KANE KM940 Hand Held Combustion Analyser
Product Information
KM940 Hand-held Combustion Analyser
Specifications
- Model: KM940
- Type: Hand-held Combustion Analyser
- Website: https://manual-hub.com/
ANALYSER LAYOUT AND FEATURES
Instrument Features and Keypad
The KM940 Hand-held Combustion Analyser comes with the following features:
- RS232 (8 pin din) port
- Infra-red emitter
- Pressure Inlet Inlet Port (side)
- Charger socket
- Store Print button
- Pump On/Off button
- Menu button
- On/Off button
- Up and Down navigation buttons
- Enter button to accept commands and menu options
Instrument Layout (Rear)
The rear of the instrument includes various ports and sockets for connectivity
and charging.
Standard Probe Configuration
The KM940 Hand-held Combustion Analyser comes with a standard probe
configuration for accurate measurements.
Analyser Connections
The analyser has various connections including an RS232 port, infra-red
emitter, pressure inlet port, and charger socket.
SAFETY WARNING
Before using the KM940 Hand-held Combustion Analyser, please read the safety
warnings provided in the user manual to ensure safe operation.
FIRST TIME USE
For first-time use of the KM940 Hand-held Combustion Analyser,please follow
the instructions provided in the user manual.
NORMAL START UP SEQUENCE
-
Every Time You Use The Analyser
Before each use, follow the normal start-up sequence outlined in the user manual. -
Automatic Calibration
The KM940 Hand-held Combustion Analyser features automatic calibration. Refer to the user manual for instructions on how to perform automatic calibration. -
Main Display Parameters
The main display of the analyser shows various parameters. The user manual provides a detailed explanation of each parameter.
MOVING THROUGH THE MENUS
-
Basic Operation
To navigate through the menus of the KM940 Hand-held Combustion Analyser, use the keypad and the Up/Down buttons. The user manual provides detailed instructions on basic operation. -
Menu Options and Settings
The analyser offers various menu options and settings. Refer to the user manual for a comprehensive list of menu options and instructions on how to configure settings.
PRINTING INFORMATION
-
Printing a ‘Live’ Test
To print a live test result, follow the instructions provided in the user manual. -
Standard Printout
The KM940 Hand-held Combustion Analyser provides a standard printout option. Refer to the user manual for instructions on how to obtain a standard printout.
STORING AND RETRIEVING DATA
-
Storing a ‘Live’ Test
To store a live test result, follow the instructions outlined in the user manual. -
Viewing and Printing a ‘Stored’ Test
To view and print a stored test result, refer to the user manual for step-by- step instructions. -
Deleting Data
If you need to delete data from the KM940 Hand-held Combustion Analyser, please follow the guidelines provided in the user manual.
MAINTENANCE
-
Emptying and Cleaning the In-line Water Trap
Regular maintenance of the in-line water trap is essential for optimal performance. The user manual provides instructions on how to empty and clean the water trap. -
Changing the Particle Filter
If the particle filter needs to be replaced, please refer to the user manual for guidance on how to change the filter.
FAQ (Frequently Asked Questions)
Q: Where can I find the user manual for the KM940 Hand-held Combustion
Analyser?
A: The user manual can be found on the official website at <https
://manual-hub.com/>.
Q: How do I perform automatic calibration?
A: Instructions on how to perform automatic calibration can be found in
the user manual under the “Normal Start Up Sequence” section.
Q: How do I navigate through the menus?
A: To navigate through the menus, use the keypad and the Up/Down buttons.
Detailed instructions can be found in the user manual under the “Moving
Through the Menus” section.
ANALYSER LAYOUT AND FEATURES
Instrument Features and Keypad
ON/OFF
MENU
Allows access to all menu functions
PUMP
Turns pump on and off
ENTER
Accepts a command ie enters a menu option
UP
Scrolls up through options ie Fuel
DOWN
Scrolls down through options
STORE
Enters data storage menu
PRINT
Prints current data
Instrument Layout (Rear)
NOTE! Do not cover exhaust port as this will severely affect analyser operation.
Standard Probe Configuration
Analyser Connections
SAFETY WARNING
This analyser extracts combustion gases that may be toxic in relatively low
concentrations. These gases are exhausted from the side of the instrument.
