AsthmaTuner AT 5.1 Digital Spirometer
Specifications
- User Manual Rev: 2.8
- Issue Date: 05 November 2021
- Model Number: 0476
- Version: 2.8.0
AsthmaTuner Spirometer
Before you use your AsthmaTuner Spirometer, please read this user manual, the labels and all the information provided with the product.
CAUTION : FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN
Thank you for choosing AsthmaTuner.
Before you use your AsthmaTuner Spirometer, please read this user manual, the
labels and all the information provided with the product.
Before connecting AsthmaTuner Spirometer to a smartphone, install the
application developed to show data measured by the device.
The package includes
- The AsthmaTuner Spirometer device
- The turbine sensor
- The plastic mouthpiece
- 2 AAA batteries
- User Manual
After removing the device from its packaging, check that there is no visible damage. If there is, do not use the device and send it straight back to MediTuner, where appropriate.
Keep the original packaging!
If your product has a problem, use the original packaging to ship it back to
MediTuner.
The manufacturer’s address is as follows:
- MediTuner AB
- Box 3161; 103 63 Stockholm, Sverige
- Webbplats: www.asthmatuner.com E-post: info@asthmatuner.com
Legal manufacturer:
MIR srl – Medical Inernational Research
Via del Maggiolino 125, 00155 Rome, Italy.
MediTuner cannot be held responsible for any damage caused by users failing to follow these instructions and/or the warnings contained in this manual.
INTRODUCTION
Intended use
AsthmaTuner Spirometer is intended to be used by a physician or by a patient
under the instruction of a physician or paramedic. The device is intended to
test lung function and can make spirometry testing in people of all ages,
excluding infants and neonates. AsthmaTuner can be used in any setting.
Usage environment
AsthmaTuner can be used in any setting. AsthmaTuner is used in the factory, in
the hospital, in the doctor’s office.
Restrictions on Use
- Analysis of the test results alone will not be enough to diagnose your clinical condition – you will need a medical examination which will take your clinical history into account as well as any other tests recommended by the doctor.
- Diagnosis and appropriate treatments are to be given only by a qualified doctor.
- The device is intended for use by one person only. If more than one person wishes to use the device, one user’s measurements must not be attributed to another. If another person intends to use the device permanently, the previous user data must be erased from the memory and the new user’s details data (date of birth, origin, weight, height, sex) must be entered.
If you wish to use the device when it has already been used by another person, make sure to disinfect the mouthpiece and turbine, as explained in the Maintenance section.
Description of product
AsthmaTuner Spirometer is a pocket-sized system for measuring the following respiratory parameters:
- PEF (Peak Expiratory Flow)
- FEV1 (Forced Expiratory Volume in 1 sec)
- FVC (Forced Vital Capacity)
- FEF2575 (Average flow between 25% and 75% of the FVC)
- FEV6 (Volume expired in the initial 6 seconds of the test)
The device connects to a smartphone via Bluetooth SMART technology. Connection is automatic once the AsthmaTuner application has been installed on the smartphone. Measurement is performed by a turbine sensor, and is based on the infrared interruption principle. This principle ensures that the measurement is accurate and reproducible.
The advantages of this type of sensor are
- Unaffected by the humidity and density of the gas
- Shockproof and unbreakable
- Inexpensive to replace
The measurements are transferred in real time from the device to the smartphone.
information on the parameters measured by AsthmaTuner Spirometer
- PEF is the maximum speed of the air when you exhale as hard as possible after filling your lungs completely.
- FEV1 is the volume of air expelled during the first second of the same exhalation.
- FVC is the volume of air expelled during the total exhalation.
- FEF2575 is the average flow between 25% and 75% of the total volume of air expelled during the total exhalation (FVC)
- FEV6 is the volume expired in the initial 6 seconds of the test.
- For each of these parameters, the result is a number shown on the smartphone screen.
A high number (associated with a green light) usually means that the air is
moving easily through your lungs. If you have asthma (or another respiratory
disease) and have an obstructive episode, the air cannot generally be expelled
as forcefully as possible, so your parameters will be lower.
AsthmaTuner Spirometer thus helps you find out what sort of obstruction you
have, if any, at a particular time.
By using the device on a regular basis, you can track any changes that may
occur in the parameters. These changes may require appropriate treatment, as
prescribed by your doctor.
