USER MANUAL
CARBON DIOXIDE TESTER
OUT OF BOX AUDIT
please check whether the instrument is damaged in transit after receiving it.
If obvious damage or abnormal operation function is found, please contact the
seller
PRODUCT OVERVIEW
The instrument adopts an NDIR sensor, which can accurately measure the concentration of carbon dioxide in the air, monitor the concentration, temperature and humidity of carbon dioxide in the environment in real-time, and make the alarm for ex0es give a concentration of carbon dioxide Therefore it’s widely used in homes, shopping malls, offices, transportation vehicles, stations and other public places Carbon dioxide is a colorless, odorless and incombustible gas at normal temperature, which is denser than air, slightly soluble in water, and reacts with water to form carbonic acid. It’s content in fresh air is about 0.03%
FEATURES
Adopt an NDIR sensor with superior performance and high precision
The battery life of a large capacity 18650 lithium battery is more than 10 hours
» Simple operation, sensitive response, and accurate detection
HD VA reverse color screen, digital and analog dual display
» Controllable sound and backlight, which makes long-term monitoring more energy-saving
Sound and screen double alarm
Back bracket and hanging buckle, for placing on the desktop or hanging on the wall for continuous monitoring
INTRODUCTION OF PRODUCT APPEARANCE
- Power
- Backlight
- Ring
- Monitor
- Charging port
- Air ventilation network
INTRODUCTION OF MONITOR
- Battery capacity indication
- Analog display
- Numeric display
- Measurement unit
- Alarm symbol
- Ring symbol
- Temperature
- Humidity
SAFE OPERATION CRITERIA
- lease, check whether the instrument is damaged before use, such as deformation of the instrument shell, poor display, and abnormal measurement.please stop using the instrument once such damage is discovered
- please keep sufficient power when the instrument is working continuously, and do not block the side and back ventilation nets in case of affecting the measurement results
3lease follow the instruction manual when measuring 4lease do not strongly impact, shake, disassemble or modify the instrument so as not to damage the instrument and affect the measurement data
INTRODUCTION TO INSTRUMENT USE
- Press the ( ) key to turn on or off the power of the instrument
- Press the( )key shortly to decrease or increase the brightness of the screen, and press it long to change /F
- Press the( )key shortly to turn off or on the alarm bell
- After the instrument is turned on and preheated for about 30 seconds, the sensor starts to enact the carbon dioxide content in the air, and the detection data is updated every 3 seconds. When the value reaches 2000ppm, the instrument starts to give the alarm prompt: an audible alarm and an ALARM flashing prompt on the screen
INSTRUMENT CHARGING INTRODUCTION
- The charging voltage is DC5V/1A
- please use the adapter required by the symbol-related safety certification for charging in an environment with good ventilation and heat dissipation and pay attention to the electrical safety regulations
- When the instrument is charged, because the heat dissipa notion of the charging circuit will affect the ambient temperature measurement, the ambient temperature measurement can be detected normally when the instrument returns to normal temperature after charging
- When the screen of the instrument flashes ( ), it means that the battery is insufficient. Please charge the instrument as soon as possible. The charging time is about 4 hours, and the actual time will be affected by various factors
CARBON DIOXIDE CONCENTRATION AND HUMAN PHYSIOLOGICAL REACTION (FOR
REFERENCE ONLY)
Concentration Content | The physiological reaction of the human body |
---|---|
350–45OPPM | General outdoor environment |
450–1000PPM | Fresh air and smooth breathing |
1000-2000PPM | Feeling the air is cloudy and starting to feel drowsy |
2000-5000PPM | Feeling headache, drowsiness, lethargy, inattention, rapid |
heartbeat, and mild nausea.
5000PPM| May lead to severe hypoxia, resulting in permanent brain injury, coma, and even death.
TECHNICAL PARAMETERS
Product name | Carbon dioxide tester |
---|---|
Detecting gas | CO2 |
Gas resolution | 1PPM |
Temperature resolution | 0.1°C/0.2°F |
Humidity resolution | 1%RH |
Gas measurement range | 400-5000ppm/(5% reading +50PPM) |
Temperature measurement range | 0-50°C/32-122°F t (1.5°C13°F) |
Humidity measurement range | 0-100% t (5%RH) |
Calibration mode | Automatic baseline calibration |
Power Supply | Built-in 3.7V/18650 lithium battery |
Charging interface | USB Type-C |
USB charging | 5V/1A |
Charging time | About 4h |
Working hours | > 10 hours (continuous work) |
Response time | About 30 seconds |
Automatic shut-down | No |
Color VA reverse display screen | Yes |
Type of alarm | Sound and screen display |
Working temperature | 0-50°C |
Working humidity | 0-70%RH |
Storage environment | -10-50C (5%—95%RH) |
Size | 110•76°35mm |
Weight( including the battery) | 186g |
MAINTENANCE
Do not fall or impact the instrument
It is forbidden to disassemble the instrument to avoid failure »/t is
forbidden to store or use the instrument under high temperature, high
humidity, and flammable, explosive, and strong electromagnetic environment >
During the work of built-in 18650 lithium battery, please avoid high
temperature or direct sunlight for instrument cleaning, please use a soft
cloth stained with a small amount of water or neutral detergent to wipe the
instrument shell, and prohibit the use of abrasives and solvents to prevent
corrosion and damage to the instrument
PACKING LIST
1. Carbon dioxide tester | x1 |
---|---|
2. Product manual | x1 |
3. USB Type-C charging cable | xi |
Automatic and manual calibration of data points Automatic calibration.
In a well-ventilated environment, make the instrument continuously works over
24 hours under the charging state and automatically do the calibration when
the lowest concentration point in 24 hours is 4009944 as the baseline
Manual calibration In the shutdown state, press ( )lk) first and then ((
).andrelease (( ‘) first and then ( ) key. The 400PPM calibration shall be
made when the 300second countdown is
displayed at the lower left and the instrument will be automati called on
once the calibration is completed
Tip; This method should be far away from animals or human bodies during
calibration. It is recommended that the weather is sunny and the environment
is well-ventilated l