Holland Green Science 10209001 Aeolus Vacuum Controller User Manual
- June 16, 2024
- Holland Green Science
Table of Contents
10209001 Aeolus
Vacuum Controller
User Manual
User Manual
LIMITED WARRANTY & DISCLAIMER
- The intellectual property rights for this manual belongs to Holland Green Science Corporation (hereinafter referred to as “the Company”).
- Products of the company are under the patent protection of China and other countries and regions (inclusive of obtained patents and patents currently under application).
- The company reserves the right to change product specifications without notice.
- The contents of the manual are subject to change without notice.
- The contents of the manual must not be copied, extracted, or modified in any form without the Company’s permission.
LIMITED WARRANTY & DISCLAIMER
Products sold by the Company are intended solely for the use(s) specified by
the manufacturer and are not to be used for any other purposes including, but
not limited to, unauthorized commercial purposes. Buyer warrants and
represents that Buyer will properly test, use, qualify, and/or validate each
Product for intended use, and, to the extent authorized, manufacture and
market any final articles made from the Company’s Products in accordance with
the practices of a reasonable person who is an expert in the field, and in
strict compliance with all applicable national, state and local laws and
regulations. Buyer acknowledges and understands that the purchase of Products
only conveys to Buyer the non-transferable right for only the Buyer to use the
Products purchased in compliance with the applicable manufacturer’s Published
User Statement, Limited User Statement, or Limited License, if any. While this
manual contains certain representations of the potential dangers from the
instrument or equipment, Buyer is solely responsible for conducting any
research necessary to learn the hazards involved for any of its intended uses
of the instrument or equipment purchased from the Company and to properly warn
its customers, employees and other personnel who may be exposed to such
instrument or equipment of any risks involved in using or handling the
instrument or equipment.
The Company warrants that this product will operate or perform substantially
in conformance with the Company’s published specifications and be free from
defects in material and workmanship, when subjected to normal, proper and
intended storage and usage by properly trained personnel, for the duration of
the manufacturer’s warranty period set forth in this product documentation,
published specifications or package inserts. If a manufacturer’s warranty
period is not specified in the product documentation, published specifications
or package inserts, the warranty period shall be one (1) year from the date of
Bill of Sale to Buyer for equipment, and the earlier of any stated expiration
dates or ninety (90) days for all other products (the “Warranty Period”). The
Company agrees during the Warranty Period, to repair or replace, at the
Company’s option, defective Products so as to cause the same to operate in
substantial conformance with said published specifications; provided that
Buyer shall (a) promptly notify the Company in writing upon the discovery of
any defect, which notice shall include the product model and serial number (if
applicable) and details of the warranty claim; and (b) after the Company’s
review, the Company will provide Buyer with service data and/or a Return
Material Authorization (“RMA”), which may include biohazard decontamination
procedures and other product-specific handling instructions, then, if
applicable, Buyer may return the defective Products to the Company with all
costs prepaid by Buyer. Replacement parts may be new or refurbished, at the
election of the Company. All replaced parts shall become the property of the
Company. Shipment to Buyer of repaired or replacement Products shall be made
in accordance with the Delivery provisions of the Terms & Conditions agreed to
by the purchase of this product. Consumables are expressly excluded from this
warranty.
Notwithstanding the foregoing, Products supplied by the Company but not
manufactured by the Company are not warranted by the Company, and the Company
disclaims all warranties associated with the accuracy of any original
manufacturer or third-party supplier Product information, but the Company
agrees to assign to Buyer any warranty rights in such Product that the Company
may have from the original manufacturer or third-party supplier, to the extent
such assignment is allowed by such original manufacturer or third-party
supplier.