It is recommended to use the device twice a day, in the morning on waking and
at bedtime. If possible, the device should also be used as soon the first
signs of respiratory problems occur, so that you can understand how serious
your respiratory problem is and/or how well your current therapy is working.
In addition to displaying the PEF, FEV1 or FVC measurement, the device also
provides a normal baseline value.
the Calculation of the graphic reading is an alternative to using the standard
baseline value, but the best way of finding out your personal baseline value
is to discuss it with your doctor. This value is normally called the personal
best value. Please refer to the section determining your baseline value for a
clear understanding of the baseline value.
Determining your baseline values
-
A measure with a high value usually means that your airflow is good.
-
The best way to determine what are a healthy parameters for you is to discuss this with your physician or other licensed healthcare professional. In fact the importance of any changes in airflow from one measuring to the next depends upon how much they are different from your baseline value you should reach when you are in healthy physical condition.
Your physician or other licensed healthcare professional will use one of two possible ways to identify your baseline value. The first method adopts the predicted value calculated according to the results of epidemiological studies of large groups of healthy subjects of your same age, height, sex, and origin. The second method adopts the personal best value you can reach when you are in the healthiest physical condition. -
The application can calculate the predicted value, i.e. the expected value for healthy people, depending on age, height, sex, and origin. Application calculates the predicted value that has been endorsed by ATS (American Thoracic Society): GLI-2012 All-Age Multi-Ethnic
-
Reference Values by Philip H. Quanjer, Sanja Stanojevic, Janet Stocks, Tim J. Cole. For PEF the predicted values are calculated according to Knudson, R. J., Slatin R. C., Lebowitz, M. D., Burrows, B., The Maximal Expiratory Flow-Volume Curve – Normal Standards, Variability, and Effects of Age, AM REV RESPIR DIS, 1976 113;587-600.
-
In this case, the predicted value becomes the baseline value for your treatment plan. If your physician or other licensed healthcare professional prefers this method, the application provides the calculation of the predicted value.
It is important to know that these predicted values are average numbers for large groups of people. You may have a higher measures than the predicted value and you may not be healthy. Or you may have a lower measures than the average and be healthy.
CAUTION: no matter which method your physician or other licensed healthcare professional prefers to use, it is important that you clearly understand the meaning of your baseline value and how it relates to your treatment plan. if you have trouble determining your baseline value, ask your physician or other licensed healthcare professional for assistance.
OPERATING THE ASTHMATUNER SPIROMETER
Inserting the batteries
Follow the instructions in the Maintenance section for correct battery
insertion.
Installing the AsthmaTuner Spirometer application
Before measuring the PEF, FEV1or FVC you need to install application on your
smartphone.
Connection between AsthmaTuner Spirometer and smartphone
Connection between the AsthmaTuner Spirometer and the smartphone is automatic. To check whether there is a connection, read the messages from the application.
Performing the test
In order to perform the test properly, please follow the instructions below.
-
Push the turbine into the slot until it stops
-
Turn the turbine clockwise until it stops
-
Insert the mouthpiece at least 0.5 cm into the turbine socket.
-
Pick up the AsthmaTuner Spirometer with your hand as if it were a cell phone.
Make sure not to obstruct the turbine with your hand.
-
Insert the mouthpiece in your mouth beyond your teeth, and close your lips tightly over it so that the air you breathe has to pass only through the mouthpiece.
To prevent turbulence that might otherwise affect the results do not put your tongue in the mouthpiece. Do not bend your neck. -
Blow out as hard as you can.
It is best to do the test standing or sitting upright. (makes no difference to test results) -
After exhalation, slowly remove the device from the mouth and check the data on the smartphone.
When AsthmaTuner Spirometer is out of your mouth, avoid sudden movements because this will push air into the turbine and a flow value will be measured that may affect the test results.
Repeat the test three times. AsthmaTuner Spirometer will save the highest value.
Evaluating the test
Tests are performed in a measurement session, after which you select to save
the value and compares it with the baseline value (personal best) set during
configuration
Results diary
The test results are automatically stored on the smartphone and can be
displayed later. Medical studies have shown that if your doctor examines the
test results on a regular basis, lung disease can be managed much better.
Important safety warnings
- Warning: indicates a potentially hazardous situation which, if not prevented, could result in minor or moderate injury to the user or patient or damage the device. Special WARNING should be given to testing elderly subjects, children and differently-able persons
- The manufacturer cannot be held responsible for damage caused by the failure of the user to follow these instructions correctly.
