GasDog GD-700 Industrial Fixed Detector User Manual
- June 17, 2024
- GasDog
Table of Contents
- Foreword
- Please read carefully before using
- Attentions
- Product Description
- Product features
- Attachment configuration list
- Technical Parameters
- Instrument installation
- Remote controller instruction
- Function Description
- Common faults and solutions
- Terms of Service
- National standards for product development, design and production
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Industrial Fixed Detector
GD-700 User Manual
Please read carefully before using
Foreword
You are welcome to use the instrument of this product, and hope that this
manual will bring you convenience when you use the instrument. If you find any
unclear, wrong or excessively lengthy place in this manual, please contact the
agent or after-sales service department in time.
Before you perform any operation on this instrument, please read this
operation manual carefully, and keep the manual properly so that you can refer
to it in time for help in the future.
It is forbidden to disseminate all the contents of this manual without
permission; this manual only provides relevant information.
The company is committed to continuous improvement of product performance, and
the company reserves the right to improve any content in the manual without
prior notice.
Please refer to the actual product for the color and style of the product and
product.
Please read carefully before using
Anyone who may use, maintain or repair the instrument should read this operating manual carefully and only follow the operating manual to achieve the design level. Otherwise the instrument will not work properly and will cause malfunction and damage to the instrument.
Warning:
- Before using the instrument, please read the product manual carefully.
- It is strictly forbidden to open the cover when the power is on at the scene.
- It is strictly forbidden to replace the sensor with power on.
- The installation, debugging, setting and other operations must be carried out by professionals.
- Gas calibration inspection should be carried out regularly, and the sensors that have exceeded the valid period of use and have faults should be replaced in time.
- It is strictly forbidden to input gas with a gas concentration higher than the measuring range to impact the sensor.
- Prevent the instrument from falling from a high place or subject to severe shocks.
- It is strictly forbidden to expose the instrument to a high-concentration corrosive gas environment to work for a long time to prevent damage to the sensor.
- It is strictly forbidden to use it under high temperature and high humidity. If the humidity is high, it must be equipped with a filter dehumidification device.
- Users are not allowed to repair or replace parts without authorization.
- Man-made damage is not covered by the warranty.
- It is not allowed to change the components or structures that affect the explosion-proof performance at will, so as not to affect the explosion-proof performance.
Attentions
- Electricity safety: Ensure that the instrument shell is properly grounded.
- For safety reasons, this product can only be operated and maintained by professionals, and this manual must be fully read and understood before operation and maintenance.
- Because the instrument is a bit heavy, at least two people are required when installing it, pay attention to it, and carry it carefully. Otherwise, it may cause physical injury.
- When carrying the instrument, please wear gloves, otherwise it may cause injury.
- During the installation process, be careful not to allow wire heads or other debris to enter the left channel port of the instrument, otherwise it may cause fire, malfunction, or malfunction.
Calibration warning:
- The gas detection instrument is a safety and life-saving measuring instrument. In order to ensure the accuracy of measurement, toxic and catalytic combustible gas sensors should be calibrated at least once every six months, and infrared sensors should be calibrated once a year.
- The gas detector needs to be carefully tested or calibrated after an alarm occurs.
Product Description
GD700 Industrial Fixed Detector is a fixed pump suction detection instrument used in complex working environments such as emission gas, oil and gas industry, pipelines,laboratories, etc. Its built-in brushless pump and suction sampling method make the detection fast and accurate. It committed to providing users with a reliable, accurate and safe gas detection program.
