Respironics 1020015 EverFlo Q Oxygen Concentrator User Manual

June 1, 2024
RESPIRONICS

Respironics 1020015 EverFlo Q Oxygen Concentrator

Respironics-1020015-EverFlo-Q-Oxygen-Concentrator-
PRODUCT

Product Information

Specifications

  • Model: EverFlo / EverFlo Q
  • Type: Oxygen Concentrator
  • Intended Use: Providing supplemental oxygen therapy
  • Power Source: AC Power
  • Class: Class II equipment
  • Features: Molecular sieve, pressure swing adsorption process

Product Usage Instructions

Introduction

  • Your healthcare professional has prescribed this oxygen concentrator for you. Do not change the flow settings without consulting your healthcare provider. Read the manual thoroughly before use.

Intended Use

  • The EverFlo / EverFlo Q Oxygen Concentrator is designed to provide supplemental oxygen therapy. It is not intended for life support.

About Your EverFlo / EverFlo Q

  • The device concentrates oxygen from room air using a molecular sieve and pressure swing adsorption process. Your home care provider will guide you on its operation.

Parts of Your Concentrator

  • Humidifier Bottle
  • Mounting Strap
  • Oxygen Outlet Port
  • Control Panel
  • Air Inlet Filter
  • Humidifier Connector Tube
  • Flow Meter
  • Filter Door

Operating Instructions

Warnings:

  • Avoid spills.
  • Contact your home care provider if the device malfunctions.

Cautions:

  • Avoid using extension cords or adapters.
  1. Select a location that allows proper airflow for the concentrator.
  2. Plug in the power cord after reading the manual.
  3. Choose between using a humidifier or not.

Step A: Without a Humidifier

Connect your nasal cannula to the Oxygen Outlet Port.

Step B: With Humidifier

  1. Open the filter door on the back of the device.

Frequently Asked Questions (FAQ)

Q: Can I use the concentrator near a heat source?

A: No, do not place the device near any heat sources.

Q: What accessories can be used with the concentrator?

A: Use only Respironics Air Inlet Filters and Humidifier Connector Tubes with this device.

Symbol Key

Respironics-1020015-EverFlo-Q-Oxygen-Concentrator-
FIG-1

Abbreviations

  • LED Light Emitting Diode
  • LPM Liters per Minute
  • OPI Oxygen Percentage Indicator

Introduction

  • Your healthcare professional has determined that supplemental oxygen is of benefit to you and has prescribed an oxygen concentrator set at a specific flow setting to meet your needs.
  • DO NOT change the flow settings unless your health care professional tells you to do so. Please read and understand this entire manual before using the device.

Intended Use

  • The EverFlo / EverFlo Q Oxygen Concentrator is intended to provide supplemental oxygen to persons requiring oxygen therapy.
  • The device is not intended to be life-supporting or life-sustaining.

About Your EverFlo / EverFlo Q

  • The device produces concentrated oxygen from room air for delivery to a patient requiring low-flow oxygen therapy. The oxygen from the air is concentrated using a molecular sieve and a pressure swing adsorption process.
  • Your home care provider will show you how to operate the concentrator and will be available to answer any questions. If you have additional questions or problems, contact your home care provider.

Parts of Your ConcentratorRespironics-1020015-EverFlo-Q-Oxygen-
Concentrator-FIG-2

Accessory Equipment and Replacement Parts

  • Contact your home care provider if you have questions about this equipment. Use only the following Respironics accessories and replacement parts with this device:
  • Air Inlet Filter
  • Humidifier Connector Tube