This instrument must only be used in well ventilated locations. It must only
be used by trained and competent persons after due consideration of all the
potential hazards.
Protection Against Electric Shock (in accordance with EN 61010-1 : 1993)
This instrument is designated as Class III equipment and should only be
connected to SELV circuits. The battery charger is designated as:
- Class II equipment
- Installation category II
- Pollution degree 2
- Indoor use only
- Altitude to 2000m
- Ambient temperature 0°C-40°C
- Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50%RH at 40°C
- Mains supply fluctuations not to exceed 10% of the nominal voltage.
FIRST TIME USE
Charge the battery for 12 hours, following this an overnight charge should be
sufficient for an average 8 hour day. See Main Parameter displays for Battery
Indicator.
The KANE940 has a rechargeable lead acid battery which uses a different
charger than other Kane analysers. Ensure the correct charger is used or
damage may occur to the instrument.
Check that you have all the items you have ordered.
Take time to read this manual fully.
When using the analyser for the first time you will need to choose from:-
- Language selection
- Calibration countdown time
- CO gas alarm
- NOx percentage for calculation
- Time and Date
- Printed header name and telephone number
The SET UP MENU (Section 5.2.5) gives details of how to change the above settings.
NORMAL START UP SEQUENCE
Every Time You Use The Analyser
- BEFORE SWITCH-ON CHECK THAT:
- the particle filter is not dirty
- the water trap and probe line are empty of water
- all hose connections, etc, are properly made
- the probe is sampling CLEAN AMBIENT air
- the water trap is correctly fitted and the instrument upright the flue temperature is connected
Switch ON the instrument by pressing
Automatic Calibration
During this sequence the analyser pumps fresh air into the sensors to allow
toxic sensors (if fitted) to be set to zero and the Oxygen sensor to be set to
20.9 %.
After switch-on the analyser will briefly display header information :
And then show the countdown screen :
The calibration time will count down in seconds to zero. Calibration time may be changed from 2 to 6 minutes. See Set-Up menu section 5.2.5.
Note ! Three minutes is recommended to allow the sensors to stabilise fully. Anything less than this may result in drift of the toxic and oxygen sensors in clean ambient air.
To obtain the quoted specification an instrument should be calibrated with clean ambient air at standard temperature and pressure (STP).
Once the time has reached zero an audible beep will be heard and will show the selected fuel on the following display:-
This zeros the toxic sensor and sets Oxygen to 20.9%. The next screen is the MAIN DISPLAY of the analyser:-
All parameters are detailed in Appendix A – MAIN DISPLAY PARAMETERS.
Main Displays
The main display can be changed to show either 4 or 8 parameters at one time. Two options are available when 4 parameters are selected.
- 4 Page Mode displays 4 lines of data in set format, each page is predefined.
- Line scroll mode allows you to customise the display to show the data you require.
- 8 Page Mode displays 8 parameters on 4 lines in set format, the bottom two can be changed.
Changing between the different modes is detailed in Display Menu Section 5.2.4.
4 Page Mode
Use the and keys to change the information that is displayed on the screen.
The following pages are available.
This screen only displayed on an analyser fitted with an NO sensor
TIP – In 4 page mode only turns the backlight ON and OFF.
Line Scroll Mode
Line scroll mode allows you to customise the display.
Use the and keys to change the bottom line of the display. Once the correct
line is displayed press to confirm and move the line up. Select the next
parameter and repeat until all lines display the desired parameters.
8 Page Mode
Displays 8 parameters on the screen at one time. Symbols used in this mode are
different to those used in 4 page and line scroll modes and are detailed in
Appendix A – MAIN DISPLAY PARAMETERS.
The bottom line of the display can be changed to display other parameters.
Use the and keys to change this line.
Sampling the Flue Gas
Once the automatic calibration procedure has been completed and the specific
fuel has been selected (See SELECT menu) the probe can be inserted into the
desired sampling point.
It is recommended that the sampling point be located at least two flue
diameters downstream of any bend and that the probe tip is in the centre of
the flue. With balanced flues and other domestic units the probe should be
positioned far enough into the flue so that no air can ‘back flush’ into the
probe. This will be indicated by a low oxygen reading and/or a low ‘Poison
Index’ reading.
The probe depth stop cone provided with the instrument allows the probe to be
used in holes whose diameters range from 8 mm to 21 mm (5/16 to 13/16 inch).
