DLAB Scientific Co Ltd SP-UV10 UV Visible Spectrophotometer User Manual
- June 16, 2024
- DLAB Scientific Co Ltd
Table of Contents
UV Visible Spectrophotometer
User Manual
SP-UV1000 SCI-UV1000
SP-V1000 SCI-V1000
SP-V1100 SCI-V1100
SP-UV1100 SCI-UV1100
SP-UV3101 SCI-3101UV
SP-UV2102 SCI-2102UV
SP-UV3102 SCI-3102UV
SP-UV2101 SCI-2101UV
SP-XUV5102 SCI-5102XUV
SP-XUV5101 SCI-5101XUV
Introduction
I. Safety
The instrument design complies with the safety specifications set forth by the
National Standard of the People’s Republic of China
GB4793.1-2007/IEC610101:2001 (i.e., Safety Requirements for Measuring, Control
and Lab Electrical Equipment: Part 1) and the National Standard of the
People’s Republic of China GB9706.1-2007/IEC60601-1:21988) (i.e., Medical
Electrical Equipment: Part 1).
II. About the device
SP-UV1000/SP-V1000 serial model visible spectrophotometer is an electronic
laboratory testing equipment.
□ Instrument use mode: Intermittent use
□ Presence or absence of overvoltage (current): No
□ Class of pollution: 2
III. Description of related symbols
| General Warning, danger
---|---
| Warning, electricity hazard
| Warning, hot surface
| Mandatory, Connect an earth terminal to the ground
| Fuse
Crossed out recycle bin, this device is to be recycled by the designated electronic equipment disposal department or vendor after use.
Overview
SP-UV1000/SP-V1000 visible spectrophotometer is characterized by wide wavelength range, high sensitivity, powerful functions, ease of use, simple structure and acceptable appearance, especially the application of large- screen LCD, highprecision A/D converter and nonvolatile memory, which makes the instrument uniquely advantageous among the products of the same grade and widely applicable chemical, pharmaceutical, biochemical, metallurgical, light industry, textile, materials, environmental protection, medical assay and education industries, as one of the important quality control instruments in the analytical testing industries and an essential instrument in conventional labs.
I. Instrument rationale and structure
The rationale of spectrophotometric analysis involves using the selective
absorption of light of different wavelengths by samples to perform qualitative
and quantitative analysis of the samples and judging the structure and
chemical composition of substances through analysis of absorption spectrum.
This instrument works on the principle of relative measurement, i.e.,
selecting a certain solvent (distilled water, air or test sample) as the
reference solution and setting its transmittance (i.e., transmission rate T)
as 100%, while the transmittance of the testing sample is derived with
reference solution as reference standard. The changes in transmittance
(transmission rate T) will have a certain functional relation with the
concentration of the test sample and it complies with the Lambert-Beer’s Law
within a certain range.
T=I/Io
A=KCL=-log I/Io
Wherein T transmittance (transmission rate)
C solution concentration
K absorption coefficient of solution
L length of liquid layer in light path
I intensity of light irradiated on the photoelectric converter through tested
sample
Io intensity of light irradiated on the photoelectric converter through
reference test sample.
SP-UV1000/SP-V1000 ultraviolet/visible spectrophotometer is a domestically
leading new-generation UV-VIS spectrophotometer. These are developed based on
this rationale, in conjunction with the modern precision optics, the latest
microelectronics and other high-end new technologies.
II. Key technical indicators
SP-UV1000/SP-V1000 serial model key technical indicators are as follows:
Item | UV Visible Spectrophotometer |
---|---|
Wavelength range | 325~1000nm |
Photometric range | -0.3~3Abs |
Spectrum bandwidth | 4nm (2nm alternative) |
Wavelength accuracy | ±1nm |
Wavelength repeatability | ≤0.2nm |
Photometric accuracy | ±0.5%T(0~100%T) |
Photometric repeatability | ≤0.2%T(0~100%T) |
Stray light | ≤0.2%T@360nm |
Baseline drift | 0.002A/h@500nm (preheated for 1hour) |
Display | 128×64 dot-matrix graphic LCD |
Data output | USB, parallel port |
Overall dimensions | 490mm×360mm×210mm |
Weight | 12kg |
III. Key functions
SP-UV1000/SP-V1000 serial model visible spectrophotometer has three major
functions.
