VTSYIQI Food Water Activity Meter Tester Monitor with Software User Guide

June 16, 2024
VTSYIQI

Food Water Activity Meter Tester Monitor with Software

VTSYIQI Food Water Activity Meter Tester Monitor with
SoftwareWe have professionally calibrated the equipment before shipping, but due to the air environment, our equipment is very sensitive, if the customer wants to be more accurate, you can calibrate it yourself after receiving the goods. The calibration method is very simple.
We offer you the following 2 options for calibration without the need to purchase additional saturated salt solution.

Simple version

Simple calibration directly with mineral water, sealed and placed for above 4 hours.
If the test result is 0.98-1, there is no need to calibrate, you can use it directly.

VTSYIQI Food Water Activity Meter Tester Monitor with Software - Simple
version

Precise version

Principle
5 grams of salt, 10 grams of water, sealed and placed for 1 day, and then measured the results of 0.75, within the error range±0.02aw is normal.
Most important:
It does not matter if the instrument is turned off during the sealing test. If it is turned off, please do not separate the instrument and the container. The instrument and the container must be sealed together for 10-30 minutes or 1day. After 10-30 minutes, the calibrated value will show the correct value when you turn on the instrument . Also i have the operation video of the accurate calibration, you can contact me the attached email, I will send it to you.

The operation way

Step 1:
Pour part of salt and part of water (purified water or distilled water is required) Generally speaking, 100 ml of water with more than 36 grams of salt, and salt particles are visible, is a saturated solution.
Step 2:
Stir evenly to a semi-melted state, and you can see salt particles. (This semi-melted state is the correct saturated salt solution). In addition, after the saturated salt solution is prepared, it needs to be sealed for one day before use.
Step 3:
Put the saturated salt solution into the container and carry out the sealing calibration measurement for 10-30 minutes.The range of readings result after calibration is 0.75±0.02
Most important:
It does not matter if the instrument is turned off during the sealing test. If it is turned off, please do not separate the instrument and the container. The instrument and the container must be sealed together for 10-30 minutes . After 10-30 minutes, the calibrated value will show the correct value when you turn on the instrument .
Calibration reading result: 0.75aw±0.02

VTSYIQI Food Water Activity Meter Tester Monitor with Software - Simple
version 1

Documents / Resources

| VTSYIQI Food Water Activity Meter Tester Monitor with Software [pdf] User Guide
Food Water Activity Meter Tester Monitor with Software, Food Water, Activity Meter Tester Monitor with Software, Meter Tester Monitor with Software, Tester Monitor with Software, Monitor with Software
---|---
| VTSYIQI Food Water Activity Meter Tester Monitor with Software [pdf] User Manual
Food Water Activity Meter Tester Monitor with Software, Food, Water Activity Meter Tester Monitor with Software, Tester Monitor with Software, Monitor with Software, Software

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>