6020 AlcoQuant Plus Alcohol Tester
General Safety and Service Instructions
Proper use of the device requires thorough knowledge of these instructions. The device is intended only for the described applications.
- Do not operate in explosive hazard areas!
- Risk of fire – Do not use in connection with easily ignitable gases (e.g., anesthetic gases).
- Wearable or mobile RF communication devices can interfere with the function of electromedical devices.
- During use on patients, connecting a PC and the device via a data cable is not permitted.
Specifications
- Manufacturer: EnviteC-Wismar GmbH a Honeywell Company
- Model: AlcoQuant 6020
- Country of Origin: Germany
- Product Number: PN Produktnummer
- Serial Number: Seriennummer
- Power Source: Batteries
- Data Interface: RS 232
Product Usage Instructions
General Safety and Service Notes
The product is equipped with safety features to ensure proper usage and
maintenance. It is important to read and follow the instructions provided in
the user manual.
Purpose and Applications
The AlcoQuant 6020 is designed for measuring alcohol content in breath. It
can be used in various settings such as law enforcement, workplace testing,
and personal use.
Operation
To operate the AlcoQuant 6020, follow these steps:
- Press the power button (EIN / AUS-Taste) to turn on the device.
- Use the selection button (Auswahl-Taste) to navigate through the menu options.
- The display (Display) will show the measurement results and other relevant information.
- The red LED (LED rot) indicates any errors or faults.
- Press the OK button (OK-Taste) to confirm selections or perform specific actions.
- The data interface RS 232 (Datenschnittstelle RS 232) allows for data transfer to external devices.
- The battery compartment (Batteriefachdeckel) houses the batteries required for operation.
Measurement Process
Before starting a measurement, ensure the following:
- Avoid tobacco smoke as it may interfere with accurate results.
- Ensure there is no residual alcohol in the mouth.
- Avoid hyperventilation before blowing into the device.
To start an active measurement:
- Follow the on-screen instructions to blow into the device.
- An acoustic and visual signal will indicate correct blowing technique.
- The device will analyze the breath sample and display the measurement result.
- The measurement value can be saved in the data storage.
Calibration and Cleaning
Regular calibration and cleaning are important for accurate and reliable
measurements:
- Refer to the user manual for detailed instructions on calibration procedures.
- Clean the device according to the provided guidelines to maintain its performance.
Battery Replacement
If the batteries need to be replaced:
- Open the battery compartment (Batteriefachdeckel).
- Remove the old batteries and dispose of them properly.
- Insert new batteries following the polarity markings.
- Close the battery compartment securely.
Troubleshooting
If any errors or faults occur during operation, refer to the following table for possible solutions:
Error | Cause | Solution |
---|---|---|
Einblasfehler | Blowing error | If the error persists, contact customer service. |
Signalgeber Fehler | Signal transmitter error | If the error persists, contact |
customer service.
Fehler Datum Zeit| Error in date and time| If the error persists, contact
customer service. The incorrect
date/time values will be stored in the memory along with the
current measurement.
Commissioning
Ensure the calibration sticker on the device has not reached or exceeded
the marked date. Calibration and maintenance must only be performed by
authorized personnel from EnviteC.
Maintenance
The device must undergo calibration and maintenance by authorized
professional personnel every six months (with protocol).
Disposal
In accordance with Directive 2002/96/EC (WEEE), the electronic and electrical equipment must be disposed of properly by the manufacturer after dismantling.
Intended Use and Application Areas
The AlcoQuant 6020 is designed for the rapid and precise determination of breath alcohol concentration and is intended for use in the following areas:
- Routine checks by the police
- Law enforcement
- Workplace safety
Device and Function Description
The AlcoQuant 6020 determines the breath alcohol content by analyzing the volume of breath provided by the subject and takes a sample of 1.2 liters (standard setting).
The device includes a specialized alcohol-only sensor with specific response characteristics and selectivity, ensuring that influences from other substances in the exhaled air such as acetone or eucalyptol are negligible due to the electrochemical measurement principle used.
The device contains the latest technologies to prevent incorrect measurements and to ensure the highest measurement accuracy. This includes constant control of functions and components such as device temperature, battery, alcohol sensor, internal memory, and calibration intervals.
Operation
Front View
No. | Description |
---|---|
1 | ON/OFF Button |
2 | Selection Button |
3 | Display |
4 | Red LED – Error Indicator |
5 | Yellow LED – Measurement Running/Attention |
6 | Mouthpiece Nozzle |
7 | Green LED – Measurement Readiness |
Back View
Includes an OK button, RS 232 data interface, outlet opening, and battery compartment access.