Product features
2.4-inch high-definition color screen, wide viewing angle, low power
consumption Built-in micro sampling pump
4-20mA current signal, RS485 digital signal, 1 set of relay signal output
Infrared remote control, convenient and quick
Supports multiple gas units switch
Attachment configuration list
Standard accessories:
- One gas detector
- One infrared remote controller
- One water and dust filter
- One manual
- Certificate of conformity, warranty card
- Packing box
Optional accessories:
- Switching power supply
- Alarm light
- High temperature probe (4) Mounting brackets
- Power adapter(24V DC)(6) Other
Technical Parameters
Product size: 28620296mm (H x W x D)
Product weight: 3.4kg (3.8kg with alarm light)
Sensor: Electrochemical principle, PID principle, infrared principle,
catalytic combustion principle Gas measurement range: To be specified
Resolution: To be specified
Precision: ≤±5%F.S or ≤±10%F.S (depend on sensor type)
Pump flow: 0.6(±0.1)L/Min
Display: 2.4-inch high-definition color display
Display content: gas molecular formula, concentration, unit, pump status, etc.
Alarm: buzzer, alarm status prompt on the display
Power supply: 12~24V DC
Installation method: wall-mounted
Working hours: 24 hours uninterrupted monitoring
Language: Chinese and English
Expansion: PC software
Humidity: 0~95%RH (non-condensing) (optional external filter will be needed if
the humidity is too high)
Temperature: -10℃~ +55℃ (optional sampling probe will be needed if the
temperature is too high) Explosion-proof grade: Ex d IIC T6 Gb
Instrument installation
Product structure
The detector installation avoids the following environmental influences:
- Environment temperature over device working temperature range.
- Environment humidity over device working humidity range.
- Environment pressure over device working pressure range
- High wind speed.
- Mechanical vibration.
- Strong electromagnetic radiation environment.
- High voltage environment.
- Electricity leakage.
Installation size
The upper and lower spacing of the three fixed holes of the instrument and the
size of the device are as shown in the figure below.
The left side is the wiring area, and the right side is the alarm.
- As shown in the figure, the housing grounding mark “ “, the instrument must be properly grounded to avoid personal safety accidents or instrument damage caused by leakage or static electricity.
Installation types
GD700 is a pump suction detector, which can be connected to tube/hose for
detection, and the installation position can be selected flexibly.
size of the hole of the detector (above), fix three 6mm expansion bolts on the
wall, fix them with screws,
Wall-mounted: This installation method is often selected for normal indoor
inspection.
According to the nuts and spring washers, and then use 6x30mm nuts and spring
washers to fix the expansion bolts on the wall, as shown.
Note: The installation location must be well grounded or insulated.
Leakage of electricity may cause the detector to work abnormally.
*Other installation methods such as horizontal pipe, vertical pipe can be adjusted according to site conditions
Sampling tube selection
The sampling tube should be made of materials that have no chemical reaction
with the gas to be detected and no adsorption, and the tube should not exceed
50 meters.
Sampling point selection
The sampling point should be close to the leak source within a distance of 1 m
or the area where workers stay.
To detect toxic gases lighter than the air index, the sampling point should be
installed above the use environment, and the smaller the detection point, the
higher.
To detect harmful gases heavier than the air index, the sampling point should
be installed below the operating environment.
Commonly, gas weight is similar to air, their sampling point installation
height should be 1.5 meters above the ground .
Wire connectioninstructions
In outdoor areas with a lot of dust, rain or water mist, dust-proof, moisture-
proof and rain-proof treatment (such as installing a rain cover) is required,
and the wiring is recommended as shown in the picture below, which can prevent
rain from pouring back and affect the internal circuit board of the instrument.
Warning: The wire connection must be operated by professionals, otherwise
it may cause electric shock or damage the instrument.
*Note: During the wire connection process, it is forbidden to power on.
Cable selection:
- RS485 transmission cable: Please use shielded twisted pair cable, the theoretical maximum transmission distance is about 1200m, and the actual transmission distance is less than the theoretical value. If long-distance transmission is required, signal transmission repeaters can be used, but no more than 8 repeaters on one wire (repeaters need to be purchased separately).
- 4-20mA transmission cable: Please use shielded cable with a core diameter of 0.75mm or more.