- Only original accessories as specified by the manufacturer must be used with the device.
- Periodically check that no impurities or foreign bodies, such as skin, hairs have deposited inside the turbine. This may cause errors in measurement or compromise the correct functioning of the device.
- Use of an unsuitable mouthpiece could also damage the turbine or harm the patient. In the event of an accident of any kind arising from use of the device, you are strongly recommended to inform your doctor so that he/she can notify the authorities as required by local legislation.
- The device is not designed to be used in direct air currents (e.g. wind), sources of heat or cold, direct sun rays or other sources of light or energy, dust, sand or chemical substances.
- Use and store the device in compliance with the environmental conditions specified in the Technical Specifications. If the device is subjected to environmental conditions other than those specified, it may malfunction and/or display incorrect results. The maintenance
- operations set out in the User Manual must be carried out with the utmost care. Failure to follow the instructions may lead to measurement errors or misinterpretation of the measured values.
- Do not modify the device without authorization from the manufacturer.
- All modifications, adjustments, repairs, reconfigurations must be performed by the manufacturer or by authorized personnel.
- In case of problems, do not try to repair the device yourself.
Data security warnings
Your smartphone stores your personal data. Potential threats such as the
following:
- Malware installation
- Physical access to the smartphone
- Interception of communications
- Physical damage to the smartphone
- Theft of the smartphone
could have an impact on the integrity or confidentiality of such data, such as
- Accessing data in memory by unauthorized persons
- Loss of data in memory
- Inability to use smartphone for communications
- The integrity check of the data is made automatically and in case of transmission error it will create a corruption of the data and the file will be illegible.
The following actions help reduce the risk of such events
- Do not open or install files from suspicious sources
- Use antivirus software
- Back up your data periodically
- Do not leave your smartphone unattended
- Use a password to access the data
- Verify the correct email address where to send the test results
- When data are transmitted Call the doctor to ask for confirmation of receipt
Warnings for use in electromagnetic environments
Due to the increasing number of electronic devices (computers, cordless
phones, cell phones, etc.) medical devices may be susceptible to
electromagnetic interference from other equipment.
Such electromagnetic interference could cause the medical device to
malfunction and create a potentially unsafe situation.
AsthmaTuner Spirometer complies with EN 60601-1-2:2007 on electromagnetic compatibility (EMC for medical devices) for both immunity and emissions.
For the device to function properly, however, the following precautions must be taken:
- Make sure that the AsthmaTuner Spirometer and the smartphone on which the application is installed are no more than 2 metres apart.
- Do not use AsthmaTuner Spirometer near other devices (computers, cordless phones, cell phones, etc.) that generate strong electromagnetic fields. Keep such equipment at a minimum distance of 7 metres.
Notes on FCC certification
AsthmaTuner Spirometer complies with Part 15 of the FCC Rules. Operation is
subject to the following conditions:
- this device may not cause harmful interference
- this device must accept any interference received, including interference that may cause undesired operation
Any modifications not expressly approved by this company could compromise use of the device by the user.
N.B.: This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference when
the equipment is operated in a residential installation. This equipment
generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful interference
to radio communications.
However there is no guarantee that interference will not occur.
If this device does cause interference to radio or television reception, which
can be determined by turning the device off and on, the user is encouraged to
correct the interference by taking one of the following measures:
- Reorient or relocate the antenna
- Increase the distance between the equipment and receiver
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/television technician for help.
CARE AND CLEANING
AsthmaTuner Spirometer is a device that requires little maintenance. The following operations are to be performed regularly:
- cleaning and disinfection of the turbine
- cleaning and disinfection of the mouthpiece
- cleaning of the device
- replacing batteries
Cleaning and disinfection of the turbine
To avoid irreparable damage to the turbine, do not use any alcoholic or oily
cleaning solutions, and do not immerse in hot water or solutions.
Do not try to sterilize the turbine in boiling water.
Never try to clean the turbine under a direct jet of water or other liquids.
If there are no liquid detergents, the turbine must at least be washed in
clean water.
Correct
operation of the turbine is guaranteed only if it is “clean” and free of
foreign objects that affect its movement. The presence of dust or foreign
bodies (such as hairs, sputum etc.) could slow or block the moving parts of
the turbine and make the result less accurate, or damage the turbine itself.
After each use, check the cleanliness of the turbine.
