ElEsa MPI-R10, MPI-R10-RF Magnetic Measuring System
Safety Instructions
The product has been designed and manufactured in accordance with the current
regulations. The product leaves the factory ready for use and complies with
the safety
standards. To maintain the product in this state, it is necessary that it is
assembled and used properly, in the closest compliance with this instructions
manual and with the following specific safety precautions. Before installing
and using the MPI-R10, read carefully this manual, which is intended as an
indispensable supplement to the existing documentation (catalogues, data
sheets). Morever, all the rules of law must be observed, in regard to accident
prevention and environmental protection.
- The use, without complying with the descriptions / specific parameters, (in combination with systems / machines / processes to be controlled), can lead to a malfunction of the product, causing:
- health hazards,
- environmental hazards,
- damage to the product and to its proper functionality.
The device must not be used:
- in explosion hazard areas;
- in medical/life support areas and equipment.
Do not open the equipment and do not tamper with it! Any tampering might have a negative impact on reliability of the device and might be dangerous. Do not attempt any repair. Return any defective equipment to the manufacturer! Any violation of the integrity of the device as delivered will cause the warranty loss. Changes or modifications, not expressly approved by the party responsible for compliance, could void the user’s authority to operate the equipment.
Setup/Commissioning
In case of any malfunction (even in case of change in operating conditions),
the device must be switched off immediately. Switch off power supply during
any installation work on the equipment. Installation and commissioning are
allowed by trained and authorised staff only. After correct setup and
commissioning, the device is ready for operation.
Maintenance/repair
Switch off the power supply of the equipment before any action. Maintenance
should be performed by trained and authorised staff only.
NOTE:
This equipment has been tested and found to comply with the limits for a Class
A digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference when
the equipment is operating in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy: if not installed and
used in accordance with the instructions manual, it may cause harmful
interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.
Description
Connected to the dedicated sensor FC-MPI, and combined with the magnetic band
M-BAND-10, the MPI-R10 is a complete system for the measurement of linear and
angular displacement. Characterized by extremely easy assembly, it allows
precise alignment and positioning, reducing time and machining procedures to a
minimum.
MPI-R10 main features are:
- Multifunction LCD with 4 function keys.
- Absolute/ incremental mode.
- Programmable offset and targets function.
- Lithium battery powered.
- Protection against accidental polarity inversion.
The sensor cable FC-MPI is made of a metallic enclosure containing the
electronic sensor, a multipolar flexible cable and a connector to be plugged
into the MPI-R10.
The sensor cable is available in different lengths. The magnetic band
M-BAND-10 is made of two separate parts: the magnetic band and the cover
strip. The magnetic band is made of a magnetic tape, a carrier strip and an
adhesive tape.The cover strip is made of a protection strip
and an adhesive tape.
Mechanical and electrical characteristics
Power supply| Lithium battery 1/2AA 3.6 V
Battery life|
3 years (2 years for MPI-R10-RF)
Display| 7-digit LCD of 12 mm height and special characters
Reading scale| -199999; 999999
Number of decimal digits| programmable
Programmable measuring unit| mm, inches, degrees (angle)
Max operating speed (1)| 1 ÷ 5 m/s programmable
Resolution (2)| 0.01 mm – 0.001 in – 0.01°
Precision (3)| ±0.03mm
Repeatability (4)| 0.0002xL mm (L is the measure in mm)
Self-diagnostic| battery check, sensor check, magnetic tape check
Protection class| IP54 or IP67
Operating temperature| 0°C ÷ +50°C
Storage temperature| -20°C ÷ +60°C
Relative humidity| max. 95% at 25°C without condensation
Operating environment| Internal use
Altitude| up to 2000 m
- The reading speed influences the battery life.
- Resolution: the smallest change in length that the system is capable of displaying.
- Precision: the maximum deviation of the value measured by the system from the actual one.
- Repeat accuracy: the degree of closeness among a series of meas-ures of the same sample, when each single measurement is carried out with the conditions unchanged.
Version – MPI-R10-RF
The MPI-R10-RF is compatible with Elesa wireless network that allows magnetic measuring system and indicators to communicate via radio with a PLC. Elesa wireless network is made by the following components:
- One control unit UC-RF
- Max 36 electronic position indicators or magnetic measuring system, such as DD51-E-RF, DD52R-E-RF or MPI-R10-RF.
M-BAND-10 Technical data
Accuracy class| ± 40 µm
Material| magnetic tape: nitrilic rubber
carrier strip: stainless steel
cover strip: stainless steel
acrylic adhesive tape
Width| magnetic band: 10 mm ± 0.20 mm
cover strip: 10 mm ± 0.20 mm
Thickness| magnetic band: 1.43 ± 0.15 mm
cover strip: 0.23 mm
Magnetic pole pitch| 5 mm
Operating and storage temperature| min -40°C max +100°C
Linear thermic expansion factor
| 17 x 10-6 / K-1
The control unit UC-RF is provided with a standard interface for the most common industrial busses to be connected to the PLC and allows the transmission of the information between the PLC and the MPI-R10-RF magnetic measuring system. The UC-RF exchanges information with the MPI-R10-RF via radio frequency and allows the setting of the target position and the control of the current position of each indicator, directly from the PLC.
Installation
Display installation
- Drill the panel according to the template dimensions reported.
- Remove all drilling burrs before fitting the MPI-R10.
- Fit the lower part of the case into the housing.
- Press onto the upper part until the case is completely snapped in.
Sensor installation
Fix the magnetic sensor by using M3 screws (not included in the supply).
During the installation, use a spacer (max 1 mm is suggested) to grant the
parallelism between the sensor and the magnetic band.
The maximum distance between the sensor and the magnetic band, to ensure a
correct reading of the displacement, is 1mm.
Magnetic band installation
To mount the magnetic band follow the instructions below:
- Clean the mounting surface carefully.
- Remove the protective film from the adhesive tape of the magnetic band.
- Stick the magnetic band on the mounting surface.
- Clean the surface of the magnetic band carefully.
- Remove the protective film from the adhesive tape of the cover strip.
- Stick the cover strip on the magnetic band. The cover strip must be put on the magnetic band to protect it from possible mechanical damages.
- In case of absence of a seat for the housing of M-BAND-10, secure the ends of the cover strip to prevent unintentional peeling.
The mounting surface must be flat. Buckles or bumps will lead to measuring
inaccuracies. To guarantee an optimal adhesion of the adhesive tapes, the
mounting surfaces must be perfectly cleaned, dry and smooth. The following
surface roughness is recommended: Ra <= 3,2 N8 (Rz <= 25). To maximize the
adhesion install the strip applying pressure. Gluing should preferably be
carried out at temperatures between 20 °C to 30 °C and in dry atmosphere.
WARNING
Once the installation is completed, the calibration procedure must be carried out as shown in cap. 8.5.2.
Display
- Absolute or relative mode indicator
- Low battery level indicator
- mm, INCH or degree unit of measure
- Target position indication
- RF connection indicator
Key functions
The function of the key changes depending on the mode of the device.
Key Or Key Combination | Operating Mode | Programming Mode |
---|---|---|
Keep pressed for 3 sec to enter the programming mode. | Parameter |
selection/Confirm of parameter change
| Keep the button pressed for 3 sec to set the origin of the
measurement. Programmable with one of the following options (see the __0
voice of the menu – cap. 8.3):
_dtArG : when a target is loaded, the display shows the actual absolute
position. Pressing the key, the target absolute position to reach appears on
the display.
_dtoG0 [DEFAULT]: when a target is loaded the display shows the distance to
reach the target position. Pressing the key, the actual absolute position
appears on the display.
OFF : the key is not assigned to any function in the operating mode.| Digit
increase / Scroll for parameters bottom-top on the menu tree
__
__
__
__
| Select the:
ABS : absolute measuring mode REL : incremental measuring mode It is
possible to choose one of
the following options (see the _0____ voice of the menu – cap. 8.3):
ArCLr [DEFAULT]: switching from ABS to REL the counter is set to zero.