The standard probe is rated at 650°C. Temperatures of up to 1100°C (2012°F)
can be accommodated using an optional high temperature probe.
TIP To conserve battery power, switch off the pump when you are not taking a
measurement. Use the key to turn ON and OFF the pump.
Taking a Pressure Reading
With the optional pressure module fitted a flue draught measurement can be
made at any time.
Connect the standard probe to the pressure sensor inlet and the probe in the flue.
The pressure reading will be displayed :-
To perform a combustion test and display draught pressure at the same time a special probe is required. Contact Kane International or Authorised Distributor for details.
Regular Checks During Sampling
Care must be taken at all times not to exceed the analysers operating
specifications, in particular ensure the following :-
- Do not exceed the maximum temperature of the flue probe.
- The analyser internal temperature does not exceed normal operating range, typically 0- 40°C.
- DO NOT PLACE THE INSTRUMENT ON A HOT SURFACE.
- The water trap is vertical at all times. Water condenses in the probe line and can quickly fill the water trap when the probe is moved. Take care and watch the water trap closely.
- The in-line particle filter is clean and does not become blocked.
Normal Shutdown Sequence
DO THIS EVERY TIME YOU USE THE ANALYSER
Remove the probe from the flue – TAKE CARE ! THE PROBE WILL BE HOT – and allow
it to cool naturally. Do not immerse the probe in water as this will be drawn
into the analyser and damage the pump and sensors.
Once the probe is removed from the flue press and the analyser will count down
from 30 to switch off.
If you have not finished but press by mistake, you can press to return to normal operation and not switch OFF.
Electromagnetic Compatibility
The European Council Directive 89/336/EEC requires that electronic equipment
does not generate electromagnetic disturbances that exceed defined levels and
has an adequate level of immunity to enable it to be operated as intended. The
specific standards applicable to this product are detailed in the appendices.
Since there are many electrical products in use that pre-date this Directive
and may emit electromagnetic radiation in excess of the standards defined in
the Directive there may be occasions where it would be appropriate to check
the analyser prior to use.
The following procedure should be adopted:
Go through the normal start up sequence in the location where the equipment is
to be used.
Switch on all localised electrical equipment that might be capable of causing
interference.
Check that all readings are as expected. (A level of disturbance in the
readings is acceptable). If not adjust the position of the instrument to
minimise interference or switch off, if possible, the offending equipment for
the duration of the test.
N.B. Maximum cable lengths must be less than 3 metres.
At the time of writing this manual (November 2005) Kane International Ltd is not aware of any field based situation where such interference has ever occurred and this advice is only given to satisfy the requirements of the Directive.
MOVING THROUGH THE MENUS
Basic Operation
Menu Options and Settings
Main Menu
The MAIN MENU consists of 4 sub menus which are shown below and detailed on
the following pages.
All sub-menus are accessed using and exited using
The and keys move the cursor within a menu and allow parameters to be changed.
TIP Holding down one of these keys scrolls through the data quicker.
Select Menu
This menu allows selections to be made for the parameters detailed below.
FUEL : Select the fuel being used by the boiler from either a standard fuel stored in the analyser or by entering the user fuel. Once the correct fuel has been selected press to view the fuel constants.
Calculation of fuel constants are detailed in the Appendix. Fuel constants will have to be calculated before a user fuel can be entered.
Use to move to the next parameter, repeat above until all parameters are
correct. Press to return to SELECT menu.
O2 Ref :
Toxic gas measurements can be referenced to defined oxygen levels.
Reference values can be set from 1-20%, to AUTO or more normally to the
default value – OFF. Setting to AUTO uses the figure in the FUEL constants
data.
Oxygen referencing is required by some regulations such as TA-LUFT. If a
reference value is selected then toxic gas measurements will be displayed with
the symbol (n) attached to the reading. i.e. CO(n)
What does Oxygen reference mean ?
If 3 % O2 reference is selected and 5 % O2 is measured in the flue then toxic
gas values will be recalculated as if 3 % were measured. The equation for
referencing is detailed in the Appendix.
Oxygen referencing prevents false readings being submitted, e.g. allowing more air into the boiler will increase the oxygen level in the flue and hence dilute any toxic gas reading. Oxygen referencing gives readings as if they were undiluted.