To clean the turbine, pull it out of the AsthmaTuner Spirometer socket by
turning it counterclockwise and simply pulling it out. To make it easier to
pull out, push the base of the turbine gently with a finger.
Weekly cleaning is recommended. Immerse the turbine flowmeter in warmy soapy
water and agitate the turbine for 2-3 minutes.
The hand dishwashing detergent that was tested includes these ingredients:
Aqua, coco glucoside, myristyl glucoside, lauryl glucoside, sodium chloride,
sodium gluconate, sodium citrate, allyl caproate, ethylene brassylate, meth
yldih hydro asmonate).
Rinse in clean water and shake gently to remove any excess water. Allow to air
dry on a towel. Store in a clean, dry place in your home.
After cleaning, insert the turbine into the socket in the direction indicated
by the screen-printed closed padlock symbol on the AsthmaTuner Spirometer. To
insert the turbine correctly, push it down and turn clockwise until it stops,
to make sure it is fully inserted into the plastic container.
Cleaning and disinfection of the mouthpiece
Make sure to clean the mouthpiece after each use. To clean the mouthpiece,
simply pull it apart from the turbine. Just as for the turbine, Immerse the
mouthpiece in warmy soapy water and agitate the mouthpiece for 2-3 minutes .
Rinse in clean water and shake gently to remove any excess water. Allow to air
dry on a towel. Store in a clean, dry place in your home.
After cleaning, insert the mouthpiece in the turbine, by pressing lightly.
Cleaning of the device
Clean the device once a day . To clean, wipe the device’s surfaces with a soft
damp cloth. Dry with a soft cloth, or allow to air dry. Ensure that all
surfaces are completely dry.
Never put the device into water or other fluids.
Replacing batteries
The device continuously monitors the battery level. A message on the
smartphone display alerts the user when the device battery is low.
-
Remove the battery cover on the back of the AsthmaTuner Spirometer
-
Remove the two batteries and replace them with two new ones, following the polarity as indicated by the symbols in the compartment
-
Reattach the battery cover
Used AsthmaTuner Spirometer batteries should only be disposed of in special containers or preferably returned to the dealer of the device or to a special collection centre. In any case, all applicable local regulations must be complied with.
ERROR MESSAGES & TROUBLESHOOTING
Error messages
If you encounter any problems when using the AsthmaTuner Spirometer, a message
will appear on the smartphone display to warn of the malfunction.
Troubleshooting
If you receive an unusually low reading, it could mean that your AsthmaTuner
Spirometer meter is broken, or it could mean that the reading is accurate and
your asthma is getting worse.
Check to make sure that the meter is not broken. You must follow directions
exactly as instructed to obtain accurate results. If your meter is not broken,
follow the instructions in your action plan for low readings and contact your
physician or other licensed healthcare professional.
If problems occur when using the device, the following points should be
checked.
ACCURACY AND RELIABILITY
This device meets the requirements of the following standard:
ATS/ERS TASK FORCE: Standardization of lung function testing (volume
26/numbers 1-5: 2005)
- Volume max 10 L
- Volume accuracy ± 3% or 0.05 L
- Flow range 960 L/minute
- Flow accuracy ± 5% or 10.2 L/minute
Time zero
At the point of peak expiratory flow (PEF), a tangent is drawn with a slope equal to PEF and its intersection on the abscissa defines the TIME ZERO. The back extrapolated volume is the volume of gas that has already been exhaled at the point of TIME ZERO as defined by back extrapolation. The method to determine the time elapsed by TIME ZERO, t0, is given by equation:
Time zero = tPEF – (VPEF/PEF)
Where
PEF is the the peak expiratory flow; tPEF is the the elapsed time at PEF; VPEF
is the the expired volume at PEF
LABELS & SYMBOLS
The label shows
Rx ONLY
Caution: Federal law restricts this device to sale by or on the order of a
physician
TECHNICAL SPECIFICATIONS
Parameters measured
- FEV1 Expiratory volume in one second of testing L
- PEF Peak expiratory flow L/minute
- FVC Forced Vital Capacity L
- Flow/volume measurement system Bi-directional turbine (rotating weel)
- Measurement principle Infrared interruption
- Dynamic resistance at 12 L/s <0.5 cm H2O/L/s
- Communication interface Bluetooth SMART (4.0 or higher)
- Power supply 2 x 1.5V AAA alkaline batteries
- Size main body 109x49x21 mm
- Weight 60.7 g (including batteries)
- Type of electrical protection Class II
- Electrical protection level BF
- IP protection level IP22
- Applicable Standards Electrical Safety IEC 60601-1