- The transmission distance of 4-20mA is determined by the load resistance. The maximum load of this detector is 500Ω (when powered by 24V). The load resistance includes the input resistance of the control system (controller, PLC, DCS, etc.) and the internal resistance of the cable.Under the condition that the input resistance of the control system remains unchanged, the transmission distance is calculated with reference to the following formula:
Reference distance = (500-Rc) ÷ Rm
*Rc-input resistance of control system, Rm-resistivity of transmission cable.
Wiring interface
J1 on left is the relay control interface: COM (common end), NO (normally open
end).
J2 terminal on left, the interfaces from top to bottom are: 24V (DC power-
positive), GND (power ground), A (RS485A-positive end), B (RS485B-negative
end).
JP5 Alarm light interface
JP2 terminal on right is the 4-20mA interface
JP3 terminal on right is the digital probe interface, from top to bottom are:
GND (power ground), followed by sensor channels COM4, COM3, COM2, COM1.
Application example: At the site, the detector detects harmful gas and reaches the alarm value. The detector screen and indicator lights prompt an alarm, and if an optional alarm light is installed, there will be an audible and visual alarm. At this time, the normally open end of the relay closes the circuit, (as shown in the figure below), which can drive the fan for ventilation or drive the alarm system.
Warning: When the instrument alarms, you must leave the workplace as soon
as possible. Otherwise, serious personal injury or personal injury may be
caused.
Please re-calibrate and check the detector afterwards to avoid the expansion
of equipment errors and cause false alarms.
Connection diagram of multiple instruments
4-20mA current signal output mode and RS485 digital signal output mode are
different in wiring; the following example power supply connection mode is for
reference only, and the power supply mode can be selected according to the
actual situation during installation.
A. 4-20mA current signal output connection mode. (must be grounded
together)
A. RS485 digital signal output connection mode 1. (can be independently grounded)
C. RS485 digital signal output connection mode 2. (can be independently grounded)
Remote controller instruction
Note: When using, please aim at the infrared receiving port above the display screen to operate. The instrument is equipped with a remote controller with a total of 11 buttons. as the picture shows.
Key | Function |
---|---|
MENU | Enter the menu |
Mute | Turn off the sound |
PUMP | Power on or off the pump |
BACK | Cancel/Return/Unit Switch |
OK | Confirm |
↑ | Shift Up |
↓ | Shift Down |
← | Shift left |
→ | Shift Up right |
+ | Plus |
- | Minus |
- MENU key: enter the menu (in the detection interface).
- Mute key: Turn off the alarm sound of the instrument (in the detection interface).
- PUMP key: switch the pump on and off (in the detection interface).
- BACK key: cancel; return to the previous menu or detection interface,Double-click the BACK button on the detection interface to switch the unit (only μmol/mol and PPM).
- OK key: confirm; enter the next level menu.
- Up key: Move the cursor up.
- Down key: Move the cursor down.
- Left key: move the cursor to the left; select the position to modify the data.
- Right key: move the cursor to the right; choose to modify the data position.
- Add key: increase the value.
- Minus key: decrease the value.
Function Description
1 Power On/Off
1.1 Power On
- In the power-off state, check whether the power cord on the left side of the instrument is in good condition and has no foreign objects.
- When connect with the power, the instrument power on.
When the instrument power on, it will initially display “Power On” (if it cannot be displayed normally, the instrument may be faulty, please contact the agent or after-sales service department at that time), then enter the self- test, and then enter the instrument preheating countdown.
After the startup is complete, the reading and various function icons will be displayed on the screen.
1.2 Power Off
In the normal measurement state or when there is power, just disconnect the
power.
2 User Interface
2.1 Detection interface description
The detection interface on the instrument displays the sensor type, indication
and other functional status information.
The top are: self-test status (passed “√”, failed “×”), air pump (pump
running-blue dynamic; pump stop-gray), alarm (on-blue; off-gray)
- Self-test status: blue √ self-test is successful and the device is normal; gray X means the device is abnormal in self-test, please contact the manufacturer.
- Air pump status: blue on, gray off.
- Sound status(Optional): blue on, gray off. It’s the optional function. If detector with alarm light, then have this function.
