20K Easy Bottle Top Dispenser User Manual
- May 15, 2024
- EASY
Table of Contents
- Introduction
- Use Instructions
- Inspection
- Control
- Assembly Instruction
- Dispensing
- Setting the volume
- Cleaning
- Cleaning and replacing the filling valve
- Autoclaving
- Checking the volume
- Calculations (for the nominal volume)
- Trouble shooting
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Bottle top dispenser EASY 20K
Introduction
Users must read this manual carefully before using the equipment, and follow
the instructions and procedures in order to be aware of all the precautions,
as well as in order to obtain the maximum benefits and duration of the
equipment.
Service
If help is needed, contact with your supplier/manufacturer or Labbox via:
www.labbox.com
Please, provide the customer service representative with the following
information:
- Serial number (on the back side)
- Description of the problem
- Your contact information
Warranty
This instrument is guaranteed to be free from defects in materials and
workmanship under normal use nd service, for a period of 24 months from the
date of invoice. The warranty is extended only to the original purchaser. It
shall not apply to any product or parts which have been damaged on account of
an improper installation, improper connections, misuse, accident or abnormal
conditions of operation. For claim under the warranty, please contact your
supplier.
Safety instructions
| Place the Bottle top dispenser EASY 20K on a flat, horizontal, and stable
table, leaving at least 30 cm of free space on each side. Make sure that the
surface is adequate for the device’s weight.
---|---
| This item must be used only by previously qualified staff that is familiar
with the equipment and knows how to operate it according to the instructions
manual.
| Carefully read the instructions in this manual before using the device.
| Use the necessary protective clothes, eye protection and gloves according to
the reagent’s manufacturer’s information.
| Pay special attention to splashes and protect yourself adequately
- Do not operate this instrument in any manner not described in this ‘User manual’.
- In case of abnormal circumstances during the manipulation of this device, please stop the operation of the equipment and contact your supplier.
- The precautions described in this ‘User manual’ are carefully developed in an attempt to cover all the possible risks. However, it is also important that you are alert for unexpected incidents.
- Use this instrument only for dispensing liquids, with strict regard to the defined operating exclusions and limitations. (chart on p.11)
- Explosive liquids are also forbidden to use with this dispenser.
- Before use, check the instrument for visual damages (piston difficult to move, sticking valves or leakage…) in case of trouble, immediately stop dispensing.
Use Instructions
This equipment is designed for dispensing liquids taking into account the following limits:
- 15 to 40 °C of temperature working range
- Steam pressure up to 500 mbar
- Density up to 2.2 g/cm3
2.1 Operating limitations
Concentrated hydrochloric acid and concentrated nitric acid, chlorinated
and fluorinated hydrocarbons, or liquids which form deposits may make the
piston difficult to move or may cause jamming.
When dispensing flammable media, make provisions to avoid static charging, e.
g., do not dispense into plastic vessels; do not wipe instruments with a dry
cloth.
Operating exclusions : Never use this instrument for:
- liquids attacking FEP, PFA, PTFE or Hastelloy
- liquids attacking borosilicate glass (hydrofluoric acid)
- liquids which are decomposed catalytically by Hastelloy
- suspensions as solids particles could clog or damage the instrument.
Inspection
3.1. Unpacking
Unpack the equipment carefully and check for any apparent damages that may
have arisen from transportation. If necessary, please contact your supplier.
Note:
If there is any apparent damage to the equipment, do not use it. Inform
your supplier.
Note:
If there is any apparent damage to the equipment, do not use it. Inform
your supplier.
3.2. Items list
The package includes the following items:
Items | Qty |
---|---|
Dispenser main body | 1 |
Discharge tube | 1 |
Filling tube | 1 |
Mounting tool | 1 |
Set of PP bottle adapters | 1 |
Instructions manual | 1 |
Please check that all the items are present and inform your supplier immediately if any of the above is missing.