SMOKE: Allows the user to enter a smoke test number from 0-9. This value will be printed on the standard printout. Default value is OFF.
RESET : Allows the user set the Oxygen to 20.9% and zero the toxic sensors without turning the analyser off.
Selecting YES and will display the following screen.
After pressing the analyser will count down for 5 seconds and then return to the main display.
WARNING : The sensors must only be reset if you are sure they have been sampling fresh air for at least 3 minutes. Errors in measurement will occur if the sensors are reset during or just after sampling.
Units Menu
Allows all displayed units to be changed.
TEMP : Choose selections from Centigrade °C or Fahrenheit °F.
GAS : Changes the toxic gas measurement units. Select from volumetric
readings, parts per million (ppm) or mass flow reading milligrams per cubic
meter (mg/m3) .
PRESS. : Flue draught can be displayed in millibar (mbar), hectaPascals
(hPa), millimeters water gauge (mmWG) or inches water gauge (in WG).
EFF. : Efficiency can be selected for Gross or Net values. Gross
efficiency assumes latent heat of vapourisation is lost in the boiler and
hence will be lower than Net efficiency. For Natural Gas the difference will
be approximately 11%.
Display Menu
Allows the configuration of the display to be changed.
- LIGHT: Choose from ON or OFF.
- MODE: Select 4 or 8 Page Mode or Line Scroll Mode as detailed in section 4.3 Main Displays.
- CONTRAST:The contrast is set to a DEFAULT value or can be adjusted ↑ LIGHTER or ↓ DARKER. Use the and keys to adjust.
Set-Up Menu
The set up menu allows the following parameters to be set/altered.
- Language.
- Automatic calibration time
- CO gas alarm
- NOx percentage for calculation
- Date and time
- Printout Header
Once an alarm has been exceeded the display will flash every two minutes warning the user of an alarm state and display the gas concentration. A similar display will be shown during a RECHARGE BATTERY and PUMP OFF alarms.
NO REF:
Displayed on the Nitric Oxide unit only. Allows the percentage P in the
following calculation to be set. The default value set is 5%. Note the
percentage allows for NO2 in a typical boiler.
OFF
1-9 %
NOx = NO + P% NO
CALENDAR: Allows the user to change the date and time. (24 hour clock).
The following screen will be shown once the parameter is entered:
FORMAT : Changes the date format for display and printing.
- dd : mm : yy
- yy : mm : dd
- mm : dd : yy
To change the time position the cursor on Time and press . The cursor will now be to the left of the 13 as shown below:
Using and scroll through the setting options i.e. 0-23.
Once the correct hour is set press to move to the next parameter, the cursor
will move to the left of minutes (53). Move to each parameter until the
correct time is set. Pressing after setting the seconds will return the cursor
to the left of the screen.
Format and Date are set in a similar manner.
Header :
Allows two lines of 20 characters to be programmed into the analyser. The
header appears on the top of the standard printout. This can be used to print
your company name and/or phone number.
The screen above shows the standard header setting with the cursor now shown
underlining the K in Kane. By using and any letter or number can be chosen.
Once the correct character is displayed, use to move right to the next. Move
along until all characters spell the desired name or phone number. If you need
to go back and change a character use to move left.
Press to return to the SET UP menu.
PRINTING INFORMATION
Supplied as accessories for the KANE940 are an infra-red thermal printer or a dot matrix serial printer. Read the manual supplied with each printer prior to operation. Connections to the KANE940 are detailed below :
- Infra-red thermal printer – this does not require a cable to transmit the data but uses an infra-red (IR) link similar to a TV remote control. The IR emitter is positioned on the top of the KANE940 and the bottom of the printer. Ensure they are pointing at each other and within 300 mm, with no obstructions in the way. Data may be lost if transmission is interrupted. Keep the KANE940 pointing at the printer until the printout has finished.
- Dot matrix serial printer – requires the supplied serial cable to transmit data. Connect the cable to the 8 pin DIN socket on the top of the KANE940 and the 25 pin D-connector on the printer.
Data can either be printed from a ‘live’ test or from stored data. Printing of stored data is detailed in STORING AND RETREIVING DATA.
Printing a ‘Live’ Test
During a combustion test the KANE940 will print data on request. With the
analyser showing the MAIN DISPLAY press and current data will be sent to the
printer.