- Gas name: The name of the gas type is in the blue box.
- Gas unit: Small black characters are gas units.
- Channel number: Corresponding to output 4-20mA interface CO1, CO2, CO3, CO4 (page 12).
- Concentration reading: The black characters are the gas concentration readings.
- Address number: detector address
When the following screen appears, please check the device.
Loss of signal: The detector is not connected or has poor contact. After the reconnection is correct, the interface still appears, please contact the manufacturer.
2.2 Main menu interface operation
In the gas detection interface, short press the “MENU” button, the instrument
will enter the main menu interface immediately; press the “BACK” button in the
main menu interface to return to the gas detection interface.
2.3 Calibration
2.3.1 Fresh air/zero calibration
In the fresh air/zero point calibration interface, press the “←” and “→” keys
to select the sensor to be calibrated, and then press the “OK” key to enter
the calibration state, and then enter the 30-second countdown. During the
calibration countdown period, you can press “BACK” Key to cancel zero
calibration..
- In the “fresh air/zero point calibration” interface, you can press the “BACK” key to return to the previous menu.
2.3.2 Span Calibration
The pump flow of YT-98H is about 0.6(±0.1)L/min. The instrument must be
connected to the standard gas cylinder through a T-shaped calibration tube
(tee joint), and the other end is connected to the flow meter. During
ventilation calibration, ensure that the emptying of the bypass flow meter is
above 100mL/min.
In the span calibration interface, press the “OK” key to enter the span
calibration sub-interface.
There is an option in the span calibration sub-interface. At the same time,
the gas type, real-time concentration value, unit and ADC value are displayed
at the top.
The span calibration process is as follows (the following example assumes that
the calibration gas is 550PPM CO):
After the instrument is turned on and enters the span calibration interface,
connect a calibration gas with a known concentration to the the instrument
through a PTFE tube.
According to the actually used calibration gas concentration value (550PPM),
use the “↑” and “↓” keys to move the cursor to select the closest calibration
point.
Press the “OK” key to enter the parameter setting mode, use the “←” and “→”
keys to move the cursor to select the pre-modified number, and the “+” and “-”
keys to modify the selected number, and set the value on the right (set fixed
value) is modified to be the same as the calibration gas used.
Open the valve of the calibration gas cylinder, and input the calibration gas
into the instrument at a flow rate of 0.9L/min (ml/min) or more. After the
real-time concentration (upper middle) displayed on the instrument is
basically stable (about 1-3 minutes, The stabilization time of different
sensor is different), press the “OK” button, then the instrument enters the
30-second countdown for calibration, the calibration operation is completed
after the screen displays ” success”.
Notice:
- During the calibration countdown period, you can press the “BACK” key to terminate the standard gas calibration.
- When performing calibration, please confirm that the standard gas pipeline has been connected. Please perform calibration after the value displayed is stable, otherwise will cause inaccurate measurement.
- Setting value: the actual concentration of the calibration gas currently input.
- When selecting the calibration point, please follow the principle of “zero point < calibration point≤ range”, otherwise the calibration will be failed.
2.4 Alarm set
2.4.1 High alarm value, low alarm value
There are 2 alarm values, high alarm value and low alarm value, take the high
reporting limit as an example. In the alarm setting interface, press the “←”
and “→” keys to move the cursor to select and enter the high alarm limit
setting interface.
High alarm value set:
- Press the “OK” key to confirm the input, and then press the “+” and “-” keys to modify the value.
- Press the “←” and “→” keys to move the cursor.
- After entering the value, press the “OK” key to confirm.
Notice:
- The high alarm value cannot be lower than the low alarm value.
- The low alarm value or the high alarm value cannot be higher than the range.
* Indicator status:
Power: After power on, the green light is on.
Low alarm : no low alarm, light is off; low alarm, red light is on.
High alarm: no high alarm, light is off; high alarm, red light is on.
Fault: no fault, no light; self-test fault, orange light on.