In no event shall the Company have any obligation to make repairs,
replacements or corrections required, in whole or in part, as the result of
(i) normal wear and tear, (ii) accident, disaster or event of force majeure,
(iii) misuse, fault or negligence of or by Buyer, (iv) use of the Product in a
manner for which they were not designed, (v) causes external to the Product
such as, but not limited to, power failure or electrical power surges, (vi)
improper storage and handling of the Product or (vii) use of the Product in
combination with equipment or software not supplied by the Company. If the
Company determines that Product for which Buyer has requested warranty
services are not covered by the warranty hereunder, Buyer shall pay or
reimburse the Company for all costs of investigating and responding to such
request at the Company’s then prevailing time and materials rates. If the
Company provides repair services or replacement parts that are not covered by
this warranty, Buyer shall pay the Company therefor at the Company’s then
prevailing time and materials rates.
ANY INSTALLATION, MAINTENANCE, REPAIR, SERVICE, RELOCATION OR ALTERATION TO OR
OF, OR OTHER TAMPERING WITH, THE PRODUCT PERFORMED BY ANY PERSON OR ENTITY
OTHER THAN THE COMPANY WITHOUT THE COMPANY’S PRIOR WRITTEN APPROVAL, OR ANY
USE OF REPLACEMENT PARTS NOT SUPPLIED BY THE COMPANY, SHALL IMMEDIATELY VOID
AND CANCEL ALL WARRANTIES WITH RESPECT TO THE PRODUCT.
THE OBLIGATIONS CREATED BY THIS WARRANTY STATEMENT TO REPAIR OR REPLACE A
DEFECTIVE PRODUCT SHALL BE THE SOLE REMEDY OF BUYER IN THE EVENT OF A
DEFECTIVE PRODUCT. EXCEPT AS EXPRESSLY PROVIDED IN THIS WARRANTY STATEMENT,
THE COMPANY DISCLAIMS ALL OTHER WARRANTIES, WHETHER EXPRESS OR IMPLIED, ORAL
OR WRITTEN, WITH RESPECT TO THE PRODUCTS, INCLUDING WITHOUT LIMITATION ALL
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE OR
USE. THE COMPANY DOES NOT WARRANT THAT THE PRODUCTS ARE ERROR-FREE OR WILL
ACCOMPLISH ANY PARTICULAR RESULT.
NOTHING IN THIS MANUAL OR IN THE TERMS AND AGREEMENT AGREED TO AT THE TIME OF
PURCHASE SHALL BE CONSTRUED TO PROVIDE FOR THE TRANSFER, ASSIGNMENT OR
ALIENATION OF THE LIMITED WARRANTIES PROVIDED BY THE COMPANY HEREIN TO ANY
OTHER PARTY OTHER THAN BUYER, INCLUDING BUT NOT LIMITED TO, SUCCESSORS, HEIRS,
SUBSIDIARIES OR ANY OTHER THIRD PERSON. THE WARRANTIES IDENTIFIED ABOVE ARE
THE SOLE AND EXCLUSIVE WARRANTIES WITH RESPECT TO THIS PRODUCT AND ARE IN LIEU
OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED. ALL OTHER WARRANTIES ARE
EXPRESSLY DISCLAIMED, INCLUDING, WITHOUT LIMITATION, THOSE OF NON-
INFRINGEMENT, MERCHANTABILITY, OR FITNESS FOR A SPECIFIC PURPOSE OR USE
(WHETHER ARISING FROM STATUTE, OR OTHERWISE IN LAW, OR FROM A COURSE OF
DEALING, OR USAGE OF TRADE). THIS WARRANTY IS NOT TRANSFERABLE FROM THE
ORIGINAL PURCHASER TO A SUBSEQUENT OWNER. FURTHER, THE COMPANY IS NOT LIABLE
IN CASES OF DELIBERATE, NEGLIGENT OR ACCIDENTAL MISUSE OF THE PRODUCTS, USE
WITH INAPPROPRIATE REAGENTS OR CONSUMABLES, DAMAGE CAUSED BY DISASTER, REPAIR
OR MODIFICATIONS DONE BY ANYONE OTHER THAN THE COMPANY. THE COMPANY DOES NOT
WARRANT THAT THE PRODUCT WILL NOT
INFRINGE THE INTELLECTUAL PROPERTY RIGHTS OF A THIRD PARTY EITHER ALONE OR IN
COMBINATION WITH OTHER PRODUCTS OR IN THE OPERATION OF ANY PROCESS. THE
COMPANY’S TOTAL LIABILITY FOR BREACH OF THESE WARRANTIES SHALL BE LIMITED TO
THE GROSS PURCHASE PRICE FOR THE PRODUCT AS STATED IN THE RELEVANT INVOICE OR
ORDER UNDER WHICH THE ACTIVITY GIVING RISE TO LIABILITY ARISES. UNDER NO
CIRCUMSTANCES SHALL THE COMPANY BE LIABLE FOR INDIRECT, INCIDENTAL,
CONSEQUENTIAL, SPECIAL OR PUNITIVE DAMAGES (INCLUDING LOSS OF BUSINESS OR
PROFITS) UNDER ANY THEORY OF LIABILITY HEREUNDER.