The display will show the following until data transmission is complete.
Standard Printout
The standard printout is shown below
STORING AND RETRIEVING DATA
The KANE940 can store up to 100 combustion tests. Once stored, the data can be viewed on the display or downloaded to a PC or printer.
Storing a ‘Live’ Test
While performing a test and viewing the data on the MAIN display access the
STORE menu as follows :-
Mode: Select from the following :-
- STORE – Allows data to be stored in memory.
- VIEW / PRINT – Stored data can be viewed or printed.
- DELETE – Clears all data in memory.
Location: Automatically allocates a location in the memory of the instrument
for the next test. On the display shown above the next location will be 3.
To store a test set MODE to STORE and press . The current readings will be
stored in the analysers memory.
Tip : Make a note of the location number for your particular test as it may be useful when downloading or printing.
Viewing and Printing a ‘Stored’ Test
Multiple tests can be printed easily with the KANE940.
Select PRINT under MODE in the STORE menu. This feature is in addition to the
VIEW/PRINT, STORE and DELETE options.
The cursor will move to the first number, use the and to select the location and start printing.
Press to move the cursor to the second number, select the last location to print.
To print the data press . In the screen shown above locations 1 to 10 will be printed.
During printing the following will be shown.
NOTE While the display above is shown (i.e. the instrument is printing a test) the keypad is disabled. To exit from printing wait until the current test has finished and the display below is shown:
Press to exit the print routine. The instrument will return to main display
Deleting Data
To delete the data in stored memory press to obtain the STORE MENU (as above)
:-
Press to access the STORE MENU
Press to access delete data screen
Press to delete data in memory, press to exit delete data screen.
MAINTENANCE
Emptying and Cleaning the In-line Water Trap
The in-line water trap should be checked and emptied on a regular basis. Water
vapour will condense and gather in the probe line. This may move suddenly to
the trap when the probe is moved. Care should be taken at all times.
Emptying of the water trap is detailed below :-
Carefully remove the end cap from the in-line housing. Dispose of the condensate in a suitable drain, care must be taken as it could be acidic. If condensate spills onto the skin or clothing, clean off immediately using fresh water, seek medical advice if problems occur.
Changing the Particle Filter
This is a very important part of the analyser and should be changed regularly.
It prevents dust and dirty particles entering the pump and sensors and hence
causing damage. The filter MUST be changed when it is discoloured.
Remove the end cap from the in-line filter housing. Carefully remove the paper filter element and dispose of it. Clean the inside of the filter housing with a suitable soft cloth. Insert a new filter element onto the spigot in the filter housing and carefully replace the end cap.
PROBLEM SOLVING
The following is a list of problems that may occur on the instrument through
its operating life.
If the cause of the fault is not easy to identify then we advise you contact
Kane International Service Department or an International Distributor for
expert advice.
Fault symptom | Causes |
---|
· Oxygen too high
· CO2 too low
| · Air leaking into probe, tubing, water trap, connectors or internal to instrument.
· Oxygen cell needs replacing.
· Oxygen Error (FAULT)
· Toxic sensor Error (FAULT)
| · Calibration time set too short and instrument not allowed to stabilise
· Instrument has been stored in a cold environment and is not at normal working temperature.
· Oxygen cell or toxic sensors needs replacing.
· Analyser not holding charge
· Analyser not charging
| · Battery exhausted.
· AC charger not giving correct output.
· Fuse blown in charger plug.
· Analyser does not respond to flue gas| · Particle filter blocked.
· Probe or tubing blocked.
· Pump not working or damaged with contaminents.
· Probe connected to pressure connector.
· Flue temperature readings erratic| · Temperature plug reversed in socket.
· Faulty connection or break in cable or plug.
· Analyser automatically switches off in operation.| · Battery below alarm level.
· Ambient temperature above 50°C.
· Battery quickly discharging and is faulty.
· Display shows dark lines and no reponse from ON/OFF key.| · Fault has occured on the instrument electronics and requires resetting. Contact Kane International or Distibutor.
ANNUAL RE-CALIBRATION
Whilst the sensors have an expected life of more than two years in normal
use it is recommended that the analyser is re-calibrated at least annually.
This is so that long term drift on the sensors and electronics can be
eliminated. Local regulation may require more frequent re-calibration and
users should check with appropriate authorities to ensure they comply with
relevant guidelines.