2.5 Monitor 2.5.1
Pump On/Off
In the main menu interface, use the “←” and “→” keys to move the cursor and
select the “Monitor” option, then press the “OK” key to enter the “Monitor”
submenu.
In the monitor interface, use the “←” and “→” keys to move the cursor to
select the “pump on/off” option, and press the “OK” key to turn on and off. In
addition, press the PUMP key will also turn on or turn off the pump on any
interface.
In the “monitor” interface, press the “BACK” key will return to the previous
menu.
2.5.2 Address
When multiple detectors are connected to the control panel, in order to
distinguish each detector, we need to set a different address value for each
detector. In the monitor setting interface, use the“↑”、“ ↓” direction keys to
move the cursor and select “address”. Then press the “OK” key to enter the
“address” setting interface. In the address setting interface, press the
“BACK” key will let you return to the main menu interface.
Address setting:
- Press the “OK” key to confirm the input, and then press the “+” and “-” keys to modify the value.
- Press the “←” and “→” keys to move the cursor.
- After entering the value, press the “OK” key to confirm..
In the “Address” menu, you can press the “BACK” key at any time to return to
the previous menu.
Note: There is only one address for each detector. Do not set the same
value for multiple detectors. The effective address range is 001~200!
For example: two four-gas detectors.
Detector 1, the address is set to 001, then the channel C1 is 001, the channel
C2 is 002, the channel C3 is 003, and the channel C4 is 004;
Detector 2, the address is set to 005, channel C1 is 005, channel C2 is 006,
channel C3 is 007, and channel C4 is 008.
By analogy, the set address is determined by the channel quantity.
2.5.3 Current Setting
In the “Current Setting” interface, you can fine-tune and calibrate the
corresponding line of hthe output signal, or adjust the output channel. In the
instrument setting interface, use the four direction keys to move the cursor
to select the gas, 4mA or 20mA option, and use the “+” and “-” keys to adjust
the parameters, and the settings will take effect immediately.
Output channel adjustment: Arrow keys to select gas, “+”, “-” keys to adjust
the gas, the change will take effect immediately. This setting is displayed
synchronously in the detection interface. Such as: C1 and C2 channel gas
exchange, return to the detection interface
After that, the channel display corresponding to CO and H2S will also be
corresponding to C2 and C1.
Current signal fine-tuning: select the 4mA or 20mA option of the channel which
has the connection error with the arrow keys, connect the multimeter to check
and verify, press the “+” and “-” keys to adjust, after the multimeter shows
that the parameters meet the requirements, press the “BACK” key to exit the
setting interface .
In the “current setting” menu, you can press the “BACK” key at any time to return to the previous menu.
2.5.4 Relay
There are 2 options in the “Relay” menu, which are “General” and “Backlash”.
Use the “↑” and “↓” keys to switch between the two options. After selected,
press “OK” to complete the setting In “Relay” setting interface, press “back”
will let you back to previous menu.
General: The detector changes from normal to alarm state, and the relay
control interface is normally closed or normally open; the detector changes
from alarm to normal state, and the relay control interface is normally open
or normally closed.
Example: When the detector alarms (low or high alarm), the relay normally open
end interface closes the circuit to drive the ventilation system or the alarm
system to remind the operator; when the detector returns to normal state, the
relay normally open end interface returns to the open circuit state and
closes the ventilation system or alarm system.
Backlash: The detector triggers the high alarm state, and the relay control
interface is normally closed or normally open; when the detector returns to
the normal state from the high alarm, the relay control interface is normally
open or normally closed.
For example: when the detector triggers a high alarm, the normally open port
of the relay closes the circuit to drive the ventilation system or alarm
system to remind the operator; the detector returns to the normal state, the
normally open port of the relay returns to the open state, and the
ventilation system or alarm system is turned off.
*Specific applications need to take into account the characteristics of special gases such as oxygen.
2.5.5 Reset
Reset device: This function can restore all settings (except the sensor
calibration parameters) to the factory settings.