-
Photometric measurement
**** The measurement modes are switchable (absorbance, transmission rate and energy) and the measurement results (up to 200 groups) can be saved in the instrument’s memory and printed. -
Quantitative measurement
Standard curve can be established using either coefficient method or standard sample calibration method, and the regression equation (up to 200 groups) and measurement results (up to 200 groups) can be saved in the instrument’s memory or printed. -
System application
Turn ON and OFF deuterium and tungsten lamps, calibrate dark current, calibrate wavelength, export data and select measurement mode and restore factory settings.
Instrument Installation
I. Environmental requirements
SP-UV1000/SP-V1000 serial model visible spectrophotometer is designed in
response to the effects of environmental factors on the instrument. To ensure
normal operation of the instrument and achieve an extended service life, check
installation environment condition before installation.
The environmental requirements that ensure normal operation of the instrument
are as follows:
-
Avoid high-temperature and high-humidity environments
The instrument shall be installed away from high-temperature and high-humidity environments. The instrument shall be used at 16~35℃ and 45-80% RH. -
Keep the instrument away from outside magnetic field interferences
Ensure to place the instrument away from magnetic/ electric field and highfrequency waves emittingf electronic devices as much as possible. -
Keep the instrument away from corrosive gases
Please don’t install the instrument in the presence of corrosive gases such as chlorine, hydrochloric acid gas, hydrogen sulfide gas and sulfurous acid gas in air significantly exceeds the applicable standards. -
Instrument must be placed on a steady surface
Place the instrument on a steady and vibration free surface. The instrument’s fan should be sufficient to allow it vent smoothly. -
Supply voltage
The power supply for normal operation of the instrument is AC 220±22V@ 50±1Hz or 110±11V@60±1 Hz. -
Power supply must be well grounded
The instrument should preferably use a standalone power socket and the power supply should be well grounded in order to prevent abnormal operation of the instrument. -
Regulated power supply
When the local voltage is unstable, please equip the instrument with regulated power supply. -
Instrument should be protected from direct sunlight.
-
Maintain dust-free environment.
II. Check supply voltage
Supply voltage varies greatly from one country to another. The input power
supply to any devices sold by DLAB in mainland China is 220V/50Hz.
Ensure to check the supply voltage before installation, in order to avoid
damages to the instrument.
III. Installation
The installation of SP-UV1000/SP-V1000 serial model visible spectrophotometer
is very simple. Please follow the following steps:
Step 1. Check the contents of package
Open the package and carefully and examine its contents against the packing
list. Please contact us or our dealer immediately if any item inside the
package is found missing or damaged.
Step 2. Place the instrument
Take out the instrument and gently place it on a steady surface.
Step 3. Install the printer (an optional accessory)
Check and ensure that the power switch on the printer is OFF and connect the
data cable of the printer in the printer port of the instrument.
Step 4. Install the power cable
Ensure to switch power OFF the instrument. Insert one end of the power cable
into the power port and the other end into the power socket of the instrument.
Step 5. Connect the power supply
Ensure whether all connecting points are well-connected and switch power ON
the instrument. The instrument is now ready for normal use after self-test.
Instrument Description
I. Instrument description
① Sample chamber cover
② Sample rack
③ Operating panel
④ Pull rod
⑤ Cooling fan
⑥ Louver cover| ⑦ Printer port software
⑧ USB(B) for
⑨ USB interface
⑩ Power socket
○11 Power switch
---|---
II. Operating panel
The operating panel of UV-1000/V-1000 serial model visible spectrophotometers
is shown as follows:
① LCD display
② Keypad
III. Key description
| Set the parameters
---|---
| Set the wavelength
| Calibrate 100%T/0Abs
| Print the data
| Function key, promoting functions according to display screen
| Roll down the menu options or turn datasheet pages
| Roll up the menu options or turn datasheet pages
Instrument Use
I. Software system
The instrument’s software system is shown as follows:
II. Basic operations
-
Measurement mode selection
On the main interface, press the corresponding function key to enter the corresponding measurement mode. -
Wavelength Setting
On the key pad, press to enter the wavelength setting interface, press or to change the entered wavelength value and press (left) to proceed to the set wavelength value and automatically calibrate 100%T/0Abs. -
Parameter Setting
Press key to enter the parameter setting interface, press or to select the settings or input parameters and press (left) to confirm. 4. -
Delete the measurement results and stored data
If the local memory is full and you want to clear all saved data, you can long press the button to delete all data in the interface of date view. -
Calibrating 100%T/0Abs
Place the reference into the light path and press key to calibrate 100%T/0Abs. -
Measuring samples
Place the sample into the light path and press (left) to measure the sample. -
Measurement results printing
On the measurement interface, short press to enter print setting, input the data number you want to print, press (left) to print the corresponding test results.