Labels and Symbols
Observe the instruction manual, manufacturing date, product number, serial
number, application part type B, and disposal instructions.
Turning On and Off
To turn on, hold the button for approximately 1 second. The device performs
a self-test of internal functions and displays the device version, software
version, and calibration date. After about 3 seconds, the device is ready for
active measurement mode.
To turn off, hold the button for approximately 3 seconds. If no measurement or
button operation occurs within 2.5 minutes, the device turns off
automatically.
Warnings
- Tobacco smoke can damage the measurement system. Wait at least 5 minutes after smoking before measuring.
- Mouth alcohol can falsify the measurement. Ensure no alcohol consumption has occurred at least 15 minutes prior to testing.
- The subject should breathe normally before measurement, not hyperventilate, as this can temporarily change the breath alcohol concentration and distort the measurement.
Menu Operation
After turning on the device, it starts in “active measurement mode”. Various menu options like “last measurement value”, “settings”, and “counter” are available for selection and confirmation.
Measurement in Active Mode
Active mode is used for accurate determination of alcohol concentration in exhaled air, ensuring the sensor only receives deep lung air. This is necessary because there is a direct correlation between the alcohol content of deep lung air and blood alcohol content.
An original EnviteC mouthpiece must be used for each measurement in active mode, ensuring hygiene and accuracy.
Possible Errors and Solutions
If the air stream during the blowing process is interrupted, an error message “Blowing error” will be displayed. The measurement must be repeated.
- For each measurement in active mode, a new mouthpiece must be used.
- The outlet opening must remain free during the blowing process, otherwise the measurement will be falsified.
- If the subject’s behavior or residual mouth alcohol is conspicuous, the device must be checked by authorized service personnel.
- Patients with reduced lung volume may not reach the necessary volume of 1.2 liters. In such cases, a passive measurement mode (without mouthpiece) can be used.
Data Memory
The device has a data memory capable of storing 9,999 measurement sets with
date, time, and test number. The memory retains its information even when
batteries are changed.
Maintenance / Device Care
Calibration
The device must be calibrated and checked every six months, including
checking the proper function of the device and calibration verification.
Cleaning and Disinfection
Cleaning should only be done with a damp cloth and mild cleaning agent. For disinfection, use a cloth moistened with isopropyl alcohol and wipe without excessive pressure according to EN 60601-1. After cleaning/disinfection with alcohol, a waiting time of at least 15 minutes must be observed before the next measurement.
Battery Replacement
A repeating signal and the display message “Change battery” indicate that the batteries need to be replaced. The device will turn off after 10 seconds. Batteries can be replaced by lightly pressing the battery compartment cover to release the locking mechanism and then removing the cover.
Disposal of Device, Alcohol Sensor, Batteries
The device and alcohol sensor must not be disposed of with household waste.
Please send both with the keyword “Disposal” back to EnviteC. Batteries must
be disposed of according to the European battery directive.
Troubleshooting and Rectification
Each error is indicated acoustically by a short double signal and visually
either by the yellow LED (error that does not impede measurement) or by the
red LED (error that impedes measurement) and leads to the corresponding error
message in the display.
- User error
- Blowing error
- Signal generator error
- LED error
- Time and date error
Technical Specifications
Feature | Description |
---|---|
Measurement Modes | Active and passive |
Measurement Range | 0 to 5.5% BAC |
Warranty
EnviteC grants a two-year warranty from the date of purchase on defects in material and workmanship. The warranty does not cover damage caused by improper handling, non-compliance with the instructions, or unauthorized interference.
Ordering Information
The AlcoQuant 6020 breath alcohol tester is available for order, including
specific configurations such as unit, language, etc., upon request.
Accessories such as carrying cases, mouthpieces, and software are also listed
with their respective order numbers.
FAQ
Q: How often should I calibrate the AlcoQuant 6020?
A: It is recommended to follow the calibration schedule provided in the
user manual. Regular calibration ensures accurate measurements.
Q: Can I use the AlcoQuant 6020 for legal purposes?
A: The AlcoQuant 6020 can be used as a screening device, but it may not be
admissible as evidence in legal proceedings. Consult local regulations for
specific requirements.
Q: Can I use the AlcoQuant 6020 for personal breath alcohol testing?
A: Yes, the AlcoQuant 6020 can be used for personal breath alcohol testing.
Follow the instructions provided in the user manual for accurate results.