that the casing is not waterproof.
    Note: Do not use cleaning agents. Solvents (thinners, acetone), acids, and bases can damage the labeling or plastic parts.

  3. Dry the product completely before using it.

4.5 Disposal
Rohde & Schwarz is committed to making careful, ecologically sound use of natural resources and minimizing the environmental footprint of our products. Help us by disposing of waste in a way that causes a minimum environmental impact.
Disposing of electrical and electronic equipment
A product that is labeled as follows cannot be disposed of in normal household waste after it has come to the end of its service life. Even disposal via the municipal collection points for waste electrical and electronic equipment is not permitted.

Figure 4-2: Labeling in line with EU directive WEEE

Rohde & Schwarz has developed a disposal concept for the eco-friendly disposal or recycling of waste material. As a manufacturer, Rohde & Schwarz completely fulfills its obligation to take back and dispose of electrical and electronic waste. Contact your local service representative to dispose of the product.

This manual describes the R&S RT-Z2T probe interface adapter (1338.0007.02).
© 2022 Rohde & Schwarz GmbH & Co. KG
Muehldorfstr. 15, 81671 Muenchen, Germany
Phone: +49 89 41 29 – 0
Email: info@rohde-schwarz.com
Internet: www.rohde-schwarz.com
Subject to change – Data without tolerance limits is not binding.
R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG.
TEKTRONIX, TEK, and TEKPROBE are registered trademarks of Tektronix, Inc.
Trade names are trademarks of the owners.
1326.4383.02 | Version 03 | R&S®RT-Z2T