PRODUCT SPECIFICATION
- Using dry gases at STP
- Calculated
APPENDICES
A – Main Display Parameters
The parameters and their meanings are detailed as follows : –
-
DATE : Analyser date. See Set-Up menu section 5.2.5 to change.
-
TIME : Analyser time. Use Set-Up menu section 5.2.5 to change.
-
BATTERY : Displays the battery level from 0-100%. The analyser will flash (BAT) RECHARGE BATTERY at less than 10 % of charge. With the charger connected the display shows AC ON.
-
NETT : Nett temperature calculated by deducting the internal AMBIENT temperature
-
(ΔT) from the measured FLUE temperature. Displays in either °C (C) or °F (F) and will display NOT FITTED (N/F) if flue probe is not connected.
If an external INLET probe is used then INLET is deducted from FLUE. -
O2 : Oxygen reading in percentage %.
-
CO : Carbon Monoxide reading indicated in ppm or mg/m3. If the figures are referenced to oxygen then the display will show CO(n). See SELECT menu 5.2.2 for oxygen reference. The display will read
-
‘O2 > 20%’ if referenced values selected and instrument is in clean ambient air.
-
EFF (G) : Combustion Efficiency calculation displayed in percentage. Gross G or Net N can be set see SELECT menu 5.2.3. The calculation is determined by fuel type see Appendix B for calculation. The efficiency is displayed during a combustion test, ‘- -’ is displayed while in fresh air.
-
CO2 : Carbon Dioxide calculation determined by the type of fuel. This only shows a reading when a combustion test is being carried out. ‘- -’ is displayed while in fresh air.
-
FLUE : Temperature measured by flue gas probe in Centigrade or (Tf) Fahrenheit. Will show ambient temperature after fresh air calibration and
-
NOT FITTED (N/F) or FAULT (FLT) if probe disconnected.
-
INLET : Temperature measured by the optional inlet air probe. This probe is (Ti) plugged into the instrument through the RS232 socket. This figure is used to calculate the NET temperature instead of
-
AMBIENT when fitted.
-
AMBIENT : Temperature measured by the internal sensor, used in the NET
-
(Ta) temperature calculation if an INLET probe is not fitted.
-
CO/CO2 R : The CO/CO2 ratio, is the ratio of measured CO divided by calculated CO2.
It gives an indication of the following :-- How good a gas sample the instrument is reading.
- How clean the boiler is running.
For example : A new or clean domestic boiler will display a ratio of less than 0.004, a unit in need of cleaning 0.004-0.008 and a unit in need of major overhaul will show greater than 0.008.
This only shows a reading when a combustion test is being carried out. ‘- -’ is displayed while in clean ambient air.
-
P INDEX : The CO/CO2 ratio expressed as a percentage %, called the ‘Poison (PI)
-
Index’ i.e. P INDEX % = 100 x CO/CO2. ‘- -’ is displayed while in clean ambient air.
-
XAIR % : Excess air calculated from the measured oxygen and type of fuel (λ) used. During a combustion test ‘O2 > 20%’ will be displayed while in clean ambient air.
-
Prs : Flue draught pressure reading. Displayed when pressure sensor fitted. See UNITS menu 5.2.3. for scales.
-
NO: Nitric Oxide reading in ppm or mg/m3. Displayed when Nitric Oxide sensor fitted. Also displayed as NO (n) when referenced to oxygen. The display will read ‘O2 > 20%’ if referenced values selected and instrument is in clean ambient air.
-
NOx : Calculated total Nitric oxides displayed in ppm or mg/m3. Where NOx =
-
NO + P%NO, note P can be set from 0-9%, default = 5%. See SELECT menu 5.2.2. Also displayed as NOx (n) referenced to oxygen. The display will read ‘O2 > 20%’ if referenced values are selected and instrument is sampling clean ambient air.
SO2 : Sulphur Dioxide reading in ppm or mg/m3. Displayed when Sulphur Dioxide sensor fitted. Also displayed as SO2 (n) referenced to oxygen. The display will read ‘O2 > 20%’ if referenced values selected and instrument is in clean ambient air. -
O2 ref % : (O2r)
Toxic gas measurements can be referenced to defined oxygen levels.
See SELECT menu 5.2.2 for details.