For claims under the warranty please contact your local supplier. You may also
send the instrument directly to manufacturer, enclosing the invoice copy and
by giving reasons for the claim.
LIMITATION OF LIABILITY
NOTWITHSTANDING ANYTHING TO THE CONTRARY CONTAINED HEREIN, THE LIABILITY OF
THE COMPANY UNDER THESE TERMS AND CONDITIONS (WHETHER BY REASON OF BREACH OF
CONTRACT, TORT, INDEMNIFICATION, OR OTHERWISE, BUT EXCLUDING LIABILITY OF THE
COMPANY FOR BREACH OF WARRANTY (THE SOLE REMEDY FOR WHICH SHALL BE AS PROVIDED
UNDER THIS SECTION) SHALL NOT EXCEED AN AMOUNT EQUAL TO THE LESSER OF (A) THE
TOTAL PURCHASE PRICE THERETOFORE PAID BY BUYER TO THE COMPANY WITH RESPECT TO
THE PRODUCT(S) GIVING RISE TO SUCH LIABILITY OR (B) ONE HUNDRED THOUSAND
DOLLARS ($100,000), WHICHEVER IS LESS.
With respect to any software products incorporated in or forming a part of the
Product hereunder Buyer intends and agrees that such software products are
being licensed and not sold, and that the words “purchase”, “sell” or similar
or derivative words are understood and agreed to mean “license”, and that the
word “Buyer” or similar or derivative words are understood and agreed to mean
“licensee”. Notwithstanding anything to the contrary contained herein, The
COMPANY or its licensor, as the case may be, retains all rights and interest
in software products provided hereunder. See the Terms & Conditions of
purchase for details regarding such license.
Safety
DANGER (may cause serious damage to property and or casualties)
- Please carefully read this User Manual prior to operating the instrument and observe the instructions on safe operation.
- The power source must be grounded reliably and away from any sources of electromagnetic interference.
- Confirm that the voltage and frequency of the power supply matches the specified voltage and frequency specified prior to use.
- The instrument should be operated on an independent power outlet.
- Do not allow the instrument to be plugged in or unplugged whilst the power switch is in the on position.
- Do not modify the power cord.
- The working environment must be free of any flammable, explosive, volatile or corrosive substances.
- The instrument is deenergised only when the power cable is disconnected from the power source.
- This instrument is for indoor use only.
- The instrument should be operated in a low humidity, well ventilated, dust and water free environment, without direct sunshine or an intense magnetic, light or heat source.
- Ensure that the instrument and its accessories are free of any potential defects.
- Ensure only standard accessories supplied by the manufacturers are used with this instrument.
WARNING (may cause property damage or personal injury)
- Prevent water from splashing on the electrical elements of this instrument.
- Before assembly, disassembly, cleaning or maintenance, the power source must be disconnected.
- Ensure the vacuum pump’s power is not more than the maximum power output of the instrument.
ATTENTION (may affect operational performance or service life)
- When pulling the power plug out from a power source, do not pull out using the power cord directly.
- Only accredited and qualified professional repair technicians can open the instrument or conduct required repairs. Persons performing repairs on the instrument other than those selected or approved by the Company shall operate to void any warranty contained hereinabove for the product
Introduction
The instrument is intended to control the vacuum level of a vacuum system with
gas as the medium in non-residential environments such as schools,
laboratories and or factories.