- Electromagnetic Compatibility IEC 60601-1-2
- ATS/ERS Standardization of spirometry 2005
- ISO 26782, ISO 23747
- Conditions of use Device for continuous use
- Storage conditions Temperature: MIN -40°C, MAX +70°C
- Humidity: MIN 10% RH; MAX 95%RH
- Transport conditions Temperature: MIN -40°C, MAX +70°C
- Humidity: MIN 10% RH; MAX 95%RH
Operating conditions Temperature:
MIN +5°C, MAX +40°C
Humidity: MIN 10% RH; MAX 95%RH
BLUETOOTH WIRELESS TECHNOLOGY INFORMATION
Bluetooth Compliance: | Version 4.0 single mode low energy |
---|---|
Operating Frequency: | 2.4 to 2.4835 GHz |
Max Output Power: | TX: -5.99 dBm; 0.25 mW |
Operating Range: | 10 meter radius (line of sight) |
Network Topology: | Star – bus |
Operation: | Server |
Antenna Type: | PCB antenna |
Modulation Technology: | FHSS |
Modulation Type: | GFSK |
Data Rate: | 1 Mbit/second |
Data Latency: | 7 – 40 ms |
Data Integrity: | Adaptive frequency hopping, Lazy Acknowledgement, 24-bitCRC, |
32-bit Message Integrity Check Data
Format:| Sends data packets once per 60 ms. Includes 3 control bytes that
allows the host to detect if packets are missing and the device toretransmit.
Quality of Service:| This device uses Bluetooth Smart technology for wireless
communications, which allows for reliable communications in electrically noisy
environments and transmits packets once per 60 ms.It includes 3 control bytes
that allows the host to detect if packets are missing and the device to re-
transmit. If the connection is lost, the App changes the connected status from
connected to disconnected and become available for connection immediately.
Bluetooth Profiles Supported:| GATT-based profile
Authentication and Encryption:| Supported
Encryption Key Size: 128-bit AES with Counter Mode CBC-MAC and application layer user defined
The Bluetooth® word mark and logo are registered trademarks owned by Bluetooth SIG, Inc.
Radio frequency (rf) communication
This device complies with the United States Federal Communications Commission
(FCC) and international standards for electromagnetic compatibility. The
following information is provided in accordance with Federal Communications
Commission (FCC) regulations.
This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions:
- This device may not cause harmful interference, and
- this device must accept any interference received, including interference that may cause undesirable operation. This device does not interfere with any radio frequency signals transmitted from outside sources. These FCC standards are designed to provide reasonable protection against excessive radio frequency interference and prevent undesirable operation of the device from unwanted electromagnetic interference.
Radio frequency (rf) interference from other wireless devices
Common consumer electronic devices that transmit in the same frequency bands
used by the AsthmaTuner Spirometer may prevent the uploader or mobile device
from receiving data. This equipment has been tested and found to comply with
the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses and
can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the
interference by increasing the separation between the equipment and receiver.
INFORMATION FOR CORRECT USE IN AN ELECTROMAGNETIC ENVIRONMENT
Guidance and manufacturer’s declaration – electromagnetic emissions
The AsthmaTuner Spirometer is intended for use in the electromagnetic environment specified below.
The customer or the user of the AsthmaTuner Spirometer should assure that it is used in such an environment.
Emissions test| Compliance| Electromagnetic environment –
guidance
RF emissions CISPR 11| Group 1| The AsthmaTuner Spirometer uses RF energy
only for its internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
| Class B| The AsthmaTuner Spirometer is suitable for use in all
establishments, including domestic establishments and those directly connected
to the public low voltage power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
| Not applicable
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
| Not applicable
Guidance and manufacturer’s declaration – electromagnetic immunity
The AsthmaTuner Spirometer is intended for use in the electromagnetic environment specified below. The customer
or the user of the AsthmaTuner Spirometer should assure that it is used in such an environment.
Immunity test| IEC 60601
test level
| Compliance level| Electromagnetic environment –
guidance
Electrostatic discharge (ESD) IEC 61000-4-2| ±6 kV contact
±8 kV air
| ±8 kV contact
±15 kV air
| Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
In the event of disruption due to ESD during oximetry test, the device recovers from any
disruption within 30 s. (according to ISO 9919).
Electrical fast transient/burst
IEC 61000-4-4
| ±1 kV for input/output
lines
| Not applicable|
---|---|---|---
Surge
IEC 61000-4-5
| ±1 kV differential mode
±2 kV common mode
| Not Applicable|
Voltage dips, short interruptions and voltage variations on power supply input
lines
IEC 61000-4-11
| <5 % U T
(>95 % dip in U T) for 0,5 cycle
**** 40 % U T
(60 % dip in U T) for 5 cycles
| Not Applicable|
| 70 % U T
(30 % dip in U T) for 25 cycles
|
| <5 % U T
(>95 % dip in U T) for 5 sec
|
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
| 3 A/m| 30 A/m| Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or hospital
environment.
NOTE U T is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The AsthmaTuner Spirometer is intended for use in the electromagnetic environment specified below.
The customer
or the user of the AsthmaTuner Spirometer should assure that it is used in such an environment.
| | | Portable and mobile RF communications equipment should be used no closer to any part of the AsthmaTuner
Spirometer , including cables, than the recommended
| | | separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
---|---|---|---
Conducted RF IEC 61000-4-6| 3 Vrms
150 kHz to
80 MHz
| NA| d not applicable
Radiated RF IEC 61000-4-3| 3 V/m
80 MHz to
2,5 GHz
| 20 V/m| d= 0,175 √P 80 MHz to 800 GHz
****d= 0,35 √P 800 MHz to 2,5 GHz
| | | where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and d is the recommended
separation distance in metres (m).
| | | Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, should be less than the compliance level in each
frequency range.
| | | Interference may occur in the vicinity of equipment marked with the
following symbol:
| | |
- NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
- NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
- ) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the AsthmaTuner Spirometer is used exceeds the applicable RF compliance level above, the AsthmaTuner Spirometer should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the AsthmaTuner Spirometer.
- Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between
portable and mobile RF communications equipment and the AsthmaTuner Spirometer
The AsthmaTuner Spirometer is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the AsthmaTuner Spirometer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the AsthmaTuner Spirometer as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
**** W
| Separation distance according to frequency of transmitter [m]
150 kHz to 80 MHz
****d= not applicable
| 80 MHz to 800 MHz
****d= 0,175 √P
| 800 MHz to 2,5 GHz
****d= 0,35 √P
0,01| not applicable| 0,017| 0,350
0,1| not applicable| 0,055| 0,110
1| not applicable| 0,175| 0,350
10| not applicable| 0,550| 1,100
100| not applicable| 0,750| 3,500
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
- NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
- NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
WARRANTY TERMS
AsthmaTuner Spirometer, together with any accessories provided, is guaranteed for a period of:
- 12 months in the case of professional use (doctor, hospital, etc.)
- 24 months where the product is purchased directly by the end-user.
The warranty period is effective from the date of purchase, which must be proven by an invoice or sales receipt.
The device must be checked at the time of purchase, or upon delivery, and any
claims must be made immediately in writing to MediTuner.
This warranty covers the repair or the replacement (at the discretion of
MediTuner) of the product or of the defective parts without charge for the
parts or for the labour.
All batteries and other consumable parts, including the turbine sensor, are
specifically excluded from the terms of this guarantee.
The product warranty shall not apply, at the discretion of MediTuner, in the following cases:
- Improper installation or operation of the device, or if the installation does not comply with current technical or safety regulations in the country of purchase
- Use of the product for purposes other than those provided or failure to follow instructions
- Repair, adaptation, modification or tampering by personnel not authorised by MediTuner
- Damage caused by lack of or incorrect maintenance
- Damage caused by abnormal physical or electrical stress
- Damage caused by defects of the mains electricity supply or of equipment to which the product has been connected
- Serial number altered, deleted, removed or rendered illegible
- The repair or replacement described in this warranty is provided for goods returned at the customers’ expense to our certified service centres. For details of these centres please contact either your local supplier or MediTuner.
- The customer shall be responsible for all transport, customs and delivery charges regarding the goods.
- Each product, or accessory, sent in for repair must be accompanied by a clear and detailed explanation of the fault. Forwarding to MediTuner requires the written permission of MediTuner himself.
- MediTuner reserves the right to replace the product or make any changes deemed necessary.