Apera Instruments PH60 Series Premium pH Testers Instruction Manual
- June 17, 2024
- APERA INSTRUMENTS
Table of Contents
Apera Instruments PH60 Series Premium pH Testers Instruction Manual
INSTALLATION INSTRUCTION
Thank you for purchasing the Apera Instruments PH60 Series Premium pH Tester Kit. Please read this manual carefully before use in order to properly use and maintain the instrument.
For video tutorials, please go to support.aperainst.com
Battery Installation
Battery Installation
Please install batteries according to the following steps. * Note the
direction of batteries:
The Positive Side (“+”) OF EVERY SINGLE Battery MUST FACE UP.
(WRONG INSTALLATION OF BATTERIES WILL CAUSE DAMAGE TO THE TESTER
AND POTENTIAL HAZARDS
- Pull the battery cap up
- Slide the battery cap along to the direction of arrow
- Open the battery cap
- Insert the batteries (ALL POSITIVE SIDES FACING UP) (see graph)
- Close the battery cap
- Slide and lock the battery cap along to the direction of arrow
- Fit the tester’s cap while making sure to push all the way down. The waterproof feature may be compromised if the cap is not fitted correctly.
Keypad Functions
Short press: < 2 seconds
Long press: > 2 seconds
|
- Short press to turn on the tester and long press to turn off the tester.
- When turned off, long press to enter parameter setting.
- In measurement mode, short press to turn on backlight.
---|---
|
- In measurement mode, short press to switch parameter from pH → ORP (Oxidation Reduction Potential), ORP probe sold separately.
- In mode setting, short press to change parameter (Unidirectional).
|
- Long press to enter calibration mode;
- In calibration mode, short press to confirm calibration;3. When measured value is locked, short press to unlock.
Complete Kit
Preparation Before Use
PH60 and PH60F: If it is first-time use or the tester hasn’t been used
for a long time, pour some 3M KCL solution to the Fill line in the probe cap
and soak the probe for about 15-30 minutes. Users can store the probe in the
3M KCL solution in the probe cap when the tester is not in use so as to keep
the sensor’s accuracy. To achieve maximum accuracy, we recommend soaking the
probe in the 3M KCL storage solution overnight (12 hours) to activate the
glass membrane thoroughly.
Storing the PH60 and PH60F dry will NOT cause any permanent damage. It will
only temporarily cause
the probe to lose its sensitivity, which can always be restored by soaking in
the storage solution.
PH60S: NEVER store the spear probe in a dry condition because permanent damage CAN be caused. The spear probe should always be stored in the 3M KCL soaking solution.
The storage solution is 3M KCL (potassium chloride). One bottle of 10mL
storage solution comes with the tester kit. If the soaking solution was
contaminated, please replace with new ones timely.
DO NOT use any other brand’s storage solutions because different chemicals
may be used and potential permanent damage could be caused to the meter.
pH Calibration
Things needed in addition to what’s in the box
A clean cup, distilled water (8-16oz), and tissue papers for rinsing and
drying the probe.
Short press to turn on.
Rinse the probe in distilled water; shake the meter in the air and use tissue paper to dap off excess water.
Pour certain amount (about half volume of the calibration bottles) of pH 7.00
pH and pH 4.00 buffer solutions in the correspondent calibration bottles.
Long press to enter calibration mode (Short press to go back to measurement
mode).
Dip the probe in pH 7.00 buffer solution, stir gently, and allow it to stand
still in the buffer solution until a stable reading is reached – when stable
icon appears and stays on the LCD screen (as shown in Diagram 2);
short press to complete 1-point calibration and the tester will return to
measurement mode. Icon will appear on the bottom left of the LCD screen,
indicating the middle point of calibration has completed.
Rinse the probe in distilled water and dry it with tissue paper. Dip the probe into pH 4.00 buffer solution, follow the steps in 5.4 to 5.5 to complete 2-point calibration. Icons will appear on the bottom left of the LCD screen. If necessary, dip the probe into pH 10.01 buffer solution (sold separately), follow the steps in 5.4~5.5, to complete 3-point calibration. Indication icon will appear on the bottom left of the LCD screen.
Notes
a) Tester can perform 1 to 3 points automatic calibration. *Please note
that pH 7.00 (USA Standard) or pH 6.86 buffer solution (NIST Standard) must be
used to conduct 1st point calibration. Then use other buffer solution to
conduct 2nd or 3rd point calibration. Tester will recognize 5 kinds of pH
buffer solutions. For details, please refer to the following table.