Warnings and Cautions

  • Warnings A warning represents the possibility of harm to the operator or patient.
  • For proper operation, your concentrator requires unobstructed ventilation. The ventilation ports are located at the rear base of the device and the side air inlet filter. Keep the device at least 15 to 30 cm away from walls, furniture, and especially curtains that could impede adequate airflow to the device. Do not place the concentrator in a small closed space (such as a closet).
  • Do not remove the covers of this device. Servicing must be referred to an authorized and trained Respironics home care provider.
  • In the event of an equipment alarm or if you are experiencing any signs of discomfort consult your home care provider and/or your health care professional immediately.
  • Oxygen generated by this concentrator is supplemental and should not be considered life-supporting or life-sustaining. In certain circumstances oxygen therapy can be hazardous; any user should seek medical advice before using this device.
  • Where the prescribing health care professional has determined that an interruption in the supply of oxygen, for any reason, may have serious consequences to the user, an alternate source of oxygen should be available for immediate use.
  • Oxygen vigorously accelerates combustion and should be kept away from heat or open flame. Not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
  • Do not smoke, allow others to smoke, or have open flames near the concentrator when it is in use.
  • Do not use oil or grease on the concentrator or its components as these substances, when combined with oxygen, can greatly increase the potential for a fire hazard and personal injury.
  • Do not use the oxygen concentrator if either the plug or power cord is damaged. Do not use extension cords or electrical adapters.
  • Do not attempt to clean the concentrator while it is plugged into an electrical outlet.
  • Device operation above or outside of the voltage, LPM, temperature, humidity, and/or altitude values specified may decrease oxygen concentration levels.
  • Your home care provider is responsible for performing appropriate preventive maintenance at the intervals recommended by the device manufacturer.
  • Cautions A caution represents the possibility of damage to the equipment.
  • Do not place liquids on or near the device.
  • If the liquid is spilled on the device, turn the power off and unplug from the electrical outlet before attempting to clean up the spill. Call your home care provider if the device does not continue to work properly.

Operating Instructions

  • Warning: Do not use extension cords or electrical adapters.
  1. Select a location that allows the concentrator to draw in room air without being restricted. Make sure that the device is at least 15 to 30 cm away from walls, furniture, and especially curtains that could impede adequate airflow to the device. Do not place the device near any heat source.
  2. After reading this entire manual, plug the power cord into an electrical outlet.
  3. Do either Step A or Step B below.
    • A. If you are not using a humidifier, connect your nasal cannula to the Oxygen Outlet Port, as shown in the top illustration on the right.
    • B. If you are using a humidifier, follow the steps below:
    • Open the filter door on the back of the device as shown.
    • Remove the humidifier connector tube from the back of the filter door and replace the filter door, as shown.
    • Fill your humidifier bottle according to the manufacturer’s instructions.
    • Mount the filled humidifier on the top of the EverFlo / EverFlo Q device inside the velcro strap, as shown in the illustration on the right.
    • Tighten the velcro strap around the bottle and secure it so it is held firmly in place.
    • Connect the humidifier connector tube (that you retrieved from the filter door) to the Oxygen Outlet Port (as shown in Step 3-A above).
    • Connect the other end of the humidifier connector tube to the top of the humidifier with the elbow in the tubing facing the front, as shown here.
    • Connect your cannula to the humidifier bottle according to the humidifier bottle manufacturer’s specifications.
  4. Press the power switch to the On [I] position. Initially, all the LEDs will illuminate and the audible alert will beep for a few seconds. After that time, only the green LED should remain lit. You can begin breathing from the device immediately even though it typically takes 10 minutes to reach oxygen purity specifications.Respironics-1020015-EverFlo-Q-Oxygen-Concentrator-FIG-6
  5. Adjust the flow to the prescribed setting by turning the knob on the top of the flow meter until the ball is centered on the line marking the specific flow rate.Respironics-1020015-EverFlo-Q-Oxygen-Concentrator-FIG-7
  6. Be sure oxygen is flowing through the cannula. If it is not, refer to the Troubleshooting Guide in this manual.
  7. Put on the cannula as directed by your home care provider.
  8. When you are not using the oxygen concentrator, press the power switch to the Off [O] position.

Cleaning & Maintenance

  • Warning: It is important to unplug the device before you perform any cleaning.
  • Caution: Excess moisture may impair the proper operation of the device.

Cleaning

  • Periodically, use a damp cloth to wipe down the exterior case of the EverFlo / EverFlo Q device. If you use medical disinfectants, be sure to follow the manufacturer’s instructions.
  • If you are using a humidifier, clean your device according to your home care provider’s or manufacturer’s instructions.

Service

  • The EverFlo / EverFlo Q Oxygen Concentrator contains no user-serviceable parts.
  • Warning: Do not remove the covers of this device. Servicing must be referred to an authorized and trained Respironics home care provider.