Reset sensor: this function can restore the instrument’s sensor to the factory
calibration settings. Reset operation:
- First press the “↑” and “↓” keys to select and press “OK” to confirm. Then a 10-second countdown will be displayed. During the countdown, press the “BACK” to terminate the reset.
In the “Reset” interface, press the “BACK” key back to previous menu.
Note: The reset operation is irreversible. After reset, the parameters set by user will be cleared and can’t be restored. Please use it carefully.
2.5.6 Language Setting
The instrument supports both Chinese and English Language.
- First press the “↑” and “↓” keys to select and press “OK” to confirm to switch.
In the “Language” setting interface, press the “BACK” key will let you back to previous menu.
2.6 Info
2.6.1 Monitor Info
Monitor info including Version and Date.
2.6.2 Sensor Info
Sensor info including gas type, measure range and ADC value
2.6.3 Firmware Upgrade
For online upgrade, you need to use the PC terminal to operate. For details,
please consult the sales staff or check the relevant operating instructions.
Common faults and solutions
Fault phenomenon | Possible cause of failure | Solution |
---|---|---|
Sensor failure | Over range used | Replace the sensor with a new one and use it |
properly
Sensor loose| Please contact dealer or manufacturer for repair
Use in non-detection environments
Personal operation error
No response to detected gas| Circuit failure
Display is not accurate| Sensor expired
Long-term uncalibrated| Please calibrate in time
Zero calibration function not available| Too much sensor drift| Calibrate or
replace sensors in time
Over range used| Please contact dealer or manufacturer for repair
When the instrument detects normally, the interface shows the full scale|
Sensor failure
Self-test failed| Sensor could not be found| 1,Restart the instrument 2,Please
contact dealer or manufacturer for repair
Terms of Service
Warranty Commitment
The company promises that all the detectors delivered by the company will be
calibrated with the relevant specific concentration of standard gas. After
purchasing the company’s products, users do not need to do span calibration
without special circumstances, and the operation must be performed by
professional technicians.
This commitment is only available for detector and does not include
accessories.
Repair time
When your machine needs to be repaired, we will repair it for you within 7
valid working days after receiving your returned machine. In case of special
circumstances, if the repair cannot be completed within 7 valid working days,
our staff will call in advance to negotiate the repair date with you.
The aforementioned restoration dates do not include shipping time.
Limited Liability Warranty
After your product is repaired by our maintenance agency, you will continue to
enjoy the promise of the original warranty period.
When you need warranty service, please present a valid warranty certificate,
including warranty card and a invoice or purchase contract.
When the situation listed in the warranty description is not covered by the
warranty, you can choose a paid maintenance services.
If the repaired parts exceed the free warranty period, please pay a certain
repair service fee. The standard of the repair service fee is provided by our
maintenance organization.
We have the right not to provide warranty service for product damage caused by
the following circumstances:
- Human-induced damage.
- Damage caused by violation of operating regulations and requirements.
- Damage caused by all natural disasters such as floods, fires, etc.
- Damage caused by harsh use environment.
- This product is repaired, altered, modified or disassembled by unauthorized service personnel.
National standards for product development, design and production
Explosion-proof standard:
GB 3836.1-2010 “Explosive Atmospheres Part 1: General Requirements for
Equipment”
GB 3836.2-2010 “Explosive Atmospheres Part 2: Equipment Protected by
Flameproof Enclosure “d””
Gas verification standards (taking conventional four gases as an example):
JJF 1421-2013 “—Outline of Type Evaluation of Carbon Oxide Alarms”
JJF 1363-2019 “Type Evaluation Outline of Hydrogen Sulfide Gas Detector”
JJG 365-2008 “Regulations for the Verification of Electrochemical Oxygen
Meters” JJG 693-2011 “Regulations for the Verification of Combustible Gas
Detection Alarms”
Other:
GB 12358-2006 “General technical requirements for ambient gas detection and
alarm instruments in workplaces”
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>