Calibration | USA Series | NIST Series | Indication icon | Recommended |
---|---|---|---|---|
1-point | 7.00 pH | 6.86 pH | Accuracy ≥ 0.1 pH | |
2-point | 7.00 pH,4.00 pH or 1.68 pH | 6.86 pH,4.01 pH or 1.68 pH | Range < |
7.00 pH
7.00 pH,10.01 pH or 12.45 pH| 6.86 pH,9.18 pH or 12.45 pH| | Range <7.00 pH
3-point| 7.00 pH, 4.00 or 1.68pH,10.01 or 12.45 pH| 6.86 pH, 4.01 or 1.68pH,
9.18 pH or 12.45 pH| | Range: 0 to 14.00 pH
b) pH 4.00 and pH 7.00 buffer solutions are included in the test kit, but pH 10.01 is not. Users can purchase it separately if needed. The buffer solutions poured into the calibration bottles are NOT for one-time use. They can be used for about ten to fifteen times as long as they are not contaminated and the bottles are covered when not in use. After that, we recommend replacing the buffer solutions in the calibration bottles with new ones that are in the buffer bottles (50ml ones) to keep the accuracy of the standard buffer solutions.
Do not pour used buffer solutions back into the buffer bottles in case of
contamination.
c) The tester has self-diagnostic functions:
Icons | Self-diagnostic information | Checking and how to fix |
---|---|---|
Wrong pH buffer solution or the range of calibration solution exceeds | ||
standard. | a) Check whether pH buffer solution is correct (1st point |
calibration must be 7.00).b) Check whether the probe is damaged.c) Check if
there is any air bubble in the glass bulb sensor
| Press key when reading is not stable during measurement.| Wait for the
smiley face icon to appear and stay, then press
-
If you find any air bubble in the glass bulb of the pH sensor, simply shake the probe for a few times
to remove it. The existence of an air bubble in the glass bulb will cause instable measurements. -
The 1st point calibration must be 7.00 pH. Perform the 2nd point calibration (4.00 pH) immediately
after the 1st point. Do NOT turn off the meter before you conduct 2nd point calibration. If the meter is
turned off after 1st point calibration, users will need to restart the calibration process with the 7.00
pH first and the 4.00 pH following after. Calibrating directly in pH 4.00 after turning meter off and
back on will cause Er1.
pH Measurement
Short press to turn on the tester. Rinse the
probe in distilled water, shake the meter and dap off excess water with tissue
paper (do not rub or wipe the glass sensor). Dip the probe in sample solution,
stir gently, and allow it to stand until a stable reading is reached. Get
readings after appears and stays on the screen.
Model/Probe | Application |
---|---|
PH60/Bulb probe | Regular water solutions’ pH measurement such as hydroponics, |
aquaculture, pools and spas, water treatment, brewing, etc.
PH60S/Spear probe| Cheese, sushi rice, meat, fruit, bread, soil, solid culture
medium and semi-solid medium measurement; also works well for regular water
solution.
PH60F/Flat probe| Flat surface measurement such as skin, paper, fabric,
leather and so on; micro sample testing; also works well for regular
watersolution.
b) PH60S Spear probe testers are widely used for solids containing water
or semi-solid
medium. When conducting such tests, pay attention to insert probe evenly, and
be careful to
prevent probe from damage. If the medium is too hard (such as meat or fruits),
please bore a
small hole with a knife before inserting the probe.
- For pH measurement of any food (such as sushi, cheese, etc.), it should be a sampling test. That means test samples should no longer be edible c) PH60F Flat probe testers are mostly for flat surface sample test.
- For skin test: skin should be without sweat or dirt, nor be overly cleaned (do not use facewash products before testing) to avoid affecting measurement results, dampen skin with some distilled water, slightly force flat probe onto the skin, get readings after value stabilized.
- For paper, fabric and leather test: add 1~2 drops of distilled water on surface, then perform measurements.
- For micro sample testing, use a container with an inner diameter<=19mm and a flat bottom. The tester can test volume >=0.5ml.
Special Notes:
a) The pH probe must be rinsed thoroughly after each use. Soap water
should be used to clean off any grease or other contaminants.
b) The PH60 series Pocket Tester will NOT give accurate or stable pH
readings when testing distilled or deionized water. This is because distilled
or deionized water do not have enough ions present for the electrode to
function properly. To measure distilled or deionized water’s pH, users need to
use a specialized instrument. Contact us at info@aperainst.com for more
details. When testing purified water like spring water or drinking water, it
will take longer for the readings to get stabilized (typically 3-5 minutes)
because there is very few ions left to be detected by the sensor in those
purified water.
c) Do NOT store probe in purified water because that will cause permanent
damage to the pH probe. Purified water is only recommended for rinsing the
probe. The probe should be stored in 3M KCL pH electrode storage solution (SKU
AI1120) for best accuracy.
d) Storing the probe dry will NOT cause permanent damage to it. It will
only temporarily cause the probe to lose its sensitivity, which can always be
restored by soaking in the 3M KCL storage solution.
