COLE PARMER 8903-8904-8906 Thermo Hygrometer Instruction Manual

June 17, 2024
Cole-Parmer

**COLE PARMER 8903-8904-8906 Thermos Hygrometer Instruction Manual

**

FUNCTION KEYS

  1. ON key – Power on
  2. OFF Key – Power off
  3. PRG Key (8721 only)
    • Switch to alarm check mode from measuring mode
    • Switch to alarm setting mode from alarm check mode
    • Save the alarm settings and calibrations
    • Switch to temperature and humidity calibration
  4. C/ F Key
    • Switch between display of C and F
  5. RESET Key
    • Reset Min/Max memory
    • Erase alarm settings
    • Skip calibration mode
    • Switch back to normal measuring mode from alarm check mode or calibration mode
  6. HOLD Key
    • Hold display
  7. MN/MX Key
    • Display minimum value of memory
    • Display maximum value of memory
    • Increase/Decrease the temperature  deviation value
    • Sleep/Non-Sleep mode switch
  8. NX/td Key (8721 only)
    • To flash next digit for alarm setting
    • To display the next alarm setting value
    • To flash next digit for calibration
    • Dew Point read out
  9. 9V DC power jack
  10. Probe socket
  11. RS232 output socket

QUICK REFERENCE

MEASURING MODE
This is the basic function to measure o o temperature ( C or F) and relative humidity

CALIBRATION MODE
With our HR33 & HR75 humidity reference, the end users can do the calibration themselves. This feature provides a DO-IT-YOURSELF accuracy maintenance and replacement of the probe.

LOW/HIGH HUMIDITY CALIBRATIONMODE
Two points’ calibration is necessary for maintaining the accuracy of the meter.

ALARM CHECK MODE
An environment monitoring system is implanted in the meter for special uses. A monitoring range can be set up earlier, the user will be warned for any out- of-range reading. This Mode provides you to check the alarm settings to see if they are under your requirement.

ALARM SETTING MODE
Set, adjust and cancel the alarm settings.

RS232 OUTPUT
To link to the computer gets records.

SLEEP MODE OPTION
We have a Non-sleep mode and a 20 minutes sleep mode. You can switch easily between two

MN/MX MODE
Maximum and Minimum monitoring. You can find out the minimum and maximum values of previous recording simply by pressing MN/MX key sequentially. The screen will display Fig. 1 & Fig. 2. Press it again to go back to normal mode or the meter will return to normal mode in 10 seconds.
MN/MX MODE MN/MX
MODE

HOLD MODE
Hold the screen for specific measuring. Simply press HOLD key when you do the measurement. The screen will display Fig. 3. Press it again to go back to normal mode.
HOLD MODE

OPERATION MODES

The 8721/8711 provides 6 operation modes, 3 output port, sleep mode switch o o and C/ F switch.

MEASURING MODE
This is the normal operation mode for measuring relative humidity and temperature in the default unit. When the power is first on, the full LCD display will appear as Fig. 4.
MEASURING MODE
Full display will appear for 1 second, then readings for humidity and temperature, (Big characters is humidity and small characters is temperature) will display.
Fig. 5
MEASURING MODE

LOW HUMIDITY CALIBRATION MODE
(The calibration solution is optional, not included in standard package.)

Two points’ calibration is necessary for the accuracy of the meter.

  1. Power off. Push ON & PRG at the same time and hold till the LCD displays CAL (Fig. 6). As ON & PRG are o o released, the C ( or F, depended on which is the default unit) will display at the rig bottom corner (Fig. 7)
    LOW HUMIDITY CALIBRATION MODE LOW HUMIDITY CALIBRATION
MODE

  2. Gently open the cover of HR33 ref. salt. Insert the probe into the container from the top until you reach the end. ( Warning: Be very careful and slow while inserting the probe or pulling it out. Any rough movement may damage the Humidity Reference Bottle).

  3. Press RESET one time, the 32.X%RH and MIN will appear on LCD (Fig. 8). The low humidity calibration will be in process automatically. This process will be completed in about 40 minutes with SA, MAX and MIN shown by LCD display (Fig. 9). (Warning: Do not touch any keys before the process is completed.)
    LOW HUMIDITY CALIBRATION MODE LOW HUMIDITY CALIBRATION
MODE
    NOTE:
    The 40 minutes is suggested for a stable temperature environment. If the temperature variation is above ±0.5°F in 5 minutes, the calibration time will be extended automatically. It is recommended to isolate or insulate the standard Humidity Reference if the environment temperature is not stable.

  4. After 40 minutes, low humidity calibration is completed. Now you can press OFF to end the process or continue to process high humidity calibration. To do this, insert the probe into HR75 and press the NX/td key, the LCD will display 75.X and MAX (Fig. 10) which indicates the high humidity calibration will soon be in process automatically.
    LOW HUMIDITY CALIBRATION MODE

NOTE :
A convenient look-up table has been built in to aid calibration.