Instrument host and Bluetooth printer online tutorial:
- Turn on the printer switch, the green light continues to flash;
- Open the instrument host, after the instrument self-test is completed, it can automatically connect with the Bluetooth printer.
- If the connection is successful, the green indicator light of the printer will be long bright.
- The operating instrument can be printed!
The specific operation process of the Bluetooth printer can be viewed in its user manual!!
III. Preparations before measurement
-
Power-ON self-test
Ensure that the light path is clear and close the sample chamber cover. Turn ON the power switch on the instrument to start self-test. The start window is like this:
Self-test
It can self-check the following items: color filter, lamp switch, detector, deuterium lamp, tungsten lamp, wavelength calibration and dark current calibration.
Self-test error
If any item has error during self-checking, it will beep and show the error item. You can press any key to continue the process. After the self-test is done, you can go to the system application for further inspections and calibrations. If errors continue, please contact the service team. -
Preheating
After self-test, the instrument enters into the preheating state. For accurate measurement, preheating time needs to be 20-30 minutes. -
Confirm colorimeter cell
Before moving the sample into the colorimeter cell, confirm the cell is clean and free of residues. If the wavelength to be tested is less than 400nm, please use quartz cell.
IV. Measurement
-
Photometric measurement
Photometric measurement is used to test the absorbance or transmittance at a fixed wavelength.
Step 1. Enter the photometric measurement interface
After the self-checking and preheating, the device will go to the main interface of photometric measurement automatically as follows.
You can select the working mode by pressing key and (left) to enter the working mode interface. Then press or to choose working
mode, like absorbance, transmittance or energy. Then confirm and press (right) to back to main interface.
On the “main interface”, press (left) to select “Basic”
Step 2. Set the wavelength to be tested
Press to enter the wavelength setting interface and or to input the wavelength value, and press (left) to proceed to the set wavelength value.
The valid WL range is 200-1000nm, if invalid number is entered, it will beep three times and go back to the main interface of photometric measurement.
Step 3. Calibrate 100%T/0Abs
Place the reference into the light path and wait until the stability indicator shows up, then press to calibrate 100%T/0Abs.
Note: remember to put in the reference sample before doing blanking, otherwise the test results will not be accurate.
Step 4. Measure the samples
Place the sample into the light path, wait until the stability indicator shows up, and press (left) to enter measurement interface, then press (test)
again to have the test results, and the readings will be displayed in the data list.
Repeat this operation to complete the sample measurements
Each page can show three data and you can see other pages via pressing or . In the measurement interface, you can set the wavelength too, but need to do the zero again, otherwise the results will not be accurate.
Step 5. Print the data
On the measurement interface, short press to enter print setting, input the data number you want to print, press (left) to print the corresponding test results.
Step 6. Delete the data
If the local memory is full and you want to clear all saved data, you can long press the button to delete all data in the interface of date view. -
Quantitative measurement
You can create the curve via standard samples with known concentration and use this curve to do quantitative measurement for similar samples. If the K and B are known in advance, you can also set up the curve via coefficient method.
Step 1. Enter the quantitative measurement interface
On the “main interface”, press (right) to select “quantitative measurement”.
Step 2. Establish the standard curve or load the stored standard curve
There are two ways to establish standard curve available for users to choose, depending upon the actual situation.
Coefficient method:
This method is a simple way of quantitative measurement, it is realized by a formula with the absorbance value tested by the spectrophotometer.
1) Select the coefficient method.
Press or to select “create curve” and press or to select “coefficient method”.
2) Set the wavelength to be measured.
Press or to enter the measured wavelength value and pres (left) to confirm and back to previous interface.
3) Set the measurement coefficient K.
Press or to enter the K value and press (left) to confirm and back to previous interface.
4) Set the measurement coefficient B.
Press or to enter the B value and press (left) to confirm and back to previous interface.
Continue to press or until Keypad become OK, press (left) to confirm and complete the coefficient method setting. Now instrument displays the newly established curve. Then you can do the measurements.
Note: before do the tests, make sure do the blanking with your reference sample first, otherwise the results will not be accurate.
Standard Sample Method
1) Select the standard sample method.
Press or to select “create curve”, press press or to select “standard Curve” and press (left) to enter standard sample method.
2) Build new standard curve.
Enter the wavelength and number of samples (up to 9), press (left) to confirm and enter the next interface. Press or to input the concentration of each sample via keypad. Then press to do blanking, after that, press (left) read the absorbance value of each sample. The details are as follows.
The standard curves established after successive completion of calibration of all standard samples will be automatically saved in the instrument’s memory.
Note: the number range of the samples is 1-9. If invalid number is entered, it will beep three time and back to setting interface.
Invoke the stored standard curve:
The newly built curves will be saved to the local memory automatically and uses can invoke the directly. It can save up to 50 curves and the latest curve is at the end of the list.
1) Select the saved curve.
Press or to select “Load Curve”, press or to select the corresponding curve equation and then press (left) to confirm
2) Enter the sample measurement interface
Press (left) to enter the measurement interface after invoking a curve.
3) Calibrate 100%T/0Abs.
Place the reference into the light path and press to calibrate 100%T/0Abs.
4) Measure the sample using the standard curve.
Place the sample into the light path and press (left) to measure it. The results will be displayed in the data list. Repeat this operation to complete measurement of all samples.
5) Print the data
On the measurement interface, short press to enter print setting, input the data number you want to print, press (left) to print the corresponding test results.
6) Delete the data
If the local memory is full and you want to clear all saved data, you can long press the button to delete all data in the interface of date view.
Remove curve:
This is used to delete the saved curves. You can go to this menu and select the curve needs to be removed, and press (left) to delete the curves.
V. System Setting
Users may set the system configuration of the instrument in the “system
application” depending upon actual use. On the “main interface”, press key to
enter the setting parameter interface.
-
Measurement mode
Press or to select “Test Mode”, press (left) to enter, press or to select “absorbance”, “transmittance” or “energy” and then press (left) to confirm. -
Turn ON and OFF the deuterium lamp
When measuring with testing wavelength within the range of 3401000nm, users can turn OFF the deuterium lamp to extend the service life of deuterium lamp.
Press or to select “D2 Lamp”, press (left) to enter, press or to select “ON” or “Off” and then press (left) to confirm.
-
Turn ON and OFF the tungsten lamp
When measuring with testing wavelength within the range of 200339nm, users can turn OFF the tungsten lamp to extend the service life of tungsten lamp.
Press or to select “D2 Lamp”, press (left) to enter, press to select “ON” or “Off” and then press or (left) to confirm.
-
Dark Current Calibration
When the working environment of the instrument changes, users may recalibrate the dark current after the instrument preheating and before sample measurement.
Press or to select “Dark Current”, press (left) to calibrate.
The sample chamber cover must be closed throughout the entire calibration process.
Return after completion.
-
Wavelength calibration
Sudden power failures while the instrument is in use, abnormal power OFF and attenuation of light source energy will all result in wavelength deviation and deviation of measurement results, where wavelength recalibration is required.
Make sure that the light path is clear and the sample chamber cover is closed. Press or to select “Reset WL” and press (left) to calibrate. The sample chamber cover may not be opened throughout the calibration process. Return after completion.
-
Version information
Press or to select “version”, press (left) to view the version information interface and then press any key to return.
-
Export data to USB
You can transfer the saved data on the spectrophotometer to the USB. Press or to select “Export”, press (left) to transfer data.
Before export the data, you must insert the USB to the USB port on the back of the device, otherwise it will give error information.
-
Date and Time Setting
Press or to select “Time Setting”, press (left) to enter the menu and set the correct date and time.
FCC Statement
This device complies with part 15 of the FCC rules. Operation is subject to
the following two conditions: (1) this device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
NOTE: This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to par t 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates uses and
can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
Important Note:
Radiation Exposure Statement
This equipment complies with FCC radiation exposure limits set forth for an
uncontrolled environment. This equipment should be installed and operated with
minimum distance 20cm between the radiator and your body.
This transmitter must not be co-located or operating in conjunction with any
other antenna or transmitter.
ISED Statement
This device complies with Industry Canada license‐exempt RSS standard(s).
Operation is subject to the following two conditions: (1) This device may not
cause interference, and (2) This device must accept any interference,
including interference that may cause undesired operation of the device.
The digital apparatus complies with Canadian CAN ICES‐3 (B)/NMB‐3(B).
This radio transmitter has been approved by Industry Canada to operate with
the antenna types listed with the maximum permissible gain indicated. Antenna
types not included in this list, having a gain greater than the maximum gain
indicated for that type, are strictly prohibited for use with this device.
Radiation Exposure Statement
This equipment complies with Canada radiation exposure limits set forth for
an uncontrolled environment. This equipment should be installed and operated
with minimum distance 20cm between the radiator & your body.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>