How to Contact Respironics

  • To have your device serviced, contact your home care provider. If you need to contact Respironics directly, call the Respironics Customer Service department at 1-724-387-4000 or Respironics Deutschland at +49 8152 93060.

You can also use the following addresses:

  • Respironics 1001 Murry Ridge Lane
  • Murrysville, PA 15668
  • USA
  • Respironics Deutschland
  • Gewerbestrasse 17
  • 82211 Herrsching
  • Germany
  • Visit the EverFlo website at: www.everflo.respironics.com

Alarms and Troubleshooting

Alarm and Indicators

  • The device has an audible alarm and three LED indicators, as shown below.

Audible Alarm / Colored LED| Possible Cause| Your Action
---|---|---
All 3 LEDs illuminate continuously and the Audible Alarm is sounding continuously.| The device has detected a system malfunction.| Immediately turn off the device, connect to a backup oxygen source, and call your home care provider.
The Audible Alarm is sounding continuously. None of the LEDs are illuminated.| The device is turned on but is not operating. Often this

indicates that the device is not plugged in or there is a power

failure.

| Check the power outlet and verify that the device is plugged in. If the problem continues, connect to a backup oxygen source and call your home care provider.
Red LED illuminates continuously and the Audible Alarm is sounding continuously.| The device has detected a system malfunction.| Immediately turn off the device and wait 5 minutes. Restart the device. If the condition persists turn the unit off, connect to a backup oxygen source, and call your home care provider.
Yellow LED illuminates continuously. The Red LED is blinking and the Audible Alarm is beeping periodically.| The device has detected an impeded oxygen flow condition.| Follow the troubleshooting guide on the next page. Connect to a backup oxygen source and call your home care provider if your troubleshooting actions fail to end this alert condition.
Yellow LED illuminates continuously. The Red LED is off and the Audible Alarm is silent.| The device has detected a low oxygen condition (OPI units only).| Continue using the unit but call your home care provider about this condition.

Troubleshooting Guide

Problem Why it Happened What to Do

The yellow LED is blinking. The Red LED is off and the Audible Alarm is beeping periodically.| The device has detected a high oxygen flow condition.| Turn the flow rate down to your prescribed level. Wait at least 2 minutes. If the condition persists turn the unit off, connect to a backup oxygen source, and call your home care provider.
Green LED illuminates continuously. The other LEDs are off and the Audible Alarm is silent.| The device is turned on and working properly.| Take no action.
The device is not working when it is turned on.

(The Audible Alarm is sounding continuously. All LEDs are off.)

| The power cord plug is not properly inserted into the electrical outlet.| Make sure the device is properly plugged into the electrical outlet.
The unit is not receiving power from the electrical outlet.| Check your household outlet fuse or circuit.
Internal part failure.| Connect to a backup oxygen source and contact your home care provider.
The device is not working when it is turned on.

(The Audible Alarm is sounding continuously and all 3 LEDs are illuminated.)

| Internal part failure.| Connect to a backup oxygen source and contact your home care provider.
Impeded oxygen flow indication is activated. (The Yellow LED illuminates continuously,

the Red LED blinks and the Audible Alarm beeps periodically.)

| The airflow to the device is impeded or blocked.| Remove any items that appear to be blocking the airflow into the device.
The flow meter knob is completely closed.| Turn the flow meter knob counterclockwise to center the ball on the prescribed LPM flow.
The oxygen tubing is kinked and blocking the delivery of oxygen.| Check to see that the tubing is not kinked or blocked. Replace if necessary.
Limited oxygen flow to the user without any fault indication.

(All LEDs are off and the Audible Alarm is silent.)

| The oxygen tubing or cannula is faulty.| Inspect and replace the items if necessary.
There is a poor connection to a device accessory.| Ensure that all connections are free from leaks.

Specifications

Environmental

| Operating| Transport & Storage
---|---|---
Temperature| 13 to 32° C| -34 to 71° C
Relative Humidity| 15 to 95%, non-condensing| 15 to 95%, non- condensing
Altitude| 0 to 2286 m| N/A

Physical

Dimensions 58 cm x 38 cm x 24 cm
Weight 14 to 15 kg

Standards Compliance

  • This device is designed to conform to the following standards:
  • IEC 60601-1 Medical Electrical Equipment, Part 1: General Requirement for Safety
  • IEC 60601-1-2 2nd edition, Medical Electrical Equipment, Part 1-2: General Requirement for Safety – Collateral Standard: Electromagnetic Compatibility – Requirements and Tests
  • ISO 8359 Oxygen Concentrators for Medical Use – Safety Requirements

Electrical, AC Power Consumption

Models 1020000, 1020001

1020002, 1020003

1020014, 1020015

| 120 VAC ±10%, 350 W, 60 Hz
---|---
1020004, 1020005| 230 VAC ±10%, 320 W, 60 Hz
1020006,1020007, 1020008

1020009, 1020010

1020011, 1020012

1020016, 1020017

| 230 VAC ±10%, <300 W, 50 Hz
1020013| 230 VAC ±10%, <300 W, 60 Hz

Oxygen

Oxygen **Concentration* (All Models except 1020007, 1020008)| 90-96% from 0.5 to 5 LPM
---|---
Models 1020007, 1020008**| 87-96% from 0.5 to 5 LPM

Device operation above or outside of the voltage, LPM, temperature, humidity, and/or altitude values specified may decrease oxygen concentration levels.

Sound Level

Models 1020000, 1020001

1020002, 1020003

1020013

| 45 dBA typical
---|---
1020004, 1020005

1020006, 1020008

1020009, 1020010

1020011, 1020012

1020016, 1020017

| 43 dBA typical
1020007, 1020014

1020015

| <40 dBA typical

Classification

The EverFlo / EverFlo Q Oxygen Concentrator is classified as:

  • IEC Class II Equipment
  • Type BF Applied Part
  • IPX1 Drip Proof
  • Not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
  • Continuous Operation

Disposal

  • Dispose of the device under local regulations.

WEEE/RoHS Recycling Directives

  • If you are subject to the WEEE/RoHS recycling directives, refer to www.respironics.com for the passport for recycling this product.

Appendix A: EMC Information

  • Guidance and Manufacturer’s Declaration – Electromagnetic Emissions: This device is intended for use in the electromagnetic environment specified below.
  • The user of this device should make sure it is used in such an environment.

Emissions t est| Compliance| Electromagnetic environment **– G uidance**
---|---|---
RF emissions CISPR 11| Group 1| The device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11| Class B| The device is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network.
Harmonic emissions IEC 61000-3-2| Class A
Voltage fluctuations/Flicker emissions IEC 61000-3-3| Complies

Guidance and Manufacturer’s Declaration – Electromagnetic Immunity:

  • This device is intended for use in the electromagnetic environment specified below.
  • The user of this device should make sure it is used in such an environment.

Immunity   t est| IEC 60601 t est l evel| Compliance   l evel| Electromagnetic environment **– G uidance**
---|---|---|---
Electrostatic Discharge (ESD) IEC 61000-4-2| ±6 kV contact ±8 kV air| ±6 kV contact ±8 kV air| Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical Fast Transient/Burst IEC 61000-4-4| ±2 kV for power supply lines

±1 kV for input-output lines

| ±2 kV for supply mains

±1 kV for input/output lines

| Mains power quality should be that of a typical home or hospital environment.
Surge IEC 61000-4-5| ±1 kV differential mode

±2 kV common mode

| ±1 kV differential mode

±2 kV for common mode

| Mains power quality should be that of a typical home or hospital environment.
Voltage dips, short interruptions, and voltage variations on power supply input lines

IEC 61000-4-11

| <5% UT

(>95% dip in UT) for 0.5 cycles 40% UT

(60% dip in UT) for 5 cycles

70% UT

(30% dip in UT) for 25 cycles

<5% UT

(>95% dip in UT) for 5 sec

| <5% UT

(>95% dip in UT) for 0.5 cycles 40% UT

(60% dip in UT) for 5 cycles

70% UT

(30% dip in UT) for 25 cycles

<5% UT

(>95% dip in UT) for 5 sec

| Mains power quality should be that of a typical home or hospital environment. If the user of the device requires continued

operation during power mains interruptions, it is recommended that the device be powered from an uninterruptible power supply or a battery.