HR 33 MgCl

TEMP 15 C 20 C 25 C 30 C 35 C
RH% 33.30 33.07 32.78 32.44 32.05

If different temperatures, say 25 C, the o RH of MGk is 32.78 and in 30 C, the R is 32.44.

The same reference says 30 C , the RH of Nacl is 75.09 RH% humidity and in o 35 C , the RH is 74.87% .(See the table)

HR75 NaCl

TEMP 15 C 20 C 25 C 30 C 35 C
RH% 75.61 75.47 75.29 75.09 74.87

The relations are recorded in the look-u table
Once the calibration mode in ON the meter will search the right humidity value according to the current temperature to get more accurate calibration

WARNING!
For Relative Humidity calibration , to ensure the longer life of the salt bottle .Keep the bottle in the bag when not in use and seal the bag tightly

a. If you are sure of that temperature is stable and humidity is in equilibrium. You can reduce the time of the above process by pressing the PRG key. A SA will be shown on the screen. Be very careful with this. If the time interval is not long enough for unit to catch the right humidity, whole measurement later will not be valid
b. To skip the on-going calibration process, press RESET. If you are not sure of the on-going process, press OFF.
c. After humidity calibration is done, if the measuring mode displays E2 instead of the RH and temperature readings, this means something is wrong. Please refer to error messages section.

HIGH HUMIDITY CALIBRATION MODE

  1. There are two ways to enter the high humidity calibration mode:
    a. Turn off the meter. Plug probe into HR75, then press ON & PRG at the same time. Hold till CAL is displayed on screen. Release ON & PRG and press RESET twice, the LCD will show 75.X% and MAX (Fig. 10). This indicates that the high humidity calibration is in automatic process.
    b. When the automatic process for low humidity calibration is completed, the LCD displays SA, MAX and MIN (Fig. 9). Insert the probe into HR75, then press NX/td. The LCD will show 75.X% and MAX (Fig. 10). This indicates that the high humidity calibration is in automatic process

  2. Once the meter starts the automatic calibration process, do not touch any key until the meter switches to measuring mode.

  3. To reduce the time for calibration process, press PRG. Be cautious with this, refer to NOTE a. of the Low Humidity Calibration

  4. To skip the on-going calibration process, press RESET. Turn off the meter if you are not sure of the on- going process.

ALARM CHECK MODE

  1. In measuring mode, press PRG key. The LCD will display the indicators LOTEMP, HITEMP, LO%RH, HI%RH plus XX.Xwith LOTEMP flashing first (Fig. 11).
    ALARM CHECK MODE

  2. XX.X is the numeric value previously set for the flashing indicator. If — instead of a reading is displayed, it indicates that there is no alarm setting for this option.

  3. Press NX/td key to check the value of HITEMP(HITEMP is flashing now), and press NX/td again to check LO%RH, and so on.

  4. In monitoring range, LOTEMP indicates the warning point for low temperature. HITEMP indicates the warning point for high temperature. LO%RH indicates the warning point for low humidity. HI%RH indicates the warning point for high humidity.

  5. Press RESET key to switch the mode back to the normal operation mode, i.e. The measuring mode.

ALARM SETTING MODE

For example, if a user needs to be alarmed when the temperature reading goes below 58°F or above 100°F and when the relative humidity reading goes below 56% or above75%, the user should set LOTEMP, HITEMP, LO%RH, and HI%RH as follows: LOTEMP=58°F; HITEMP=100°F LO%RH=56%; HI%RH =75%; Once the meter readings go out of the range, the alarm will start to beep. The LCD will also display which indicator is alarmed. Steps to set up the alarm are as follows

  1. Power on the meter. When the %RH reading and the temperature reading are displayed (Fig. 5). Press PRG to switch from the measuring mode to alarm check mode (Fig.11 w/ LOTEMP flashing).

  2. Press PRG key, it will display XX°F as Fig. 12 with “X” flashing. LOTEMP means we are in low temperature alarm setting screen.
    ALARM SETTING MODE

  3. Press NX/td key to change the digital flash. Press MN/MX and change the number until it displays 58°F as Fig. 13.
    ALARM SETTING MODE

  4. 4. Press PRG key until it displays SA, release it, the display will show as Fig. 14 with LOTEMP flashing.
    ALARM SETTING MODE

  5. Press the NX/td key until HITEMP flash.

  6. Press the PRG key. Repeat step 2, 3. Change the number to 45°C. Press PRG to save the setting. It will show as Fig. 15 with HITEMP flashing.
    ALARM SETTING MODE

  7. Press NX/td until LO%RH flash as Fig. 16.
    ALARM SETTING MODE

  8. Press PRG, it will display as Fig. 17 with ” 5 ” flashing.
    ALARM SETTING MODE

  9. By Pressing NX/td, MN/MX to change the number to 56. Then press PRG key to save. It shows as Fig. 18 with LO%RH flashing.
    ALARM SETTING MODE

  10. Press NX/td one time and it will flash HI%RH.

  11. Do step 8, 9 but change the number to 75. It will show Fig. 19 with HI%RH flashing.
    ALARM SETTING MODE

  12. Press OFF, turn off the meter or press RESET (after hearing the beep, release the key) and goes back to measuring mode.