Ar : switching from ABS to REL the counter is not set to zero.
OFF : the key is not assigned to any function in the operating mode.| Digit
decrease / Scroll for
parameters top-bottom on the menu tree
__
__
__
__
| Press the key to select the unit of measure needed. The options available
are: millimeters, inches and degrees. It is possible to choose one of the
following options (see the _0__ voice of the menu – cap. 8.3):
ALL [DEFAULT]: selectable units of measure: mm, inch, degrees nod EG :
selectable units of measure: mm, inch
OFF : the key does not allow the unit of measure conversion| Programming
mode exit / Digit selection
| Programmable with one of the following options (see the
_0____ 0 voice of the menu – cap. 8.3): PrOGOrG [DEFAULT]: show and
set the OriGin parameter PrOGOFS : show and set the OFFS parameters
OFF : the key combination is not assigned to any function in the operating mode.
| NA
Key Or Key Combination| Operating Mode| Programming Mode
---|---|---
| Programmable with one of the following options (see the 00_ voice of
the menu – cap. 8.3): LOAdOrG [DEFAULT]: the key combination sets the
absolute value to the sum of the parameters Origin and Offset.
OFF : the key combination is not assigned to any function in the operating
mode.| NA
| Programmable with one of the following options (see the _____00 voice of
the menu – cap. 8.3): tArGEtS : the keys combination allows to load/program
one of the 32 target positions. See 8.4 OFF [DEFAULT]: the key is not
assigned to any function in the operating mode| NA
|
mm-inch Turn the indicator on hold then SET press the key . After the start- up sequence the indicator is ready to be used (see chap. 4).
| NA
| | When the reset parameter is selected, press the SET key . At this point,
press the ABS-REL button and then press the key mm-inch ; the display will
turn off and the indicator will go into low power mode of the battery (see
chap. 4).
System Switching on/off
Switching on the system
After you have read the section “Safety Instructions”, proceed by switching on
the indicator.
To switch the indicator on:
- hold the key
- press the key .
The display will light up and the indicator is ready to be used.
Switching off the system
To switch the system off :
- enter the programming mode,
- select rESEt (see chap. 8.3)
- press
- press
- press
The display will turn off and the indicator will go into low power mode.
Operating mode
Absolute/incremental measuring mode selection
Press the key to select the absolute or incremental measuring mode. The
measuring mode selected is shown on the display by the symbols:
- ABS: absolute measuring mode.
- REL: incremental measuring mode.
It is possible to change the key function by the menu voice 0__ (see chap.8.3). The available options are:
- ArCLr (default): when relative measurement is selected, the value is always reset to zero.
- Ar: passing from ABS to REL, the relative measurement is not reset to zero.
In this case, the counter is set to zero by pressing
- OFF: the key is disabled and does not allow changing the selected measuring mode.
To program the parameters listed above, see chap. 8.3.
Unit of measure selection
Press the key to select the unit of measure.
The options available are millimeters, inches and degrees. The measuring mode
selected is shown on the display by the symbols: mm for millimeters, inch for
inches and with the ° suffix for degrees. It is possible to change the key
function by the menu voice 0__ .
The available options are:
- ALL (default): units of measure that can be selected: mm, inch, degree.
- nodEG : units of measure that can be selected: mm, inch.
- OFF: the key is disabled and does not allow changing the selected measuring mode.
To program the parameters listed above, see chap. 8.3
Setting the origin
Press the key for three seconds to set the origin of the measurement. Keeping
pressed the ORG key for three seconds, the display will show the question Set
Org. At this point press the key to confirm or another key to cancel.
Confirming the setting of the origin, the display will be reset to zero: this
position of the sensor must be considered as the origin of the following
measurements.
Setting the absolute reference
After having selected the absolute measuring mode and stopped the sensor in
the starting or in the reference position, press the key combination + to
set the absolute value to the sum of the values of the parameters Origin
(absolute value of reference) and the selected OFFS (compensation value). The
value of compensation (OFFSET) allows you to adjust the value shown on the
display in such a way that takes into account, for example, wear or tool
change. The system allows you to store up to 10 values of compensation.