COMBUSTION EFFICIENCY CALCULATION
The efficiency calculation is based upon British Standard BS845.
This identifies three sources of loss associated with fuel burning:
- Losses due to flue gasses:
- Dry Flue gas loss,
- Moisture and hydrogen
- Sensible heat of water vapour
- Unburned gas
- Losses due to refuse: Combustible in ash
- Combustible in riddlings
- Combustible in dust
- Other losses:
- radiation
- convection
- conduction
- other unmeasured losses
Net efficiency calculations assume that the energy contained in the water vapour (formed as a product of combustion and from wet fuel) is recovered and the wet loss term is zero. Gross efficiency calculations assume that the energy contained in the water vapour is not recovered.
Since the fuel air mixture is never consistent there is the possibility of
unburned/partially unburned fuel passing through the flue. This is represented
by the unburned carbon loss.
Losses due to combustible matter in ashes, riddlings, dust and grit,
radiation, convection and conduction are not included.
Efficiency Calculation:
-
Known Data – Fuel: Qgr = Gross Calorific Value (kJ/kg)
- Qnet = Net Calorific Value (kJ/kg)
- K1 = Constant based on Gross or Net Calorific Value:
- K1g = ( 255 x %Carbon in fuel )/Qgr
- K1n = ( 255 x %Carbon in fuel )/Qnet
- K2 = % max theoretical CO2 (dry basis)
- K3 = % Wet Loss
- H2 = % Hydrogen
- H2 O = % Water
-
Measured Data:
- Tf = Flue Temperature
- Ti = Inlet Temperature
- O2 m = % Oxygen in flue gas
- O2 r = Oxygen reference %
-
Calculated data: Tnet = Net Temperature
- % CO2 content in flue gas
- % Dry Flue Gas losses
- % Wet losses
- % Unburned carbon loss
- % Efficiency
-
Tnet = Flue Temperature – Inlet Temperature
-
Dry flue gas loss % = 20.9 x K1 x (Tnet) / K2 x (20.9 – O2m)
-
Wet loss % = 9 x H2 + H2O / Qgr x [2488 + 2.1Tf – 4.2 Ti]
-
simplified = [(9 x H2 + H2O) / Qgr] x 2425 x [1 + 0.001 Tnet]
-
Wet loss % = K3(1+0.001xTnet)
-
Where K3 = [(9 x H2 + H2O) / Qgr] x 2425
-
Net Efficiency % = 100 – dry flue gas losses
= 100 – 20.9 x K1n x (Tnet) / K2 x (20.9 – O2m) -
Gross Efficiency % = 100 – {dry flue gas losses + wet losses}
= 100 – {[20.9 x K1g x (Tnet) / K2 x (20.9 – O2m)]+ -
[K3 x (1 + 0.001 x Tnett)]}
-
Excess Air = [(20.9% / (20.9% – 02m%)) – 1] x 100%
-
CO2% = [(20.9 – O2m) x K2 / 20.9]
-
Unburned fuel Loss % = K4 x CO / ( CO + CO2 ) Note: CO scaled in %
-
Where K4 = 70 for coke
- = 65 for anthracite
- = 63 for Bituminous coal
- = 62 for coal tar fuel
- = 48 for liquid petroleum fuel
- = 32 for natural gas
The formula for K4 is based on the gross calorific value Qgr. To obtain the loss based on net calorific value multiply by Qgr/Qnet. Since this loss is usually small this conversion has been ignored. This loss is subtracted from the efficiency.
- Oxygen Reference CO(n) = CO x (20.9 – O2r)
(20.9 – O2m)
CALCULATION OF FUEL DATA
For any fuel not specified by Kane International the net calorific value,
gross calorific value and composition should be obtained from the fuel
supplier.
The following fuel data has been calculated with reference to the efficiency
calculation.
Example 1:
The fuel values to program into the Analyser are as follows:
Assumed values in the absence of supplied data. See previous appendix for other fuels.
ELECTROMAGNETIC COMPATABILITY (CE) STATEMENT
This product has been tested for compliance with the following generic
standards:
EN 61000-6-3
EN 61000-6-1
and is certified to be compliant
Specification
EC/EMC/KI/KANE940 details the specific test configuration,performance and
conditions of use.
Please Note: Batteries used in this instrument should be disposed of in accordance with current legislation and local guidelines.