Power frequency (50/60 Hz) magnetic field IEC 61000-4-8| 3 A/m| 3 A/m| Power frequency magnetic fields should be at levels characteristic of a typical location in a typical hospital or home environment.
Note: UT is the a.c. mains voltage before application of the test level.

  • Guidance and Manufacturer’s Declaration – Electromagnetic Immunity: This device is intended for use in the electromagnetic environment specified below.
  • The user of this device should make sure it is used in such an environment.

Immunity   t est| IEC 60601 t est l evel| Compliance   l evel| Electromagnetic environment **– G uidance**
---|---|---|---
| | | Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Conducted RF IEC 61000-4-6| 3 Vrms

150 kHz to 80 MHz

| 3 Vrms| Recommended separation distance:

d = 1.2      150 kHz to 80 MHz

Radiated RF IEC 61000-4-3| 3 V/m

80 MHz to 2.5 GHz

| 3 V/m| d = 1.2      80 MHz to 800 MHz

d = 2.3      800 MHz to 2.5 GHz

Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).

| | | Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey a , should be less than the compliance level in each frequency range b.
| | | Interference may occur in the vicinity of equipment marked with the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

a: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcasts, and TV broadcasts cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the device.

b: Over the frequency range of 150 kHz to 80 MHz, the field strengths should be less than 3 V/m.

  • Recommended Separation Distances between Portable and Mobile RF Communications Equipment and This Device: The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
  • The customer or the user of this device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and this device as recommended below, according to the maximum output power of the communications equipment

r ated M axiMuM p ower o utput of t ransMitter (w)| s eparation d istance according to the f requency of t ransMitter ( M )
---|---
150 kHz to 80 Mh z d = 1.2| 80 Mh z to 800 Mh z d = 1.2| 800 Mh z to 2.5 Gh z d = 2.3
0.01| 0.12| 0.12| 0.23
0.1| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

Limited Warranty

  • Respironics, Inc. warrants that the system shall be free from defects of workmanship and materials and will perform under the product specifications for three (3) years from the date of sale by
  • Respironics, Inc. to the dealer. Respironics warrants that the EverFlo / EverFlo Q units serviced by Respironics, or an authorized service center, will be free from defects in serviced materials for 90 days and free from defects in workmanship for 90 days from the time of service. Respironics accessories are warranted to be free of defects in materials and workmanship for 90 days from the time of purchase.
  • If the product fails to perform under the product specifications, Respironics, Inc. will repair or replace – at its option – the defective material or part. Respironics, Inc. will pay customary freight charges from
  • Respironics, Inc. to the dealer location only. This warranty does not cover damage caused by accident, misuse, abuse, alteration, or other defects not related to material or workmanship.
  • Respironics, Inc. disclaims all liability for economic loss, loss of profits, overhead, or consequential damages that may be claimed to arise from any sale or use of this product. Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you.
  • This warranty is given in place of all other express or implied warranties, including the implied warranties of merchantability and fitness for a particular purpose. In addition, in no event shall Respironics be liable for lost profits, loss of goodwill, or incidental or consequential damages, even if Respironics has been advised of the possibility of the same. Some states or provinces do not allow the exclusion or limitation of implied warranties or the disclaimer of incidental and consequential damages. Accordingly, the laws of your state or province may give you additional protections.
  • To exercise your rights under this warranty, contact your local authorized Respironics, Inc. dealer or contact Respironics, Inc. at:
  • Respironics
  • 1001 Murry Ridge Lane
  • Murrysville, PA 15668 USA
  • Respironics Deutschland Gewerbestrasse 17 82211 Herrsching
  • Germany 1-724-387-4000 +49 8152 93060
  • EverFlo is a trademark of Respironics, Inc. and its affiliates.
  • © 2008 Respironics, Inc. and its affiliates. All rights reserved.

Documents / Resources

| Respironics 1020015 EverFlo Q Oxygen Concentrator [pdf] User Manual
1020015 EverFlo Q Oxygen Concentrator, 1020015, EverFlo Q Oxygen Concentrator, Oxygen Concentrator, Concentrator
---|---

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals