EasyPulse POC Portable Oxygen Concentrator
Model: PM4130 Series
RECEIVING / INSPECTION
Remove the Precision Medical, Inc. EasyPulse Portable Oxygen Concentrator (POC) from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider.
INTENDED USE
The Precision Medical EasyPulse POC is intended to provide supplemental oxygen to persons requiring oxygen therapy. The Precision Medical POC can be used in a home, institution, vehicle and for transportable use.
READ ALL INSTRUCTIONS BEFORE USING
This manual is provided for your safety and to prevent damage to the POC.
If you do not understand this manual, DO NOT USE the POC and contact your
Provider.
EXPLANATION OF ABBREVIATIONS
mL/min | Milliliters Per Minute |
---|---|
LED | Light Emitting Diode |
POC | Portable Oxygen Concentrator |
Indicator light Colors and Meaning
Red | Warning – Immediate response by the OPERATOR is required |
---|---|
Amber | Caution – Prompt response by the OPERATOR is required |
Green | Ready for use |
SAFETY INFORMATION – WARNINGS AND CAUTIONS
DANGER
Indicates an imminently hazardous situation which, if not avoided, will
result in death or serious injury.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could
result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may
result in minor or moderate injury.
CAUTION
Used without the safety alert symbol indicates a potentially hazardous
situation which, if not avoided, may result in property damage.
| PM4130 WITH RESPECT TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN
ACCORDANCE WITH ANSI/AAMI ES60601-1 (2005, 3rd ed), CAN/ CSAC22.2 No. 60601-1
(2008) 62NA
---|---
| Caution! U.S. Federal Law restricts this device to sale by or on the order
of a physician.
| This device may contain electrical components that are hazardous to the
environment. DO NOT dispose device into standard trash. Contact your local
waste Management for disposal of Electronic Equipment.
| Authorized by FAA for Use Onboard Aircraft Docket #: FAA-2012-0928
| The patient cannula is intended for single patient use only. Use on more
than one patient may cause cross contamination.
| General Warning Sign
| Follow Instructions for Use
| General Mandatory Action Sign
| NO SMOKING
| DO NOT Disassemble
| NO Oil or Grease
| The Oxygen Concentrator, gas pathways, components and accessories do not
contain any natural rubber latex.
| Protected from touch by fingers and objects greater than 12 millimeters. \
Protected from condensation.
| Class II Equipment
| General Alarm
| Power OFF
| Type BF Equipment
****| Direct Current
| Excessive Heat Warning
DANGER
This product is not intended as a life-sustaining or life-supporting
device.
DO NOT smoke, allow others to smoke or have open fl ames near the POC
when in use.
DO NOT use near any type of fl ame or fl ammable/explosive substances.
DO NOT allow POC to exceed “Operating or Storage Temperature Specifi
cations”.
Excessive heating of the Carry Bag can occur if intake vents are blocked.
Equipment not suitable for use in the presence of fl ammable anesthetic.
TO REDUCE THE RISK OF ELECTRICAL SHOCK:
DO NOT use while bathing.
DO NOT reach for the POC that has fallen into water. Unplug immediately.
DO NOT attempt to repair POC. Any attempt could result in electrical
shock.
POC is designed to be used in dry conditions, the POC is not to be submerged,
operated under water or used while swimming.
WARNING
Use the POC for its “Intended Use” as described in this manual.
This POC contains magnetic, ferrous material that may affect the results of an
MRI.
In the event of an equipment alarm or if you are experiencing any signs of
discomfort, connect to another source of oxygen and contact your Provider and
/ or Health Care Professional immediately.
In certain circumstances oxygen therapy can be hazardous; any user should seek
medical advice prior to using this POC.
Where the prescribing Health care professional has determined that an
interruption in the supply of oxygen, for any reason, may have serious
consequences to the user, an alternative source of oxygen should be available
for immediate use.
Place the POC on a secure flat surface when not transporting to avoid
accidental damage to the POC.
ALWAYS keep at least 6 in. (15.24 cm) away from walls, furniture, and
especially curtains that could prevent adequate airflow to the POC.
DO NOT place the POC in a small closed space such as a closet.
DO NOT use extension cords or electrical adapters.
The POC is designed to operate with a single lumen, adult cannula with a
maximum length of 7 ft (2.1 m).
DO NOT use Components / Accessories not approved by Precision Medical, see
“ACCESSORIES / REPLACEMENT PARTS”.
The POC delivers oxygen in a concentrated form when a breath is taken. An
oxygen enrichment of the surrounding area is unlikely.
TO REDUCE RISK OF BURNS, ELECTROCUTION, FIRE, OR INJURY TO PERSONS:
Keep away from heat or open flame.
DO NOT use an AC or DC Power Adapter with a damaged power cord or plug.
DO NOT use if POC has been dropped or damaged.
DO NOT use oil or grease on the POC.
DO NOT cover the air vents, the POC requires proper ventilation.
DO NOT remove the cover on the POC. Servicing must be performed by a
trained professional.
Use only water-based lotions or salves that are oxygen-compatible before and
during oxygen therapy. Never use petroleum or oil-based lotions or salves to
avoid the risk of fire and burns.
Do not lubricate fittings, connections, tubing, or other accessories of the
oxygen concentrator to avoid the risk of fire and burns.
Use only spare parts recommended by Precision Medical to ensure proper
function and to avoid the risk of fire and burns.
Use of this device at an altitude above or below sea level to 14,000 ft (4267
m) or outside a temperature of 41°F to 104°F (5°C to 40°C) or a relative
humidity above 95% Non-Condensing is expected to adversely affect the flowrate
and the percentage of oxygen and consequently the quality of the therapy.
Geriatric, pediatric or any other patient unable to communicate discomfort can
require additional monitoring and or a distributed alarm system to convey the
information about the discomfort and or the medical urgency to the responsible
care giver to avoid harm.
Smoking during oxygen therapy is dangerous and is likely to result in facial
burns or death. Do not allow smoking within the same room where the oxygen
concentrator or any oxygen carrying accessories are located. If you intend to
smoke, you must always turn the oxygen concentrator off, remove the cannula
and leave the room where the oxygen concentrator is located. If unable to
leave the room, you must wait 10 minutes after you have turned the oxygen
concentrator off before smoking.
Open flames during oxygen therapy are dangerous and is likely to result in
fire or death. Do not allow open flames within 2 meters of the oxygen
concentrator or any oxygen carrying accessories.
To ensure receiving the therapeutic amount of oxygen delivery according to
your medical condition, the Oxygen Concentrator must be used with the specific
combination of parts and accessories that are in line with the specification
of the concentrator manufacturer and that were used while your settings were
determined.
To ensure receiving the therapeutic amount of oxygen delivery according to
your medical condition, the Oxygen Concentrator must be used only after one or
more settings have been individually determined or prescribed for you at your
specific activity levels by your healthcare provider.
If you feel discomfort or are experiencing a medical emergency while
undergoing oxygen therapy, seek medical assistance immediately to avoid harm.
There is a risk of fire associated with oxygen enrichment during oxygen
therapy. Do not use the oxygen concentrator or accessories near sparks or open
flames.
CAUTION
The POC is designed to be used indoors or outdoors, in the outdoors
environment the POC should be used in a “dry location” with Battery only.
The proper placement and positioning of the prongs of the nasal cannula in the
nose is critical to the amount of oxygen delivered to the respiratory system
of the patient.
DO NOT place any liquids on or near the POC. If any liquid gets on the
POC, immediately turn “OFF” the power switch and unplug from the electrical
outlet.
DO NOT use while sleeping.
GENERAL SPECIFICATIONS
When measuring any published tolerance, be sure to consider the tolerance
of accuracy of the measuring equipment.
Gas volume and flowrate specifications for gas delivered to the patient are
expressed at ATPD (ambient temperature and pressure, dry).
Precision Medical Inc, and your equipment provider are accountable for
ensuring the compatibility of the oxygen concentrator and all of the parts or
accessories used to connect to the patient before use.
The Oxygen Concentrator, its parts or accessories do not contain known
phthalates which are classified as carcinogenic, mutagenic or toxic.
The Oxygen Concentrator, its parts and accessories are intended for use by a
single patient and should be cleaned\disinfected before use on a new patient.
Note: The nasal cannula cannot be cleaned and should be deposed of.
SPECIFICATIONS
Dimensions: | Height: | 8.5 in (21.6 cm) |
---|---|---|
Width: | 6.5 in (16.5 cm) | |
Depth: | 4.5 in (11.4 cm) | |
Weight: | 4.9 lbs (2.22 kg) | |
Shipping Weight: | 10.8 lbs (4.91 kg) | |
Altitude: | Sea level to 14,000 ft (4267 m) | |
Operating Temperature: | 41˚F to 104˚F (5˚C to 40˚C) | |
Storage / Transport Temperature: | -4˚F to 140˚F (-20˚C to 60˚C) | |
Operating / Storage / Transport Relative Humidity: | Up to 95% Non-Condensing | |
Start Up Time: | 87% O2 Concentration ≤ 10 min | |
Oxygen Minute Volume*: | Setting 1: | 240 +/-15% mL/min |
Setting 2: | 380 +/-15% mL/min | |
Setting 3: | 520 +/-15% mL/min | |
Oxygen Concentration: | 87% to 95% on all settings | |
Mode of Delivery: | Pulsed Dose only | |
Trigger Sensitivity: | 0.50 cmH2O maximum | |
Trigger Delay: | 0.11 seconds typical | |
Power: | AC Power Adapter: | 100-240 VAC (1.0 A @ 120 VAC) |
DC Power Adapter: | 11-16 VDC (5.0 A @ 12 VDC) |
Battery Duration (Approximate):
Setting 1: 5.5 hrs
Setting 2: 4.0 hrs
Setting 3: 3.0 hrs
Times may vary based on atmospheric conditions and battery age.
Sound (@ Setting 2): 42 dBA
Cannula: Maximum 7 ft (2.1 m) long standard adult single lumen nasal
cannula.
Equipment Classification:
Class II – with respect to protection from electrical shock
Type BF – degree of protection against electrical shock
IP21 – Protected from touch by fingers and objects greater than 12
millimeters. \ Protected from condensation.
Volatile Organic Compound (VOC) and Particulate Requirements:
The oxygen delivered from the EasyPulse POC meets the following requirements
for particulate levels, VOC levels, carbon monoxide levels, carbon dioxide
levels and ozone levels.
EPA PM 2.5 | Particulate Matter |
---|---|
ASTM D5466 | Ozone Levels |
21 CFR 801.415 | Carbon Monoxide Levels |
EPA NAAQS – Carbon Monoxide | Carbon Dioxide Levels |
OSHA Permissible Exposure Limits
Standard Test Method for Determination of Volatile Organic Chemicals in
Atmospheres (Canister Sampling Methodology)
- At Sea level
Specifications are subject to change without prior notice.
COMPONENT DESCRIPTION
CAUTION
Missing or illegible labels must be replaced; contact Precision Medical,
Inc.
OPERATING INSTRUCTIONS
WARNING
Read this User Manual before operating the POC.
The POC shall be used by an adult or supervised by an adult when being used by
a child.
Additional monitoring or attention may be required for patients using this POC
who are unable to hear or see alarms or communicate discomfort.
The Setting Knob does not rotate 360°. Rotating the knob less than the “OFF (
) position” or over “setting 3” will damage the Concentrator. (Dial version
only)
CAUTION
ALWAYS confirm prescribed dose before administering to patient and monitor on
a frequent basis.
For Proper Oxygen Flow, be sure that cannula is not kinked or obstructed
before or during use.
DO NOT use humidifier bottles with the POC.
DO NOT use pediatric, low flow nasal cannulas or oxygen masks with the POC.
CAUTION
Inspect the POC for visual damage before use, DO NOT USE if damaged.
To prevent damage to the POC, DO NOT operate the POC without the inlet filter
or while the inlet filter is wet.
Using the EasyPulse POC
-
Verify battery status (level) prior to each use, charge as necessary.
-
Connect the end of a standard adult single lumen oxygen nasal cannula with a maximum length of 7 feet (2.1 m) to the POC’s oxygen outlet fitting.
To ensure proper device operation, use only an adult single lumen cannula maximum 7 ft (2.1 m). -
Place the cannula over your ears and position the prongs in your nose as instructed by your Equipment Provider or cannula manufacturer.
-
Prior to operation, ensure the inlet and exhaust vents on the POC are clear.
Any blockage of the vents can inhibit performance. -
Choose a power option: AC Power Adapter, DC Power Adapter or Internal Battery (Refer to the “POWER SUPPLY” section of the user manual for details on these 3 options).
-
Turn ON the POC by selecting the setting for your prescribed flow. The setting number will be illuminated green when properly selected.
-
Breathe normally through your nose. A measured pulse of oxygen will be delivered each time the POC senses an inhalation. To show the POC is triggering properly the illuminated setting number will turn off briefly each time it senses an inhalation.
-
To turn OFF the POC, select the “OFF” ( ) position.
WARNING
ISO 80601-2-69 (Standard for Oxygen Concentrators) highly recommends that the
patient cannula that delivers gas to the patient from the oxygen concentrator
should include a Fire Stop Check Valve to stop the flow of gas towards the
patient in the case that the cannula becomes ignited. The Fire Stop Check
Valve should be located as close to the patient as reasonably practicable.
Fire Stop Check Valve – Precision Medical Part # 507706
2” Tubing Connector – Precision Medical Part # 507707
Operating Position
The POC can be operated in an upright position with or without the carry bag
as shown below.
The POC can be operated lying on its back ONLY while in the carry bag as shown below.
Using the Carry Bag
The carry bag is intended to make the POC easier to transport, prevent
movement due to vibration and protect the device from minor mishaps.
CAUTION
The carry bag is not intended to make the POC waterproof, or to protect the
device from major falls or accidents/abuse.
When inserting the POC into the carry bag make sure of the following:
- Before placing POC inside bag make sure bag is clean and empty.
- Make sure the control panel is aligned with the clear window on the bag.
- Make sure bottom exhaust vents are aligned with the mesh opening along the bottom of the bag.
WARNING
| The mesh on the carry bag must be aligned with the exhaust vents on the POC
to keep the POC cool and operating properly.
Be sure the exhaust vents are facing away from body while being carried.
If the outside of the bag feels hot or causes any discomfort, set the POC down
and check to be sure all vents are unobstructed.
DO NOT cover Air Vents, the POC requires proper ventilation.
The carry handle can be placed in either of the positions below with the flap secured.
POWER SUPPLY
To prevent damage to the device, use only Precision Medical approved external power sources.
AC Power Adapter:
The AC power adapter connects the POC to a 100-240 VAC, 50/60 Hz wall outlet.
The power adapter converts AC voltage to DC voltage in order to power the POC.
The AC adapter will power the POC and simultaneously charge the internal
battery.
Connection: To connect the AC Power Adapter to the POC, insert the DC Plug Connector into the Power Input jack on the POC and insert the AC Wall Plug into a wall socket.
DC Power Adapter:
The DC power adapter connects the POC to an automobile’s 12 VDC outlet. The DC
adapter will power the POC and simultaneously charge the internal battery.
Connection: To connect the DC Power Adapter to the POC, insert the DC plug connector into the power input jack on the POC and insert the other end into your automobile’s 12 V Accessory Jack.
Internal Battery Pack:
The POC contains an internal rechargeable lithium ion battery pack. The
Battery Level Indicator in the upper right corner of the control panel
continuously displays the charge level of the internal battery while the POC
is on. When the battery charge level drops to 15%, the last LED will turn
solid red and a brief audible beep will be emitted once every minute. When the
battery charge level drops to 5% the last LED starts to flash red and a brief
audible beep is emitted once every 20 seconds to signal that the battery is
almost depleted and you will need to connect the POC to another power source.
ENSURE BATTERY IS FULLY CHARGED PRIOR TO FIRST USE.
Charging: The internal battery can be charged by connecting either the AC
Power Adapter or DC Power Adapter to the POC. The internal battery will begin
charging as soon as power is connected. You will also be able to run the POC
while the internal battery charges.
Lithium Ion Battery Guidelines
Proper Use: | Storage: |
---|
In order to prolong life of the battery, it is best to charge the battery
frequently
rather than waiting until the battery is fully discharged.| When storing for
long periods of time without use (> 1 month) be sure to charge the battery to
around 50% charge level to prolong life of battery.
ACCESSORIES / REPLACEMENT PARTS
Only the items listed below are to be used with the POC.
Description | Part # |
---|---|
AC Power Cord | 506961 |
AC Power Supply | 506629 |
Carry Bag (PM4130) | 507947 |
Cannula | 504833 |
DC Power Supply | 506750 |
External Battery | 506760 |
Inlet Filter | 505847 |
Use of accessories or replacement parts not listed in this instruction may
cause adverse effects to basic safety or essential performance of the Oxygen
Concentrator.
Other Power Cord / Plug combination for the AC Power Adapter available through
Precision Medical, Inc. ONLY.
ALARM / INDICATORS
PM4130 Series
Green | Ready for use |
---|---|
Amber | Caution – Prompt response by the OPERATOR is required |
Red | Warning – Immediate response by the OPERATOR is required |
The Audible Alarm and LED Indicators are listed in the chart.
Visual Indicator/ Problem| Audible Alarm| Probable
Cause| Remedy
---|---|---|---
The last LED on the internal Battery Level Indicator turns red|
Repetitive beep| Internal battery has dropped below 15% charge remaining|
Connect the POC to the AC Power Adapter, DC Power Adapter, or external battery
pack
The last LED on the internal Battery Level Indicator flashes red|
Repetitive beep| Internal battery has dropped below 5% charge remaining| 1.
Connect the POC to the AC Power Adapter, DC Power Adapter, or external battery
pack
2. Change to alternate oxygen source if power source is not available
Amber alarm indicator is illuminated and setting number is not
flashing during inhalation to signal a detected breath| Repetitive
beep| No breath detected for 1 minute| 1. Verify the cannula is not kinked
2. Verify the cannula is properly connected to oxygen outlet
3. Verify the cannula is properly positioned in the nose
Red alarm indicator is illuminated and setting number is not flashing
during inhalation to signal a detected breath| Repetitive beep| No
breath detected for 2 minutes| 1. Verify the cannula is not kinked
2. Verify the cannula is properly connected to oxygen outlet
3. Verify the cannula is properly positioned in the nose
4. If alarm persists change to alternate oxygen source and contact your
Equipment Provider
Amber alarm indicator is illuminated| Repetitive beep| Breath rate
exceeding 35 breaths / minute| Reduce level of activity to allow breath rate
to slow
Red alarm indicator is illuminated| Repetitive beep| Breath rate
exceeding 35 breaths / minute for longer than 1 minute| 1. Reduce level of
activity to allow breath rate to slow
2. If alarm persists, change to alternate oxygen source and contact your
Equipment Provider
Red alarm indicator is illuminated, unit is not running and setting
numbers 1 & 2 are flashing| Repetitive beep| Abnormal compressor operation
detected| 1. Select the “OFF” position
2. Select the prescribed setting
3. If alarm persists, change to alternate oxygen source and contact your
Equipment Provider
Red Alarm and 2 & 3 Setting Numbers **Illuminated and device Shuts
Down| Single Beep| Cannula Occluded/ Kinked| 1. Ensure the Cannula and/or
Outlet are Not Blocked
2. Select the “OFF” position and wait for at least 15 seconds.
3. Select the prescribed Setting
4. If Alarm persists, change to alternate Oxygen source and Contact your
Equipment Provider.
All other Alarm Conditions**| Unknown
1. Switch to another source of Oxygen.| Unknown
1. Switch to another source of Oxygen.| 1. Switch to another source of
Oxygen.
2. Contact Equipment Provider.
MAINTENANCE / CLEANING
WARNING
Prior to cleaning, ensure the POC is turned off and unplug any external power
sources.
DO NOT disassemble or attempt to repair. There are no user serviceable parts
inside. Contact your Equipment Provider for service.
DO NOT spray or apply any cleaners directly onto the case.
DO NOT allow any liquids to drip inside any case openings.
DO NOT use harsh and/or flammable chemicals to clean the device.
Perform cleaning as required.
Cabinet:
- The patient should connect to an alternate Oxygen source.
- Select the “OFF” ( ) position.
- Unplug the external power source before cleaning.
- Clean exterior surfaces of the POC with a cloth dampened with disinfectant solution. Example: Sporicidin® Brand Disinfectant Solution.
- Allow POC to air dry.
- DO NOT place inside carry bag or run the POC until it is thoroughly dry.
- When not in use, store the POC in a clean area free from grease, oil and other sources of contamination.
Inlet Filter:
- Remove the inlet filter by pulling on the protruding tab.
- Wash filter with mild detergent. Rinse thoroughly with water and allow to dry completely.
- Once filter is dry, push the filter back into the case so only the small tab is protruding.
RETURNS
Returned products require a Returned Goods Authorization (RGA) number, contact
Precision Medical, Inc. All returns must be packaged in sealed containers to
prevent damage. Precision Medical, Inc. will not be responsible for goods
damaged in transit. Refer to Precision Medical, Inc. Return Policy available
on the Internet, www.precisionmedical.com.
Manuals available on our Website:
www.precisionmedical.com
DISPOSAL INSTRUCTIONS
This device may contain electrical components that are hazardous to the environment. DO NOT dispose device into standard trash. Contact your local waste Management for disposal of Electronic Equipment.
LIMITED WARRANTY AND LIMITATION OF LIABILITY
Precision Medical, Inc. warrants that the Portable Oxygen Concentrator POC, (the Product), will be free of defects in workmanship and/or material for the following period:
Portable Oxygen Concentrator | three (3) years from date of shipment |
---|---|
Battery | one (1) year from date of shipment |
Precision Medical, Inc. is NOT responsible for normal wear and tear, or any
neglect or abuse of the product.
The customer is responsible for the shipping costs of repairs back to
Precision Medical, Inc.
Precision Medical, Inc. will have in its sole and absolute discretion, the
final determination if your product is covered under this limited warranty.
Should any failure to conform to this warranty appear within the applicable period, Precision Medical, Inc. shall, upon written notification thereof and substantiation that the goods have been stored, installed, maintained and operated in accordance with Precision Medical, Inc.’s instructions and standard industry practice, and that no modifications, substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.
The representatives of Precision Medical, Inc. or any retailers are not
authorized to make oral warranties about the merchandise described in this
contract, and any such statements shall not be relied upon and are not part of
the contract for sale. Thus, this writing is a final, complete and exclusive
statement of the terms of that contract.
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.
Precision Medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property. Correction of non- conformities as provided above shall constitute fulfillment of all liabilities of Precision Medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision Medical, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or specifications without notice.
Precision Medical, Inc. reserves the right to correct clerical or typographical errors without penalty.
**300 Held Drive
Northampton PA 18067, USA
www.precisionmedical.com
T: (+001) 610-262-6090 • F: (+001) 610-262-6080
ISO 13485 Certified
** For technical specifications call Precision Medical or visit
wwwprecisionmedical.com
506640 Rev18 10/19 (10M) Printed in USA