  13. In measuring mode, if the meter reads 75.1% and 74.9°F, it will continue beeping and display as shown in Fig. 20.LOTEMP, it means temperature is below our setting. HI%RH, it means the humidity is above our setting.
    ALARM SETTING MODE

  14. Cancel setting: when in step 4 (or step 9), if you press RESET and hold it for 2 seconds and hear beeps. This means the setting has been cancelled and it will display as Fig. 21.
    ALARM SETTING MODE

NOTE :
You can also cancel the whole setting by the following: Turn off the meter. Press RESET and ON at the same time the meter will display OAL one second as shown in Fig. 22 with two beeps, then turns to the normal mode. Thus LOTEMP, HITEMP , LO%RH,and HI%RH settings will be cancelled at one operation.
ALARM SETTING MODE

RS232 OUTPUT(Optional)
Plug the earphone jack of the cable VZRS232M into RS232 socket on the meter and connect 9-pin D-sub to the computer’s available COM port. Press ON key to start measurement. The VZRS232M is optional accessory. Contact with the store for more cable & software purchasing information.

SLEEP MODE OPTION

  1. In default, if there is no key pressed in 20 minutes, the meter will go off automatically. A warning continuous beep will sound shortly before 20 minutes, press any key can restart the 20 minutes counts.
  2. You can switch the meter to non- sleep mode by first turn the meter off, then push and hold MN/MX and ON at the same time until null is displayed as shown in Fig. 23. Release the key and you are in Non-Sleep mode.
    SLEEP MODE OPTION

°C/°F SWITCH MODE
You can change the °C/°F display anytime you want by pressing the °C/°F key during operation. You can also change the default display by the following:
For example, if the meter default to °F, while turning off the meter, press PRG & ON at the same time until CAL is shown in one second, then release the keys. °F will display on the right bottom corner of the LCD. See Fig. 24
°C/°F SWITCH MODE

Press MN/MX to change F to C ,then press PRG until it displays SA in one second and XX.X%RH and MIN will appear on the left corner , turn off the meter. Now when you turn on the meter, o it will default to C

ERROR MESSAGES

  1. Display OP with continuous beeps as shown in Fig. 25:
    Probe is not connected in the right position.
    ERROR MESSAGES

  2. Display 0.0%, 99.9% for a long period of time:
    Probe damaged. Please replace a new probe.

  3. Display Err 1, Err 2, Err 3 and Err 4 with a beep as shown in Fig. 26:
    Err 1, Err 3 and Err 4: Circuit error
    Er 2: Improper calibration; Probe damaged
    ERROR MESSAGES

  4. Screen flashing:
    It means battery is weak, the batteries needs to be replaced.

TECHNICAL SPECIFICATION

  1. .Humidity measurement range:
    8711 : 5%-95%
    8721 : 0%-100%
    Temperature measurement range:
    8711 : -10~50 C
    8721 : -20~50 C

  2. Humidity Accuracy:
    +2% from 0~95% at 25°C (8721)
    ±3% from20~90% at 25℃ (8711)
    Temperature Accuracy:
    ±1°C from -20~50°C

  3. RS232 output: Baud rate: 1200bps
    Data bit : 7,
    Stop bit :1,
    Parity : None
    Format : TXXX.XC(F):HXX.X%

  4. Storage Temperature: -20°C to 60°C

  5. Operating Temperature: 0°C to 50°C

  6. Power requirements: Single 9V battery

  7. Battery life: 100 hours typical (alkaline)

  8. Optional accessories:
    a. Calibration salt
    VZ0033AZ or VZ0033AZ1 for 33%
    VZ0075AZ or VZ0075AZ1 for 75%
    b. RS232 cable with software CD.

  9. his package contains:
    a. Meter x1
    b. Hard carry case x 1
    c. Battery 9 volt x 1
    d. Operation instruction x 1

RETURN AUTHORIZATION

Authorization must be obtained from the supplier before returning items for any reason.
When requiring a RA (Return Authorization), please include data regarding the defective reason, the meters are to be returned along with good packing to prevent any damage in shipment and insured against possible damage or loss.

WARRANTY

The meter is warranted to be free from defects in material and workmanship for a period of one years from the date of purchase.

This warranty covers normal operation and does not cover batteries, misuse, abuse, alteration, tampering, neglect, improper maintenance, or damage resulting from leaking batteries. Proof of purchase is required for warranty repairs.

Accuracy, the Zenith of
Measuring / Testing Instruments !
More products available !
Hygrometer / Psychrometer
Thermometer
Anemometer
Sound Level Meter
Air Flow meter
Infrared Thermometer
K type Thermometer
K.J.T. type Thermometer
K.J.T.R.S.E. type Thermometer
pH Meter
Conductivity Meter
T.D.S. Meter
D.O. Meter
Saccharimeter
Manometer
Tacho Meter
Lux / Light Meter
Moisture Meter
Data logger
Temp. / RH transmitter
Wireless Transmitter

More products available !

2006/07,1000

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals