WIELAND Zenotec Select Hybrid with Without Material Changer
Product Information
Specifications
- Product Name: Zenotec select hybrid
- Model: With/Without material changer
- Issue Date: March 2016
Product Usage Instructions
Installation and Initial Start-Up
To ensure proper installation and start-up of the Zenotec select hybrid, follow these steps:
Installation and Orientation
- Choose a suitable installation site that meets the requirements mentioned in the user manual.
- Set up the machine in the selected location.
Compressed Air Connection
- Connect the compressed air supply to the machine.
Connecting the Maintenance Unit
- Connect the maintenance unit to the machine.
- Set the air pressure for the maintenance unit according to the provided guidelines.
Operating the Machine
To operate the Zenotec select hybrid, follow these instructions:
Cutters
- Insert the appropriate cutter into the machine.
- Set the desired parameters for cutting.
- Start the cutting process.
Grinders
- Insert the grinder into the machine.
- Set the desired parameters for grinding.
- Start the grinding process.
Changing Discs (Zenotec select hybrid without material changer)
If your Zenotec select hybrid does not have a material changer, follow these steps to change discs:
Material Changer (Zenotec select hybrid with material changer)
If your Zenotec select hybrid has a material changer, follow these steps to use the material changer:
FAQs
- Q: Where can I find the list of parts for the Zenotec select hybrid?
- A: The list of parts can be found in section 1 of the user manual.
- Q: How do I connect the maintenance unit to the machine?
- A: Please refer to section 5.4 of the user manual for detailed instructions on connecting the maintenance unit.
- Q: Can I use different cutters for different materials?
- A: Yes, you can insert different cutters depending on the material you are working with. Refer to section 7.4.1 for more information.
OPERATING INSTRUCTIONS
Zenotec select hybrid with / without material changer
Operating Instruction Issue March 2016
List of Parts
1 x Power cord
1 x USB cable
8 x Disc holders
1 Set of extra screws for the disc holder and the tool measuring plate
1 x Drill 2.8 mm for tool changer inserts
1 x Machine dongle
1 x Calibration set consisting of a micrometer gauge and calibration discs
1 x Maintenance unit for compressed air operation
6
1 x Spindle Maintenance Kit
1 x Allen key 2.5 mm for disc change
1 x Key for emergency release of the front flap
1 x Hose connection for external extraction
1 x Compressed air hose Ø 6 mm
2 x Tool changer inserts made of rubber
1 x Allen key 3 mm to tighten blocks in the IPS e.matrix
7
Description of Zenotec select hybrid components
A E F G
C D
B
Figure 1: Zenotec select hybrid – description
K
H
L
I
Figure 2: Milling chamber – description 8
2.1 Introduction
Dear Customer Thank you for having purchased the Zenotec select hybrid milling
system. It is a modern 5-axis milling system for dental applications. The
milling system has been designed according to the latest industry standards.
Inappropriate use may damage the equipment and be harmful to personnel. Please
observe the relevant safety instructions and read the Operating Instructions
carefully.
Enjoy working with the Zenotec select hybrid.
No.
Component description
A
Front flap with safety contact and safety interlock
B
Working chamber or milling chamber
C
Synchronous spindle SFS300P
D
Disc holder with tool magazine
E
Measuring sensor for tool length measurement
F
Material changer chamber
G
Release for disc holder
H
A-axis: Rotation range +/- 360°
I
B-axis: Rotation range +/- 30°
K
Measuring sensor for tool length measurement
L
Tool changer for 16 tools
9
2.2 Signs and symbols
The signs and symbols in these Operating Instructions facilitate locating
important points. They have the following meanings:
Symbol
Note
General warnings. Failure to observe them may lead to injury
Notes on preventing a hazard
HAZARD describes a dangerous situation which may lead to severe bodily injury or death
WARNING describes a dangerous situation which may lead to severe bodily injury or death
CAUTION describes a dangerous situation which may lead to light or medium bodily injury
NOTE describes a situation which may lead to damage to the product or in the vicinity of the product
HAZARD indicates immediate electrical danger
Note on disconnecting the machine from its power source
Note on the risk of stumbling and falling
Note on the risk of injury by cutting and crushing
Note on the risk of hearing damage due to loud noise
Note on specific conditions ensuring trouble-free operation of the machine
10
Background information Notes on making work easier Application notes and other important information Notes on disposal
2.3 Action requests
M
Manual action step
S
Action step in the software
11
2.4 Notes regarding the Operating Instructions
Machine concerned: Zenotec select hybrid with and without material changer
Target group: Dental technicians, dental health professionals, CNC technicians
This machine may only be operated by trained specialists. These Operating
Instructions facilitate the correct, safe and economical use of the milling
system. Should you lose these Operating Instructions, you can order extra
copies from the respective service centre or download them from www.wieland-
dental.de.
2.5 Different voltage versions
The system is equipped with a multi-voltage power pack and can be operated in
power networks with the following specifications: AC network 100 240 V 50
60 Hz Country-specific power plug
In these Operating Instructions, the milling system is powered via a Schuko
plug, as an example. Please note that the plug shown in the Figures may differ
depending on the country-specific version of your milling system.
2.6 Figures in the Operating Instructions
All Figures and illustrations in these Operating Instructions are used for
exemplification. The details are not authoritative for the construction of the
milling system. They are symbols which may slightly differ from the original,
e.g. due to simplification.
12
Safety
This chapter is especially important for individuals who work with the Zenotec
select hybrid milling system or who have to carry out maintenance or repair
work. This chapter and the corresponding instructions must be read and
followed carefully!
3.1 General safety notes
Carefully read these Operating Instructions before you operate the machine.
Follow the listed safety instructions to eliminate risks and prevent any
possible severe injuries.
The operating personnel must be informed about the intended use of the
machine. They must observe the stipulated safety instructions and protective
regulations (see chapter 3). Inspect the machine in general and the protective
equipment in particular for any damage. Damaged protective equipment or parts
must be repaired or replaced by a service technician authorized by Wieland
Dental + Technik, if not otherwise indicated in the Operating Instructions.
Apart from during maintenance procedures, Zenotec select hybrid with and
without material changer must only be operated with the front flap closed and
with the safety interlock activated. It is forbidden to circumvent the
existing safety equipment of the machine or to shut them down.
Children and animals have to be kept away from the machine. Do not operate the
machine unsupervised if the operating conditions described in chapter 3 and 5
cannot be fulfilled. Operating the machine in the service mode is only
permitted for personnel authorized by Wieland Dental + Technik. Please also
observe the information regarding tool selection and setting of the tool
parameters (see chapter 7 and 8).
Figure 3: Zenotec select hybrid with closed front flap
13
3.2 Warnings
Use of damaged cables Mortal danger by electric shock!
Disconnect the machine from the power source and protect the machine from
accidental operation. Consult Customer Service. Replace the damaged cables
with original replacement cables.
Troubleshooting during operation Mortal danger by electric shock!
Consult Wieland Service before you rectify a malfunction. Disconnect the
machine from the power source and protect the machine from accidental
operation.
Operating the machine in the service mode with the front flap open. Risk of
injury by cutting and crushing! Hazard due to ejected shavings! Apart from
during maintenance procedures, Zenotec select hybrid must only be operated
with the front flap closed and with the safety interlock activated. It is
forbidden to circumvent the existing safety equipment of the machine or to
shut them down.
Milling of materials hazardous to health Respiratory disorders by inhalation
of hazardous substances!
Work only with the extraction system switched on. Use only materials that do
not present a hazard to health provided the extraction system is properly
used. Always operate the extraction system with the filter intended for this
purpose by the manufacturer.
14
Contamination of processing materials by other materials after material change
Thoroughly clean the machine before and after processing CoCr alloys, so as to
avoid damage to the subsequently milled restorations.
Loud operating noise of the machine Damage to hearing caused by loud noise!
Examine the fixation of the work pattern, the condition of the tool and the
processing materials. If loud operating noise cannot be prevented, use hearing
protectors during milling.
Open pneumatic connections Hazard from loose pneumatic components under
compressed air!
Disconnect the machine from the compressed air supply. Consult Customer
Service.
Mess at the workplace Risk of stumbling and falling!
Keep the work station clean. Store peripherals, such as Zenotec cyclon, in a
safe place.
Ergonomics of the work station Consequential damage by continuous one-sided
physical stress!
Arrange the work station in an ergonomic fashion. Ensure optimum seat height,
screen position and lighting.
Mounting of milling tools Cutting injuries by sharp cutting edges!
Grasp the milling tools by the shafts only, not the cutting edges. Make sure
not to injure yourself or others when handling the tools.
Disposal The machine must not be disposed in the normal domestic waste. Please
correctly dispose of old devices according to the corresponding EU Council
Directive. Information regarding disposal can also be found on the respective
national Wieland Dental + Technik website. The packaging may be discarded with
the regular household refuse.
15
3.3 Indications
Zenotec select hybrid is designed for easy to medium-hard milling procedures
in the field of dental technology. Please note that only the indications,
materials and milling strategies approved by the manufacturer ensure
processing without damage to the machine. In-house modifications and thus
possible damage to the machine void all warranty services provided by the
manufacturer. The machine has been designed for milling the following
materials:
Wax Telio CAD® for Zenotec, PMMA Cast and other commercially available dental
resins Composites Zirconium dioxide Glass-ceramics* (IPS e.max® CAD for
Zenotec, IPS Empress CAD® for Zenotec) CoCr alloys (Zenotec NP)
- Basically, all Wieland Dental + Technik and Ivoclar Vivadent materials of the respective categories are suitable for processing with the machine. The processing of materials from other manufacturers with this system is not recommended due to the lack of coordination of the milling strategies and tool harmonization.
