ROHDE SCHWARZ ZN-ZE1xx Calibration Units User Manual
- June 12, 2024
- ROHDE SCHWARZ
Table of Contents
ROHDE SCHWARZ ZN-ZE1xx Calibration Units
Product Information
- Product Name: Calibration Unit
- Model Number: 1179.5224.02
- Manufacturer: dataTec AG
- Contact Information: info@datatec.eu
- Website: www.datatec.eu
- Target Audience: Developers and Technicians
Product Usage Instructions
Safety Information
The product documentation provides instructions for safe and efficient
usage of the product. It is important to follow these instructions to prevent
personal injury or damage caused by dangerous situations. Safety information
can be found in the multilingual printed Safety Instructions that are
delivered with the product. Throughout the documentation, safety instructions
are provided when necessary.
Preparing for Use
Unpacking and Checking
- Unpack the product carefully.
- Retain the original packing material. Use it when transporting or shipping the product later.
- Using the delivery notes, check the equipment for completeness.
- Check the equipment for damage. If the delivery is incomplete or the equipment is damaged, contact Rohde & Schwarz.
ESD Protective Measures
To protect the calibration unit and vector network analyzer(VNA) against
electrostatic discharge (ESD) damage, use the wrist strap and grounding cord
supplied with the VNA. Connect yourself to the GND connector of the VNA. Refer
to the Getting Started guide of your VNA for more details.
USB Connection
A direct USB connection is recommended. You can connect several calibration
units to different USB ports of the analyzer and also connect other USB
devices simultaneously (e.g., mouse, keyboard, memory stick). Some important
information regarding
USB connection:
- The length of the connecting cable must not exceed 3 m. Use the USB cable delivered with the calibration unit, if possible.
- The calibration unit relies on a USB voltage supply of 5V. If you connect it via a USB hub, use an active one.
- During a firmware update of the VNA, the calibration unit must be disconnected.
- Do not disconnect the unit while data is being transferred between the analyzer and the unit.
RF Connections
The calibration unit provides two RF connectors (numbered 1 to 2), which
should be connected to the test ports of the analyzer. The connector type is
the same for all RF ports of a calibration unit, either type N or PC 3.5 mm
equivalent.
Important considerations:
- The maximum RF input power of the calibration unit is beyond the RF output power range of the analyzer. If directly connected to the test ports of the analyzer, there is no risk of damaging the calibration unit.
- If using an external power amplifier, ensure that the maximum RF input power of the calibration unit mentioned in the data-sheet is never exceeded.
Preparing for use
Here, you can find basic information about setting up the product for the first time.
Unpacking and checking
- Unpack the product
- Retain the original packing Use it when transporting or shipping the product later.
- Using the delivery notes, check the equipment for
- Check the equipment for
If the delivery is incomplete or the equipment is damaged, contact Rohde & Schwarz.
ESD protective measures
To protect the calibration unit and vector network analyzer (VNA) against
electrostatic discharge (ESD) damage, use the wrist strap and grounding cord
supplied with the VNA. Connect yourself to the GND connector of the VNA. For
details, refer to the “Getting Started” guide of your VNA.
USB connection
The calibration unit is equipped with a USB Type-C connector at the rear.
The USB connection is used to power supply and control the unit from a VNA of
the R&S® ZVx or R&S® ZNx family. A direct USB connection is recommended. You
can connect several calibration units to different USB ports of the analyzer
and you can connect other USB devices (mouse, keyboard, memory stick etc.)
simultaneously.
In general, an unused calibration unit can remain connected to the USB port of the analyzer and you can connect or disconnect it while the VNA is operating. However, note the following restrictions.
- The length of the connecting cable must not exceed 3 Use the USB cable delivered with the calibration unit, if possible.
- The calibration unit relies on a USB voltage supply If you connect it via a USB hub, you have to use an active one.
- During a firmware update of the VNA, the calibration unit must be disconnected
- Do not disconnect the unit while data is being transferred between the analyzer and the unit.
RF connections
The calibration unit provides two RF connectors (numbered 1 to 2), to be
connected to the test ports of the analyzer. The connector type is the same
for all RF ports of a calibration unit, either type N or PC 3.5 mm equivalent.
The maximum RF input power of the calibration unit is beyond the RF output
power range of the analyzer. If the device is directly connected to the test
ports of the analyzer, there is no risk of damaging the calibration unit. If
you use an external power amplifier, make sure that the maximum RF input power
of the calibration unit quoted in the data sheet is never exceeded.
Working with R &S®ZN-ZE1xx
A calibration unit enables the automatic calibration of several VNA ports in
one simple procedure. It contains calibration standards that are
electronically switched by the analyzer firmware when a calibration is
performed. The characteristic data of these standards is stored in the
calibration unit so that the analyzer can calculate the error terms and apply
the calibration without any further input.
Automatic calibration is faster and less error-prone than manual calibration:
- There is no need to connect calibration standards manually, which is particularly time-saving if you want to calibrate multiple ports.
- Invalid calibrations due to operator errors (e.g. wrong standards or improper connections) are almost excluded.
- There is no need to handle the calibration kit
- The internal standards do not wear out because they are switched electronic-
Establishing the USB connection
To establish the USB connection between the calibration unit and VNA,
proceed as follows:
- Switch on and start up your
- NOTICE! ESD protective Wear a grounded wrist strap whenever you connect the calibration unit to the VNA via USB
- Connect the USB Type-A connector of the USB cable to any of the USB Type-A connectors on the front or rear panel of the analyzer. You can also connect the unit before switching on the analyzer.
