Cole-Parmer HP-200-AR Acid Resistant Hotplate Instruction Manual
- June 13, 2024
- Cole-Parmer
Table of Contents
HP-200 Series
Acid Resistant
Hotplate Instruction Manual
98030(ES) Version 2.4
Introduction
Thank you for purchasing this piece of ColeParmer equipment. To get the best
performance from the equipment and for your own safety, please read these
instructions carefully before use. Before discarding the packaging check that
all parts are present and correct.
This equipment is designed to operate under the following conditions:
- For indoor use only
- Use in a well ventilated area
- Ambient temperature range +5ºC to +40ºC
- Altitude to 2000m
- Relative humidity not exceeding 80%
- Mains supply fluctuation not exceeding 10%
- Over-voltage category II IEC60364-4-443
- Pollution degree 2
- Use with a minimum distance all around of 200mm from walls or other items
If the equipment is not used in the manner described in this manual and with accessories other than those recommended by Cole-Parmer, the protection provided may be impaired.
Electrical Installation
THIS EQUIPMENT MUST BE EARTHED
Before connection please ensure that the line supply corresponds to that shown
on the rating plate.
Model | Supply requirements |
---|---|
HP-200-AR | 230V, 50 or 60Hz,~, single phase |
The unit is supplied with two mains leads fitted with IEC plugs for connection
to the control box. One has a U.K. 3 pin plug and the other has a 2 pin
“Shuko” plug for connection to the mains. Choose the lead appropriate for your
electrical installation and discard the other. Should neither lead be
suitable, take the lead with the U.K. plug and replace the plug with a
suitable alternative.
This involves cutting off the moulded plug, preparing the cable and connecting
to the rewireable plug in accordance with its instructions.
THIS OPERATION SHOULD ONLY BE UNDERTAKEN BY A QUALIFIED ELECTRICIAN
The wires in the mains cable are coloured as follows:
BROWN | – | LIVE |
---|---|---|
BLUE | – | NEUTRAL |
GREEN/YELLOW | – | EARTH |
The control unit is fitted with an IEC socket at the rear for connection of
the mains lead.
The appropriate mains lead should be connected BEFORE connection to the mains
supply.
Should the mains lead need replacement, a cable of 1mm2 of harmonised code
H05W-F connected to an IEC320 plug should be used.
NOTE: Refer to the equipment’s rating plate to ensure that the plug and
fusing are suitable for the voltage and wattage stated.
The UK mains lead is protected by a 10A fuse mounted in the plug top.
Model | Power | Fuse |
---|---|---|
HP-200-AR | 900W | T5A |
Fuse fitted in both live and neutral lines.
IF IN DOUBT CONSULT A QUALIFIED ELECTRICIAN
Safety Advice Before Use
-
Do not use to heat flammable materials.
-
Position the instrument on a firm level surface away from any heat sensitive or flammable material.
-
Never move or carry the unit while it is in use or connected to the mains supply.
Switch off and allow to cool for at least 30 minutes. -
The unit should be carried using both hands with fingers under each side frame.
-
In the case of mains interruption, a fault or failure, the unit WILL continue to operate on removal of the fault.
-
NEVER immerse the hotplate unit in acid or any other liquid. The unit has breather holes in the base which will allow liquid access that will destroy sensitive components.
General Description
The acid resistant hotplate consists of a ceramic tile and heater assembly mounted in a block of high grade PTFE. The hotplate is controlled via a separate control unit which is connected by a 2 metre lead (partly covered with a PTFE sleeve). This allows the hotplate to be located in a fume hood and the controller outside, away from the corrosive environment.
Controls
Front of control unit
Operation
- Position the hotplate unit on a firm level surface away from any heat sensitive or flammable material.
- Plug the hotplate lead into the hotplate inlet socket of the control unit.
- Connect the control unit to the mains supply (the green mains indicator light will illuminate).
- Position the control unit in the area it is to be used (e.g. outside the fume hood).
- Turn the control unit ON using the temperature control knob and select the temperature of the hotplate by using the printed scale around the knob. The heater on light will illuminate.
WARNING: When the hotplate surface is hot (>50ºC) the hot warning light will begin to flash red. This will continue to flash while the plate temperature is above 50ºC even after the hotplate has been switched off (the control unit must remain connected to the mains supply).
Maintenance
WARNING: Before attempting to clean the unit, ensure that the hotplate is
cool and that the control unit is disconnected from the mains electricity
supply.
The control unit should be cleaned using a damp cloth and mild detergent
solution. Do not use harsh or abrasive cleaning agents.
The hotplate body can be cleaned with almost any cleaning agent compatible
with PTFE. A damp cloth will remove most types of contamination. For more
difficult stains a domestic cleaner is recommended.
Cleaning is made easier if spillages are attended to promptly. In any case,
spillages of alkali, phosphoric acid and hydrofluoric acid MUST be removed
immediately as these chemicals can attack and damage the glass ceramic. Ensure
the appropriate safety precautions are observed.
NEVER immerse the hotplate unit in acid or any other liquid. The unit has
breather holes in the base which will allow liquid access that will destroy
sensitive components.
WARNING: A ceramic top which is scratched, chipped, chemically etched or
damaged must not be used.
Servicing
This product does not require any routine servicing.
NOTE: There are no internal userreplaceable parts.
In the event of a product failure any repairs or replacement of parts MUST
only be undertaken by suitably qualified personnel.
For advice, please contact the Service Department, quoting both the model
number and serial number:
Email: cpservice@antylia.com
Tel: +44 (0)1785 810475
Only spare parts supplied or specied by Antylia Scientific or its agents
should be used.Fitting of non-approved parts may affect the performance of the
safety features designed into the instrument. If in any doubt, please contact
the Service Department of Antylia Scientific or the point of sale.
For any other technical enquiries, please contact the Technical Support
Department:
Email: cptechsupport@antylia.com ,
Tel: +44 (0)1785 810433.
Technical Specification
HP-200-AR
Plate material (hotplate) | PTFE |
---|---|
Body material (hotplate) | 300 x 300mm |
Plate dimensions | 200 x 200mm |
Heated area | 320 x 360 x 60mm |
Heater power | 150 x 160 x 65mm |
Maximum plate temperature | Glass Ceramic |
Hotplate dimensions (w x d x h) | 900W |
Control unit dimensions (w x d x h) | 400ºC |
Lead length | 230V, 50-60Hz |
Net weight | 2m |
Electrical supply | 11kg |
NOTES…………….
This product meets the applicable CE Directives and UKCA Legislation for
radio frequency interference and may be expected not to interfere with, or be
affected by, other equipment with similar qualifications. We cannot be sure
that other equipment used in its vicinity will meet these standards and so we
cannot guarantee that interference will not occur in practise. Where there is
a possibility that injury, damage or loss might occur if equipment
malfunctions due to radio frequency interference, or for general advise before
use, contact the manufacturer.
Declaration of Conformity is available to view online at
www.coleparmer.com
EU Representative address
Antylia Scientific GmbH
Futtererstraße 16
97877 Wertheim
Deutschland
Tel: +49 9377 9203-0
Email: sales@coleparmer.de
UK Representative address
Antylia Scientific
9 Orion Court
Ambuscade Road
Colmworth Business Park
St. Neots
PE19 8YX
United Kingdom
Tel: +44 (0) 1480 277339
Email: enquiries@antylia.com
Ordering Information
Order No. | Series | Model | Legacy SKU |
---|---|---|---|
04801-08 | HP-200 | HP-200-AR | CP300 |
Warranty Registration
http://www.coleparmer.com/product-warranty-
registration?utm_source=cp&utm_medium=qrcode&utm_campaign=warranty
Antylia Scientific Ltd
Beacon Road,
Stone,
Staffordshire,
ST15 0SA,
United Kingdom
UK
T: +44 (0) 1480
272279 E:
uk.sales@antylia.com
W: coleparmer.co.uk
References
- Cole-Parmer Canada, an Antylia Scientific company - Lab equipment and supplies for life sciences.
- Cole-Parmer United Kingdom, an Antylia Scientific company - Lab equipment and supplies for life sciences.
- Cole-Parmer US, an Antylia Scientific company - Lab equipment and supplies for life sciences
- Cole-Parmer Germany, an Antylia Scientific company - Lab equipment and supplies for life sciences.
- Cole-Parmer India, an Antylia Scientific company - Lab equipment and supplies for life sciences.
- Cole-Parmer US, an Antylia Scientific company - Lab equipment and supplies for life sciences
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>