Pressing the key combination + , the screen will display the last compensation
value used (eg OFS 0). It’s possible to choose the desired compensation value
by pressing the key or , and then pressing the key to confirm.
The screen will display the absolute value equal to the sum of the values of
the parameters Origin and OFF Set. To program the offset values, see parameter
OFF Set of cap. 8.3.
It is possible to change the function of the keys combination + choosing one
of the available options in the menu voice __0___0
The available options are:
LOAd_OrG: the key combinations allow to choose an offset compensation and
to set the origin value
OFF: the keys combination + is not associated to any function in the
operating mode.
For programming the parameters listed above see paragraph 8.3
Direct programming of the absolute reference value and compensation
values
The keys combination + can be programmed to allow direct access to the
programming of the OrIGIn or OFF SEt parameters. It is possible to change the
function of the keys combination choosing one of the available options in the
menu voice 0_____0 .
The available options are:
- PrOGOrG: direct programming of the absolute reference value (OrIGIn parameter).
- PrOGOFS: direct programming of the compensation value (OFFSEt parameter).
- OFF: the keys combination is not associated to any function in the operating mode.
Targets
MPI-R10 allows to set up to 32 target positions to store relevant machine
configuration setting.
To program the targets:
- select tArGEtS in the main menu (see cap 8.3).
- select PrOG_TG (see chap. 8.4).
- select the required memory location (PtrG 01 to PtrG 32).
- press the key to select.
- Follow the instructions in cap. 8.1 to set the required value. To load a target:
- select tArGEtS in the main menu (see cap 8.3).
- select LOAd_TG (see chap. 8.4).
- select the required target value (LtrG 01 to LtrG 32) using the keys ABS-REL and
- press the key to select.
- The value of the selected target is shown.
- Press again to confirm or press to go back to the target selection list.
The keys combination + allows direct access to the programming or loading of
targets depending on the value assigned to parameter ___0_0 .
If enabled, the key combination allows to choose between the two following
operations:
- LOAd_tG: choose one of the 32 available target SET positions, then press to confirm.
- PrOG_tG: choose to program one of the 32 available SETtarget positions, then press to start programming.
It is possible to change the function of the keys combination choosing one of
the available options in the menu voice ___0_0 .
The available options are:
- Target: enable the direct load or program targets functions.
- OFF: the keys combination is not linked to any function in the operating mode.
Reaching the target position
When a target is selected by the PLC (version MPI-R10-RF), the device will
suggest the direction of movement of the sensor to reach the target by means
of the symbols . It is possible to set the FLIP_tG parameter (see chap. 8.2)
to adapt the target position indication to the actual sensor configuration. It
is possible to set the tolerance of the target as absolute difference from the
set value by means of the P_tOLL parameter (see chap. 8.2).
The target position indicators will work, depending from the FLIP_tG and
P_tOLL parameters, as in the following table.
| FLIP| FLIP
---|---|---
M < T – Toll| |
T – Toll <= M < T| |
M = T| |
T < M < = T + Toll| |
M > T + Toll| |
- T = set target
- M = measured value
- Toll = tolerance (see P_toll)
If a target is selected, it is possible to cancel it by pressing the keys
combination + and to confirm the StOP_ tG command by pressing the key . To
keep the target mm-inch
selection press the key .
Target display mode
Press the key to show the present or the target position depending on the
settings of the device.
It is possible to change the function of the key and the target mode choosing
one of the available options in the menu voice __0 .
The available options are:
-
d_tArG (default): when a target is loaded, the display shows the actual absolute position and the indication to reach the target as explained before in cap.7.6.1. Pressing ORG
the key , the set target position is shown. -
d_to_Go : when a target is loaded, the display shows the distance to the set target and the indication to reach the target as explained before in cap.7.6.1. If the target is ORG
not reached, the display blinks. Pressing the key the display shows the actual absolute position.
Target tolerance
Set the value of P_toll parameter to define the tollerance allowed for target
(see chap. 8.2 for details).
Angular measurement
MPI-R10 allows to measure angular displacements. To obtain the correct
measurement, it is needed to set the parameter “Radius” with the measure of
the radius of the arc where the magnetic band is placed.
Version – MPI-R10-RF
Programming the network parameter (nEt id) and the channel parameter (nEt
ch)
The system radio network is defined by the following two parameters:
- nEt id: id 00/99
- nEt ch: ch 01/36
These parameters can be configured in the Radio menu of the indicator (see chap. 8.3) and must be set according to the PLC recipe to guarantee a perfect communication between UC-RF and MPI-R10-RF.
Warning
For MPI-R10-RF with firmware release equal to 5.1 or higher, channel 1 is
equivalent to channel 4 of the previous version. Consider it when used in old
system with UC-RF with firmware release lower than 5.1.
Targets
Using MPI-R10-RF, target positions can be sent from the PLC to the indicators
through the control unit. When a target is set, the behaviour is the same as
described in cap 7.6.
Programming mode
Press the key for 3 seconds to enter the programming mode. Depending on the
setting of PASS parameter (see chap. 8.2), the system may require you to enter
a password.
Press the key and to scroll through the list of parameters and select the
required one by pressing Press the key to exit the programming mode. The
programming mode is automatically dropped after 30 seconds of inactivity.
Programming parameters with numeric values
Press the key mm-inch to select the digit to change. Then use and to
respectively decrease or increase the flashing digit. Press the key to confirm
the value and go back to the list of parameters. The numeric values of the
parameters must be inserted taking into account the selected unit of measure.
When a parameter is changed from its stored value, by confirming it, the
display will show the message Changed. When exiting from the programming mode,
the parameter are stored in the internal memory. If a parameter was changed,
the display will show the message Stored.
Device parameters (in alphabetical order)
Parameter | Description | Available options | Default |
---|---|---|---|
Deg corr | Angular scale correction | Programmable value: _0.000001 |
+/-_ 9.999999. 0.00000 cannot be accepted (the coefficient is
automatically set to 1.00000).| 1.000000
Deg res| Resolution of the angular measurement| The parameter allows to define
the resolution of angular measurement. The available options are: 1 ;
0.1 ; 0.01| 0.01
Dir| Measurement direction Set direction of the positive axis| dir– ( )
– -dir ( )
| dir–
( )
FLIPp_tG| Arrow to tar- get indicators direction| or The parameter set the
direction of the arrow indicators when the target is not reached| **
Lin corr| Linear scale correction| Programmable value: 0.000001
+/- 9.999999. 0.00000 cannot be accepted (the coefficient is
automatically set to 1.00000).| 1.000000
Parameter| Description| Available options| Default**
---|---|---|---
Offset| Offset Value| Programmable value
Res = 1:-999999 ÷ 999999
Res = 0.1:-99999.9 ÷ 99999.9
Res = 0.01:-9999.99 ÷ 9999.99
Res =0.001:-999.999 ÷ 999.999
The system allows you to store up to 10 compensation values:
OFS 0 … OFS 9| 0000.00
Origin| Reference value| Programmable value
Res = 1:-999999 ÷ 999999
Res = 0.1:-99999.9 ÷ 99999.9
Res = 0.01:-9999.99 ÷ 9999.99
Res =0.001:-999.999 ÷ 999.999| 0000.00
Pass| Password| ON : the system requires the password 22011 to enter the
programming mode. OFF [DEFAULT]: the system does not require a password to
enter the programming mode.| OFF
P toll| Tolerance of target position| 0.01÷9.99
The parameter value depends on the unit of
measure selected.| 0.10
Radius| Radius of the circumference where the reading sensor moves|
Programmable value: 0.01- 9999.99
The parameter allows to define the radius of the arc where the magnetic band
is placed for angular measurement.| 100.00
Res| Resolution| The parameter allows defining the resolution of the measure.
The available options are:
mm: 1; 0.1; 0.01
inches: 1; 0.1; 0.01;
0.001| mm: 0.01
inches: 0.01
Speed| Reading max speed| Programmable values 1;2;3;4;5
The parameter sets the maximum speed of the movement in m/s that can be
correctly read.| 3
Targets| Target value| Programmable value
Res = 1:-999999 ÷ 999999
Res = 0.1:-99999.9 ÷ 99999.9
Res = 0.01:-9999.99 ÷ 9999.99
Res =0.001:-999.999 ÷999.999 _
_The system allows you to store up to 32 compensation values: LtG01 …
LtG32.
The parameter value depends on the unit of measure and resolution set.| 0
Main menu tree
6 The parameter value depends on the unit of measure and
resolution set.
7 The symbols on the display related to the target feature are used.
8 See chap. 5
Target menu tree
Additional features
Reset
To reset the device to its factory setup:
- select the voice RESEt from the main menu (see chap. 8.3).
- elect YES pressing the key
- press the key to confirm.
Calibration
The Calibration voice in the main menu activates the CALIBRATION MODE and the
display shows GO. At this point, the user must slowly move the sensor in one
direction along the magnetic band. After the GO, a progress bar is displayed
that will grow as long the sensor is moved. The procedure is completed when
the measurement position is shown again by the display. This operation allows
the sensor to be accurately bound to the magnetic tape and has to be done
every time after the installation of the sensor.
Test LCD
The LcdtESt voice in the main menu allows to switch on all the display
segments.
Correction coefficients
To improve the correctness of the measurement, MPI-R10 allows to set two
correction factors that take into account the differences between ideal and
actual installation of the magnetic band:
Lin Corr: is the ratio between the actual measurement and the value
measured by the device in linear measurement.
Ang Corr: is the ratio between the actual measurement and the value
measured by the device in angular measurement.
To calculate the correction factor, set it to 1 then read the value measured
(M) in a reference point (K). The correction factor will be equal to K/M.
Verify that the measurement done in the reference point and/or other known
points is correct.
Revision
The release data of the device are shown, starting with the r letter, as last
voice in the main menu. These data can be scrolled pressing the key .
- Please note these values and communicate them to Elesa in case of support needed.
Battery replacement
The internal lithium 1/2 AA – 3.6 V battery ensures over 3 years battery life (RF version – 2 years). The symbol is shown on the display when the battery replacement is required. The replacement is made by simply removing the cover on the back. By replacing the battery in less than 5 seconds, all the measurements and settings wil not be lost. If more time is required and the display turns off, the settings of the device have to be set or verified again.
Display messages and troubleshooting
Message on the display | Description | Action |
---|---|---|
Sensor | The sensor is not connected | Connect the sensor or verify the cable |
and the connector
no tAPE| The magnetic tape is not detected| Verify if the sensor is correctly
mounted near the magnetic tape
Speed X| The sensor is moving too fast according to the value of the setting
in the Speed parameter. X is the present setting of the Speed parameter.|
Press to go back to the value reading and re-set the absolute reference.
****Flashing battery symbol| Low Battery| Replace the battery (see chap. 9).
EU DECLARATION OF CONFORMITY {DoC)
COMPANY NAME: Elesa S.p.a.
POSTAL ADDRESS: Via Pompei 29
POSTCODE AND CITY: 20900 Monza
TELEPHONE NUMBER: +39 039 28111
E-MAIL ADDRESS: info@elesa.com
Declare that the DoC is issued under our sole responsibility and belongs to
the following product:
PRODUCT: Magnetic measuring system
APPARATUS MODEL: MPI-R10
TRADE MARK: Elesa
The object of the Declaration described above is in conformity with the
relevant Union harmonization legislation:
2014/30/EU (EMC): Electromagnetic Compatibility Directive 2011/65/UE (RoHS):
Restriction of the use of certain Hazardous Substances in electrical and
electronic equipment
The following harmonized standards and technical specifications have been
applied:
EN 61326-1 :2013