If the machine is used for any purpose other than the above designated uses, the system may be damaged. The protective equipment must not be circumvented or suspended. The indications also include observing these Operating Instructions and following the maintenance instructions therein. In order to ensure the product safety and warranty services, the product must be exclusively operated with the original accessories from Wieland Dental + Technik. The user bears the risk of using not approved accessories. Wet grinding is only permitted in conjunction with Zenotec hydro. Should you make any modifications to the system or remove the housing without prior written consent by a Wieland Dental + Technik Service Centre, all warranty claims are void. The machine may be operated unsupervised, provided the national and local laws and provisions allow for such action and provided that they are observed. Furthermore, the requirements of the respective insurance company must be met.
The following requirements must be met to ensure unsupervised operation:
Never operate the machine with the milling chamber uncleaned. The machine must be protected from intervention by unauthorized individuals. The room in which the machine is operated must be equipped with an automatic fire detection system.
The operation of Zenotec select hybrid requires a compressed air connection and a RCD breaker.
Dry milling is only admissible in combination with Zenotec cyclon.
Insufficient extraction and/or cooling during dry/wet processing results in machine damage.
16
Product description
4.1 General aspects
Zenotec select hybrid is a state-of-the-art 5-axis milling system for dental
applications. The Zenotec milling system is suitable for the fabrication of
digital restorations. It has been designed in such a way that various
materials can be processed. The milling procedure is controlled via an
electronic system with the corresponding software. The Zenotec select hybrid
system consists at least of the following components:
CNC milling machine CNC software with PC Zenotec cyclon extraction system
Optional: Zenotec hydro (see chapter 11)
4.2 Technical data
4.2.1 Basic system Dimensions (w/d/h): approximately 490 x 445 x 540 mm
(Zenotec select hybrid), approximately 692 x 445 x 540 mm (Zenotec select
hybrid with material changer) 5-axis mechanics; traversing range (x/y/z): 140
x 98 x 76 mm Precision ball screw spindles in the 3 linear axes with 4 mm
spindle pitch Repeat accuracy ± 0.003 mm Complete housing of the working
chamber with safety contact and safety interlock at the front flap Working
chamber lighting Negative pressure sensor for monitoring the extraction system
Disc dimensions: Ø 98.5 mm (± 0.2 mm), with shoulder Disc height: 10 mm to 30
mm Weight: approximately 75 kg (Zenotec select hybrid), approximately 95 kg
(Zenotec select hybrid with material changer) Compressed air consumption of
max. 80 l/min
4.2.2 Rotation axes Clearance-free rotation axes with Harmonic-Drive® Rotation
angle: 360 degrees (A axis), ± 30 degrees (B axis) Automatic measurement of
the rotation axis and automatic axis compensation for precise results Integral
disc holder for 98.5 mm discs (Zenotec select hybrid) Open disc holder for
98.5 mm discs (Zenotec select hybrid with material changer)
17
4.2.3 Spindle
Synchronous spindle with speeds of up to 60,000 rpm Nominal capacity under
permanent operation: 300 Watt Maximum output (Pmax): 600 Watt 4-fold bearing
Hybrid ceramic ball bearings Pneumatic collet Ø 3 mm 1 compressed air nozzle
for dry processing 3 liquid nozzles for wet processing Cone cleaning Sealing
air
4.2.5 Electronic controls
5-axis microstep electronic controls Simultaneous activation of 5 axes Smooth
running and high precision due to microstep operation Quick processing due to
exponential acceleration ramps and automatic switchover to full-step operation
Continuous path due to look-ahead function Included control software for the
convenient output of the milling data generated by the postprocessor Travel
speeds: 0.003 mm/s to 55 mm/s
4.2.4 Tool changer Automatic tool changer for 16 tools Automatic tool length
measurement and fracture monitoring by means of a measuring sensor Use of
diamond tools possible Compressed air monitoring for automatic tool change
4.2.6 Material changer
(only Zenotec select hybrid with material changer) Mounting of up to 8
material holders Safety contact at the material changer front flap Automatic
delivery of the selected material into the work chamber Pneumatically driven
slide flap between the work chamber and the material changer
4.2.7
Sound emission
The sound emission strongly varies depending of the processed material and the
milling conditions. If the machine is too loud, change the milling conditions.
Examine the fixation of the work pattern, the condition of the tool and the
processing materials.
If loud operating noise cannot be prevented, use hearing protectors during milling.
Measuring conditions: Tool (service life already elapsed) Measuring distance to the sound source 1 m Extraction system activated Measurement according to ISO 3746, accuracy class 3
Operating status
Maximum value of A-rated noise intensity levels
Milling of the measuring disc (milling conditions see above) 77 dB(A)
All other action movements
(tool change, axis movement, < 70 dB(A)
etc)
18
4.3 Hazardous areas and safety equipment
Description of the hazardous areas of the machine:
Hazardous area Type of risk
Machine chamber / Milling chamber Risk of injury
Hazardous area Type of risk
Opening and closing mechanism Risk of crushing
Hazardous area Type of risk
Electrical components Risk of electrical shock
Description of the safety equipment of the machine:
Safety equipment Protective conductor Electrical fuses Machine housing and hood
Protective effect Protection from electrical shock Protection from electrical shock Protection from electrical shock, burning and crushing
19
Installation and initial start-up
5.1 Installation and orientation
At least two people are recommended for the installation of the system.
Installation and orientation must be performed by a service technician
authorized by Wieland Dental + Technik.
5.1.1 Requirements placed upon the installation site The location for the
machine must meet the following requirements:
The milling system should neither be placed nor operated in areas where there
is an explosion hazard.
Place the machine on a stable, even surface constructed to hold the weight of
the machine. The control computer should, if possible, be connected to a
separately secured circuit (supply voltage fluctuations caused by other
equipment should be prevented, if possible). The room temperature should be
between 18 °C (64.4 °F) and 25 °C (77 °F); maximum room temperature 32 °C
(89.6 °F). The machine location should be free of dust, since ambient air is
aspirated. Maximum relative humidity 80%, non-condensing A/C power connection
with 100 240V and 50 / 60 Hz In-bound compressed air at least 6.5 to 7 bar
(94.27 to 101.5 psi), free of oil and water There must be sufficient clearance
to the wall on the right side of the machine to prevent the air vents from
being blocked (at least 10 cm). On the left and the rear of the machine, there
must be enough clearance to access the machine connections. Required
connections: Compressed air connection, as well as one socket each for the
machine, extraction system and laptop and/or PC
Connect the machine to a separately secured circuit or make sure that no
devices are connected that cause severe supply voltage fluctuations when
switched on. These fluctuations interfere with the electronic controls and may
cause failure of the system.
20
5.2 Setting up the machine
Use only the original cables intended for the system for the connections at
the switching outputs. The system must exclusively be controlled via the
Wieland CAM computer and the control must not be modified. The Zenotec cyclon
is recommended for optimum utilization of the system.
Figure 4: Machine connections
M1. M2.
M3. M4. M5. M6. M7. M8. M9. M10.
Unpack the milling machine. Keep the packaging of the machine to use it if you
have to send in the machine for maintenance or repair.
Connect the supplied maintenance unit to the standard compressed air
connection of your compressed air supply (see chapter 5.3).
The blue compressed air hose must be inserted into the connection marked in
blue at the lower left of the connection panel of the machine (see Figure 4).
If the compressed air hose is not connected, the machine does not start!
When connecting the compressed air hose, make sure to insert it only into the
socket provided for that purpose.
Check the pressure indicated on the manometer. If the indicated pressure is
not 6.5 – 7 bar (94.27 – 101.5 psi), set the compressed air supply accordingly
(see chapter 5.4.1).
Connect the Wieland CAM computer using the USB port of the machine.
Attach the extraction hose to the connection of the machine intended for this
purpose.
Insert the machine dongle in a USB port of the CAM computer.
Connect the power cord with the machine.
Connect the machine to a separately secured circuit or make sure that no
devices are connected that cause severe supply voltage fluctuations when
switched on (e.g. compressor, etc).
Make sure that the front flap is closed.
Switch on the machine with the master switch.
21
5.3 Compressed air connection
The SFS 300P spindle requires a compressed air connection. This compressed air
connection must supply 6.5 – 7 bar (94.27 – 101.5 psi) during non-stop
operation. The spindle with automatic tool change requires the compressed air
in order to activate the pneumatic collet. Furthermore, sealing air is
required to prevent foreign objects from entering the spindle. The air
consumption of the machine is 80 l/min.
The delivery form of your machine also contains a maintenance unit for the
compressed air supply. The incoming pressure for the machine can be regulated
at the maintenance unit. The built-in water separator with a filter cartridge
should prevent air contaminated with humidity or dirt particles from entering
the machine.
The incoming compressed air must be dry and oil-free according to ISO 8573-1,
since the water separator can only filter out minor residual quantities.
Compressed air that does not meet these requirements may lead to failure of
the bearings and electrical damage to the spindle.
Air purity according to ISO 8573-1:
Solid matter particles Water Total oil content
Class 3 Class 4 Class 3
Filter rate better than 5 µm for solid matters Maximum pressure condensation point + 3°C Maximum oil content 1 mg/m³
5.4 Connecting the maintenance unit
The maintenance unit can either be mounted to the housing or directly
connected to the compressed air supply. For the choice of mounting, it is
important to consider that the maintenance unit must be clearly visible during
daily operation to enable regular inspection. The following text describes the
mounting of the maintenance unit to the housing. Please observe the steps M3.
to M5. for connecting the maintenance unit directly to the compressed air
supply.
The maintenance unit must always be installed upright. If this is not done, the function of the water separator cannot be ensured.
M1.
Remove the two blind screws below the air vents on the side panel of
the milling machine.
M2.
Attach the maintenance unit in an upright position using the cylinder-
head screws supplied.
Figure 5: Maintenance unit
Risk of injury by escaping compressed air or uncontrolled compressed air hoses. In a first step, connect the maintenance unit with the machine, before connecting it to the compressed air supply. Otherwise there is a risk of injury.
22
M3.
Shorten the compressed air hose supplied as required.
M4.
Connect the right compressed air connection of the
maintenance unit with the pneumatic input of the machine.
M5.
Connect the compressed air supply with the standard compressed air
connection of the maintenance unit (left side).
5.4.1 Setting the air pressure Setting the air pressure is only necessary, if
the pressure indicated by the manometer is not between 6.5 and 7 bar (94.27
and 101.5 psi).
To set the air pressure, the maintenance unit must be connected to the
compressed air supply.
M1.
Pull the control knob on top of the maintenance unit slightly out.
M2.
Regulate the pressure (in the direction of the arrow “+” you increase the pressure, in the direction of the arrow
“-“, you reduce it) until the value indicated is 6.5 and 7 bar (94.27 and 101.5 psi).
M3.
Subsequently, press the knob down again.
The compressed air setting is now locked in place and cannot be accidentally changed.
23
Installing the software
The control software can only be installed and the milling machine operated
with the dongle connected.
For the installation of the software, the CAM computer must be connected with
the machine and switched on.
S1.
Insert the dongle supplied in a USB port of the CAM computer.
S2.
Run setup.exe.
S3.
Follow the installation steps of setup.exe.
Further information about the CNC software can be found in chapter 8 under: Operating the CNC software.
24
Operating the machine
7.1 Introduction to the operation
Once you have made sure that all cables and hoses are correctly connected, you
can switch on the machine by means of the master switch at the rear of the
machine. For a milling job, you only need the discs you want to use, as well
as the corresponding tools. As the machine is operated via the software, all
the additional information regarding the operation of the machine can be found
in the software manual. The machine is equipped with a safety contact and
safety interlock at the front flap. Given the safety interlock, the front flap
cannot be opened during the milling procedure.
The machine may only be operated if the safety interlock is activated.
7.2 Extraction system
You can find all information regarding the operation, installation and
maintenance of the extraction system in the corresponding operating
instructions. Improper handling or use for purposes other than intended may
detrimentally affect the safe operation of the system.
Operate the machine only with the extraction system switched on and make sure
that the connection of the extraction system is not obstructed. Otherwise,
flying shavings may reach sensitive parts of the machine, such as bearings or
ball screws, and cause damage! It is imperative to set the extraction system
to the maximum level for operation.
The machine is equipped with a negative pressure sensor to measure the
extraction output. Make sure that the extraction system is switched on and
that the connection of the extraction system is not obstructed. The integrated
sealing air concept of the work chamber prevents dust and shavings from
entering the mechanical and electronic components of the machine and thus
reduces wear and maintenance efforts.
25
7.3 SFS 300P synchronous spindle
For the operation, installation and maintenance of the spindle, the provisions
of occupational safety apply. Improper handling or use for purposes other than
the ones intended considerably compromise the safety during operation!
The SFS 300P synchronous spindle is a precision component, which must not be
subject to the use of force, e.g. striking! Remember that the submersion depth
and travel speed must not be too high during processing. This may result in
the work pattern being ripped out and the tool getting jammed, which may cause
irreparable damage to the machine and/or the spindle. For optimum milling
results, use the milling strategies provided in Zenotec CAM from Wieland
Dental + Technik. If other or modified milling strategies are used, Wieland
Dental + Technik will not accept any liability. Do not use tools with a large
unbalance (single-blade cutters, gravers) in combination with a high spindle
speed. The unbalance puts very high strain on the ball bearings, which may
cause damage to the bearings.
Please note that wet grinding will reduce the service life of the spindle due
to the harder materials processed and the higher spindle speed.
7.4 Tools
The tools for the individual materials are predefined in the CNC software (see
chapter 8.1.6 and 8.2.2 ).
For operation, it is imperative that the corresponding tools are inserted in
the respective, defined slots.
In addition to the standard cutters, what is known as sister tools can be
inserted into the tool holder as an option. If a standard tool has reached its
wear limit, the corresponding sister tool is automatically used for further
milling.
26
7.4.1 Cutters
The standard tools for the dry processing of wax, PMMA, zirconium dioxide,
CoCr and model material are listed in the table below.
Please use the designations and article numbers of the tools for your orders.
F flat cutter C conic
N non-precious alloys B ball
Figure 6: Cutters for dry processing
Material (Dry processing) Wax PMMA Zirconium dioxide
Model
CoCr
Designation
Zenotec special cutter 0.3 select Zenotec basic cutter 0.7 select Zenotec
basic cutter 1.0 select Zenotec basic cutter 2.5 select Zenotec special cutter
1.5 F select Zenotec special cutter 2.4 F select Zenotec special cutter 0.6 N
mini / select Zenotec special cutter 1.0 N mini / select Zenotec special
cutter 1.5 N mini / select Zenotec special cutter 2.0 N mini / select
Article Number
662917 662916 662915 662914 662920 662921 679927 679928 679929 679930
- Only suitable for the dry processing of Zenotec® NP material and for the following disc heights: 10 mm, 12 mm, 13.5 mm, 15 mm Indications: Crowns and bridges with up to 6 units (ISO 22674, type 3 alloy)
7.4.2 Grinders
The standard tools for the wet processing of glass-ceramics are listed in the table below. Please use the designations and article numbers of the tools for your orders.
Material (Wet processing)
Lithium disilicate glass-ceramic block
Designation
Zenotec special grinder 0.5 C select Zenotec special grinder 1.0 B select
Zenotec special grinder 2.0 B select Zenotec special grinder 3.0 B
select
Article Number
667023 667019 667021 667022
** Only suitable for the wet processing of IPS e.max CAD for Zenotec / IPS Empress CAD for Zenotec
Figure 7: Grinders for wet processing
27
7.5 Changing discs (Zenotec select hybrid without material changer)
The discs are secured in the clamping fixture by means of a clamp washer. When the machine is in its home position, the disc can be changed as follows (See Figure 8).
1. Loosen the 4 screws (A).
2. Rotate the clamp washer (B) and lift it off.
A
The screws may remain in the clamping fixture.
The disc can now be exchanged.
3. Position the clamp washer (B) and centre it by
aligning it with the screws (A).
4. Position the screws (A) and tighten them
crosswise.
B
Figure 8: Changing discs at Zenotec select hybrid without material changer
Tightening the screws too much may result in tensions, which might damage the disc!
7.6 Material changer (Zenotec select hybrid with material changer)
7.6.1 Loading the disc holder
1. Loosen the 4 screws (A) at the disc holder. 2. Position the disc. 3. Turn
in the hold-down bracket (B). 4. Position the screws (A) and tighten them
crosswise.
A B
Figure 9: Changing discs at Zenotec select hybrid with material changer
Tightening the screws too much may result in tensions, which might damage the
disc!
Make sure to mark the position of the disc in the disc holder in order to
avoid tool fractures and damage to the spindle when inserting the disc again.
Only insert the material intended for the respective milling job into the disc
holder. Inserting the wrong material may result in tool fracture or increased
load on the spindle.
28
7.6.2
Mounting the disc holder
To operate the material changer in the software, please observe the
corresponding chapter in the software manual.
The front flap of the material changer is equipped
with a safety contact.
The material changer can only move if the flap is
closed.
E
If the machine is switched on and both front flaps are closed, you can position the disc holder in the material changer as follows:
1. Select a disc position in the software.
The material changer turns until the holder
changing position is reached.
2. Open the front flap of the material changer.
D
3. Hold the disc holder by its grip area (C) opposite
the clamping pin (D).
Risk of crushing by movement of the clamping lever! Hold the disc holder only by its grip area. Do not hold the disc holder in the area of the clamping pin!
C
G F
Figure 10: Functions of the material changer (only Zenotec select hybrid with
material changer)
4. Press the holder release (E) and keep it pressed down.
The clamping lever (F) at the holder changing position (G) opens.
5. Insert the disc holder at the holder changing position (perpendicular
position). 6. You can now let go of the holder release.
The clamping lever closes and secures the disc holder.
29
7.7 Procedure in case of a power failure (Zenotec select hybrid with material
changer)
If a short power failure occurs while the material changer is not in use, it
is sufficient to just switch on the machine again.
If a power failure occurs during processing, however, follow the instructions
of the software to enable referencing of the machine. In case of a prolonged
power failure or a defect in the power line, you can deactivate the safety
interlock with the emergency key to remove the disc from the machine.
If the power supply is interrupted while the material changer is in use, it is
not sufficient to just switch on the machine again. In such a case, the disc
changer would trigger an emergency-off signal before the referencing of the
machine. This is a safety feature to prevent damage to the machine.
If the machine fails while the material changer is in use, the following steps
are required to bring the machine back to its home position.
1. Switch off the machine using the master switch and
disconnect the machine from the power supply.
2. Set the pressure at the maintenance unit to 0 bar/psi or
disconnect the compressed air line from the main line.
3. Open the front flap of the material changer.
4. Insert the key for the emergency-release of the front flap into
the opening on the right of the material changer LED.
5. Deactivate the safety interlock of the front flap by turning the
key in an anti-clock-wise direction until it stops.
The front flap of the working chamber can now be opened.
6. Open the front flap of the working chamber.
7. Open the gripper and remove the disc holder, provided it is
not locked in place in the milling chamber (see Figure 13).
Figure 11: Deactivate safety interlock
Figure 12: Open the Zenotec select hybrid front flap
Figure 13: Closed gripper (Zenotec select hybrid with material changer)
30
8. Close the gripper (Figure 14).
If you cannot reach the gripper, move the release slide towards the material
changer until you can reach the gripper.
Slide the gripper to the left stop only while it is closed to prevent damage to the machine!
9. Hold the slide flap open with one hand and push the release
slide towards the material changer with the other hand until it stops.
Figure 14: Opened gripper (Zenotec select hybrid with material changer)
10. Push the gripper away from the working chamber up to the left stop.
11. Open the gripper.
The release slide is in its home position.
12. Close the front flap of the working chamber.
13. Activate the safety interlock of the front flap by turning the key
in a clock-wise direction until it stops.
Make sure that the safety interlock has again been activated. The machine may
only be operated if the safety interlock is activated!
14. Switch on the machine again.
The machine executes a reference run and is then again ready for operation.
Figure 15: Close front flap (Zenotec select hybrid)
Figure 16: Activate safety interlock of the front flap
31
Operating the CNC software
The CNC software can be directly started as an independent program and is also
used as the machine control program. With this software, the data calculated
by the CAM software are transmitted to the machine.
By right-clicking the button in the software, the function of the button is
indicated in the information bar at the bottom of the window.
8.1 Milling job
This is the main page of the program and it is displayed when the program is
started.
It shows the details of the milling job, an overview of the tools in the tool
changer as well as various indicators and functions, which will be explained
in more detail below.
32
8.1.1 Details of the milling job
Button
Function
Milling of the selected job The milling job can be started with this button.
If the status indicator of the selected job is green, all the details for this
job are correct (inserted tools, disc height, etc.) and milling can be started
immediately. If the indicator is red, no suitable tools have been allocated
for this job.
Milling cannot be started if the indicator is red!
Air-cleaning of the disc after milling With this button, you can define
whether or not the milled object will be blasted with compressed air to remove
rough milling dust. Once the spindle has deposited the tool, it moves over the
work pattern and air-cleans it via the air jet of the machine as well as the
air jet for cone cleaning.
Water pump This button activates (blue) or deactivates the water pump for wet
grinding. The water pump is automatically controlled by the software during
milling.
Deleting a job The current job can be deleted using this button. The job will
be completely deleted from the software and cannot be restored.
33
8.1.2 Tool changer
The picture above shows the allocated tools in the tool manager of the CNC software. The sequence of the tool positions shown in the CNC software corresponds with the actual sequence of the positions in the machine. The tools required for the individual milling jobs are marked by the green status indicator. The progress bars represent the operating counter, while the number behind it indicates the respective operating hours. As soon as the defined maximum service life of a tool has been reached, the indicator turns red. If a sister tool has been placed at another position, it will automatically be used.
Button
Function
Replacing tools, resetting the operating counter Once you have replaced a tool
by a new one, the operating hours can be reset to the maximum value by means
of the icon behind the operating hours of every tool.
Only reset the operating counter if you have replaced the worn down tools by
new ones.
Loading the tool magazine The tool magazine can be loaded via this symbol. It opens a window with the positions of the tool magazine. They have already been selected for the respective allocated positions. Simply select the positions you would like to have loaded. It is also possible that you have to select one preselected position after the other.
34
8.1.3 Functions
Depending on the machine type connected (e.g. machines with or without material changer), buttons that are not required remain invisible. Most of the functions have to be manually performed only during maintenance work.
Button
Function
Machine status Connected with the machine.
Machine status Not connected.
Machine status Machine running.
Machine status Machine hood open.
Machine status Machine idle.
Machine status Emergency-off activated.
35
Home position If this button is activated, the machine returns to its home
position. Here, the discs are changed and the tool changer can be loaded.
Extraction system The extraction system can be switched on and off by means of
this button.
The extraction system is automatically controlled by the software during
milling.
Cleaning the working chamber The spindle and the disc holder are moved into a
position which renders the working chamber easier to access for a cleaning
procedure.
Tool removal / spindle maintenance The spindle automatically runs to the
centre of the processing chamber and opens the collet after confirmation.
The tool must be removed manually. Depending on the degree of contamination,
the collet must be cleaned in the respective intervals. As soon as the collet
is opened, the tool drops and can be damaged!
Use special caution when changing the tools manually! Cutting injuries or
burns by the tools may occur during tool removal.
36
8.1.4 Inlets
Button
Function
Collet The collet can be manually opened and closed by means of this button.
Furthermore, the status of the collet is indicated (black icon = closed / blue
icon = open).
Sealing air Activates (blue button) or deactivates the sealing air. The sealing air is automatically controlled by the software and prevents shavings from entering the spindle.
Air jets Activates (blue button) or deactivates the air jets. The air jets are automatically controlled by the software during milling.
Cone cleaning Activates (blue button) or deactivates the cone cleaning. For that purpose, air is automatically blown through the open collet during tool change.
Water pump Activates (blue button) or deactivates the water pump for wet grinding. The water pump is automatically controlled by the software during milling.
Material contact of the spindle Indicates the material contact of the spindle (blue button). This is only for your information. This symbol can light up in various situations, such as during calibration, etc.
Tool measuring sensor Indicates the activation of the button by the tool (button turns blue). This is only for your information. The measuring sensor button lights up during tool length measurement as well as during tool fracture monitoring.
Pressure sensor Indicates the air pressure.
Negative pressure sensor Indicates the existence of the required negative pressure in the working chamber.
Milling or grinding is only possible with the extraction system switched on!37
8.1.5 Operating the material changer
Mounting the material holder By clicking the desired position in the material
changer, the selected position is moved to the material holder changing
position. The current position of the material holder changing position is
indicated in green.
The mounting of the material holders is described in detail in chapter 7.5 and
7.6.
Loading the material holder into the working chamber There are two methods to load the desired material into the working chamber:
Button
Function
Start symbol With this button, the material holder is loaded into the working
chamber and the desired milling job is started.
Loading / unloading the material holders By clicking the loading/unloading
button and a subsequent click on the desired material changing position, the
material holder is loaded into the working chamber or unloaded accordingly.
After that, the respective position number is displayed in blue in the centre
of the material changer.
Virtual loading / unloading of discs When loading discs virtually, you assign
the disc holder currently loaded in the working chamber a certain position in
the material changer, without the Zenotec select hybrid actually loading it in
the allocated position. When unloading discs virtually, you set the status of
the working chamber in the CNC software to “Empty”, without the Zenotec select
hybrid actually unloading the disc holder.
Before loading a disc virtually, make sure there is no disc holder in the
respective material changing position.
Improper virtual loading may result in damage to the Zenotec select hybrid.
38
8.1.6 Tool manager
The tools that have been successfully generated are shown on the left side of
the tool manager. The right side shows the tool magazine of the machine.
The tools may be physically dragged from the left side and dropped into the
tool magazine on the right side (keeping the left mousebutton depressed). In
this way, the tool changer can be loaded in the software. You can generate
three magazines one for zirconium tools, once for resin tools and a mixed
magazine.
39
Button
Function
Removing tools from the tool magazine If a tool is no longer required, you can
return it to the Library using the arrow symbol. Once there, it can be removed
via the recycle bin symbol.
Replacing tools / Resetting the operating counter If you replace a worn down
tool by a new one with identical tool geometry, you can reset the counter for
the operating hours to the respective maximum value.
New tool The software can manage up to thirty tools. Use this icon to add
tools.
40
8.1.7 Milling protocol
The most important data of the milling jobs are logged in the milling
protocol. This includes the job ID, any possible error codes, start/
completion time, milling duration and material type.
Button
Function Restoring a job Milling jobs can be restored within the defined filing period and can be milled again.
41
8.2 Software settings
8.2.1 User-defined program settings
Language
User profile
Cleaning reminder Number of days of data storage Mark disk position
Tool measurement with lower tolerance (WDT tool tolerance)
Selecting the program language.
After a change, the program has to be restarted.
The standard user profile is “User”.
The service mode is only used by service technicians.
If this option is activated, a dialogue window with the reminder that the
machine should be cleaned appears once a day at the most.
Define how long the data of completed milling jobs should be stored. After the
indicated period, the data will be deleted.
If this function is activated, the discs are marked at the disc edge. This
facilitates reclamping the disc in the same position. By clicking this button,
the tool measuring tolerance is set to 0.05 mm. By default, the tool measuring
tolerance is set to 0.4 mm. The activation of this button is recommended when
processing Zenotec NP or other grinding materials in order to avoid tool
fractures.
42
Button
Function Port number Setting-up of the USB connection between the machine and the computer upon connecting the machine.
8.2.2 CNC tool allocation
Here, all the tools are listed that are offered by Wieland Dental + Technik for the connected machine. If a new tool is added to the tool manager, it will appear in the drop-down menu shown above.
43
8.2.3 Machine geometry
All possible technical data regarding the connected machine are indicated under machine geometry. This includes, for example, the measuring sensor position, tool position and the position of the axes. These data are used either by service technicians or the machine, for example, for the axis compensation.
Button
Function
Exporting the machine geometry data
With this button, the machine geometry data can be exported to a text file.
This file is then stored in the installation directory of the software.
Manual tool change
Manual tool change can be activated for the respective position by means of
the hand symbol. In this way, the selected position can be manually loaded
with tools, for example, in case of defective rubber inserts in the tool
magazine. If the manual tool change is activated for a position, the
respective hand symbol is shown in blue.
44
8.2.4 Machine control
Button
Function
Milling a calibration corpus
A calibration corpus is milled to determine deviations of the spatial axis
position, which is subsequently adjusted by software compensation (see chapter
8.3).
Drilling tool changer inserts
With the help of this button, new rubber inserts for your tool changer can be
drilled, if the previous inserts are worn down. The procedure is described in
chapter 8.4: Drilling fixing plates for tool changer.
Displaying the dongle Shows the details of the dongle used.
These details are only required when requested by the service technician.
45
Searching for updates This button can be used to search for software updates.
Please note that this function requires an internet connection.
Firmware version The current firmware version is shown when this button is
clicked.
8.3 Milling a calibration corpus
In order to mill a calibration corpus, select “Program settings” in the
“Settings” window and click the “Milling calibration corpus” button. The disc
position must be manually selected in advance. The software will then guide
you through the process.
Button
Function Milling a calibration corpus By clicking the button, the already
calculated milling file for the calibration corpus is loaded as a milling
order.
Before the start, make sure that a 10 mm calibration disc has been inserted
and that the required tools have been loaded.
After clicking the milling calibration corpus button, the following screen is shown:
46
You can now start the milling procedure for the calibration corpus by means of
the start button. Subsequently, the progress indicator will show a list of the
processes being executed. After the milling procedure, remove the calibration
corpus from the machine and measure the indicated points with the help of the
micrometer gauge that is included in the delivery. Enter the measured values
in the corresponding fields of the following window and confirm to continue.
Calibration should only be conducted by trained personnel. Make sure that the
values entered are correct, as the machine executes the axis compensation on
the basis of the values entered! A saved calibration cannot be cancelled.
47
8.4 Drilling fixing plates for tool changer
Before you start the procedure, return the machine to its home position.
Loosen the four Allen screws and remove the cover plate of the tool changer.
Remove the four worn rubber inserts of the tool changer, position the new
undrilled inserts and replace and secure the cover plate.
Click the button “Drilling tool changer insert” in the window “Settings”
“Program Settings”. The software will then guide you through the process.
Button
Function Drilling fixing plates for tool changer The fixing plates for tool changer are drilled by clicking this button.
Once the button is clicked, the following screen is shown.
Start the drilling procedure using the start button. The procedure can be
aborted with the Stop symbol. The machine switches to the emergency mode and a
prompt appears, whether or not the procedure should be continued or stopped.
A window opens, in which you are prompted to manually insert the drilling tool
(2.8 mm drill) into the spindle.
48
Continue to hold the tool. Confirm the next window to continue and close the
chuck. Remove your hand from the milling chamber and confirm the notice
“Next”.
Subsequently, the progress indicator will show the drilling procedure in the
form of a list of processes being executed.
Once the drilling procedure has been completed, the drilling tool has to be
manually removed. Confirm the following dialogue to open the collet chuck in
order to remove the tool.
49
Maintenance and cleaning
Maintenance and cleaning are necessary to ensure smooth operation of the
machine. Furthermore, cleaning positively affects the service life of your
tools.
Wieland Dental + Technik reserves the right not to accept liability for
warranty claims in case of failure to observe the cleaning procedures. Please
observe the prepared cleaning plan in the Appendix and also use it as a
monitoring instrument.
9.1 Cleaning and maintenance cycles
The following cleaning and maintenance procedures have to be performed:
Component Disc holders Measuring sensor
Collet
Interval Prerequisite/remark
Daily (more frequently, if required)
Weekly (more frequently, if required)
Maintenance unit
Weekly
Material changer
Weekly
Housing
Monthly
Calibration of the axes
Only, if required!
Maintenance by Wieland Dental + Technik Service
Annually
Working chamber
Contaminated interior
Out-of-round running of the spindle, poor milling results Condensate in the
water separator, filter cartridge dirty Cleaning with a vacuum cleaner and
brush Cleaning with a cloth and, if necessary, a cleaning agent
Contact your Service Centre. The corresponding addresses can be found in the
jacket.
50
9.2 Cleaning of the interior (working chamber, disc holder, measuring sensor,
material changer)
Never use compressed air to clean the interior, since this may result in
shavings reaching sensitive components, such as the linear guides or the
spindle bearings. Clean the machine regularly to prevent dirt or shavings from
accumulating in the guide rails and other sensitive machine components. Use
only a vacuum cleaner and brush for cleaning.
If you only use a brush for cleaning, switch on the extraction system in the
software so that the shavings are immediately removed.
Cleaning of the working chamber and disc holders Make sure that you remove all
material residues in the working chamber with the brush and/or vacuum cleaner.
Cleaning of the measuring sensor The measuring sensor should be separately
cleaned using a brush to remove all contaminants.
Cleaning of the material changer (only Zenotec select hybrid with material
changer)
To clean the material changer, remove all the disc holders from the machine
and clean them. Clean the interior of the material changer using a vacuum
cleaner and a moist cloth. The release slide is best reached by lifting the
slide flap from the working chamber.
Use the cleaning function in the software for that purpose.
51
9.3 Cleaning the spindle / collet
Only use the original spindle maintenance set for cleaning the spindle or the
collet chuck.
Figure 17: Spindle maintenance set
Cleaning of the collet is described in detail in the Instructions for Use of
the Maintenance Kit.
Never use compressed air to clean the spindle, as this will damage the
bearings of the spindle.
Clean the collet once a week as follows: 1. Click the button “Remove tool” in
the CNC module of the software. 2. Confirm the prompt to open the collet. 3.
Open the front flap. 4. Manually unscrew the collet using the knurled screw of
the Spindle Maintenance Kit.
Make sure that a measuring pin is in place while unscrewing the collet to
avoid a deformation of the collet.
5. Clean the inner cone of the spindle with the cleaning cone of the
Maintenance Kit. 6. Clean the collet using the brush of the Maintenance Kit
and make sure that the sealings in the collet do not
fall off. 7. Take a small quantity of the collet grease from the Maintenance
Kit and apply it on the slopes of the collet. Make sure that no grease enters
the longitudinal slots and the interior of the collet and use only the collet
grease supplied with the Maintenance Kit to avoid damaging the spindle.
52
8. Screw the collet into the spindle until it stops using the screw-in piece.
Make sure that a measuring pin is in place while screwing the collet into the
spindle to avoid a deformation of the collet.
9. Click the button “Basic settings”.
9.4 Cleaning of the exterior (housing)
Basically, powder-coated surfaces should be cleaned with a soft, dry cloth in
a first step. If any contamination cannot be removed in this way, the cloth
may also be moistened. If necessary, a cleaning agent with a neutral pH may be
used.
Basically, contact of alkaline or acidic substances with the powder coating
must be prevented. Metallic powder coating is particularly sensitive. During
cleaning, particular attention should be paid to the stick-on label to make
sure that it is not peeled off. The adhesive letters are particularly
sensitive to friction and strong cleaning agents.
9.5 Compressed air maintenance
Check on a weekly basis if water has accumulated in the water separator and
empty the container, if necessary, by opening the discharging screw (see
Figure 4 on Page 21). The condensate is then blown out under pressure.
Condensate in the separator usually indicates that the compressed air is
insufficiently dry. In this context, please observe our requirements regarding
air purity in chapter 5. In case of severe contamination, the filter cartridge
in the water separator must be cleaned or replaced. A severely contaminated
cartridge may lead to a decline in the compressed air output.
53
1. Disconnect the machine from the compressed air supply. 2. Unscrew the bowl of the water separator. 3. Unscrew the filter screw below the filter cartridge. 4. Pull out the filter cartridge. 5. Check the filter cartridge, clean it or replace it, if necessary. 6. Insert the new or cleaned filter cartridge.
If the compressed air meets the purity requirements of ISO 8573-1, the filter
cartridge usually does not have to be replaced. If the filter cartridge is
contaminated, please check the purity level of your compressed air.
9.6 Calibration
The machine is supplied in a calibrated condition. As long as your milling
results are impeccable, renewed calibration is not necessary. If inaccuracies
occur, however, calibration is required. In case of a problem, please consult
a service technician before you recalibrate the machine. Calibration is time-
consuming and may lead to poorer milling results if performed incorrectly.
Therefore, check the operating conditions first. Check the fixation of the
work pattern, the condition of the tool and the material to be processed
before you execute a calibration procedure. Details on conducting the
calibration procedure can be found in the Software Manual.
9.7 Machine maintenance (by Wieland Dental + Technik Service)
Wieland Dental + Technik recommends having a service technician regularly
perform maintenance on the machine. Machine maintenance should be conducted at
least once a year.
54
10 Notes
Safety notes
The Zenotec select hybrid milling system has been designed based on the
following standards:
Applicable EC Directives – 2006/42/EC Machine Directive – 2004/108/EC EMC
Directive
Applied harmonized standards – EN 61326-1:2006 – EN 61326-2-1:2006 – EN 61000-3-2:2006 + A1:2008 + A2:2009 – EN 61000-3-3:2008
– IEC 60204-1:2005 – IEC 61010-1:2010 – EN ISO 12100:2010
Radio protection, electromagnetic compatibility: – EMC tested
10.1 Acceptable operating conditions
Acceptable ambient temperature range: The room temperature should be between
18 °C (64.4 °F) and 25 °C (77 °F); maximum room temperature 32 °C (89.6 °F)
Acceptable humidity range: Max. relative humidity 80% for temperatures up to
31 °C (87.8 °F), condensation excluded. Acceptable ambient pressure: The
furnace is tested for use at altitudes of up to 2000 m above sea level.
10.2 Acceptable transportation and storage conditions
Acceptable temperature range: 10 °C to 35 °C (50 °F to 95 °F) Acceptable
humidity range: Max. 80% relative humidity Acceptable ambient pressure: 500
mbar to 1060 mbar Use only the original packaging together with the
corresponding foam material for shipping purposes.
55
11 Zenotec hydro
56
12 List of parts (Zenotec hydro)
A B C E D
No.
Component description
A
Control cable for membrane pump
B
Power pack
C
Power cord
D
3 Fine filter cartridges
E
3 Coarse filter pads
F
Fluid hose
G
Assembly tool for filter change
H
Metal sealing plug
I
Set of rollers
K
Connecting hose
F
G
K I H
57
12.1 Indications of Zenotec hydro
This manual enables you to use the Zenotec hydro from Wieland Dental +
Technik, as well as the corresponding accessories, in a safe and reliable way.
Please read these Operating Instructions carefully and observe the notes
contained therein. Should you have any questions, please contact Wieland
Dental + Technik directly. Zenotec hydro may only be operated with the housing
and in conjunction with the Zenotec select hybrid models (with/without
material changer) and Zenotec cyclon from Wieland Dental. Only the materials
listed in these Operating Instructions may be processed.
When using Zenotec hydro, a residual current-operated protective device (RCD)
must be installed in the electric circuit of Zenotec select hybrid, Zenotec
cyclon and Zenotec hydro.
Do not move the Zenotec hydro during operation! If the wet-grinding module is
equipped with rollers, the roller brakes must be activated when operating the
wetgrinding module.
If the machine is used for any purpose other than the above designated use,
the system may be damaged. The protective equipment must not be circumvented
or suspended. The indications also include observing these Operating
Instructions and following the maintenance instructions therein. In order to
ensure product safety and the validity of the warranty, the product must be
exclusively operated with the original accessories from Wieland Dental. The
user bears the risk of using nonapproved accessories. Zenotec hydro must be
operated exclusively with the lubricating coolant Synergy 905. Please also
observe the notes regarding the disposal of the milling emulsion (see Product
Data Sheet). Should you make any modifications to the device or remove the
housing of the electronic controls without prior written consent by a Wieland
Service Centre, all warranty claims are void. Zenotec hydro may be operated
unsupervised. In this context, please observe the conditions for the
unsupervised operation of Zenotec select hybrid and the notes on operating
Zenotec hydro.
The space where the pump and controls of the Zenotec hydro are located must
remain dry. Remove any accumulated fluid immediately.
58
12.2 Warnings
Use of damaged cables Mortal danger by electric shock!
Disconnect the machine from the power source and protect the machine from
accidental operation. Consult Customer Service. Replace the damaged cables
with original replacement cables.
Troubleshooting during operation Mortal danger by electric shock!
Consult Wieland Service before you rectify a malfunction. Disconnect the
machine from the power source and protect the machine from accidental
operation.
Operation of the Zenotec hydro without a residual current protective device
(RCD) Mortal danger by electric shock!
When using Zenotec hydro, a residual current-operated protective device (RCD)
must be installed in the electric circuit of Zenotec select hybrid and Zenotec
hydro. Consult Wieland Service if an RCD is not installed.
Incorrect installation of water-carrying lines Mortal danger by electric
shock!
Make sure that no lines containing water are installed across electrical lines
or electrical devices. Consult Wieland Service if an RCD is not installed.
Use of the wrong lubricating coolant Injuries to health!
Only use the lubricating coolant Synergy 905 recommended by Wieland Dental +
Technik.
59
Mess at the workplace Risk of stumbling and falling!
Keep the work station clean. Store the Zenotec hydro in a safe place. If the
Zenotec hydro is equipped with rollers, the roller brakes must be activated
during operation.
Liquids escaping at the work station Risk of stumbling and falling!
Keep your work station clean. Store the Zenotec hydro in a safe place. Wipe
off escaping liquid immediately.
Skin contact with lubricating coolant Erythema or irritation of the skin! Upon
contact of the lubricating coolant or milling emulsion with the skin,
thoroughly rinse under running tap water.
Drinking of lubricating coolant Nausea and diarrhoea! Store lubricating
coolant in clearly marked containers in order to avoid that it is confused
with foodstuffs.
Mounting of the module insert Danger of crushing!
When mounting the module insert, always grasp it by the handles. Take care not
to jam the finger of any other person when inserting the module.
Disposal The machine must not be disposed in the normal domestic waste. Please
correctly dispose of old devices according to the corresponding EU Council
Directive. Information regarding disposal can also be found on the respective
national Wieland Dental + Technik website. The packaging may be discarded with
the regular household refuse.
60
12.3 Transport and storage
Zenotec hydro must always be transported upright. Always transport Zenotec
hydro separately. Modules must not be stacked on top of each other. If the
liquid container is full, remove the liquid container and transport it
separately. The ambient temperature for storage and transport must be between
10 °C (50 °F) and 35 °C (95 °F). Store Zenotec hydro only with the liquid and
filter containers empty. Store the module insert always upright to prevent the
housing foot from breaking off.
Weight of the individual components: Housing: approx. 15 kg Module insert:
approx. 16 kg Accessories: approx. 3 kg
Keep the packaging of the module to use it if you have to send the module in
for maintenance or repair.
61
12.4 Components of Zenotec hydro
A B C D
E
62
F G H
I K L
No.
Component description
A
Handles
B
Module insert
C
Display
D
Housing
E
Housing foot / roller
F
Coarse filter drawer
G
Fine filter container
H
Membrane pump
I
Flat filter pad
K
Measuring station
L
Leg support
63
12.5 Technical data
12.5.1 Basic system Dimensions with housing feet (w/d/h): approx. 382 x 455 x
580 mm Dimensions with rollers (w/d/h): approx. 382 x 455 x 640 mm Empty
weight of the module: approx. 34 kg Separation of grinding dust and moisture
Coarse filter drawer for coarse dirt Removable liquid container Labyrinth air
cooling All metal parts are made of stainless steel Ultrasound sensor to
measure the liquid level Power pack with an output voltage of 24 V DC, 60 W
Input current membrane pump: 2A
12.5.2 Controls Integrated RGB display Automatic control of the membrane pump
via the software possible Integrated operating hours counter Liquid volume
indicator (in per cent) Deactivation of the membrane pump if the liquid level
is low
12.5.3 Filter Coarse filter pad for coarse dirt Flat filter pad Fine filter
cartridge with a pore size of 5 m
64
12.6 Synergy 905
Synergy 905 is a synthetic, water-soluble lubricating coolant that is free of
chlorine, mineral oil and ester oil. It can be used with soft and hard water
(0-20 dH).
Prevent the new or used water emulsion from entering the soil, water or the
sewage water system (see the Safety Data Sheet for more information).
12.7 Sound emission
The sound emission strongly fluctuates depending on the processed material and
the milling conditions. If the machine is too loud, check the milling
conditions. Examine the fixation of the work pattern, the condition of the
tool and the materials processed.
If loud operating noise cannot be prevented, use hearing protectors during
milling.
65
Installation
13.1 Conditions governing location
The location for the wet-grinding module requires the following conditions:
The conditions required for the location of Zenotec select hybrid also apply
to Zenotec hydro. The device should neither be placed nor operated in areas
where there is an explosion hazard.
Place the machine on a stable, even surface constructed to hold the weight of
Zenotec hydro. The room in which the module is set up must be well-ventilated
as humid air is released into the room. The room temperature should be between
18 °C (64.4 °F) and 25 °C (77 °F); maximum room temperature 32 °C (89.6 °F)
The machine has been designed for use in altitudes of up to 2000 m above sea
level Maximum relative humidity 80%, non-condensing A/C power connection with
100-240 V and 50 / 60 Hz
Use of damaged cables Mortal danger by electric shock!
Disconnect the machine from the power source and protect the machine from
accidental operation. When using Zenotec hydro, a residual current-operated
protective device (RCD) must be installed in the electric circuit of Zenotec
select hybrid, Zenotex cyclon and Zenotec hydro. Consult Customer Service.
Replace the damaged cables with original replacement cables.
66
13.1 Mounting the rollers
The wet-milling module is delivered ex works with stable feet. Wieland Dental
- Technik recommends mounting the rollers if the wetgrinding module is stored
readily accessible on the floor or beneath a table. In this way, the module is
easier to reach for refilling the liquid or if the filter needs to be
replaced.
During mounting of the rollers there is a risk of injury.
1. Remove the module insert and the liquid container from the housing. 2. Set
the housing aside so that you can reach the bottom side of Zenotec hydro. 3.
Remove the nuts and the housing feet. 4. Attach the rollers instead of the
housing feet. Make sure to attach the two rollers with the built-in brakes
pointing
to the front.
5. Mount each roller with four washers each and four nuts.
Make sure that the roller brakes are always engaged during operation.
After moving the Zenotec hydro, check all connections and supply lines for
possible damage.
67
13.3 Placing Zenotec hydro
Make sure that no electrical devices are located beneath the wet-grinding
module. Do not install any lines containing water across electrical lines or
electrical devices.
In order to prevent any suction performance loss, the connecting hose must not
be longer than 1.5 metres. This restricts the distance between Zenotec select
hybrid and Zenotec hydro.
M1.
Unpack Zenotec hydro. Please keep the packaging of the machine in case you need to send in the machine for
maintenance or repair.
M2.
Set up Zenotec hydro. The following images show possible set-up options.
The hookup of the connecting hose at Zenotec hydro may not be set up higher than the suction hose connection of the Zenotec select hybrid!
68
13.4 Connections of Zenotec hydro
Use only the original cables provided with the system for the connections at
the switching outputs. Make sure to connect the liquid hose to the socket
intended for that purpose.
Do not set Zenotec hydro into operation before you have made sure that all the
cables and hoses are correctly connected.
E F
A G
B
H
I
C
D
No.
Component description
A
Switching output for membrane pump
B
Liquid connection for the membrane pump
C
Extraction connection
D
Liquid discharge (closed, as not used)
E
Connection for the extraction hose
F
Hookup for the connecting hose
G
Liquid connection
H
Switching input for membrane pump
I
Connection for the power pack
69
13.5 Mounting the components
The connections described in this paragraph may also remain connected when
Zenotec select hybrid is used in the dry-milling mode. Attach the components
with the help of the table below. The letters of the connections refer to the
connection overview.
Component Control cable
Connection Zenotec select hybrid
Connection Zenotec hydro
Switching output external pump (A)
Switching input for membrane pump (H)
Picture
Liquid hose
Liquid connection (B)
Liquid connection (G)
Metal sealing plug
Liquid discharge (D)
When removing or replacing the liquid hose, always disconnect the hose from the Zenotec hydro first. Hold the end of the hose over a liquid collection container to drain the hose and prevent a backflow of liquid into the milling machine.
70
13.6 Installing the connecting hose
The connecting hose must not be allowed to sag, as this will result in liquid
accumulating in the hose and thus reduce the suction performance.
13.6.1 Length of the connecting hose Connect the connecting hose with the
suction opening of the Zenotec select hybrid and hold the hose to the hookup
for the connecting hose at the Zenotec hydro for verification. Check if the
hose sags. If it does, shorten the connecting hose. 13.6.2 Shortening the
connecting hose Only if the connecting hose sags!
Mark the area where the connecting hose reaches the hookup at the Zenotec
hydro. Add approximately 3 cm excess length. Cut the connecting hose at the
marked area using a sharp knife. Connect the connecting hose with the Zenotec
hydro. 13.6.3 Securing the connecting hose at the Zenotec hydro Loosen the
screw-type hose connection of the Zenotec hydro. Push the connecting hose
through the screw-type hose connection with a distance of approximately 3 cm.
Make sure that the black sealing ring does not fall off.
Hold the connecting hose in place with one hand and tighten the screw-type hose connection with the other hand.
71
13.6.4 Filling in Synergy 905 Make sure that the Zenotec hydro is disconnected
from the power supply when filling the liquid container.
1. Remove the module insert. 2. Fill the liquid container (max. capacity 22
litres) with 1.5 litres of Synergy 905 and 20.5 litres of tap water. 3. You
now have 22 litres of an emulsion consisting of lubricating coolant and water
(7% solution). 4. Replace the module insert. 5. Connect the power pack with
Zenotec select hybrid and the power supply. 6. Zenotec hydro is now ready for
operation.
13.6.5 Refilling Loss of water must be expected during milling.
Regularly check the liquid level of the Zenotec hydro. Make sure to observe
the indications on the display in this context.
Generally, the liquid container is topped up with pure water. When topping up
while the extraction system is running, the water can also be filled directly
into the milling chamber.
You may top up the liquid container with water until you are reminded by the
Zenotec CNC software that the filter needs to be replaced.
72
Operating Zenotec hydro
14.1 Switching from dry operation to wet operation
Before every wet-grinding procedure, make sure that Zenotec hydro is ready for
operation. Operate the Zenotec hydro always with the Zenotec cyclon switched
on to prevent damage to the machines.
When switching the operation mode, follow the workflow of the Zenotec CNC
software.
14.1.1 Zenotec select hybrid without material changer Remove the disc holder
from the milling chamber. Clean the entire milling chamber of the Zenotec
select hybrid with a vacuum cleaner and a brush, if necessary. Ensure that
there is no dust left in the milling chamber after cleaning.
Remove the suction hose from the Zenotec select hybrid. Attach the suction
hose to the connection of the Zenotec hydro intended for this purpose. Connect
the Zenotec hydro connecting hose with the Zenotec select hybrid. Empty the
dust drawer of the Zenotec cyclon.
Figure 18: Zenotec cyclon dust drawer Check the liquid volume at the Zenotec
hydro screen and top up to 100% if necessary.
You must check the coolant circuit before you start milling. Insert IPS
e.matrix®.
73
14.1.2 Zenotec select hybrid with material changer Remove all the disc holders
from the material changer. Clean the entire milling chamber of the Zenotec
select hybrid with a vacuum cleaner and a brush, if necessary. Remove the
suction hose from the Zenotec select hybrid. Attach the suction hose to the
connection of the Zenotec hydro intended for this purpose. Connect the Zenotec
hydro connecting hose with the Zenotec select hybrid.
Wet processing
ENOTEC cyclon Empty the dust drawer of the Zenotec cyclon.
Cooling liquid Air Dust
Figure 19: Zenotec cyclon dust drawer
Check the liquid volume at the Zenotec hydro screen and top up to 100% if
necessary. Insert IPS e.matrix® in the material changer.
74
14.1.3 CNC workflow From dry to wet
ZENOTEC – From dry to wet
Remove disc-holder from the material changer
ZENOTEC – From dry to wet
Clean milling chamber follow the steps described in instructions for use
ZENOTEC – From dry to wet
Remove suction hose from the Zenotec select and connect to Zenotec hydro
ZENOTEC – From dry to wet
Connect the hose of the Zenotec hydro with the Zenotec select hybrid
ZENOTEC – From dry to wet
Empty dust drawer of the Zenotec cyclon
ZENOTEC – From dry to wet
Check liquid volume at the Zenotec hydro screen and refill if necessary
ZENOTEC – From dry to wet
Close the cover and wait until the button is active again. The coolant circuit
is checked for correct functioning.
ZENOTEC – From dry to wet
Insert IPS e.matrix© in the material changer
75
14.2 Switching from wet operation to dry operation
Before switching the operating mode, activate the suction system in order to
ensure that as much liquid as possible is removed from the milling chamber of
the Zenotec select hybrid.
Open the front flap of the Zenotec select hybrid. Remove the IPS e.matrix®
from the Zenotec select hybrid. Rinse the milling chamber and remove any
possible milling residue with the help of a spatula. Remove the remaining
liquid from the milling chamber of the Zenotec select hybrid using a cloth.
Only plastic spatulas should be used for cleaning. The use of e.g. metal
spatulas may lead to damage in the interior of the machine.
Remove the connecting hose from the Zenotec select hybrid. Remove the Zenotec
cyclon suction hose from the Zenotec hydro. Attach the Zenotec cyclon suction
hose to the Zenotec select hybrid.
Dry processing
ENOTEC cyclon
Cooling liquid Air Dust
Moist air or residual water in the milling chamber of the Zenotec select
hybrid may reach the electrical components of the machine and condensate
there! Therefore, it is imperative to activate the Zenotec cyclon during and
after wet grinding.
Leave the front flap open for at least 15 minutes for drying purposes.
76
14.2.1 CNC workflow From wet to dry
ZENOTEC – From wet to dry
Remove all IPS e.matrix© from the material changer
ZENOTEC – From wet to dry
Clean milling chamber follow the steps described in instuction for use
ZENOTEC – From wet to dry
Switch on the Zenotec cyclon and complete the 15 minutes drying process …
14:50
ZENOTEC – From wet to dry
Insert disc holder in the material changer
77
15 Display
The display of the Zenotec hydro provides information on the liquid volume, operating hours, version of electronic controls and the status of the membrane pump. Only use the Zenotec hydro if the display is fully operational.
B
C
A
D
E
Figure 20: Display of Zenotec hydro
No.
Component description
A
Liquid volume in %
B
Serial number of the Zenotec hydro electronic controls
C
Operating hours
D
Version of electronic controls
E
Status of the membrane pump
78
The display of Zenotec hydro changes its colour depending on the fill level.
Fill level [mm] 60 70 80 90 100 110 120 130 140 150 160 170 180 190 200 210 220 230 240 250 260
Quantity [litres] 4.8 5.6 6.4 7.2 8.1 8.9 9.7 10.6 11.4 12.2 13.1 13.9 14.8 15.7 16.5 17.4 18.3 19.1 20.0 20.9 21.8
Display
Figure 21: Zenotec hydro display options
79
16 Operating the CNC software
If you switch from dry operation to wet operation or vice versa, a separate
window opens in the Zenotec CNC software. It shows the defined workflow for
switching between the operation modes illustrated with pictures.
To switch from dry operation to wet operation or vice versa, follow the
illustrated work-flow of the working mode in the CNC software.
The workflow for switching from one operation mode to another can be found in
chapter 14.
80
Maintenance and cleaning of Zenotec hydro
Maintenance and cleaning are necessary to ensure smooth operation of the machine. Furthermore, cleaning positively affects the service life of your tools. Wieland Dental + Technik reserves the right not to accept liability for warranty claims in case of failure to observe the cleaning procedures. Please observe the prepared cleaning plan in the Appendix and also use it as a monitoring instrument. Always disconnect the wet-grinding module from the power source before you remove the module insert!
17.1 Cleaning and maintenance cycles
Component
Interval
Prerequisite/remark
Power lines / liquid lines
Once per week
Visual control
Liquid container
If the display lights up in yellow (more frequently, if required)
Top up with tap water
Clean the coarse filter
If required
Replace the filter cartridge
After 70 h or if the flow rate is too low (notice CNC)
Replace liquid
Zenotec select hybrid working chamber Zenotec hydro maintenance by Wieland
Dental + Technik Service
incl. filter
If required (at least when switching from wet grinding to dry milling and vice
versa)
Every 2 years
Zenotec select hybrid flow rate drops
Cleaning with a vacuum cleaner and brush
Cleaning with a cloth and, if necessary with cleaning agent
Contact your Service Centre. The corresponding addresses can be found in the
jacket.
81
17.2 Cleaning the coarse filter and liquid container
1. Remove the coarse filter drawer and dispose of the residue. 2. Rinse the
coarse filter pad. As long as the pores of the coarse filter pad are not
clogged and liquid may pass
through, the pad may be reused.
3. Replace the clean coarse filter pad or place a new coarse filter pad in
the coarse filter drawer. 4. Clean the liquid container using a cloth.
17.3 Changing the fine filter cartridge
1. Use the assembly tool for the filter change and unscrew the fine filter
container by turning it clockwise. 2. Dispose of the liquid in the fine filter
container. 3. Remove the fine filter cartridge and dispose of it in a proper
manner. 4. Clean the fine filter container if it is dirty. 5. Insert a new
fine filter cartridge.
Check the filter cartridge and replace it, if necessary. Make sure to centre
the fine filter cartridge and resecure the fine filter container.
Disposal
Disposal The machine must not be disposed in the normal domestic waste. Please correctly dispose of old devices according to the corresponding EU Council Directive. Information regarding disposal may also be found on the respective national Wieland Dental + Technik website. The packaging may be discarded with the regular household refuse.
82
19 Appendix
Ivoclar Vivadent worldwide
Ivoclar Vivadent AG Bendererstrasse 2 9494 Schaan Liechtenstein Tel. +423 235
35 35 Fax +423 235 33 60
www.ivoclarvivadent.com
Ivoclar Vivadent Pty. Ltd. 1 5 Overseas Drive P.O. Box 367 Noble Park, Vic.
3174 Australia Tel. +61 3 9795 9599 Fax +61 3 9795 9645
www.ivoclarvivadent.com.au
Ivoclar Vivadent GmbH Tech Gate Vienna Donau-City-Strasse 1 1220 Wien Austria
Tel. +43 1 263 191 10 Fax: +43 1 263 191 111
www.ivoclarvivadent.at
Ivoclar Vivadent Ltda. Alameda Caiapós, 723 Centro Empresarial Tamboré CEP
06460-110 Barueri SP Brazil Tel. +55 11 2424 7400 Fax +55 11 3466 0840
www.ivoclarvivadent.com.br
Ivoclar Vivadent Inc. 1-6600 Dixie Road Mississauga, Ontario L5T 2Y2 Canada
Tel. +1 905 670 8499 Fax +1 905 670 3102
www.ivoclarvivadent.us
Ivoclar Vivadent Shanghai Trading Co., Ltd. 2/F Building 1, 881 Wuding Road,
Jing An District 200040 Shanghai China Tel. +86 21 6032 1657 Fax +86 21 6176
0968 www.ivoclarvivadent.com
Ivoclar Vivadent Marketing Ltd. Calle 134 No. 7-B-83, Of. 520 Bogotá Colombia
Tel. +57 1 627 3399 Fax +57 1 633 1663
www.ivoclarvivadent.co
Ivoclar Vivadent SAS B.P. 118 F-74410 Saint-Jorioz France Tel. +33 4 50 88 64
00 Fax +33 4 50 68 91 52
www.ivoclarvivadent.fr
Ivoclar Vivadent GmbH Dr. Adolf-Schneider-Str. 2 D-73479 Ellwangen, Jagst
Germany Tel. +49 7961 889 0 Fax +49 7961 6326
www.ivoclarvivadent.de
Wieland Dental + Technik GmbH & Co. KG Lindenstraße 2 D-75175 Pforzheim
Germany Tel. +49 7231 3705 0 Fax +49 7231 3579 59 www.wieland-
dental.de
Ivoclar Vivadent Marketing (India) Pvt. Ltd. 503/504 Raheja Plaza 15 B Shah
Industrial Estate Veera Desai Road, Andheri (West) Mumbai, 400 053 India Tel.
+91 22 2673 0302 Fax +91 22 2673 0301
www.ivoclarvivadent.in
Ivoclar Vivadent s.r.l. Via Isonzo 67/69 40033 Casalecchio di Reno (BO) Italy
Tel. +39 051 6113555 Fax +39 051 6113565
www.ivoclarvivadent.it
Ivoclar Vivadent K.K. 1-28-24-4F Hongo Bunkyo-ku Tokyo 113-0033 Japan Tel. +81
3 6903 3535 Fax +81 3 5844 3657
www.ivoclarvivadent.jp
Ivoclar Vivadent Ltd. 12F W-Tower, 1303-37 Seocho-dong, Seocho-gu, Seoul
137-855 Republic of Korea Tel. +82 2 536 0714 Fax +82 2 596 0155
www.ivoclarvivadent.co.kr
Ivoclar Vivadent S.A. de C.V. Av. Insurgentes Sur No. 863, Piso 14, Col.
Napoles 03810 México, D.F. México Tel. +52 55 5062 1000 Fax +52 55 5062 1029
www.ivoclarvivadent.com.mx
Ivoclar Vivadent BV De Fruittuinen 32 2132 NZ Hoofddorp Netherlands Tel. +31
23 529 3791 Fax +31 23 555 4504
www.ivoclarvivadent.com
Ivoclar Vivadent Ltd. 12 Omega St, Rosedale PO Box 303011 North Harbour
Auckland 0751 New Zealand Tel. +64 9 914 9999 Fax +64 9 914 9990
www.ivoclarvivadent.co.nz
Ivoclar Vivadent Polska Sp. z o.o. Al. Jana Pawla II 78 00-175 Warszawa Poland
Tel. +48 22 635 5496 Fax +48 22 635 5469
www.ivoclarvivadent.pl
Ivoclar Vivadent Marketing Ltd. Prospekt Andropova 18 korp. 6/ office 10-06
115432 Moscow Russia Tel. +7 499 418 0300 Fax +7 499 418 0310
www.ivoclarvivadent.ru
Ivoclar Vivadent Marketing Ltd. Qlaya Main St. Siricon Building No.14, 2nd
Floor Office No. 204 P.O. Box 300146 Riyadh 11372 Saudi Arabia Tel. +966 11
293 8345 Fax +966 11 293 8344
www.ivoclarvivadent.com
Ivoclar Vivadent S.L.U. Carretera de Fuencarral nº24 Portal 1 Planta Baja
28108-Alcobendas (Madrid) Spain Telf. +34 91 375 78 20 Fax: +34 91 375 78 38
www.ivoclarvivadent.es
Ivoclar Vivadent AB Dalvägen 14 S-169 56 Solna Sweden Tel. +46 8 514 939 30
Fax +46 8 514 939 40 www.ivoclarvivadent.se
Ivoclar Vivadent Liaison Office Tesvikiye Mahallesi Sakayik Sokak Nisantas’
Plaza No:38/2 Kat:5 Daire:24 34021 Sisli Istanbul Turkey Tel. +90 212 343
0802 Fax +90 212 343 0842
www.ivoclarvivadent.com
Ivoclar Vivadent Limited Ground Floor Compass Building Feldspar Close Warrens
Business Park Enderby Leicester LE19 4SE United Kingdom Tel. +44 116 284 7880
Fax +44 116 284 7881
www.ivoclarvivadent.co.uk
Ivoclar Vivadent, Inc. 175 Pineview Drive Amherst, N.Y. 14228 USA Tel. +1 800
533 6825 Fax +1 716 691 2285
www.ivoclarvivadent.us
Version: 2
Date information prepared: 01/2015 The apparatus has been developed solely for
use in dentistry. Start-up and operation should be carried out strictly
according to the Operating Instructions. Liability cannot be accepted for
damages resulting from misuse or failure to observe the Instructions. The user
is solely responsible for testing the apparatus for its suitability for any
purpose not explicitly stated in the Instructions. Descriptions and data
constitute no warranty of attributes and are not binding. © Wieland Dental +
Technik GmbH & Co. KG
83
LEADING DIGITAL ESTHETICS
Wieland Dental, a company of the Ivoclar Vivadent Group, was established in
Pforzheim in 1871 and is one of the world’s leading suppliers in the field of
dental technology. Future-ready integrated technologies and materials are the
hallmarks of our expertise and prowess. Wieland Dental opens up the way to
progress in dental technology. wieland-dental.de
Wieland Dental + Technik GmbH & Co. KG Lindenstraße 2 | 75175 Pforzheim |
Germany Fon +49 7231-3705 0 | info@wieland-dental.de
wieland-dental.de
673411 / EN / 2015-12-04 / Rev. 002
References
- apotheke.de - Ihre zuverlässige Online Apotheke
- Modern Dental Solutions for Dentists & Technicians | Ivoclar AU
- Ivoclar Österreich
- 이보클라 코리아
- Modern Dental Solutions for Dentists & Technicians | Ivoclar NZ
- Modern Solutions for Dentists & Technicians | Ivoclar GB
- Modern Dental Solutions for Dentists & Technicians | Ivoclar USA
- Modern Dental Solutions for Dentists & Technicians | Ivoclar AU
- ivoclarvivadent.com.br
- Ivoclar Latinoamérica
- Moderne Lösungen für Dentalexpertinnen und Dentalexperten | Ivoclar
- Ivoclar Spain
- Solutions dentaires modernes pour chirurgiens-dentistes et prothésistes dentaires | Ivoclar
- Ivoclar India
- Ivoclar Italia
- Ivoclar Japan - イボクラールビバデント
- Ivoclar Polska
- Ivoclar Nordic
- Modern Dental Solutions for Dentists & Technicians | Ivoclar USA
- Modern Dental Solutions for Dentists & Technicians | Ivoclar USA