Control
4.1 Bottle top dispenser Components
‘
- Piston seat
- Volume adjustment knob
- Scale pointer
- Cylinder sleeve
- Piston
- Glass cylinder
- Seal
- Discharge valve
- Discharge tube sleeve
- Discharge tube
- Closure cap
- Filling seal
- Filling valve
- Filling tube
- Air vent cap
- Valve block
Assembly Instruction
5.1 Tighten the discharge valve with tools.
5.2 Push the discharge tube as far as possible into the discharge valve. 5.3
Connecting the filling tube
- Cut off the bottom of the filling tube diagonally.
- Push the filling tube into the valve block as far as possible – with the diagonal end down.
5.4 Mounting the instrument on the bottle
- Can be done by screwing directly onto a GL 45 screw thread. For other bottles use the included adapters.
- Hold the discharge tube and carefully remove the cap
- Slide the cap lock backwards on the tube support, away from the discharge tube.
Attention:
- The adapters supplied with the instrument are made of polypropylene (PP), and can only be applied for media which does not attack PP.
- For small bottles use a bottle stand to prevent tipping over
- Reagent may drop off the discharge tube end and cap, be careful.
Dispensing
6.1 Hold an appropriate collecting vessel under the discharge tube opening.
6.2 Gently pull the piston up, then press it down rapidly.
6.3 Repeat this procedure until most of the air bubbles in the glass cylinder
and discharge tube have been removed.
6.4 Before using the instrument for the first time, ensure it is rinsed
carefully or discard first few samples dispensed
Attention:
- The discharge tube must always point away from the user or any other person.
- Never press down the piston when the closure cap is on. The reagent could spill.
Setting the volume
7.1 Place an appropriate collecting vessel under the discharge tube.
7.2 Pull the piston up slowly and evenly to the upper stop to draw reagent
into the glass cylinder.
7.3 Push the piston down gently and evenly to dispense the reagent into the
vessel
Cleaning
The instrument must be cleaned in the following situations to assure correct operation:
- If the piston becomes stiff and difficult to move
- After dispensing liquids that can form deposits
- Before changing the reagent
- Prior to long term storage
- If some liquid has accumulated in the closure cap
- Prior of sterilization
- Before valves replacement
8.1 Ensure that the cap is locked
8.2 Empty and unscrew the instrument from the bottle and lift the instrument
so that the filling tube is out of the liquid but in the bottle.
8.3 Carefully tap the filling tube against the bottle from the inside so that
the reagent runs back into the bottle.
8.4 Remove the closure cap, and dispense the remaining reagent into the bottle
by repeated piston operation.
8.5 Mount the instrument onto another bottle filled with an appropriate
cleaning solution.
8.6 Flush the instrument.
8.7 Empty the bottle, flush it and fill it with water. Repeat step 6 and 7.
8.8 Empty the instrument. Figure 10
8.9 Unscrew the piston seat (1).
8.10 Carefully pull the piston (5) out of the glass cylinder (6) by pulling on
the piston seat only.
8.11 Carefully remove deposits at the edge of the glass cylinder with a screw
driver. Clean piston and cylinder with a bottle-brush.
8.12 Carefully reassemble the instrument in the reverse order. Flush
instrument with distilled water.
Cleaning and replacing the filling valve
9.1 Remove the filling tube
9.2 Unscrew the filling valve using the mounting tool. Remove the valve with
its
sealing washer. Make sure the sealing washer does not remain in the valve
block.
9.3 Rinse the filling valve in the cleaning solution, and clean with a soft
brush.
If the ball gets stuck, push it with a pointed instrument to release the valve
ball.
9.4 Screw the cleaned/new filling valve together with its sealing washer into
the valve block and tighten it firmly using the mounting tool.
Cleaning/replacing the discharge valve
10.1 Unscrew the discharge tube.
10.2 Remove the valve locking ring.
10.3 Unscrew the discharge valve using the mounting tool. Remove the valve
with its sealing valve block.