Environment requirements are as follows:
- Voltage fluctuation does not exceed nominal value by ±10%.
- Place the instrument 30cm away from the wall and at least 10 cm from any other instruments.
1. After Sales Support
If problems are encountered or technical support is required when installing
or using the instrument, please contact
serviceusa@hollandgreenscience.com
The Company may provide technical assistance and information regarding the
instrument or equipment or service without charge at its sole discretion.
Buyer assumes sole responsibility for any reliance on or use of such
assistance and information, and the
Company makes no warranty thereon.
Upon contact the following information is required:
- Product serial number (located on the instrument nameplate)
- Description of issue or problem
- Method and or operating steps you have undertaken towards resolution.
- Your contact details inclusive of telephone number and email address.
2. Proper Use
The instrument is designed for non-residential use and is to be used in
conjunction only with accessories recommended within this manual and by the
manufacturer.
Technical Specifications
Model | 10209001 |
---|---|
Product Name | Aeolus |
Control Range | 1-1000mbar |
Measurement Range | 1-1000mbar |
Control Operation Modes | Single-point control operation mode, program |
operation mode
Control Program| Up to 5 Programs.
Each program contains five-step control (vacuum degree and time can be set).
Interface| Touch Screen setting
Display Screen| 5-inch TFT
Sensor pressure limit| 1500mbar
Connecting Tube Interface Diameter| 8mm
Materials in Contact with Process| PTFE, PP, Silica Gel and Ceramics
Process Temperature| 5℃ to 40℃
External Dimension D x W x H| 189mm×207mm×193mm
Weight| 4.0kg
Input Voltage| AC110—240V, 50Hz/60Hz
Ambient Temperature| 10℃-40℃, ≤80%RH
Input Power| 600W
Maximum Output (Control) Power| 500W(220V), 300W(110V)
Inspection
Packing List
Unpack the equipment carefully and check for any damage which may have arisen
during transport. In the event of identified damage, please contact
serviceusa@hollandgreenscience.com
The package includes the following items:
Item Description.
|
Quantity.
---|---
Main Unit| 1
Power Cable| 1
Quick Setup Card| 1
Note:
If you find any visible damage on the product, please do not connect the
product to the power supply.
Operating Instructions
1. Overview
Name
|
Description
---|---
Touch Screen| Setting parameters and displaying settings and operation
information
Three-in-One Switch| Unit power input, with a switch controlling the power
supply to the entire unit.
Pump Connector| Connects the vacuum pump
Air Vent| Connects with air or inert gas if necessary.
Application Connector| Connects with the application.
Power Output Socket| Provides power (control) to vacuum pump.
- Instrument Preparation
- Place the instrument on a stable and well-ventilated surface, clear of flammables or explosives.
- Ensure the instrument voltage matches the power supply voltage before connection.
- Ensure that the power socket is grounded properly.
- Connect the vacuum hose of the pump and application.
- Connect the power cable of the vacuum pump to the instrument and turn ‘ON’ the vacuum pump switch. Do not exceed the rated output power of the controller.
- Connect the power cable of the instrument and turn ‘ON’ the power switch.
3. Start-up Interface Description
Button Name
|
Function
---|---
VENT| Press VENT to vent and release to stop venting.
PROGRAM| Choose the control mode and set the control pressure value and
running time.
SETTINGS| Set the delay time to ‘open and close’ the pump and valve. Set
the local atmospheric pressure and then adjust the Pressure sensor calibration
parameters.
START| Press START to commence the control program, the icon will change
to , press to ‘STOP’ the control program.
****| ‘PROGRAM’ and ‘SETTINGS’ buttons cannot be operated if the control
program starts.
4. Instrument Configuration
Press the ‘SETTINGS’ button to display the ‘PASSWORD SOFT KEYBOARD’, enter
the password. The default initial password is 6666, then press ‘ENTER’,
confirm the password is correct and then ‘SETTING INTERFACE’ displays
Name
|
Function
---|---
Pump Delay (ms)| The delay time after ‘opening or closing’ the vacuum pump.