Parameter Setting
Setup Menu
Symbol | Contents | Parameter | Factory Default |
---|---|---|---|
P1 | Select pH buffer solution | USA – NIST | USA |
P2 | Low value measurement alarm setting | 0 ~ 14.00pH | 0 |
P3 | High value measurement alarm setting | 0 ~ 14.00pH | 14.00 |
P4 | Select automatic lock | Off – On | Off |
P5 | Select backlight | Off – 1 – On | 1 |
P6 | Select temperature unit | ℃ – ℉ | ℃ |
P7 | Restore to factory default | No – Yes | No |
Parameter setup method
When turned off, long press to enter parameter setting→Short
press
to switch P1-P2-
…P7→Short parameter flashing→Short press to choose parameter→Short press
to confirm→Long press to go back to measurement
mode.
Parameter setting instruction
-
Select standard pH buffer solution (P1):
There are two options of standard buffer solutions: USA series and NIST series. Factory default is USA series, for details see clause 5.7.
b) Heads-Up Function (P2&P3) Examples: -
Alert when measured value ≤ 3.20 pH:
Preset lowest value (P2) = 3.20 pH, highest value (P3) = 14.00 pH, when measured value ≥ 3.20 pH (stable displays on LCD); LCD displays red backlight. -
Alert when measured value ≥ 8.60 pH:
Preset highest value (P3) = 8.60 pH, lowest value (P2) = 0.00 pH, when measured value ≤ 8.60 pH (stable displays on LCD); LCD displays red backlight. -
Alert when measured value ≤ 3.20 pH or ≥ 8.60 pH
Preset lowest value (P2) = 3.20 pH, highest value (P3) = 8.60 pH, when measured value is lower than 3.20 pH or higher than 8.60 pH (stable displays on LCD), LCD displays red backlight.
c) Automatic Lock (P4)
Select “On” to activate auto lock function. When reading is stable for more than 10 seconds, the tester will lock the value automatically, and HOLD icon will display on LCD. Press key to cancel reading hold.
(d) Backlight (P5) “Off”-turn off backlight, “On”-always turn on backlight, “1”- backlight will last for 1 minute.
e) Temperature Unit (P5)
Select C˚ or F˚, the factory default is ˚F.
f) Factory default setting (P7)
Select “Yes” to recover instrument calibration to the theoretical value (pH value in zero potential is 7.00pH, slope is 100%), parameter setting return to initial value. This function can be used when instrument does not work properly in calibration or measurement. Calibrate and measure again after recovering the instrument to factory default status.
ORP Measurement
ORP stands for Oxidation-Reduction Potential. ORP is a measure of the
cleanliness of the water &
its ability to break down contaminants. Refer to Clause 11 to replace ORP
probe (to be purchased separately), press key to enter ORP mode. Rinse the
probe in distilled water and dry it. Dip the probe in sample solution, stir
gently, and allow it to stand still until a stable reading is reached. Get
readings after appears and stays.
Technical Specifications
**** pH | Measuring Range | -2.00 – 16.00 pH |
---|---|---|
Resolution | 0.01pH | |
Accuracy | ±0.01pH ±1 digit | |
Calibration Points | 1 – 3 points | |
Automatic Temperature Compensation (ATC) | 0 – 50˚C (32 – 122˚F) | |
**** ORP (mV) | Measuring Range | ± 1000mV |
Resolution | 1mV | |
Accuracy | ±0.2% F.S | |
**** Temp. | Measuring Range | 0 – 50˚C (32-122˚F) |
Resolution | 0.1˚C | |
Accuracy | ±0.5˚C |
Other Specifications
** LCD| 3-color LCD screen,Blue: Measurement; Green: Calibration; Red:
Heads-Up
---|---
Reading Lock| HOLD
Low-Voltage Warning| ** Flashing, reminder of battery replacement needed
Auto. Power-Off| In 8 minutes without operation
Water Proof Rating| IP67
Power| DC3V, AAA batteries×4
Battery Life| Continuous operation>2000 hours
Dimension/Weight| Tester: 40×40×178mm/133g; Case: 255×210×50mm/700g;
Probe Replacement
Twist off the probe ring, unplug the probe, plug in new probe (pay attention to probe’s position), and twist on the probe ring.
The model numbers of replacement probes that are compatible with the PH60 Series testers are:
- PH60-E (Regular pH glass bulb probe)
- PH60S-E (Spear pH probe for solids/semi-solids pH testing)
- PH60F-E (Flat pH probe for surface pH testing)
- ORP60-E (ORP probe)
Warranty
We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at option of APERA INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS, LLC for a period of two years from the delivery (a six-month limited warranty applies to probes). This warranty does not apply to defects resulting from actions such as misuse (violation of the instructions in this manual or operations in the manner not specified in this manual), improper maintenance, or unauthorized repairs. Warranty period is the time limit to provide free service for the products purchased by customers, not the service life of the tester or probe.
PERA INSTRUMENTS, LLC
Address: 977 Old Henderson Rd, Columbus, Ohio 43220
Tel: 1-614-285-3080
Email: info@aperainst.com
Website: www.aperainst.com
References
- Support : Apera Instruments
- Apera Instruments® Meters and Sensors for pH, Conductivity, TDS, Salinity, ORP, DO, Turbidity, ISE
- Manual-Hub.com – Free PDF manuals!