10.4 Rinse the discharge valve in the cleaning solution, and clean with soft
brush. If the ball gets stuck, push it with a pointed instrument to release
the valve ball (e.g., disposable pipette tip), and release the valve ball.
10.5 Screw the cleaned/new discharge valve together with its sealing washer
into the valve block and tighten it firmly using the mounting tool.
10.6 Slide on the valve locking ring making sure the two prongs properly align
with the notches.
Autoclaving
This instrument withstands steam sterilization at 121 °C, 2 bar absolute (15
psi) for 20 minutes in accordance with widely used guidelines. It is the
user’s responsibility to ensure effective autoclaving.
11.1. Preparation prior to the autoclaving
-
Clean the instrument before sterilization.
-
Remove the discharge tube (10) with removed closure cap (11), filling tube (14), and put all parts on a towel.
Avoid contact with hot metal surfaces. -
Lay the instrument with piston in down position on the towel.
-
Autoclave all parts
Note: Allow to cool down in the closed autoclave to avoid deformation through too rapid cooling. After every autoclaving, inspect all parts for deformities or damage. If necessary, replace them. Do not reassemble the instrument until it has cooled down to room temperature (cooling time approx. 2 hours). Mounting tool can be steam sterilized at 121 °C.
Checking the volume
The accuracy and coefficient of variation of the instrument are determined gravimetrically as follows:
- Set to the nominal volume.
- Dispense distilled H2O.
- Weigh the dispensed quantity on an analytical balance.
- Calculate the dispensed volume taking the temperature into account.
- Perform at least 10 dispensing and weighing operations.
- Calculate the accuracy (A%) and the coefficient of variation (CV%) by means of the formular used in statistical quality control. The proceeding is described e.g., in DIN EN ISO 8655-6. Observe the operating manual of the balance manufacturer and the corresponding standards.
Calculations (for the nominal volume)
Mean value xi= results of weighings
n= number of weighings
Mean volume
Standard deviation
Z= correction factor
Accuracy
V0= nominal volume
Coefficient of variation
Trouble shooting
Problem | Possible cause | Corrective action |
---|---|---|
Piston with difficulty to move | Dirt, possible formation of crystals | -Stop |
dispensing immediately. Loosen the piston by circular motion without
disassembling.
Follow all the cleaning instructions.
Filling not possible| -Volume adjusted to minimum setting
-Sticking filling valve| -Set the required volume
-Clean the filling valve and replace with sealing washer if necessary.
Instrument does not dispense liquid| – Closing cap is on
-Discharge tube not firmly connected or damaged| -Remove closing cap
-Push on the closing cap. Hold the piston in place, and push the original manufacturer’s discharge tube on firmly.
Replace a deformed or damaged discharge tube.
Air bubbles in the instrument| – Reagent with high vapor pressure has been
poured too quickly.
-Seal not inserted, knurled locking nuts not firmly connected
-Air not expelled from the instrument
-Filling tube not firmly connected or damaged
-Valves dirty, not firmly installed or damaged| -Pour the reagent slowly.
-Double check that seal (12) is in place, and that the knurled locking nuts at the discharge tube are firmly seated and finger tightened
-Priming the instrument
-Push the filling tube on firmly.
If necessary cut off approx. 1 cm of tube at the upper end and reconnect it or
replace filling tube.
-Cleaning procedure. Tighten the valves using the mounting tool. If necessary, replace the valves and sealing washers.
Dispensed volume is too low| -Discharge tube not firmly connected or damaged
-Filling tube not firmly connected or damaged
-Valves not firmly installed or damaged
**** -Sticky filling valve| -Push the original manufacturer’s tube on firmly.
-Cleaning procedure. Push the filling tube on firmly. If this does not suffice, cut off approx. 1 cm of tube at the upper end and re-connect it or replace filling tube.
-Clean the instrument. Tighten the valves using the mounting tool, if necessary replace the valves and sealing washers.
– Clean or replace discharge tube, if necessary.