Valve Delay (ms)| The delay time after ‘opening or closing the solenoid valve
(inside the instrument).
K, B| Pressure sensor calibration parameters
P0 (mbar)| The local atmospheric pressure value
Reset| Restore to factory settings
After setting, press the ‘OK’ key to return to the start-up interface.
Note:
Before pressing the ‘Reset’ button, record the pressure sensor calibration
parameters (K and B). After restoring the default setting, write (K and B)
into the instrument otherwise, the pressure measurement accuracy will be
adversely affected.
5. Operating Mode – Selection and Configuration
Press the ‘PROGRAM’ button to display the ‘Program Selection Interface’
5.1 Single-Point Control Operation Mode
Press ‘VACUUM CONTROL’ to refresh the current page
Name
|
Function
---|---
Set Vacuum (mbar)| Set the vacuum value within a range of 1 – local
atmospheric pressure (P0)
Hysteresis (mbar)| Hysteresis value affects the pressure fluctuation
range and the ‘ON/OFF’ frequency of the vacuum pump.
Setting range 1-300mbar
Time (min)| Time can be set within a range of 0-5999 min
After setting the parameters, press the ‘OK’ key to return to the ‘Start-Up
Interface’.
Note :
The larger the hysteresis value set, the larger the pressure fluctuation, the
smaller the hysteresis value set, the smaller the pressure fluctuation, and
the more frequently the vacuum pump starts up and shuts down. The recommended
hysteresis value is set at 10mbar.
Note:
The sum of the set vacuum value and hysteresis value cannot exceed (P0-10)
mbar and the set vacuum value is larger than the hysteresis value.
5.2 Program Operation Mode
Press ‘VACUUM CONTROL’ button to display the ‘Program Selection Interface’
Name|
Function
---|---
Open| Select the program number
Hysteresis (mbar)| Hysteresis value affects the pressure fluctuation
range and the ‘ON’ frequency of the vacuum pump Setting range 1-300mbar
After setting the parameters, press the ‘OK’ key to enter the program
operation
5.3 Program Editing
Press the ‘Edit’ button in the ‘Program Selection Interface’ to display the
‘Program Editing Interface’
Name|
Function
---|---
NO| Sequence No. of program execution
Time (min)| Timing, set within a range of 0-5999min
Vac (mbar)| Set the vacuum value within a range of 1 – local atmospheric
pressure (P0)
After setting the parameters, press the ‘Program Number’ to save until the upper part of the interface indicates ‘SAVE’
Note:
If the set vacuum value or timing is blank, it indicates that the current step
will be skipped to the next control program.
5.4 Ventilation
Press the ‘VENT’ button to reduce the vacuum.
Fault
|
Cause Analysis
| Solution
---|---|---
The vacuum pump does not work.| Check the power switch on the vacuum pump is
‘ON’.
Check the power cable connection.| Switch the vacuum pump, power switch ‘ON’.
Connect the cable.
An error message displays when setting the
pressure value| Check whether the SET value exceeds the limit.| Adjust the SET
value.
The solenoid valve and pump switch ON / OFF
frequently.| Check the pipeline for air tightness.
Check whether the hysteresis value is too low.| Connect the pipeline properly.
Increase the hysteresis value.
Maintenance and Cleaning
Keep the instrument dry and clean in routine operations.
Clean the outer surface with a non-abrasive cleanser and only connect the
power supply when the entire instrument surface is dry.
If liquid or moist solids enter the instrument, please immediately disconnect
the power supply and contact
supportusa@hollandgreenscience.com
- Surface stains on the instrument should be cleaned only with a clean, soft rag and detergent.
- The power must be disconnected before any maintenance or cleaning.
- Do not clean the instrument with any corrosive cleaning solutions.
- If the instrument is left unused for a long period, switch ‘OFF’ the power and store it on a dry, clean, level, and stable surface at normal temperature.
Caution!
Before any maintenance or inspection, the power cable MUST be removed from the
power socket.