- Wait until the operating system has recognized and initialized the new hardware. After completing the initialization, the status LED will switch to green
The unit is ready to use, as outlined below.
The function of the status LED
The LED on top of the Calibration Unit informs about the actual status of the
device. The different states have the following meanings:
Performing an automatic calibration
OFF | The calibration unit is not connected or defective. |
---|---|
Red | The microcontroller is running but there is no USB communication with the |
VNA.
With a current VNA firmware and error-free operation, this status only appears for a short time after connecting the calibration unit.
Blinking red| During the boot sequence of the VNA, the LED possibly shows a
fast blinking red until the operating system is started. The LED switches
between red and green until the firmware comes up.
Green| The calibration unit is ready to use.
Blinking orange| Data transfer between the calibration unit and the VNA. Do
not disconnect the USB cable.
Blinking blue| Calibration in progress. Do not disconnect the USB cable.
Performing an automatic calibration
Once the USB connection between the calibration unit and VNA has been established, an automatic calibration can be performed as follows:
- Start the automatic calibration function of your
- Configure the automatic In particular:
- Select the test ports that you want to calibrate and the desired calibration
- Select the appropriate R&S®ZN-ZE1xx model (see Chapter 2.2, “ESD protective measures”
- Configure the required RF port assignments between the analyzer and the calibration unit.
- NOTICE! ESD protective measures. Wear a grounded wrist strap whenever you establish an RF connection between the calibration unit and VNA see Chapter 3.2, “Function of the status LED
For each RF port assignment:
- Establish the RF connections between the VNA and the calibration unit, as depicted in the analyzer GUI.
- Acquire calibration data
- Proceed with the next port assignment or dissolve the RF connections, as depicted in the analyzer GUI.
Accuracy considerations
To ensure an accurate calibration, please observe the following items:
- Do not use adaptors between the calibration unit and the test
- After connecting the unit to the USB port, allow for a sufficient warm-up time (see data sheet) before starting the calibration.
- To ensure the best accuracy, the analyzer automatically reduces the source power to –10 dBm. If the test setup contains a large attenuation, deactivate Automatic Power Reduction for Calibration Unit in the Calibration tab of the System Configuration dialog. Also, ensure an input power of –10 dBm at the ports of the calibration unit (please also refer to the datasheet).
The calibration type depends on the number of ports and the analyzer type. If a single port is calibrated, the analyzer uses a reflection calibration type (e.g. Full One Port / Refl OSM). For 2 ports, you can choose among several calibration types.
User characterization
The calibration unit offers the possibility to store multiple user
characterization data files either on its internal flash memory or on the
microSD card. Use the “Characterize Cal Unit” function of your VNA for this
purpose. For details, refer to the analyzer’s help system.
Microsd card On the rear of the calibration units, there is a microSD card slot. The microSD card can be used to store user characterization data. The factory data is always stored in the internal memory.
- Use the microSD card shipped with the calibration unit, if possible
- Before removing or inserting the microSD card, disconnect the USB connection between the calibration unit and the VNA
Contacting customer support
Technical support
Where and when you need it
For quick, expert help with any Rohde & Schwarz product, contact our customer support center. A team of highly qualified engineers provides support and works with you to find a solution to your query on any aspect of the operation, programming, or applications of Rohde & Schwarz products.
Contact information
Contact our customer support center at www.rohde-
schwarz.com/support, or follow this QR
code
Figure 4-1: QR code to the Rohde & Schwarz support page
This User Manual describes the following R&S®ZN-ZE1xx models:
- R&S®ZN-ZE104, 2 ports, 5 kHz to 4.5 GHz (1350.8040.04)
- R&S®ZN-ZE109, 2 ports, 5 kHz to 9 GHz (1350.8040.09)
- R&S®ZN-ZE118, 2 ports, 5 kHz to 18 GHz (1350.8040.18)
- R&S®ZN-ZE126, 2 ports, 5 kHz to 5 GHz (1350.8040.26)
- The R&S®ZN-ZE1xx models are equipped with two of the following port options:
- R&S®ZN-ZE1-B170, N (f), port 1
- R&S®ZN-ZE1-B171, N (m), port 1
- R&S®ZN-ZE1-B270, N (f), port 2
- R&S®ZN-ZE1-B271, N (m), port 2
- R&S®ZN-ZE1-B130, 3.5 mm (f), port 1
- R&S®ZN-ZE1-B131, 5 mm (m), port 1
- R&S®ZN-ZE1-B230, 3.5 mm (f), port 2
- R&S®ZN-ZE1-B231, 3.5 mm (m), port 2
The adapters are not removable. Mixing of port types is not possible but mixing of port gender is possible. N-type connectors are not available for frequencies > 18 GHz. For option names and ordering information, see the R&S ®ZN-ZE1xx data sheet.
© 2022 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Muenchen, Germany Phone: +49 89 41 29 – 0 Email: info@rohde-schwarz.com Internet: www.rohde-schwarz.com
Subject to change
Data without tolerance limits is not binding. R&S® is a registered trademark
of Rohde & Schwarz GmbH & Co. KG. Trade names are trademarks of the owners.
1179.5224.02 | Version 02 | R&S®ZN-ZE1xx R&S®ZN-Z1xx is abbreviated as R&S ZN-
Z1xx
Ihr Ansprechpartner / Your Partner: dataset AG E-Mail: info@datatec.eu >>> www.datatec.eu
References
- Die Experten für Mess- & Prüftechnik | dataTec
- Service and Support Request 2022 - GLORIS Rohde & Schwarz
- Service and Support Request 2022 - GLORIS Rohde & Schwarz
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>